Four Way Crossover Closed Loop With Exercise Detection

August 16, 2018 updated by: Jessica Castle, Oregon Health and Science University

A Randomized, Four-way, Cross-over Outpatient Study to Assess the Efficacy of a Dual-hormone Versus Insulin Alone Closed-loop System With Exercise Detection vs a Predictive Low Glucose Suspend System vs Current Care

Closed-loop systems are an emerging technology that automate hormone delivery. They are quickly paving the way to revolutionize the treatment of type 1 diabetes. Several categories have emerged: dual-hormone (insulin and glucagon) closed-loop systems and closed-loop systems with insulin only, one variety of which is the low glucose suspend safety feature now available from Medtronic (MiniMed 530G with Enlite). The study described within this protocol is designed to test the efficacy of a new closed-loop algorithm for managing blood glucose in people with type 1 diabetes before and after exercise. The new algorithm will have 3 modes: a single hormone insulin only mode, a dual-hormone insulin and glucagon mode and an insulin only mode with predictive low glucose suspend, all with an exercise detection algorithm. The purpose of this study is to determine whether a dual hormone AP with an exercise detection algorithm outperforms both single hormone AP and a low glucose suspend algorithm and sensor augmented pump therapy using the subject's own insulin pump.

Study Overview

Detailed Description

Subjects will undergo four approximately 84 hour studies. The first day of each study will be an approximately 12 hour inpatient visit to include activities of daily living and an exercise period with 2 days spent as an outpatient. The subject will come back to the research center on the fourth day to complete another approximately 12 hour inpatient visit to include activities of daily living and an exercise period followed by removal of all devices. In randomized order, subjects will have glucose controlled using the following systems: 1) dual-hormone closed-loop system, 2) insulin only closed-loop system, 3) predictive low glucose suspend system and 4) sensor augmented pump therapy with subject using their own insulin and insulin pump. Both of the closed-loop system algorithms have an exercise detection algorithm that uses inputs from a heart rate monitor and accelerometer, the Zephyrlife BioPatch. After exercise detection, insulin is turned off for the first 30 minutes, the total insulin infusion rate is adjusted by an exercise multiplier, and for the dual-hormone system, the target glucose for glucagon is increased along with the maximum dose of glucagon allowed. The predictive low glucose suspend system will utilize the patient's optimized basal rates, correction factor, and carb ratio. The patient will bolus for meals and hyperglycemia as usual under PLGS, but will have the additional safety net of the pump suspending insulin when it predicts a hypoglycemic event. The Dexcom CGM system and the Tandem t:slim pumps will be used for all three intervention visits.

Subjects will arrive at the CTRC at 7am at the start of intervention visits. Subjects will eat breakfast, lunch and dinner in the research center. Subjects will eat breakfast at 8:30am. After breakfast, subjects will complete activities of daily living (sitting on a chair, lying on a bed, washing dishes, sweeping the floor, etc.). Subjects will eat lunch at 12pm. Subjects will exercise for 45 minutes on a treadmill around 2pm. Subjects will eat dinner at 6pm and be discharged. The subject will then go home for Day 2 and Day 3 and return on Day 4 to complete another 12 hour inpatient visit with the schedule of activities identical to Day 1.

Study Type

Interventional

Enrollment (Actual)

25

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • Oregon
      • Portland, Oregon, United States, 97239
        • Oregon Health and Science University

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

21 years to 45 years (Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Description

Inclusion Criteria:

  • Diagnosis of type 1 diabetes mellitus for at least 1 year.
  • Physically willing and able to perform 45 min of exercise (as determined by the investigator after reviewing the subjects activity level)
  • Current use of an insulin pump.
  • Lives with another person age 18 or older who will be present while subject exercises at home and that can attend the training on using the system with the subject.
  • Lives within 20 miles of OHSU.
  • A1C<10%
  • Willingness to follow all study procedures, including attending all clinic visits.
  • Willingness to sign informed consent and HIPAA documents.

Exclusion Criteria:

