Wireless MicroPremVitals Button Sensor in the NICU
Miniaturized Vital Sign Monitoring for Preterm Infants in the Neonatal Intensive Care Unit
Aperçu de l'étude
Statut
Les conditions
Intervention / Traitement
Description détaillée
Building on prior collaborative work between the Smart Hospital Project and the Dr. John Rogers Laboratory (QSIB- Northwestern University, Chicago, USA) in skin-integrated, soft, flexible electronics for premature infants, these button-sized sensors are designed to provide continuous ECG, oxygen saturation (SpO₂), and skin temperature monitoring without the bulky wires and large adhesives that impact the most fragile infants. By miniaturizing the hardware and tailoring form factor, weight, and interface materials for fragile neonatal skin, the system aims to extend wireless, clinical-grade monitoring down to the smallest, sickest babies.
In the NICU, the possible use of wireless sensors has received positive feedback from parents and HCPs. Overall, the perception is that eliminating the wires and cables will facilitate sensors application and monitoring, improve patient comfort, and skin-to-skin contact between infants and parents.
This proposal aims to:
- Investigate the feasibility, safety, and accuracy of the Wireless MicroPremVitals Button Sensors monitoring system specifically developed for use in extremely preterm neonates.
- Assess parents and HCPs perspectives on standard and wireless vital sign monitoring systems in extremely preterm infants.
Type d'étude
Inscription (Estimé)
Phase
- N'est pas applicable
Contacts et emplacements
Coordonnées de l'étude
- Nom: Alyssa Maximov
- Numéro de téléphone: 78229 5149341934
- E-mail: alyssa.maximov@muhc.mcgill.ca
Lieux d'étude
-
-
Quebec
-
Montreal, Quebec, Canada, H4A 3H9
- Recrutement
- Montreal Children's Hospital
-
Contact:
- Alyssa Maximov
- Numéro de téléphone: 78229 5149341934
- E-mail: alyssa.maximov@muhc.mcgill.ca
-
Chercheur principal:
- Guilherme Sant'Anna, MD, PhD
-
-
Critères de participation
Critère d'éligibilité
Âges éligibles pour étudier
- Enfant
- Adulte
- Adulte plus âgé
Accepte les volontaires sains
La description
Inclusion Criteria:
- Infants admitted to the NICU with a gestational age of ≥ 22+0 weeks and ≤ 29+6 weeks at birth,
- AND >72h of age at time of study,
- AND clinically stable,
- AND with the Neonatal Nurse Practitioner (i.e., NNP) and Healthcare Team in agreement for the patient to participate in the study.
Exclusion Criteria:
- Infants with a skin abnormality that would potentially increase the risk of injury,
- patient too unstable or family experiencing too much stress as determined by an attending physician or bedside care team to not be approached for recruitment.
- Infants admitted to the NICU with a gestational age of < 22+0 weeks or > 29+6 weeks at birth,
- and/or ≤ 72h of age at time of study,
- and/or not meeting all requirements of clinical stability,
- and/or with the Neonatal Nurse Practitioner (i.e., NNP)/Healthcare Team in disagreement for the patient to participate in the study.
Plan d'étude
Comment l'étude est-elle conçue ?
Détails de conception
- Objectif principal: Faisabilité de l'appareil
- Répartition: Non randomisé
- Modèle interventionnel: Affectation parallèle
- Masquage: Aucun (étiquette ouverte)
Armes et Interventions
Groupe de participants / Bras |
Intervention / Traitement |
|---|---|
|
Expérimental: Extremely Preterm Infant
The Wireless MicroPremVitals Button Sensor System consists of three (3) sensors: a chest unit, a forehead unit, and a limb unit. In this study, the MicroPremVitals Button forehead sensor will be mounted on the cerebral area of neonates at the level of the forehead. The chest unit will be mounted on the chest, between the nipples, or slightly below the nipple-line as placement allows. The limb unit will be mounted on the outer surface of the hand, wrist, or ankle of any limbs, as placement with standard wired equipment allows. |
These button-sized sensors are designed to provide continuous ECG, oxygen saturation (SpO₂), and skin temperature monitoring without the bulky wires and large adhesives that impact the most fragile infants.
