Reliability of the Freestyle Libre CGM in the Inpatient Setting During the COVID-19 Surge

1 mars 2021 mis à jour par: Northwell Health
The FDA has provided an emergency waiver for the use of non-invasive continuous glucose monitors (CGM) in hospitals, so frontline health care providers (in hospitals) can remotely monitor patients with diabetes thus reducing patient's discomfort, limiting exposure to COVID-19 and preserving critical personal protective equipment (PPE). The FreeStyle Libre 14-day system is a continuous glucose monitoring system consisting of a handheld reader (smart phone may be used) and a sensor applied with adhesive to back of the upper arm. In order to evaluate the reliability of the Freestyle Libre CGM for in-patient use, we propose a study which will examine the correlation between the libre CGM data and capillary blood glucose test - the current standard of care taken by the Accuchek Inform II platform.

Aperçu de l'étude

Description détaillée

In hospitalized patients diagnosed with COVID-19, diabetes mellitus is associated with a higher mortality. Patients with controlled blood glucose have a lower mortality rate than those who are uncontrolled. However, a rigid blood glucose control - which may lead to hypoglycemia- is associated with higher mortality rates. Thus, adequate glucose monitoring is important for hospitalized diabetic patients.

Previous studies have shown that CGMS, in the outpatient setting, engender better glycemic control thereby reducing risk of hypoglycemia. CGMS offer a replacement for finger stick blood glucose monitoring which may not identify an incidence of hypoglycemia until symptoms drive the patient to test. Furthermore, finger stick blood glucose monitoring is painful and time consuming thus causing many patients to avoid this aspect of diabetes self-care management.

Freestyle Libre CGMs sample interstitial fluid glucose subcutaneously. Measurement frequencies typically range from 1 to 15 minutes and most commonly are every 5 minutes.

Previous studies in hospitalized patients living with diabetes mellitus on basal bolus regimen have shown:

• Use of the CGM increased detection of both hypoglycemic and hyperglycemic events versus those sensed by POCT.

One investigation identified 88 postprandial hyperglycemic excursions in patients with CGM use as opposed to 61 in those with POCT. Another noted that when comparing the CGM use with that of POCT, the former identified 55 hypoglycemic events whereas the latter only found 12.

• No difference in mean daily glucose CGM and POCT readings. In a study of 38 patients, the investigators noted no difference in mean daily glucose between the CGM and POCT.

Type d'étude

Interventionnel

Inscription (Réel)

52

Phase

  • N'est pas applicable

Contacts et emplacements

Cette section fournit les coordonnées de ceux qui mènent l'étude et des informations sur le lieu où cette étude est menée.

Lieux d'étude

    • New York
      • New York, New York, États-Unis, 10022
        • Lenox Hill Hospital

Critères de participation

Les chercheurs recherchent des personnes qui correspondent à une certaine description, appelée critères d'éligibilité. Certains exemples de ces critères sont l'état de santé général d'une personne ou des traitements antérieurs.

Critère d'éligibilité

Âges éligibles pour étudier

18 ans à 99 ans (Adulte, Adulte plus âgé)

Accepte les volontaires sains

Non

Sexes éligibles pour l'étude

Tout

La description

Inclusion Criteria:

  • Participants acquired from among all non-critical care in patients who are living with Diabetes Mellitus. Because the patients will have to be instructed on the use of the CGM, this study will be limited to English speaking/reading adults (18 years of age or older), who possess the cognitive and physical ability to participate.

Exclusion Criteria:

  • Minors, pregnant women, and critically ill.

Plan d'étude

Cette section fournit des détails sur le plan d'étude, y compris la façon dont l'étude est conçue et ce que l'étude mesure.

Comment l'étude est-elle conçue ?