  • Female of childbearing potential who is pregnant or intending to become pregnant or breast-feeding, or is not using adequate contraceptive methods. Acceptable contraception includes birth control pill / patch / vaginal ring, Depo-Provera, Norplant, an IUD, the double barrier method (the woman uses a diaphragm and spermicide and the man uses a condom), or abstinence.
  • Any cardiovascular disease, defined as a clinically significant EKG abnormality at the time of screening or any history of: stroke, heart failure, myocardial infarction, angina pectoris, or coronary arterial bypass graft or angioplasty. Diagnosis of 2nd or 3rd degree heart block or any non-physiological arrhythmia judged by the investigator to be exclusionary.
  • Renal insufficiency (GFR < 60 ml/min, using the MDRD equation as report by the OHSU laboratory).
  • Liver failure, cirrhosis, or any other liver disease that compromises liver function as determined by the investigator.
  • Hematocrit of less than or equal to 34%.
  • Hypertensive subjects with systolic blood pressure >= 160 mmHg or diastolic blood pressure >= 100 mmHg despite treatment or who have treatment-refractory hypertension (e.g. requiring four or more medications).
  • History of severe hypoglycemia during the past 12 months prior to screening visit or hypoglycemia unawareness as judged by the investigator. Subjects will complete a hypoglycemia awareness questionnaire. Subjects will be excluded for four or more R responses.
  • Adrenal insufficiency.
  • Any active infection.
  • Known or suspected abuse of alcohol, narcotics, or illicit drugs.
  • Seizure disorder.
  • Active foot ulceration.
  • Severe peripheral arterial disease characterized by ischemic rest pain or severe claudication.
  • Major surgical operation within 30 days prior to screening.
  • Use of an investigational drug within 30 days prior to screening.
  • Chronic usage of any immunosuppressive medication (such as cyclosporine, azathioprine, sirolimus, or tacrolimus).
  • Bleeding disorder, treatment with warfarin, or platelet count below 50,000.
  • Allergy to aspart insulin.
  • Allergy to glucagon.
  • Insulin resistance requiring more than 200 units per day.
  • Need for uninterrupted treatment of acetaminophen.
  • Current administration of oral or parenteral corticosteroids.
  • Any life threatening disease, including malignant neoplasms and medical history of malignant neoplasms within the past 5 years prior to screening (except basal and squamous cell skin cancer).
  • Beta blockers or non-dihydropyridine calcium channel blockers.
  • Current use of any medication intended to lower glucose other than insulin (ex. use of liraglutide).
  • Diagnosis of pheochromocytoma, insulinoma, or glucagonoma, personal or family history of multiple endocrine neoplasia (MEN) 2A, MEN 2B, neurofibromatosis or von Hippel-Lindau disease.
  • History of severe hypersensitivity to milk protein.
  • Current use of any medication with strong anticholinergic properties, such as antihistamines, sleep aids, and antidiarrheal medications.
  • Current use of indomethacin.
  • Conditions that may result in low levels of releasable glucose in the liver and an inadequate reversal of hypoglycemia by glucagon such as prolonged fasting, starvation or chronic hypoglycemia as determined by the investigator.
  • A positive response to any of the questions from the Physical Activity Readiness Questionnaire.
  • Any chest discomfort with physical activity, including pain or pressure, or other types of discomfort.
  • Any clinically significant disease or disorder which in the opinion of the Investigator may jeopardize the subject's safety or compliance with the protocol.

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Treatment
  • Allocation: Randomized
  • Interventional Model: Crossover Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: Predictive Low Glucose Suspend Arm
Subjects will undergo an 84 hour study with 24 hours inpatient and 60 hour outpatient. The predictive low glucose suspend system will run through the artificial pancreas controller in predictive low glucose suspend mode and utilize the patient's optimized basal rates, correction factor, and carb ratio, but it will have the additional safety net of the pump suspending insulin when it predicts a hypoglycemic event.
The artificial pancreas controller contains an algorithm for managing blood glucose in people with type 1 diabetes which includes an exercise detection component. The new algorithm will have 3 modes: a single hormone closed-loop insulin only mode, a dual-hormone closed-loop insulin and glucagon mode and an insulin only mode with predictive low glucose suspend. Closed-loop Artificial Pancreas Controller includes insulin delivery algorithm implemented on a smart phone, utilizing sensor glucose values from a Dexcom G4 sensor and sending delivery commands to one Tandem t:slim insulin pumps filled with insulin.
Experimental: Dual Hormone Closed-loop Arm
Subjects will undergo an 84 hour study with 24 hours inpatient and 60 hour outpatient using the closed-loop artificial pancreas controller in dual hormone mode to manage blood sugar that includes an exercise detection component that includes a reduction in insulin delivery and an increase in glucagon delivery upon detection.
The artificial pancreas controller contains an algorithm for managing blood glucose in people with type 1 diabetes which includes an exercise detection component. The new algorithm will have 3 modes: a single hormone closed-loop insulin only mode, a dual-hormone closed-loop insulin and glucagon mode and an insulin only mode with predictive low glucose suspend. Closed-loop Artificial Pancreas Controller includes insulin and glucagon delivery algorithm implemented on a smart phone, utilizing sensor glucose values from a Dexcom G4 sensor and sending delivery commands to two Tandem t:slim insulin pumps, one filled with insulin and one with glucagon.
Other Names:
  • APC
Experimental: Single Hormone Closed-loop Arm
Subjects will undergo an 84 hour study with 24 hours inpatient and 60 hour outpatient using the closed-loop artificial pancreas controller in single hormone mode to manage blood sugar that includes an exercise detection component that includes a reduction in insulin delivery upon detection.
The artificial pancreas controller contains an algorithm for managing blood glucose in people with type 1 diabetes which includes an exercise detection component. The new algorithm will have 3 modes: a single hormone closed-loop insulin only mode, a dual-hormone closed-loop insulin and glucagon mode and an insulin only mode with predictive low glucose suspend. Closed-loop Artificial Pancreas Controller includes insulin delivery algorithm implemented on a smart phone, utilizing sensor glucose values from a Dexcom G4 sensor and sending delivery commands to one Tandem t:slim insulin pumps filled with insulin.
Active Comparator: Sensor Augmented Pump Therapy arm
Subjects will undergo an 84 hour study with 24 hours inpatient and 60 hour outpatient with subject's insulin pump and glucose sensor, if he/she typically uses one. Subjects will still wear a heart rate monitor uploading to a smart phone.
Subject will continue on their subcutaneous delivery of insulin on his/her own insulin pump using their own basal rates and carb ratios for meal boluses, managing their blood sugar as they normally would.

