[Evaluation of Bispectral Index time delay in response to anesthesia induction: an observational study]

Ana Leitão Ferreira, Joaquim Gabriel Mendes, Catarina Sofia Nunes, Pedro Amorim, Ana Leitão Ferreira, Joaquim Gabriel Mendes, Catarina Sofia Nunes, Pedro Amorim

Abstract

Background and objectives: According to the manufacturer, the Bispectral Index (BIS) has a processing time delay of 5–10 s. Studies addressing this have suggested longer delays. We evaluated the time delay in the Bispectral Index response.

Methods: Based on clinical data from 45 patients, using the difference between the predicted and the real BIS, calculated during a fixed 3 minutes period after the moment the Bispectral Index dropped below 80 during the induction of general anesthesia with propofol and remifentanil.

Results: The difference between the predicted and the real BIS was in average 30.09 ± 18.73 s.

Conclusion: Our results may be another indication that the delay in BIS processing may be much longer than stated by the manufacture, a fact with clinical implications.

Keywords: Awareness; Bispectral index monitor; Consciência; Depth of anesthesia; Monitor do índice bispectral; Profundidade da anestesia; Tempo de atraso; Time delay.

Copyright © 2019 Sociedade Brasileira de Anestesiologia. Publicado por Elsevier Editora Ltda. All rights reserved.

Figures

Figure 1
Figure 1
Anesthesia setup.
Figure 2
Figure 2
Block diagram of BIS modulation.
Figure 3
Figure 3
Data results from all 45 patients: (A) Effect-site concentration of propofol and remifentanil for each patient during the first 15 min of anesthesia induction; (B) Real versus modeled BIS values.
Figure 4
Figure 4
Time delay (Δt) between the real and the modeled BIS for: (A) Patient no. 5; (B) Patient no. 19; (C) Patient no. 22 and (D) Patient no. 27.
Figure 5
Figure 5
Time delay (Δt) between the real and the modeled BIS for all patients.

References

    1. Lobo F.A., Schraag S. Limitations of anaesthesia depth monitoring. Curr Opin Anaesthesiol. 2011;24:657–664.
    1. Dahaba A.A. Different conditions that could result in the bispectral index indicating an incorrect hypnotic state. Anesth Analg. 2005;101:765–773.
    1. Covidien. [Online]. Available: .
    1. Nunes R.R., Chaves I.M., de Alencar J.C., et al. Bispectral index and other processed parameters of electroencephalogram: an update. Rev Bras Anestesiol. 2012;62:105–117.
    1. Rampil I.J. A primer for EEG signal processing in anesthesia. Anesthesiology. 1998;89:980–1002.
    1. Dryden J. Preventing memories of surgery 2011. [Online]. Available: .
    1. Lang J. 2013. Awakening Atl.
    1. Zanner R., Pilge S., Kochs E.F., et al. Time delay of electroencephalogram index calculation: analysis of cerebral state, bispectral, and Narcotrend indices using perioperatively recorded electroencephalographic signals. Br J Anaesth. 2009;103:394–399.
    1. Pilge S., Zanner R., Schneider G., et al. Time delay of index calculation: analysis of cerebral state, bispectral, and narcotrend indices. Anesthesiology. 2006;104:488–494.
    1. Ferreira A.I.L., Nunes C.S., Gabriel J., et al. In: Jaffray D., editor. Vol. 51. IFMBE Proceedings; Toronto, Canada: 2015. The influence of two different drug infusion profiles on the pharmacodynamics model performance; pp. 874–880. (World congress on medical physics and biomedical engineering, June 7–12).
    1. Minto C.F., Schnider T.W., Short T.G., et al. Response surface model for anesthetic drug interactions. Anesthesiology. 2000;92:1603–1616.
    1. Chernik D.A., Gillings D., Laine H., et al. Validity and reliability of the Observer's Assessment of Alertness/Sedation Scale: study with intravenous midazolam. J Clin Psychopharmacol. 1990;10:244–251.
    1. Schnider T.W., Minto C.F., Gambus P.L., et al. The influence of method of administration and covariates on the pharmacokinetics of propofol in adult volunteers. Anesthesiology. 1998;88:1170–1182.
    1. Minto C.F., Schnider T.W., Egan T.D., et al. Influence of age and gender on the pharmacokinetics and pharmacodynamics of remifentanil. I. Model development. Anesthesiology. 1997;86:10–23.
    1. Minto C.F., Schnider T.W., Shafer S.L. Pharmacokinetics and pharmacodynamics of remifentanil. II. Model application. Anesthesiology. 1997;86:24–33.
    1. Bruhn J., Bouillon T.W., Radulescu L., et al. Correlation of approximate entropy, bispectral index, and spectral edge frequency 95 (SEF95) with clinical signs of “anesthetic depth” during coadministration of propofol and remifentanil. Anesthesiology. 2003;98:621–627.
    1. Shepherd J., Jones J., Frampton G.K., et al. Clinical effectiveness and cost-effectiveness of depth of anaesthesia monitoring (E-Entropy, Bispectral Index and Narcotrend): a systematic review and economic evaluation. Health Technol Assess (Rockv) 2013;17:1–124.
    1. Purdon P.L., Pavone K.J., Akeju O., et al. The Ageing Brain: age-dependent changes in the electroencephalogram during propofol and sevoflurane general anaesthesia. Br J Anaesth. 2015;115:i46–i57.
    1. Myles P.S., Williams D.L., Hendrata M., et al. Patient satisfaction after anaesthesia and surgery: results of a prospective survey of 10,811 patients. Br J Anaesth. 2000;84:6–10.
    1. Leslie K., Chan M.T.V., Myles P.S., et al. Posttraumatic stress disorder in aware patients from the B-Aware trial. Anesth Analg. 2010;110:823–828.
    1. Myles P.S., Leslie K., McNeil J., et al. Bispectral index monitoring to prevent awareness during anaesthesia: the B-Aware randomised controlled trial. Lancet. 2004;363:1757–1763.
    1. Johansen J.W. Update on bispectral index monitoring. Best Pract Res Clin Anaesthesiol. 2006;20:81–99.
    1. Luginbühl M., Schnider T.W. Detection of awareness with the bispectral index: two case reports. Anesthesiology. 2002;96:241–243.
    1. Leslie K., Myles P.S., Forbes A., et al. The effect of bispectral index monitoring on long-term survival in the B-aware trial. Anesth Analg. 2010;110:816–822.
    1. Kertai M.D., Pal N., Palanca B.J.A., et al. Association of perioperative risk factors and cumulative duration of low bispectral index with intermediate-term mortality after cardiac surgery in the b-unaware trial. Anesthesiology. 2010;112:1116–1127.
    1. Lindholm M.-L., Träff S., Granath F., et al. Mortality within 2 years after surgery in relation to low intraoperative bispectral index values and preexisting malignant disease. Anesth Analg. 2009;108:508–512.

Source: PubMed

3
S'abonner