Intraosseous blood samples for point-of-care analysis: agreement between intraosseous and arterial analyses

Milla Jousi, Simo Saikko, Jouni Nurmi, Milla Jousi, Simo Saikko, Jouni Nurmi

Abstract

Background: Point-of-care (POC) testing is highly useful when treating critically ill patients. In case of difficult vascular access, the intraosseous (IO) route is commonly used, and blood is aspirated to confirm the correct position of the IO-needle. Thus, IO blood samples could be easily accessed for POC analyses in emergency situations. The aim of this study was to determine whether IO values agree sufficiently with arterial values to be used for clinical decision making.

Methods: Two samples of IO blood were drawn from 31 healthy volunteers and compared with arterial samples. The samples were analysed for sodium, potassium, ionized calcium, glucose, haemoglobin, haematocrit, pH, blood gases, base excess, bicarbonate, and lactate using the i-STAT® POC device. Agreement and reliability were estimated by using the Bland-Altman method and intraclass correlation coefficient calculations.

Results: Good agreement was evident between the IO and arterial samples for pH, glucose, and lactate. Potassium levels were clearly higher in the IO samples than those from arterial blood. Base excess and bicarbonate were slightly higher, and sodium and ionised calcium values were slightly lower, in the IO samples compared with the arterial values. The blood gases in the IO samples were between arterial and venous values. Haemoglobin and haematocrit showed remarkable variation in agreement.

Discussion: POC diagnostics of IO blood can be a useful tool to guide treatment in critical emergency care. Seeking out the reversible causes of cardiac arrest or assessing the severity of shock are examples of situations in which obtaining vascular access and blood samples can be difficult, though information about the electrolytes, acid-base balance, and lactate could guide clinical decision making. The analysis of IO samples should though be limited to situations in which no other option is available, and the results should be interpreted with caution, because there is not yet enough scientific evidence regarding the agreement of IO and arterial results among unstable patients.

Conclusions: IO blood samples are suitable for analysis with the i-STAT® point-of-care device in emergency care. The aspirate used to confirm the correct placement of the IO needle can also be used for analysis. The results must be interpreted within a clinical context while taking the magnitude and direction of bias into account.

Keywords: Blood gas analysis; I-STAT; Intraosseous access; Point-of-care.

Conflict of interest statement

Authors’ information

MJ and JN are anaesthesiologists and emergency doctors, and both work as HEMS doctors. SS is a registered nurse and paramedic. He is a main lecturer of paramedics at Saimaa University of Applied Sciences, Lappeenranta, Finland. JN is the head of the research unit of Department of Emergency Medicine and Services of Helsinki University Hospital.

Ethics approval and consent to participate

The research was conducted according to the principles of the Declaration of Helsinki, and the study protocol was approved by the Coordinating Ethics Committee of Helsinki University Hospital (ref no 250/13/03/00/15). Written informed consent was obtained from all the participants, and follow-up was organised in cases of complications.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Figures

Fig. 1
Fig. 1
Bland-Altman graphs of intraosseous and arterial samples (left panels), and the comparison of the biases between intraosseous samples and venous samples in reference to arterial samples (right panels) for POC measurements of potassium, sodium, and ionised calcium. The dotted lines on the Bland-Altman graphs indicate bias and the limits of agreement (+/− 1.96 SD). The error bars on the right panels represent biases (95% confidence intervals). IO1 – first intraosseous sample; IO2 – second intraosseous sample after 2 ml of waste blood was drawn; V – venous sample; A – arterial sample
Fig. 2
Fig. 2
Bland-Altman graphs of intraosseous and arterial samples (left panels), and the comparison of the biases between intraosseous samples and venous samples in reference to arterial samples (right panels) for POC measurements of pH, base excess, bicarbonate, pO2, and pCO2. The dotted lines on the Bland-Altman graphs indicate bias and the limits of agreement (+/− 1.96 SD). The error bars on the right panels represent biases (95% confidence intervals). IO1 – first intraosseous sample; IO2 – second intraosseous sample after 2 ml of waste blood was drawn; V – venous sample; A – arterial sample
Fig. 3
Fig. 3
Bland-Altman graphs of intraosseous and arterial samples (left panels), and the comparison of the biases between the intraosseous samples and venous samples in reference to arterial samples (right panels) for POC measurements of haemoglobin, haematocrit, glucose, and lactate. The dotted lines on the Bland-Altman graphs indicate bias and the limits of agreement (+/− 1.96 SD). The error bars on right panels represent biases (95% confidence intervals). IO1 – first intraosseous sample; IO2 – second intraosseous sample after 2 ml of waste blood was drawn; V – venous sample; A – arterial sample

