A Photoplethysmography-Based Machine Learning Algorithm for Early Atrial Fibrillation Detection: A Prospective Validation Study (HeartCore AF)
Prospective Validation of a Machine-Learning Algorithm Using Photoplethysmography Signals for Early Detection of Atrial Fibrillation During Remote Telemonitoring
Studieoversigt
Status
Betingelser
Intervention / Behandling
Detaljeret beskrivelse
Atrial fibrillation (AF) and heart failure (HF) frequently coexist and share a bidirectional causal relationship; their concurrence is associated with worse clinical outcomes. Early detection of AF may enable timely intervention and improve outcomes. This study is prospectively validating a machine-learning algorithm for AF detection from PPG signals, intended for integration into the Seerlinq remote monitoring platform. This algorithm builds on the same core PPG signal-processing technology as Seerlinq's HeartCore device (a CE-certified, Class IIb device under the EU MDR that monitors left ventricular filling pressures in heart failure patients). It is a stand-alone algorithm designed specifically to detect clinically relevant (≥ 30s) atrial fibrillation.
Validation of the algorithm will proceed in three stages: (1) internal cross-validation; (2) external validation against an independent cohort with paired PPG-ECG recordings, to confirm generalizability; and (3) validation in a cohort of patients with paroxysmal atrial fibrillation and frequent sinus-AF transitions, to assess performance during clinically challenging rhythm changes.
The study is enrolling toward an estimated 1,000 unique PPG recordings. A 12-lead ECG is used to confirm cardiac rhythm classification (gold standard) as the reference for evaluating algorithm performance.
Undersøgelsestype
Tilmelding (Anslået)
Kontakter og lokationer
Studiekontakt
- Navn: Marta Kollárová, MSc., PhD.
- Telefonnummer: +421 950 896 026
- E-mail: marta.kollarova@premedix.org
Studiesteder
-
-
-
Bratislava, Slovakiet
- Rekruttering
- PreMedix
-
Kontakt:
- Allan Bohm, M.D., MSc. PhD.
- Telefonnummer: +421 907 411 499
- E-mail: allan.bohm@premedix.org
-
Ledende efterforsker:
- Allan Bohm, M.D., MSc., PhD.
-
-
Deltagelseskriterier
Berettigelseskriterier
Aldre berettiget til at studere
- Voksen
- Ældre voksen
Tager imod sunde frivillige
Prøveudtagningsmetode
Studiebefolkning
Beskrivelse
Inclusion Criteria:
- Adults ≥18 years with a diagnosis of heart failure (HFrEF, HFmrEF, or HFpEF)
- 12-lead ECG performed to confirm cardiac rhythm classification (AF vs. non-AF)
Exclusion Criteria:
- Missing a valid PPG recording
Studieplan
Hvordan er undersøgelsen tilrettelagt?
Design detaljer
Kohorter og interventioner
Gruppe / kohorte |
Intervention / Behandling |
|---|---|
|
Documented AF
HF patients with a history of permanent/paroxysmal AF and AF documented on 12-lead ECG at enrollment
|
The PPG-based atrial fibrillation detection algorithm is a non-invasive signal processing approach that analyzes photoplethysmographic waveforms obtained during remote monitoring.
The algorithm evaluates pulse-to-pulse variability, waveform characteristics, and signal quality parameters to identify irregular rhythm patterns associated with atrial fibrillation and provide early detection of potential arrhythmic events.
|
|
Non-AF
HF patients in sinus rhythm on the index 12-lead ECG with no prior documented AF episodes
|
The PPG-based atrial fibrillation detection algorithm is a non-invasive signal processing approach that analyzes photoplethysmographic waveforms obtained during remote monitoring.
The algorithm evaluates pulse-to-pulse variability, waveform characteristics, and signal quality parameters to identify irregular rhythm patterns associated with atrial fibrillation and provide early detection of potential arrhythmic events.
|
Hvad måler undersøgelsen?
Primære resultatmål
Resultatmål |
Tidsramme |
|---|---|
|
Diagnostic accuracy (area under the ROC curve) of the PPG-based machine-learning algorithm for detecting clinically relevant AF (≥ 30s), compared with gold-standard 12-lead ECG
Tidsramme: Through study completion (estimated November 2026)
|
Through study completion (estimated November 2026)
|
Sekundære resultatmål
Resultatmål |
Foranstaltningsbeskrivelse |
Tidsramme |
|---|---|---|
|
Sensitivity and specificity of the algorithm at the Youden-optimal threshold
Tidsramme: Through study completion (estimated November 2026)
|
Through study completion (estimated November 2026)
|
|
|
Positive predictive value and negative predictive value
Tidsramme: Through study completion (estimated November 2026)
|
Through study completion (estimated November 2026)
|
|
|
Average precision
Tidsramme: Through study completion (estimated November 2026)
|
area under the precision-recall curve
|
Through study completion (estimated November 2026)
|
|
Model calibration
Tidsramme: Through study completion (estimated November 2026)
|
e.g., calibration curve / Brier score
|
Through study completion (estimated November 2026)
|
|
Matthews correlation coefficient
Tidsramme: Through study completion (estimated November 2026)
|
Through study completion (estimated November 2026)
|
|
|
Overall classification accuracy
Tidsramme: Through study completion (estimated November 2026)
|
Through study completion (estimated November 2026)
|
|
|
Specificity and false-positive rate in the subgroup with frequent atrial/ventricular extrasystoles
Tidsramme: Through study completion (estimated November 2026)
|
Through study completion (estimated November 2026)
|
|
|
Accuracy of AF detection during sinus-AF transitions at the individual patient level
Tidsramme: Through study completion (estimated November 2026)
|
Through study completion (estimated November 2026)
|
Samarbejdspartnere og efterforskere
Sponsor
Datoer for undersøgelser
Studer store datoer
Studiestart (Faktiske)
Primær færdiggørelse (Anslået)
Studieafslutning (Anslået)
Datoer for studieregistrering
Først indsendt
Først indsendt, der opfyldte QC-kriterier
Først opslået (Faktiske)
Opdateringer af undersøgelsesjournaler
Sidste opdatering sendt (Faktiske)
Sidste opdatering indsendt, der opfyldte kvalitetskontrolkriterier
Sidst verificeret
Mere information
Begreber relateret til denne undersøgelse
Nøgleord
Yderligere relevante MeSH-vilkår
Andre undersøgelses-id-numre
- HeartCoreAF01
Plan for individuelle deltagerdata (IPD)
Planlægger du at dele individuelle deltagerdata (IPD)?
IPD-planbeskrivelse
Lægemiddel- og udstyrsoplysninger, undersøgelsesdokumenter
Studerer et amerikansk FDA-reguleret lægemiddelprodukt
Studerer et amerikansk FDA-reguleret enhedsprodukt
Disse oplysninger blev hentet direkte fra webstedet clinicaltrials.gov uden ændringer. Hvis du har nogen anmodninger om at ændre, fjerne eller opdatere dine undersøgelsesoplysninger, bedes du kontakte register@clinicaltrials.gov. Så snart en ændring er implementeret på clinicaltrials.gov, vil denne også blive opdateret automatisk på vores hjemmeside .