Three Way Crossover Closed Loop Study With Xeris Glucagon

May 13, 2020 updated by: Jessica Castle, Oregon Health and Science University

A Randomized, Three-way, Cross-over Study to Assess the Efficacy of a Dual-hormone Closed-loop System With XeriSol™ Glucagon vs Closed-loop System With Insulin Only vs a Predictive Low Glucose Suspend System

Closed-loop systems are an emerging technology that automate hormone delivery. They are quickly paving the way to revolutionize the treatment of type 1 diabetes. Several categories have emerged: dual-hormone (insulin and glucagon) closed-loop systems and closed-loop systems with insulin only, one variety of which is the low glucose suspend safety feature now available from Medtronic (MiniMed640G with SmartGuard). The study described within this protocol is designed to test the efficacy of a new closed-loop algorithm for managing blood glucose in people with type 1 diabetes before and after exercise. The new algorithm will have 3 modes: a single hormone insulin only mode, a dual-hormone insulin and glucagon mode and an insulin only mode with predictive low glucose suspend, all with an exercise detection algorithm. The purpose of this study is to determine whether a dual hormone AP using a stable investigational glucagon with an exercise detection algorithm outperforms both single hormone AP and a low glucose suspend algorithm.

Study Overview

Detailed Description

Subjects will undergo three approximately 76 hour studies. During each of these intervention visits, subjects will wear an Omnipod to deliver insulin and a Dexcom G5 CGM to measure glucose. In the dual-hormone closed-loop study, participants will wear a second Omnipod filled with XeriSol™ glucagon. The first day of intervention visit will be a 12 hour inpatient visit with the subsequent time spent as an outpatient. The subject will come back to the research center on Day 4 in the morning to remove all devices. The blood glucose control system is called the Artificial Pancreas Controller (APC). During one of the studies, glucose will be controlled using the dual hormone mode. During another study, glucose will be controlled using the single hormone mode. The single and dual hormone modes of the APC controller determine insulin only and insulin with glucagon delivery rates based on proportional and derivative error and contain an exercise detection component. During the other study, glucose will be controlled using a predictive low glucose suspend (PLGS) mode. In the PLGS mode, the APC will be programmed with the subject's basal profile(s) which will be transmitted to the Omnipod Personal Diabetes Manager (PDM) and allow for boluses to be inputted by the subjects for meals and corrections using their typical carb ratios and correction factors. However, this system will have a safety feature to suspend insulin delivery when sensor glucose is predicted to go below a threshold. Treatment order will be randomized.

Subjects will arrive at the clinic at approximately 7 or 8am for the inpatient visits. Subjects will eat breakfast and lunch at approximately 8:30am, noon and 6pm respectively. The dinner meal will be consumed while the subject is at home. Subjects will exercise on a treadmill for 45 minutes. Subjects will be discharged 9 hours after admission. The subject will then go home for the remainder of the study visit. Studies will be remotely monitored. Subjects will return to OHSU on Day 4 for removal of all devices.

During each study, the subject will wear one subcutaneous DexcomTM G5 continuous glucose monitoring (CGM) system. The CGM system will provide sensed glucose data every 5 minutes. The accuracy of the sensed data will be obtained by reference measurements of capillary blood glucose. Sensed glucose data will be wirelessly transmitted via Bluetooth Low Energy (BTLE) from the Dexcom G5 transmitter to the Nexus 5 master controller every five minutes. The controller for all 3 modes is a Google Nexus 5 phone. The smart phone will wirelessly communicate via BTLE to an Omnipod through a PDM (Insulet Corp. ) for automated insulin delivery or in the case of dual-hormone AP, one used for automated insulin delivery and one for automated glucagon delivery.

A physician or nurse practitioner will be present for study start-up, will be on campus for exercise on Day 1 and will be immediately available on call at all other times.

Study Type

Interventional

Enrollment (Actual)

23

Phase

  • Phase 1

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • Oregon
      • Portland, Oregon, United States, 97239
        • Oregon Health and Science University

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

21 years to 50 years (Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Description

Inclusion Criteria:

  • Diagnosis of type 1 diabetes mellitus for at least 1 year.
  • Male or female subjects 21-50 years of age.
  • Physically willing and able to perform 45 min of exercise.
  • Current use of an insulin pump.
  • Lives with another person age 18 or older who will be present while subject exercises at home and that can attend the training on using the system.
  • Lives within 40 miles of OHSU main campus.
  • HbA1c <= 10% at screening.
  • Total daily insulin requirement is less than 139 units/day.
  • Current use of a phone or other device so can be contacted by study staff off-campus.
  • Willingness to follow all study procedures, including attending all clinic visits.
  • Willingness to sign informed consent and HIPAA documents.

