Four-year clinical follow-up of the first-in-man randomized comparison of a novel sirolimus eluting stent with abluminal biodegradable polymer and ultra-thin strut cobalt-chromium alloy: the INSPIRON-I trial

Marcos Danillo Peixoto Oliveira, Expedito E Ribeiro, Carlos M Campos, Henrique B Ribeiro, Bruno L R Faillace, Augusto C Lopes, Rodrigo B Esper, George X Meirelles, Marco A Perin, Alexandre Abizaid, Pedro A Lemos, Marcos Danillo Peixoto Oliveira, Expedito E Ribeiro, Carlos M Campos, Henrique B Ribeiro, Bruno L R Faillace, Augusto C Lopes, Rodrigo B Esper, George X Meirelles, Marco A Perin, Alexandre Abizaid, Pedro A Lemos

Abstract

Background: The Inspiron™ sirolimus-eluting stent (SES) is a low-dose, ultra-thin-strut cobalt-chromium stent abluminally coated with biodegradable polymers (BP). Previous results from the INSPIRON-I trial, a first-in-man study, have proven the efficacy of the novel stent in reducing neointimal proliferation. The present report aims at evaluating the long-term clinical outcomes of patients enrolled into the INSPIRON-I trial (Clinical Trials Gov. identifier: NCT01093391).

Methods: A total of 57 patients (60 lesions) were randomly allocated in a 2:1 ratio to treatment with the Inspiron™ SES vs. its equivalent Cronus™ bare metal stent (BMS, both by Scitech Medical™, Aparecida de Goiânia, Goiás, Brazil), in four tertiary centers. The primary endpoint of the present analysis was the occurrence of major adverse cardiac events (MACE) [death, myocardial infarction (MI), target vessel revascularization (TVR) and/or target lesion revascularization (TLR)] at 4 years.

Results: Baseline clinical and angiographic characteristics of both groups were similar. After 4 years, the primary endpoint occurred in 7.9% vs. 23.5% of patients in Inspiron and control groups respectively (P=0.11). The rate of death/MI was similar between the groups, but there was a significant decrease in the risk of repeat revascularization in the Inspiron group compared to the control arm TLR (0.0% vs. 23.5% respectively, P=0.02). There were no stent thromboses in the study population.

Conclusions: The novel Inspiron™ SES showed a sustained safe and effective clinical profile after 4-year of follow-up, with very low adverse events and null stent thrombosis (ST) occurrence.

Keywords: Biodegradable polymer (BP); long-term outcomes; percutaneous coronary intervention (PCI); sirolimus-eluting stent (SES).

Conflict of interest statement

Conflicts of Interest: The authors have no conflicts of interest to declare.

Figures

Figure 1
Figure 1
Design of the Inspiron™ SES (Scitech Medical™, Aparecida de Goiânia, Goiás, Brazil)—see text for detailed information.
Figure 2
Figure 2
The Kaplan-Meier curves for the adverse events. MACE, major adverse cardiac events; TLR, target lesion revascularization.