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Percent of Time With Sensed Glucose < 3.9 mmol/L
Time Frame: entire 84 hour study
Assess the percent of time that the Dexcom G4 Share reported sensor glucose values less than 3.9 mmol/L using Dexcom sensor downloads across all four arms.
entire 84 hour study
Percent of Time With Sensed Glucose < 3.9 mmol/L
Time Frame: From 14:00-18:00 for each 12 hour inpatient visit
Assess the percent of time that the Dexcom G4 Share reported sensor glucose values less than 3.9 mmol/L using Dexcom sensor downloads across all four arms.
From 14:00-18:00 for each 12 hour inpatient visit
Percent of Time With Sensed Glucose Between 3.9-10 mmol/L
Time Frame: entire 84 hour study
Assess the percent of time that the Dexcom G4 Share reported sensor glucose values between 3.9-10 mmol/L using Dexcom sensor downloads across all four arms.
entire 84 hour study
Percent of Time With Sensed Glucose Between 3.9-10 mmol/L
Time Frame: From 14:00-18:00 for each 12 hour inpatient visit
Assess the percent of time that the Dexcom G4 Share reported sensor glucose values between 3.9-10 mmol/L using Dexcom sensor downloads across all four arms.
From 14:00-18:00 for each 12 hour inpatient visit

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Mean Sensed Glucose
Time Frame: entire 84 hour study
Assess the mean sensor glucose using Dexcom sensor downloads across all four arms.
entire 84 hour study
Number of Carbohydrate Treatments
Time Frame: entire 84 hour study
Assess the number of rescue carbohydrate treatments per day across all four arms.
entire 84 hour study
Percent of Time With Sensed Glucose < 3.0 mmol/L
Time Frame: entire 84 hour study
Assess the percent of time that the Dexcom G4 Share reported sensor glucose values less than 3.0 mmol/L using Dexcom sensor downloads across all four arms.
entire 84 hour study
Percent of Time With Sensed Glucose > 10 mmol/L
Time Frame: entire 84 hour study
Assess the percent of time that the Dexcom G4 Share reported sensor glucose values greater than 10 mmol/L using Dexcom sensor downloads across all four arms
entire 84 hour study
Number of Events With Capillary Blood Glucose < 3.9 mmol/L
Time Frame: entire 84 hour study
Assess number of events with capillary blood glucose < 3.9 mmol/L using downloads from a Contour Next blood glucose meter across all four arms. .
entire 84 hour study
Number of Events With Capillary Blood Glucose < 3.0 mmol/L
Time Frame: entire 84 hour study
Assess number of events with capillary blood glucose < 3.0 mmol/L using downloads from a Contour Next blood glucose meter across all four arms.
entire 84 hour study
Mean Amount of Insulin Delivered
Time Frame: 24 hours
Assess the average amount of insulin delivered per day in units/kg as documented through the artificial pancreas controller across all four arms.
24 hours
Mean Amount of Glucagon Delivered
Time Frame: 24 hours
Assess the average amount of insulin delivered per day in units/kg as documented through the artificial pancreas controller in dual hormone arm.
24 hours

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Publications and helpful links

The person responsible for entering information about the study voluntarily provides these publications. These may be about anything related to the study.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

October 1, 2016

Primary Completion (Actual)

August 1, 2017

Study Completion (Actual)

January 31, 2018

Study Registration Dates

First Submitted

July 28, 2016

First Submitted That Met QC Criteria

August 5, 2016

First Posted (Estimate)

August 11, 2016

Study Record Updates

Last Update Posted (Actual)

September 14, 2018

Last Update Submitted That Met QC Criteria

August 16, 2018

Last Verified

August 1, 2018

More Information

Terms related to this study

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

NO

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Type 1 Diabetes

Clinical Trials on Artificial Pancreas Controller in Predictive Low Glucose Suspend Mode

Subscribe