Exclusion Criteria:

  • Female of childbearing potential who is pregnant or intending to become pregnant or breast-feeding, or is not using adequate contraceptive methods. Acceptable contraception includes birth control pill/patch/vaginal ring, Depo-Provera, Norplant, an IUD, the double barrier method (the woman uses a diaphragm and spermicide and the man uses a condom), or abstinence.
  • Any cardiovascular disease, defined as a clinically significant EKG abnormality at the time of screening or any history of: stroke, heart failure, myocardial infarction, angina pectoris, or coronary arterial bypass graft or angioplasty. Diagnosis of 2nd and 3rd degree heart block or any non-physiological arrhythmia judged by the investigator to be exclusionary.
  • Renal insufficiency (GFR < 60 ml/min, using the MDRD equation as report by the OHSU laboratory).
  • Liver failure, cirrhosis, or any other liver disease that compromises liver function as determined by the investigator.
  • Hematocrit of less than 36% for men, less than 32% for women.
  • Hypertensive subjects with systolic blood pressure >= 160 mm Hg or diastolic blood pressure >= 100 mm Hg despite treatment or who have treatment-refractory hypertension (e.g. requiring four or more medications).
  • History of severe hypoglycemia during the past 12 months prior to screening visit or hypoglycemia unawareness as judged by the investigator. Subjects will complete a hypoglycemia awareness questionnaire. Subjects will be excluded for four or more R responses.
  • Adrenal insufficiency.
  • Any active infection.
  • Known or suspected abuse of alcohol, narcotics, or illicit drugs.
  • Seizure disorder.
  • Active foot ulceration.
  • Severe peripheral arterial disease characterized by ischemic rest pain or severe claudication.
  • Major surgical operation within 30 days prior to screening.
  • Use of an investigational drug within 30 days prior to screening.
  • Chronic usage of any immunosuppressive medication (such as cyclosporine, azathioprine, sirolimus, or tacrolimus).
  • Bleeding disorder, treatment with warfarin, or platelet count below 50,000.
  • Allergy to aspart insulin.
  • Allergy to glucagon.
  • Need for uninterrupted treatment of acetaminophen.
  • Current administration of oral or parenteral corticosteroids.
  • Any life threatening disease, including malignant neoplasms and medical history or malignant neoplasms within the past 5 years prior to screening (except basal and squamous cell skin cancer).
  • Beta blockers or non-dihydropyridine calcium channel blockers.
  • Current use of any medication intended to lower glucose other than insulin (e.g. use of liraglutide).
  • Diagnosis of pheochromocytoma, insulinoma, or glucagonoma, personal or family history of multiple endocrine neoplasmia (MEN) 2A, MEN 2B, neurofibromatosis or von Hippel-Lindau disease.
  • History of severe hypersensitivity to milk protein.
  • Current use of any medication with strong anticholinergic properties, such as antihistamines, sleep aids, and antidiarrheal medications.
  • Current use of indomethacin.
  • Conditions that may results in low levels of releasable glucose in the liver and an inadequate reversal of hypoglycemia by glucagon such as prolonged fasting, starvation or chronic hypoglycemia as determined by the investigator.
  • A positive response to any of the questions from the Physical Activity Readiness Questionnaire with one exception: subject will not be excluded if he/she takes a single blood pressure medication that doesn't impact heart rate and blood pressure is controlled on the medication (blood pressure is less than 140/90 mmHg).
  • Any chest discomfort with physical activity, including pain or pressure, or other types of discomfort.
  • Any clinically significant disease or disorder which in the opinion of the investigator may jeopardize the subject's safety or compliance with the protocol.

