The accuracy of pulse oximetry in measuring oxygen saturation by levels of skin pigmentation: a systematic review and meta-analysis

Chunhu Shi, Mark Goodall, Jo Dumville, James Hill, Gill Norman, Oliver Hamer, Andrew Clegg, Caroline Leigh Watkins, George Georgiou, Alexander Hodkinson, Catherine Elizabeth Lightbody, Paul Dark, Nicky Cullum, Chunhu Shi, Mark Goodall, Jo Dumville, James Hill, Gill Norman, Oliver Hamer, Andrew Clegg, Caroline Leigh Watkins, George Georgiou, Alexander Hodkinson, Catherine Elizabeth Lightbody, Paul Dark, Nicky Cullum

Abstract

Background: During the COVID-19 pandemic, there have been concerns regarding potential bias in pulse oximetry measurements for people with high levels of skin pigmentation. We systematically reviewed the effects of skin pigmentation on the accuracy of oxygen saturation measurement by pulse oximetry (SpO2) compared with the gold standard SaO2 measured by CO-oximetry.

Methods: We searched Ovid MEDLINE, Ovid Embase, EBSCO CINAHL, ClinicalTrials.gov, and WHO International Clinical Trials Registry Platform (up to December 2021) for studies with SpO2-SaO2 comparisons and measuring the impact of skin pigmentation or ethnicity on pulse oximetry accuracy. We performed meta-analyses for mean bias (the primary outcome in this review) and its standard deviations (SDs) across studies included for each subgroup of skin pigmentation and ethnicity and used these pooled mean biases and SDs to calculate accuracy root-mean-square (Arms) and 95% limits of agreement. The review was registered with the Open Science Framework ( https://osf.io/gm7ty ).

Results: We included 32 studies (6505 participants): 15 measured skin pigmentation and 22 referred to ethnicity. Compared with standard SaO2 measurement, pulse oximetry probably overestimates oxygen saturation in people with the high level of skin pigmentation (pooled mean bias 1.11%; 95% confidence interval 0.29 to 1.93%) and people described as Black/African American (1.52%; 0.95 to 2.09%) (moderate- and low-certainty evidence). The bias of pulse oximetry measurements for people with other levels of skin pigmentation or those from other ethnic groups is either more uncertain or suggests no overestimation. Whilst the extent of mean bias is small or negligible for all subgroups evaluated, the associated imprecision is unacceptably large (pooled SDs > 1%). When the extent of measurement bias and precision is considered jointly, pulse oximetry measurements for all the subgroups appear acceptably accurate (with Arms < 4%).

Conclusions: Pulse oximetry may overestimate oxygen saturation in people with high levels of skin pigmentation and people whose ethnicity is reported as Black/African American, compared with SaO2. The extent of overestimation may be small in hospital settings but unknown in community settings. REVIEW PROTOCOL REGISTRATION: https://osf.io/gm7ty.

Keywords: Arterial blood oxygen saturation; Ethnicity; Measurement bias; Pulse oximetry; Skin pigmentation; Systematic review.

Conflict of interest statement

NC reports research grants from the National Institute for Health Research and the Accelerated Access Collaborative at NHS England and NHS Improvement and payments were made to the University of Manchester. PD is the National Deputy Medical Director, NIHR Clinical Research Network Coordinating Centre, UK, developing and delivering clinical research in the subject area and contributing to developing and implementing NIHR’s Equality, Diversity and Inclusion strategy. All other authors have declared that no competing interests exist.

© 2022. The Author(s).

Figures

Fig. 1
Fig. 1
The study selection flowchart. This flowchart shows the number of records and studies at each stage of the study selection process
Fig. 2
Fig. 2
Risk of bias assessment results. The left section of this figure shows risk of bias judgements for each domain of the QUADAS-2 tool for each study and the right section shows applicability judgements for each concern domain of the QUADAS-2 tool for each study. Please see Additional file 4: Box S3 for all signalling questions used in the QUADAS-2 assessment and further considerations

