Comparison of the paediatric blade of the Pentax-AWS and Ovassapian airway in fibreoptic tracheal intubation in patients with limited mouth opening and cervical spine immobilization by a semi-rigid neck collar: a randomized controlled trial

D H Kim, J Y Yoo, S Y Ha, Y J Chae, D H Kim, J Y Yoo, S Y Ha, Y J Chae

Abstract

Background: We compared the performances of the paediatric blade of a Pentax Airway Scope and an Ovassapian airway in fibreoptic tracheal intubation in patients whose necks were stabilized by semi-rigid neck collars.

Methods: Ninety patients were enrolled in this prospective, open-label, randomized controlled trial. Patients were randomly allocated to one of two groups (Group OVA-FOB and Group AWS-FOB). The time to tracheal intubation, success rate of tracheal intubation, number of optimization manoeuvres (jaw thrust), and difficulty of manipulation of the fibreoptic bronchoscope were compared between the groups.

Results: The time to tracheal intubation was significantly shorter (32 vs 50 s; median difference 19 s; 95% confidence interval 14-25 s; P<0.001) and manipulation of the fibreoptic bronchoscope was significantly easier for Group AWS-FOB. Optimization manoeuvres were rarely required to facilitate fibreoptic tracheal intubation in Group AWS-FOB [jaw thrust, 0 (0%); jaw thrust with anterior neck collar removal, 1 (2%)] compared with that required in Group OVA-FOB [jaw thrust, 39 (87%); jaw thrust with anterior neck collar removal, 2 (4%)]. There was no significant difference in the success rate of tracheal intubation on the first attempt between groups [Group AWS-FOB, 45 (100%); Group OVA-FOB, 44 (98%)].

Conclusions: Combined use of the paediatric blade of a Pentax Airway Scope and a fibreoptic bronchoscope enabled rapid tracheal intubation, minimizing the use of external manoeuvres of the airway, in patients with limited mouth opening and cervical spine immobilization by semi-rigid neck collars, compared with use of the Ovassapian airway and the fibreoptic bronchoscope.

Clinical trial registration: NCT02827110.

Keywords: airway management; bronchoscopes; laryngoscopes.

© The Author 2017. Published by Oxford University Press on behalf of the British Journal of Anaesthesia. All rights reserved. For Permissions, please email: journals.permissions@oup.com