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Treatment
  • Allocation: Randomized
  • Interventional Model: Crossover Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: Dual Hormone Closed-loop Arm
Subjects will undergo a 76 hour study with 9 hours inpatient and 67 hours outpatient using the closed-loop artificial pancreas controller in dual hormone mode and XeriSol glucagon to manage blood sugar that includes an exercise detection component that includes a reduction in insulin delivery and an increase in glucagon delivery upon detection.
The artificial pancreas controller contains an algorithm for managing blood glucose in people with type 1 diabetes which includes an exercise detection component. The algorithm has 3 modes: a single hormone closed-loop insulin only mode, a dual-hormone closed-loop insulin and glucagon mode and an insulin only mode with predictive low glucose suspend. The Closed-loop Artificial Pancreas Controller includes insulin and glucagon delivery algorithm implemented on a smart phone, utilizing sensor glucose values from a Dexcom G5 sensor and sending delivery commands to two Omnipod insulin pumps, one filled with insulin and one with glucagon.
XeriSol glucagon is an investigational drug that is a stable, soluble liquid glucagon formulation that can be injected or administered through an insulin pump. XeriSol glucagon will be used to fill the Omnipod to deliver glucagon for the Artificial Pancreas Controller in Dual Hormone mode.
Experimental: Single Hormone Closed-loop Arm
Subjects will undergo a 76 hour study with 9 hours inpatient and 67 hours outpatient using the closed-loop artificial pancreas controller in single hormone mode to manage blood sugar that includes an exercise detection component that includes a reduction in insulin delivery upon detection.
The artificial pancreas controller contains an algorithm for managing blood glucose in people with type 1 diabetes which includes an exercise detection component. The new algorithm will have 3 modes: a single hormone closed-loop insulin only mode, a dual-hormone closed-loop insulin and glucagon mode and an insulin only mode with predictive low glucose suspend. The Closed-loop Artificial Pancreas Controller includes insulin delivery algorithm implemented on a smart phone, utilizing sensor glucose values from a Dexcom G5 sensor and sending delivery commands to one Omnipod pump filled with insulin.
Experimental: Predictive Low Glucose Suspend Arm
Subjects will undergo a 76 hour study with 9 hours inpatient and 67 hours outpatient using the closed-loop artificial pancreas controller in predictive low glucose suspend mode. The system will run through the closed-loop system but will utilize the patient's optimized basal rates, correction factors and carb ratios as they would normally run on their own insulin pump. But the mode will have the additional safety net of the pump suspending insulin when it predicts a hypoglycemic event.
The artificial pancreas controller contains an algorithm for managing blood glucose in people with type 1 diabetes which includes an exercise detection component. The new algorithm will have 3 modes: a single hormone closed-loop insulin only mode, a dual-hormone closed-loop insulin and glucagon mode and an insulin only mode with predictive low glucose suspend. The Closed-loop Artificial Pancreas Controller includes insulin delivery algorithm implemented on a smart phone, utilizing sensor glucose values from a Dexcom G5 sensor and sending delivery commands to one Omnipod insulin pump filled with insulin.

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Percent of Time With Sensed Glucose < 70 mg/dl
Time Frame: 4 hours post exercise on Day 1
Assess the percent of time that the Dexcom G5 or G6 reported sensor glucose values less than 70 mg/dl using Dexcom sensor across all three arms.
4 hours post exercise on Day 1

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Percent of Time With Sensed Glucose Between 70-180 mg/dl
Time Frame: entire 76 hour study
Assess the percent of time that the Dexcom G5 or G6 reported sensor glucose values between 70-180 mg/dl using Dexcom sensor across all three arms.
entire 76 hour study
Number of Carbohydrate Treatments
Time Frame: 24 hours
Assess the number of rescue carbohydrate treatments across all three arms per day.
24 hours
Percent of Time With Sensed Glucose < 54 mg/dl
Time Frame: entire 76 hour study
Assess the percent of time that the Dexcom G5 or G6 reported sensor glucose values less than 54 mg/dl using Dexcom sensor across all three arms.
entire 76 hour study
Percent of Time With Sensed Glucose > 180 mg/dl
Time Frame: entire 76 hour study
Assess the percent of time that the Dexcom G5 or G6 reported sensor glucose values greater than 180 mg/dl using Dexcom sensor across all three arms.
entire 76 hour study
Mean Amount of Insulin Delivered in One Day
Time Frame: 24 hours
Assess the mean amount of insulin delivered in units/kg as documented through the artificial pancreas controller across all three arms.
24 hours
Mean Amount of Glucagon Delivered in One Day
Time Frame: 24 hours
Assess the mean amount of XeriSol glucagon delivered in mcg/kg as documented through the artificial pancreas controller in dual hormone arm.
24 hours

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Publications and helpful links

The person responsible for entering information about the study voluntarily provides these publications. These may be about anything related to the study.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

March 29, 2018

Primary Completion (Actual)

May 17, 2019

Study Completion (Actual)

May 17, 2019

Study Registration Dates

First Submitted

January 31, 2018

First Submitted That Met QC Criteria

January 31, 2018

First Posted (Actual)

February 6, 2018

Study Record Updates

Last Update Posted (Actual)

May 27, 2020

Last Update Submitted That Met QC Criteria

May 13, 2020

Last Verified

May 1, 2020

More Information

Terms related to this study

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

Yes

Studies a U.S. FDA-regulated device product

Yes

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Type1diabetes

Clinical Trials on Artificial Pancreas Controller in Dual Hormone Mode

3
Subscribe