Détails de conception

  • Objectif principal: Autre
  • Répartition: Non randomisé
  • Modèle interventionnel: Affectation croisée
  • Masquage: Aucun (étiquette ouverte)

Armes et Interventions

Groupe de participants / Bras
Intervention / Traitement
Autre: Freestyle Libre 14-day CGM
The FreeStyle Libre 14 day system is a continuous glucose monitoring system consisting of a handheld reader and a sensor worn on the back of the upper arm.
The FreeStyle Libre 14 day system is a continuous glucose monitoring system consisting of a handheld reader and a sensor worn on the back of the upper arm.
The Accu-Chek Inform II system is designed to deliver data with speed and efficiency, helping to provide optimal care. Wireless meter quickly transfers data. The Accu-Chek Inform II system is the first point-of-care blood glucose device to offer truly wireless technology at the meter level.
Autre: Accuchek Inform II meter
ACCU-CHEK INFORM II system quantitatively measures glucose in fresh venous, arterial, neonatal heel stick and capillary whole blood from the finger and is used as an aid in monitoring the effectiveness of glucose control
The FreeStyle Libre 14 day system is a continuous glucose monitoring system consisting of a handheld reader and a sensor worn on the back of the upper arm.
The Accu-Chek Inform II system is designed to deliver data with speed and efficiency, helping to provide optimal care. Wireless meter quickly transfers data. The Accu-Chek Inform II system is the first point-of-care blood glucose device to offer truly wireless technology at the meter level.

Que mesure l'étude ?

Principaux critères de jugement

Mesure des résultats
Description de la mesure
Délai
Comparison of values between two devices
Délai: Through end of patient hospitalization, an average of 14 days
Correlation between blood glucose results from the Freestlye Libre 14-day CGM and Accuchek Inform II platform using the Bland-Altman method to determine whether the two glucose methods agree
Through end of patient hospitalization, an average of 14 days

Collaborateurs et enquêteurs

C'est ici que vous trouverez les personnes et les organisations impliquées dans cette étude.

Parrainer

Les enquêteurs

  • Chercheur principal: Renee Murray-Bachmann, Ed.D, MSN, Northwell Health

Publications et liens utiles

La personne responsable de la saisie des informations sur l'étude fournit volontairement ces publications. Il peut s'agir de tout ce qui concerne l'étude.

Dates d'enregistrement des études

Ces dates suivent la progression des dossiers d'étude et des soumissions de résultats sommaires à ClinicalTrials.gov. Les dossiers d'étude et les résultats rapportés sont examinés par la Bibliothèque nationale de médecine (NLM) pour s'assurer qu'ils répondent à des normes de contrôle de qualité spécifiques avant d'être publiés sur le site Web public.

Dates principales de l'étude

Début de l'étude (Réel)

12 juin 2020

Achèvement primaire (Réel)

30 novembre 2020

Achèvement de l'étude (Réel)

30 novembre 2020

Dates d'inscription aux études

Première soumission

1 juin 2020

Première soumission répondant aux critères de contrôle qualité

3 juin 2020

Première publication (Réel)

4 juin 2020

Mises à jour des dossiers d'étude

Dernière mise à jour publiée (Réel)

2 mars 2021

Dernière mise à jour soumise répondant aux critères de contrôle qualité

1 mars 2021

Dernière vérification

1 mars 2021

Plus d'information

Termes liés à cette étude

Autres numéros d'identification d'étude

  • 20-0455

Plan pour les données individuelles des participants (IPD)

Prévoyez-vous de partager les données individuelles des participants (DPI) ?

NON

Description du régime IPD

The team has not decided whether to share the results with other researchers. The team will consider publishing results after study completion

Informations sur les médicaments et les dispositifs, documents d'étude

Étudie un produit pharmaceutique réglementé par la FDA américaine

Non

Étudie un produit d'appareil réglementé par la FDA américaine

Oui

produit fabriqué et exporté des États-Unis.

Oui

Ces informations ont été extraites directement du site Web clinicaltrials.gov sans aucune modification. Si vous avez des demandes de modification, de suppression ou de mise à jour des détails de votre étude, veuillez contacter register@clinicaltrials.gov. Dès qu'un changement est mis en œuvre sur clinicaltrials.gov, il sera également mis à jour automatiquement sur notre site Web .