A Photoplethysmography-Based Machine Learning Algorithm for Early Atrial Fibrillation Detection: A Prospective Validation Study (HeartCore AF)
Prospective Validation of a Machine-Learning Algorithm Using Photoplethysmography Signals for Early Detection of Atrial Fibrillation During Remote Telemonitoring
Studie Overzicht
Toestand
Conditie
Interventie / Behandeling
Gedetailleerde beschrijving
Atrial fibrillation (AF) and heart failure (HF) frequently coexist and share a bidirectional causal relationship; their concurrence is associated with worse clinical outcomes. Early detection of AF may enable timely intervention and improve outcomes. This study is prospectively validating a machine-learning algorithm for AF detection from PPG signals, intended for integration into the Seerlinq remote monitoring platform. This algorithm builds on the same core PPG signal-processing technology as Seerlinq's HeartCore device (a CE-certified, Class IIb device under the EU MDR that monitors left ventricular filling pressures in heart failure patients). It is a stand-alone algorithm designed specifically to detect clinically relevant (≥ 30s) atrial fibrillation.
Validation of the algorithm will proceed in three stages: (1) internal cross-validation; (2) external validation against an independent cohort with paired PPG-ECG recordings, to confirm generalizability; and (3) validation in a cohort of patients with paroxysmal atrial fibrillation and frequent sinus-AF transitions, to assess performance during clinically challenging rhythm changes.
The study is enrolling toward an estimated 1,000 unique PPG recordings. A 12-lead ECG is used to confirm cardiac rhythm classification (gold standard) as the reference for evaluating algorithm performance.
Studietype
Inschrijving (Geschat)
Contacten en locaties
Studiecontact
- Naam: Marta Kollárová, MSc., PhD.
- Telefoonnummer: +421 950 896 026
- E-mail: marta.kollarova@premedix.org
Studie Locaties
-
-
-
Bratislava, Slowakije
- Werving
- PreMedix
-
Contact:
- Allan Bohm, M.D., MSc. PhD.
- Telefoonnummer: +421 907 411 499
- E-mail: allan.bohm@premedix.org
-
Hoofdonderzoeker:
- Allan Bohm, M.D., MSc., PhD.
-
-
Deelname Criteria
Geschiktheidscriteria
Leeftijden die in aanmerking komen voor studie
- Volwassen
- Oudere volwassene
Accepteert gezonde vrijwilligers
Bemonsteringsmethode
Studie Bevolking
Beschrijving
Inclusion Criteria:
- Adults ≥18 years with a diagnosis of heart failure (HFrEF, HFmrEF, or HFpEF)
- 12-lead ECG performed to confirm cardiac rhythm classification (AF vs. non-AF)
Exclusion Criteria:
- Missing a valid PPG recording
Studie plan
Hoe is de studie opgezet?
Ontwerpdetails
Cohorten en interventies
Groep / Cohort |
Interventie / Behandeling |
|---|---|
|
Documented AF
HF patients with a history of permanent/paroxysmal AF and AF documented on 12-lead ECG at enrollment
|
The PPG-based atrial fibrillation detection algorithm is a non-invasive signal processing approach that analyzes photoplethysmographic waveforms obtained during remote monitoring.
The algorithm evaluates pulse-to-pulse variability, waveform characteristics, and signal quality parameters to identify irregular rhythm patterns associated with atrial fibrillation and provide early detection of potential arrhythmic events.
|
|
Non-AF
HF patients in sinus rhythm on the index 12-lead ECG with no prior documented AF episodes
|
The PPG-based atrial fibrillation detection algorithm is a non-invasive signal processing approach that analyzes photoplethysmographic waveforms obtained during remote monitoring.
The algorithm evaluates pulse-to-pulse variability, waveform characteristics, and signal quality parameters to identify irregular rhythm patterns associated with atrial fibrillation and provide early detection of potential arrhythmic events.
|
Wat meet het onderzoek?
Primaire uitkomstmaten
Uitkomstmaat |
Tijdsspanne |
|---|---|
|
Diagnostic accuracy (area under the ROC curve) of the PPG-based machine-learning algorithm for detecting clinically relevant AF (≥ 30s), compared with gold-standard 12-lead ECG
Tijdsspanne: Through study completion (estimated November 2026)
|
Through study completion (estimated November 2026)
|
Secundaire uitkomstmaten
Uitkomstmaat |
Maatregel Beschrijving |
Tijdsspanne |
|---|---|---|
|
Sensitivity and specificity of the algorithm at the Youden-optimal threshold
Tijdsspanne: Through study completion (estimated November 2026)
|
Through study completion (estimated November 2026)
|
|
|
Positive predictive value and negative predictive value
Tijdsspanne: Through study completion (estimated November 2026)
|
Through study completion (estimated November 2026)
|
|
|
Average precision
Tijdsspanne: Through study completion (estimated November 2026)
|
area under the precision-recall curve
|
Through study completion (estimated November 2026)
|
|
Model calibration
Tijdsspanne: Through study completion (estimated November 2026)
|
e.g., calibration curve / Brier score
|
Through study completion (estimated November 2026)
|
|
Matthews correlation coefficient
Tijdsspanne: Through study completion (estimated November 2026)
|
Through study completion (estimated November 2026)
|
|
|
Overall classification accuracy
Tijdsspanne: Through study completion (estimated November 2026)
|
Through study completion (estimated November 2026)
|
|
|
Specificity and false-positive rate in the subgroup with frequent atrial/ventricular extrasystoles
Tijdsspanne: Through study completion (estimated November 2026)
|
Through study completion (estimated November 2026)
|
|
|
Accuracy of AF detection during sinus-AF transitions at the individual patient level
Tijdsspanne: Through study completion (estimated November 2026)
|
Through study completion (estimated November 2026)
|
Medewerkers en onderzoekers
Sponsor
Studie record data
Bestudeer belangrijke data
Studie start (Werkelijk)
Primaire voltooiing (Geschat)
Studie voltooiing (Geschat)
Studieregistratiedata
Eerst ingediend
Eerst ingediend dat voldeed aan de QC-criteria
Eerst geplaatst (Werkelijk)
Updates van studierecords
Laatste update geplaatst (Werkelijk)
Laatste update ingediend die voldeed aan QC-criteria
Laatst geverifieerd
Meer informatie
Termen gerelateerd aan deze studie
Trefwoorden
Aanvullende relevante MeSH-voorwaarden
Andere studie-ID-nummers
- HeartCoreAF01
Plan Individuele Deelnemersgegevens (IPD)
Bent u van plan om gegevens van individuele deelnemers (IPD) te delen?
Beschrijving IPD-plan
Informatie over medicijnen en apparaten, studiedocumenten
Bestudeert een door de Amerikaanse FDA gereguleerd geneesmiddel
Bestudeert een door de Amerikaanse FDA gereguleerd apparaatproduct
Deze informatie is zonder wijzigingen rechtstreeks van de website clinicaltrials.gov gehaald. Als u verzoeken heeft om uw onderzoeksgegevens te wijzigen, te verwijderen of bij te werken, neem dan contact op met register@clinicaltrials.gov. Zodra er een wijziging wordt doorgevoerd op clinicaltrials.gov, wordt deze ook automatisch bijgewerkt op onze website .