References

    1. Swigris JJ, Zhou X, Wamboldt FS, Du Bois R, Keith R, Fischer A, et al. Exercise peripheral oxygen saturation (SpO2) accurately reflects arterial oxygen saturation (SaO2) and predicts mortality in systemic sclerosis. Thorax. 2009;64(7):626–630. doi: 10.1136/thx.2008.111393.
    1. NHS England. COVID Oximetry@home. 2022. . Accessed 21 Apr 2022.
    1. World Health Organization (WHO). COVID-19 clinical management: living guideline (updated 25.1.21). 2021. . Accessed 21 Apr 2022.
    1. The Royal Australian College of General Practitioners . Managing COVID-19 at home with assistance from your general practice: a guide, action plan and symptom diary for patients. East Melbourne: RACGP; 2021.
    1. Luks AM, Swenson ER. Pulse oximetry for monitoring patients with COVID-19 at home: potential pitfalls and practical guidance. Ann Am Thorac Soc. 2020;17:1040–1046. doi: 10.1513/AnnalsATS.202005-418FR.
    1. NHS Race and Health Observatory . Pulse oximetry and racial bias: recommendations for national healthcare, regulatory and research bodies. 2021.
    1. Sjoding MW, Dickson RP, Iwashyna TJ, Gay SE, Valley TS. Racial bias in pulse oximetry measurement. N Engl J Med. 2020;383(25):2477–2478. doi: 10.1056/NEJMc2029240.
    1. Jensen LA, Onyskiw JE, Prasad NG. Meta-analysis of arterial oxygen saturation monitoring by pulse oximetry in adults. Heart Lung. 1998;27(6):387–408. doi: 10.1016/S0147-9563(98)90086-3.
    1. Page MJ, Moher D, Bossuyt PM, Boutron I, Hoffmann TC, Mulrow CD, et al. PRISMA 2020 explanation and elaboration: updated guidance and exemplars for reporting systematic reviews. BMJ. 2021;372:n160. doi: 10.1136/bmj.n160.
    1. British Standards Institution (BSI) BS EN ISO 80601–2–61:2019: Medical electrical equipment – Part 2–61: Particular requirements for basic safety and essential performance of pulse oximeter equipment (ISO 80601–2–61:2017, Corrected version 2018–02) London: British Standards Institution; 2019.
    1. Toffaletti J, Zijlstra WG. Misconceptions in reporting oxygen saturation. Anesth Analg. 2007;105(6):S5–9. doi: 10.1213/01.ane.0000278741.29274.e1.
    1. Whiting PF, Rutjes AW, Westwood ME, Mallett S, Deeks JJ, Reitsma JB, et al. QUADAS-2: a revised tool for the quality assessment of diagnostic accuracy studies. Ann Intern Med. 2011;155(8):529–536. doi: 10.7326/0003-4819-155-8-201110180-00009.
    1. Wan X, Wang W, Liu J, Tong T. Estimating the sample mean and standard deviation from the sample size, median, range and/or interquartile range. BMC Med Res Methodol. 2014;14(1):1–3. doi: 10.1186/1471-2288-14-135.
    1. Hedges LV, Tipton E, Johnson MC. Robust variance estimation in meta-regression with dependent effect size estimates. Res Synth Methods. 2010;1(1):39–65. doi: 10.1002/jrsm.5.
    1. Pustejovsky JE, Tipton E. Meta-analysis with robust variance estimation: Expanding the range of working models. Prev Sci. 2021;7:1–4.
    1. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10. doi: 10.1016/S0140-6736(86)90837-8.
    1. Campbell M, McKenzie JE, Sowden A, Katikireddi SV, Brennan SE, Ellis S, et al. Synthesis without meta-analysis (SwiM) in systematic reviews: reporting guideline. BMJ. 2020;368:l6890. doi: 10.1136/bmj.l6890.
    1. Schünemann HJ, Mustafa RA, Brozek J, Steingart KR, Leeflang M, Murad MH, et al. GRADE guidelines: 21 part 1. Study design, risk of bias, and indirectness in rating the certainty across a body of evidence for test accuracy. J Clin Epidemiol. 2020;122:129–41. doi: 10.1016/j.jclinepi.2019.12.020.
