Relevance of Artificial Intelligence-Assisted Echocardiography for Left Ventricular Ejection Fraction Assessment in Geriatric Patients (REFERENCE-AI)

27 juillet 2026 mis à jour par: Gérond'if

Heart failure (HF) is the leading cause of hospitalization among adults aged 80 years and older and represents a major diagnostic challenge in geriatric medicine due to frequently atypical clinical presentations and the presence of multiple comorbidities. Although transthoracic echocardiography (TTE) with measurement of left ventricular ejection fraction (LVEF) remains the gold standard for cardiac functional assessment, access to echocardiography is often limited in geriatric wards.

Recent advances in artificial intelligence (AI) have enabled the development of portable ultrasound devices and automated image analysis software capable of providing reliable and reproducible LVEF measurements. AI-assisted automated LVEF assessment (AutoEF-AI) may therefore represent a valuable alternative to conventional echocardiography for the cardiac evaluation of older patients with heart failure.

Aperçu de l'étude

Description détaillée

Prospective, single-center interventional study comparing two methods of left ventricular ejection fraction measurement in patients aged 75 years and older hospitalized for acute heart failure.Each hemodynamically stable participant will undergo two echocardiographic examinations performed within 24 hours:

  • Standard Echocardiography (Reference Method)
  • AI-assisted automated LVEF assessment (AutoEF-AI)

Type d'étude

Interventionnel

Inscription (Estimé)

130

Phase

  • N'est pas applicable

Contacts et emplacements

Cette section fournit les coordonnées de ceux qui mènent l'étude et des informations sur le lieu où cette étude est menée.

Coordonnées de l'étude

Sauvegarde des contacts de l'étude

Lieux d'étude

      • Paris, France, 75013
        • Recrutement
        • Broca Hospital
        • Contact:

Critères de participation

Les chercheurs recherchent des personnes qui correspondent à une certaine description, appelée critères d'éligibilité. Certains exemples de ces critères sont l'état de santé général d'une personne ou des traitements antérieurs.

Critère d'éligibilité

Âges éligibles pour étudier

  • Adulte plus âgé

Accepte les volontaires sains

Non

La description

Inclusion Criteria:

  • Age ≥75 years.
  • Hospitalization in a geriatric unit for acute heart failure according to the 2021 ESC diagnostic criteria.
  • Hemodynamic stability at the time of echocardiographic examination.
  • Ability to understand study information, provide written informed consent, and willingness to participate.
  • Affiliation with, or beneficiary of, a health insurance/social security scheme.

Exclusion Criteria:

  • Hemodynamic instability preventing echocardiographic assessment.
  • Contraindication to transthoracic echocardiography.
  • Patients under legal protection (guardianship, curatorship, or judicial protection measures) or unable to provide informed consent.
  • Refusal to participate in the study.

Plan d'étude

Cette section fournit des détails sur le plan d'étude, y compris la façon dont l'étude est conçue et ce que l'étude mesure.

Comment l'étude est-elle conçue ?

Détails de conception

  • Objectif principal: Diagnostique
  • Répartition: Non randomisé
  • Modèle interventionnel: Affectation croisée
  • Masquage: Seul

Armes et Interventions

Groupe de participants / Bras
Intervention / Traitement
Autre: Standard Echocardiography (Reference method)
The standard echocardiography will be performed by a cardiologist as part of routine clinical care.
The standard echocardiography will be performed by an expert cardiologist as part of routine clinical care and will serve as the gold standard
Expérimental: AutoEF-AI Assessment
AI-assisted automated LVEF assessment

Performed by a geriatrician who has completed a one-day practical training session and combines:

  • A handheld ultrasound device providing real-time image acquisition guidance for obtaining apical views.
  • Us2.ai software for automated LVEF analysis. The geriatrician is blinded to results of the gold standard echocardiography performed by the cardiologist.

Que mesure l'étude ?

Principaux critères de jugement

Mesure des résultats
Description de la mesure
Délai
Agreement between LVEF measured using AutoEF-AI and LVEF measured using standard echocardiography.
Délai: At baseline

Agreement will be assessed using:

  • Intraclass Correlation Coefficient (ICC)
  • Spearman correlation coefficient
  • Bland-Altman analysis
  • Weighted kappa coefficient for the classification of patients according to LVEF categories (≤40%, 41-49%, and ≥50%)
At baseline

Mesures de résultats secondaires

Mesure des résultats
Description de la mesure
Délai
Diagnostic performance of AutoEF-AI for the detection of LVEF <50%, including sensitivity, specificity, positive predictive value, negative predictive value, and overall accuracy.
Délai: At baseline
Calculation of the sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), and overall accuracy of AutoEF-AI for identifying left ventricular ejection fraction (LVEF) below 50%, compared with the standard reference method.
At baseline
Analysis of factors associated with agreement between the two methods
Délai: At baseline
Multiple linear regression model will be used to identify clinical and technical variables (age, sex, cardiovascular history, presence of a pacemaker, cardiac arrhythmias, image quality, comorbidities, etc.) associated with a significant discrepancy between LVEF measurements obtained using AutoEF-AI and standard echocardiography.
At baseline
Feasibility of AutoEF-AI use by a geriatrician after a short training program
Délai: At baseline
  • Success rate of valid image acquisition (i.e acquizition of analyzable apical views suitable for automated analysis)
  • Quality of the acquired images (image quality score and proportion of non-analyzable images)
  • Difficulties encountered during the use of the AutoEF-AI device
At baseline

Collaborateurs et enquêteurs

C'est ici que vous trouverez les personnes et les organisations impliquées dans cette étude.

Parrainer

Les enquêteurs

  • Chercheur principal: Olivier HANON, MD PhD, Geriatric Department, Broca hospital

Dates d'enregistrement des études

Ces dates suivent la progression des dossiers d'étude et des soumissions de résultats sommaires à ClinicalTrials.gov. Les dossiers d'étude et les résultats rapportés sont examinés par la Bibliothèque nationale de médecine (NLM) pour s'assurer qu'ils répondent à des normes de contrôle de qualité spécifiques avant d'être publiés sur le site Web public.

Dates principales de l'étude

Début de l'étude (Réel)

7 juillet 2026

Achèvement primaire (Estimé)

30 juin 2028

Achèvement de l'étude (Estimé)

30 juin 2028

Dates d'inscription aux études

Première soumission

1 juillet 2026

Première soumission répondant aux critères de contrôle qualité

8 juillet 2026

Première publication (Réel)

14 juillet 2026

Mises à jour des dossiers d'étude

Dernière mise à jour publiée (Réel)

28 juillet 2026

Dernière mise à jour soumise répondant aux critères de contrôle qualité

27 juillet 2026

Dernière vérification

1 juillet 2026

Plus d'information

Termes liés à cette étude

Informations sur les médicaments et les dispositifs, documents d'étude

Étudie un produit pharmaceutique réglementé par la FDA américaine

Non

Étudie un produit d'appareil réglementé par la FDA américaine

Non

Ces informations ont été extraites directement du site Web clinicaltrials.gov sans aucune modification. Si vous avez des demandes de modification, de suppression ou de mise à jour des détails de votre étude, veuillez contacter register@clinicaltrials.gov. Dès qu'un changement est mis en œuvre sur clinicaltrials.gov, il sera également mis à jour automatiquement sur notre site Web .

S'abonner