|
|
Expérimental: Micro Preterm Infant
The Wireless MicroPremVitals Button Sensor System consists of three (3) sensors: a chest unit, a forehead unit, and a limb unit. In this study, the MicroPremVitals Button forehead sensor will be mounted on the cerebral area of neonates at the level of the forehead. The chest unit will be mounted on the chest, between the nipples, or slightly below the nipple-line as placement allows. The limb unit will be mounted on the outer surface of the hand, wrist, or ankle of any limbs, as placement with standard wired equipment allows. |
These button-sized sensors are designed to provide continuous ECG, oxygen saturation (SpO₂), and skin temperature monitoring without the bulky wires and large adhesives that impact the most fragile infants.
|
Que mesure l'étude ?
Principaux critères de jugement
Mesure des résultats |
Description de la mesure |
Délai |
|---|---|---|
|
HR Accuracy
Délai: 6 months
|
Comparison of heart rate measured by the Wireless MicroPremVitals Button Sensor System and the Philips IntelliVue MX800.
|
6 months
|
|
Oxygen Saturation Accuracy
Délai: 6 months
|
Comparison of SpO₂ measured by the Wireless MicroPremVitals Button Sensor System and the Philips IntelliVue MX800.
|
6 months
|
|
Skin Temperature Accuracy
Délai: 6 months
|
Comparison of skin temperature measured by the Wireless MicroPremVitals Button Sensor System and the Philips IntelliVue MX800.
|
6 months
|
Mesures de résultats secondaires
Mesure des résultats |
Description de la mesure |
Délai |
|---|---|---|
|
Perspectives
Délai: 6 months
|
Clinician and parent/guardian perspectives on both the Wireless MicroPremVitals Button Sensor System and the wired vital sign monitoring system, will be assessed using a custom study questionnaire.
The questionnaire includes Likert-scale items, multiple-choice questions, and open-ended responses evaluating usability, perceived safety, accuracy, comfort, and acceptability.
Responses will be summarized descriptively by questionnaire item.
|
6 months
|
Collaborateurs et enquêteurs
Les enquêteurs
- Chercheur principal: Guilherme Sant'Anna, MD, PhD, The Research Institute of the McGill University Health Center
Dates d'enregistrement des études
Dates principales de l'étude
Début de l'étude (Réel)
Achèvement primaire (Estimé)
Achèvement de l'étude (Estimé)
Dates d'inscription aux études
Première soumission
Première soumission répondant aux critères de contrôle qualité
Première publication (Réel)
Mises à jour des dossiers d'étude
Dernière mise à jour publiée (Réel)
Dernière mise à jour soumise répondant aux critères de contrôle qualité
Dernière vérification
Plus d'information
Termes liés à cette étude
Termes MeSH pertinents supplémentaires
- Maladies urogénitales
- Processus pathologiques
- Maladies urogénitales féminines et complications de la grossesse
- Travail obstétrique, prématuré
- Complications du travail obstétrical
- Complications de grossesse
- Infections
- Maladies des voies respiratoires
- État septique
- Syndrome de réponse inflammatoire systémique
- Inflammation
- Troubles respiratoires
- Nourrisson, nouveau-né, maladies
- Signes et symptômes respiratoires
- Maladies et anomalies congénitales, héréditaires et néonatales
- Conditions pathologiques, signes et symptômes
- Signes et symptômes
- Septicémie néonatale
- Naissance prématurée
- Apnée
Autres numéros d'identification d'étude
- Wireless MicroPremVitals
Informations sur les médicaments et les dispositifs, documents d'étude
Étudie un produit pharmaceutique réglementé par la FDA américaine
Étudie un produit d'appareil réglementé par la FDA américaine
Ces informations ont été extraites directement du site Web clinicaltrials.gov sans aucune modification. Si vous avez des demandes de modification, de suppression ou de mise à jour des détails de votre étude, veuillez contacter register@clinicaltrials.gov. Dès qu'un changement est mis en œuvre sur clinicaltrials.gov, il sera également mis à jour automatiquement sur notre site Web .