    1. Schünemann HJ, Mustafa RA, Brozek J, Steingart KR, Leeflang M, Murad MH, et al. GRADE guidelines: 21 part 2. Test accuracy: inconsistency, imprecision, publication bias, and other domains for rating the certainty of evidence and presenting it in evidence profiles and summary of findings tables. J Clin Epidemiol. 2020;122:142–52. doi: 10.1016/j.jclinepi.2019.12.021.
    1. Wolfgang (), Metafor package: bias and sensitivity diagnostics. 7 June 2015. . Accessed 5 Jan 2022.
    1. Abrams GA, Sanders MK, Fallon MB. Utility of pulse oximetry in the detection of arterial hypoxemia in liver transplant candidates. Liver Transpl. 2002;8(4):391–396. doi: 10.1053/jlts.2002.32252.
    1. Adler JN, Hughes LA, Vivilecchia R, Camargo CA., Jr Effect of skin pigmentation on pulse oximetry accuracy in the emergency department. Acad Emerg Med. 1998;5(10):965–970. doi: 10.1111/j.1553-2712.1998.tb02772.x.
    1. Avant MG, Lowe N, Torres A., Jr Comparison of accuracy and signal consistency of two reusable pulse oximeter probes in critically ill children. Respir Care. 1997;42(7):698–704.
    1. Bickler PE, Feiner JR, Severinghaus JW. Effects of skin pigmentation on pulse oximeter accuracy at low saturation. Anesthesiology. 2005;102(4):715–719. doi: 10.1097/00000542-200504000-00004.
    1. Bothma PA, Joynt GM, Upman J, Hon H, Mathala B, Scribante J, et al. Accuracy of pulse oximetry in pigmented patients. S Afr Med J. 1996;86(5):8914569.
    1. [preprint] Brooks JC, Raman S, Gibbons K, et al. Transcutaneous oxygen saturation accuracy in critically ill children. Research Square 2020. DOI: 10.21203/rs.2.21938/v1.
    1. Ebmeier SJ, Barker M, Bacon M, Beasley RC, Bellomo R, Chong CK, et al. A two centre observational study of simultaneous pulse oximetry and arterial oxygen saturation recordings in intensive care unit patients. Anaesth Intensive Care. 2018;46(3):297–303. doi: 10.1177/0310057X1804600307.
    1. Escourrou PJ, Delaperche MF, Visseaux A. Reliability of pulse oximetry during exercise in pulmonary patients. Chest. 1990;97(3):635–638. doi: 10.1378/chest.97.3.635.
    1. Feiner JR, Severinghaus JW, Bickler PE. Dark skin decreases the accuracy of pulse oximeters at low oxygen saturation: the effects of oximeter probe type and gender. Anesth Analg. 2007;105(6):S18–23. doi: 10.1213/01.ane.0000285988.35174.d9.
    1. Foglia E, Whyte R, Chaudhary A, Mott A, Chen J, Propert K, et al. Accuracy and precision of pulse oximetry in hypoxemic infants. J Pediatr. 2017;182:375–377. doi: 10.1016/j.jpeds.2016.11.043.
    1. Gabrielczyk MR, Buist RJ. Pulse oximetry and postoperative hypothermia: an evaluation of the Nellcor N-100 in a cardiac surgical intensive care unit. Anaesthesia. 1988;43(5):402–404. doi: 10.1111/j.1365-2044.1988.tb09025.x.
    1. Harris BU, Char DS, Feinstein JA, Verma A, Shiboski SC, Ramamoorthy C. Accuracy of pulse oximeters intended for hypoxemic pediatric patients. Pediatr Crit Care Med. 2016;17:315–320. doi: 10.1097/PCC.0000000000000660.
    1. Harris BU, Stewart S, Verma A, Hoen H, Stein ML, Wright G, et al. Accuracy of a portable pulse oximeter in monitoring hypoxemic infants with cyanotic heart disease. Cardiol Young. 2019;29(8):1025–1029. doi: 10.1017/S1047951119001355.
    1. Harskamp R, Bekker L, Himmelreich J, De Clercq L, Karregat EP, Sleeswijk ME, et al. Performance of popular pulse oximeters compared with simultaneous arterial oxygen saturation or clinical-grade pulse oximetry: a cross-sectional validation study in intensive care patients. BMJ Open Respir Res. 2021;8(1):e000939. doi: 10.1136/bmjresp-2021-000939.
    1. Hinkelbein J, Genzwuerker HV, Sogl R, Fiedler F. Effect of nail polish on oxygen saturation determined by pulse oximetry in critically ill patients. Resuscitation. 2007;72(1):82–91. doi: 10.1016/j.resuscitation.2006.06.024.