References

    1. Soar J, Nolan JP, Böttiger BW et al. European Resuscitation Council Guidelines for Resuscitation 2015. Resuscitation 2015;95:100-147.
    1. Eriksson M, Strandberg G, Lipcsey M, Larsson A. Troponin I can be determined in Intraosseous aspirates in a porcine shock model. Clin Lab. 2015;61(7):825–829.
    1. Strandberg G, Lipcsey M, Eriksson M, Larsson A. Intraosseous samples can be used for creatinine measurements - an experimental study in the anaesthetised pig. Clin Lab. 2014;60(10):1587–1591.
    1. Veldhoen ES, De Vooght K, Slieker MG, Versluys AB, Turner N. Analysis of bloodgas, electrolytes and glucose from intraosseous samples using an i-STAT point-of-care analyser. Resuscitation. 2014;85:359–363. doi: 10.1016/j.resuscitation.2013.12.002.
    1. Strandberg G, Larsson A, Lipcsey M, Berglund L, Eriksson M. Analysis of intraosseous samples in endotoxemic shock – an experimental study in the anaesthetised pig. Acta Anaesthesiol Scand. 2014;58:337–344. doi: 10.1111/aas.12274.
    1. Larsson T, Strandberg G, Eriksson M, Bondesson U, Lipcsey M, Larsson A. Intraosseous samples can be used for opioid measurements - an experimental study in the anaesthetized pig. Scand J Clin Lab Invest. 2013;73:102–106. doi: 10.3109/00365513.2012.744088.
    1. Strandberg G, Eriksson M, Gustafsson MG, Lipcsey M, Larsson A. Analysis of intraosseous samples using point of care technology - an experimental study in the anaeshetised pig. Resuscitation. 2012;83:1381–1385. doi: 10.1016/j.resuscitation.2012.04.007.
    1. Miller LJ, Philbeck TE, Montez D, Spadaccini CJ. A new study of Intraosseous blood for laboratory analysis. Arch Pathol Lab Med. 2010;134:1253–1260.
    1. Greco SC, Talcott MR, LaRegine MC, Eisenbeis PE. Use of intraosseous blood for repeated hematologic and biochemical analyses in healthy pigs. Am J Vet Res. 2001;62:43–47. doi: 10.2460/ajvr.2001.62.43.
    1. Hurren JS. Can blood taken from intraosseous cannulations be used for blood analysis? Burns. 2000;26:727–730. doi: 10.1016/S0305-4179(00)00034-6.
    1. Voelckel WG, Lindner KH, Wenzel V, et al. Intraosseous blood gases during hypothermia: correlation with arterial, mixed venous, and sagittal sinus blood. Crit Care Med. 2000;28:2915–2920. doi: 10.1097/00003246-200008000-00038.
    1. Abdelmoneim T, Kissoon N, Johnson L, Fiallos M, Murphy S. Acid-base status of blood from intraosseous and mixed venous sites during prolonged cardiopulmonary resuscitation and drug infusions. Crit Care Med. 1999;27:1923–1928. doi: 10.1097/00003246-199909000-00034.
    1. Kissoon N, Idris A, Wenzel V, Murphy S, Rush W. Intraosseous and central venous blood acid-base relationship during cardiopulmonary resuscitation. Ped Emerg Care. 1997;13:250–253. doi: 10.1097/00006565-199708000-00002.
    1. Ummenhofer W, Frei FJ, Urwyler A, Drewe J. Are laboratory values in bone marrow aspirate predictable for venous blood in paediatric patients? Resuscitation. 1994;27:123–128. doi: 10.1016/0300-9572(94)90004-3.
    1. Kissoon N, Peterson R, Murphy S, Gayle M, Ceithaml E, Harwood-Nuss A. Comparison of pH and carbon dioxide tension values of central venous and intraosseous blood during changes in cardiac output. Crit Care Med. 1994;22:1010–1015. doi: 10.1097/00003246-199406000-00021.
    1. Kissoon N, Rosenberg H, Gloor J, Vidal R. Comparison of the acid-base status of blood obtained from intraosseous and central venous sites during steady- and low-flow states. Crit Care Med. 1993;21:1765–1769. doi: 10.1097/00003246-199311000-00028.
    1. Brickman KR, Krupp K, Rega P, Alexander J, Guinness M. Typing and screening of blood from Intraosseous access. Ann Emerg Med. 1992;21:414–417. doi: 10.1016/S0196-0644(05)82661-7.
    1. Grisham J, Hastings C. Bone marrow aspirate as an accessible and reliable source for critical laboratory studies. Ann Emerg Med. 1991;20:1121–1124. doi: 10.1016/S0196-0644(05)81388-5.
    1. Orlowski JP, Porembka DT, Gallagher JM, Van Lente F. The bone marrow as a source of laboratory studies. Ann Emerg Med. 1989;18:1348–1351. doi: 10.1016/S0196-0644(89)80274-4.
    1. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–310. doi: 10.1016/S0140-6736(86)90837-8.
    1. Hanneman SK. Design, analysis and interpretation of method-comparison studies. Adv Crit Care. 2008;19(2):223–234.
    1. Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res. 1999;8:135–160. doi: 10.1177/096228029900800204.
    1. Bland JM, Altman DG. Comparing methods of measurement: why plotting difference against standard method is misleading. Lancet. 1995;346:1085–1087. doi: 10.1016/S0140-6736(95)91748-9.
    1. Koo TK, Li MY. A guideline of selecting and reporting Intraclass correlation coefficients for reliability research. J Chiropr Med. 2016;15(2):155–163. doi: 10.1016/j.jcm.2016.02.012.
    1. Kottner J, Audigé L, Brorson S, Donner A, Gajewski BJ, Hróbjartsson A, et al. Guidelines for reporting reliability and agreement studies (GRRAS) were proposed. J Clin Epidemiol. 2011;64(1):96–106. doi: 10.1016/j.jclinepi.2010.03.002.
    1. Bland JM, Altman DG. A note on the use of the intraclass correlation coefficient in the evaluation of agreement between two methods of measurement. Comput Biol Med. 1990;20(5):337–340. doi: 10.1016/0010-4825(90)90013-F.

Source: PubMed

3
S'abonner