- ICH GCP
- US Clinical Trials Registry
- Clinical Trial NCT07739810
Miniaturized Vital Sign Monitoring for Newborns in the Delivery Room
July 28, 2026 updated by: Guilherme Sant'Anna, MD, McGill University Health Centre/Research Institute of the McGill University Health Centre
This prospective observational study aims to evaluate the accuracy, safety, and feasibility of the Wireless MicroPremVitals Button Sensor System for continuous monitoring of electrocardiography (ECG), heart rate, respiratory rate, oxygen saturation, and skin temperature in newborn infants in the delivery room.
Measurements obtained from the wireless sensor system will be compared with standard-of-care monitoring.
The study will also assess skin tolerability, signal quality, and parent and healthcare provider perceptions of the device.
Study Overview
Status
Not yet recruiting
Intervention / Treatment
Study Type
Observational
Enrollment (Estimated)
5
Contacts and Locations
This section provides the contact details for those conducting the study, and information on where this study is being conducted.
Study Contact
- Name: Alyssa Maximov, BSc.
- Phone Number: 78229 5149341934
- Email: alyssa.maximov@muhc.mcgill.ca
Participation Criteria
Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.
Eligibility Criteria
Ages Eligible for Study
- Child
Accepts Healthy Volunteers
Yes
Sampling Method
Non-Probability Sample
Study Population
Healthy newborn infants in the delivery room at the study site who are eligible for routine delivery room monitoring and whose parent(s) or legal guardian provide informed consent for study participation.
Description
Inclusion Criteria:
- Infants born by C-Section or vaginal delivery, with ≥35 weeks of gestational age and judged clinically stable immediately after birth.
Exclusion Criteria:
- Infants with skin abnormality would potentially increase the risk of injury.
- Family experiencing too much stress as determined by an attending physician or bedside care team, to not be approached for recruitment.
- Infants born at <35 weeks gestational age.
Study Plan
This section provides details of the study plan, including how the study is designed and what the study is measuring.
How is the study designed?
Design Details
Cohorts and Interventions
Group / Cohort |
Intervention / Treatment |
|---|---|
|
Newborn Infants
Healthy newborn infants enrolled in the study who will undergo simultaneous monitoring with the Wireless MicroPremVitals Button Sensor System and standard-of-care monitoring in the delivery room to evaluate the accuracy, safety, and feasibility of the wireless monitoring system.
|
A wireless wearable sensor system applied to newborn infants in the delivery room to continuously measure electrocardiography (ECG), heart rate, respiratory rate, oxygen saturation, and skin temperature.
Measurements will be compared with standard-of-care monitoring for validation purposes only and will not be used for clinical decision-making.
|
What is the study measuring?
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Electrocardiography (ECG) Accuracy
Time Frame: During two 15-minute monitoring periods within the first 90 minutes after birth
|
Agreement between ECG measurements obtained using the Wireless MicroPremVitals Button Sensor System and the standard-of-care Dräger Infinity M540 monitor.
|
During two 15-minute monitoring periods within the first 90 minutes after birth
|
|
Skin Temperature Accuracy
Time Frame: During two 15-minute monitoring periods within the first 90 minutes after birth.
|
Agreement between skin temperature measurements obtained using the Wireless MicroPremVitals Button Sensor System and the standard-of-care Dräger Infinity M540 monitor.
|
During two 15-minute monitoring periods within the first 90 minutes after birth.
|
|
Oxygen Saturation (SpO₂) Accuracy
Time Frame: During two 15-minute monitoring periods within the first 90 minutes after birth.
|
Agreement between oxygen saturation (SpO₂) measurements obtained using the Wireless MicroPremVitals Button Sensor System and the standard-of-care Dräger Infinity M540 monitor.
|
During two 15-minute monitoring periods within the first 90 minutes after birth.
|
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Heart Rate Accuracy
Time Frame: During two 15-minute monitoring periods within the first 90 minutes after birth.
|
Heart rate will be calculated offline from raw optical and motion data collected by the Wireless MicroPremVitals Button Sensor System.
Agreement between calculated heart rate and the standard-of-care Dräger Infinity M540 monitor will be evaluated.
|
During two 15-minute monitoring periods within the first 90 minutes after birth.
|
|
Heart Rate Variability (HRV)
Time Frame: During two 15-minute monitoring periods within the first 90 minutes after birth.
|
Heart rate variability will be calculated offline from raw optical and motion data collected by the Wireless MicroPremVitals Button Sensor System.
Derived HRV measurements will be compared with HRV derived from the standard-of-care Dräger Infinity M540 monitor.
|
During two 15-minute monitoring periods within the first 90 minutes after birth.
|
Collaborators and Investigators
This is where you will find people and organizations involved with this study.
Investigators
- Principal Investigator: Guilherme Sant'Anna, MD, PhD, RI-MUHC
Study record dates
These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.
Study Major Dates
Study Start (Estimated)
September 1, 2026
Primary Completion (Estimated)
September 1, 2027
Study Completion (Estimated)
September 1, 2027
Study Registration Dates
First Submitted
July 28, 2026
First Submitted That Met QC Criteria
July 28, 2026
First Posted (Actual)
July 31, 2026
Study Record Updates
Last Update Posted (Actual)
July 31, 2026
Last Update Submitted That Met QC Criteria
July 28, 2026
Last Verified
July 1, 2026
More Information
Terms related to this study
Keywords
Other Study ID Numbers
- Button Sensors in the DR
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
NO
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
No
Studies a U.S. FDA-regulated device product
No
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.