    1. Hinkelbein J, Koehler H, Genzwuerker HV, Fiedler F. Artificial acrylic finger nails may alter pulse oximetry measurement. Resuscitation. 2007;74(1):75–82. doi: 10.1016/j.resuscitation.2006.11.018.
    1. Jubran A, Tobin MJ. Reliability of pulse oximetry in titrating supplemental oxygen therapy in ventilator-dependent patients. Chest. 1990;97:1420–1425. doi: 10.1378/chest.97.6.1420.
    1. Lee KH, Hui KP, Tan WC, Lim TK. Factors influencing pulse oximetry as compared to functional arterial saturation in multi-ethnic Singapore. Singapore Med J. 1993;34:385–387.
    1. McGovern JP, Sasse SA, Stansbury DW, Causing LA, Light RW. Comparison of oxygen saturation by pulse oximetry and co-oximetry during exercise testing in patients with COPD. Chest. 1996;109(5):1151–1155. doi: 10.1378/chest.109.5.1151.
    1. Muñoz X, Torres F, Sampol G, Rios J, Martí S, Escrich E. Accuracy and reliability of pulse oximetry at different arterial carbon dioxide pressure levels. Eur Respir J. 2008;32(4):1053–1059. doi: 10.1183/09031936.00126507.
    1. Pilcher J, Ploen L, McKinstry S, Bardsley G, Chien J, Howard L, et al. A multicentre prospective observational study comparing arterial blood gas values to those obtained by pulse oximeters used in adult patients attending Australian and New Zealand hospitals. BMC Pulm Med. 2020;20(1):1–9. doi: 10.1186/s12890-019-1042-0.
    1. Ploen L, Pilcher J, Beckert L, Swanney M, Beasley R. An investigation into the bias of pulse oximeters. Respirology. 2016;21(Suppl 2):6.
    1. Ries AL, Farrow JT, Clausen JL. Accuracy of two ear oximeters at rest and during exercise in pulmonary patients. Am Rev Respir Dis. 1985;132(3):685–689.
    1. Ries AL, Prewitt LM, Johnson JJ. Skin color and ear oximetry. Chest. 1989;96(2):287–290. doi: 10.1378/chest.96.2.287.
    1. Ross PA, Newth CJ, Khemani RG. Accuracy of pulse oximetry in children. Pediatrics. 2014;133(1):22–29. doi: 10.1542/peds.2013-1760.
    1. Schallom M, Prentice D, Sona C, Arroyo C, Mazuski J. Comparison of nasal and forehead oximetry accuracy and pressure injury in critically ill patients. Heart Lung. 2018;47(2):93–99. doi: 10.1016/j.hrtlng.2017.12.002.
    1. Smyth RJ, D’Urzo AD, Slutsky AS, Galko BM, Rebuck AS. Ear oximetry during combined hypoxia and exercise. J Appl Physiol. 1986;60(2):716–719. doi: 10.1152/jappl.1986.60.2.716.
    1. Stewart KG, Rowbottom SJ. Inaccuracy of pulse oximetry in patients with severe tricuspid regurgitation. Anaesthesia. 1991;46(8):668–670. doi: 10.1111/j.1365-2044.1991.tb09720.x.
    1. Thrush D, Hodges MR. Accuracy of pulse oximetry during hypoxemia. South Med J. 1994;87(4):518–521. doi: 10.1097/00007611-199404000-00019.
    1. Valbuena VS, Barbaro RP, Claar D, Valley TS, Dickson RP, Gay SE, et al. Racial bias in pulse oximetry measurement among patients about to undergo extracorporeal membrane oxygenation in 2019–2020: a retrospective cohort study. Chest. 2022;161(4):971–978. doi: 10.1016/j.chest.2021.09.025.
    1. Vesoulis Z, Tims A, Lodhi H, Lalos N, Whitehead H. Racial discrepancy in pulse oximeter accuracy in preterm infants. J Perinatol. 2021;42(1):79–85. doi: 10.1038/s41372-021-01230-3.
    1. Wiles MD, El-Nayal A, Elton G, Malaj M, Winterbottom J, Gillies C, et al. The effect of patient ethnicity on the accuracy of peripheral pulse oximetry in patients with COVID-19 pneumonitis: a single-centre, retrospective analysis. Anaesthesia. 2022;77(4):489–491. doi: 10.1111/anae.15656.
    1. Zeballos RJ, Weisman IM. Reliability of noninvasive oximetry in black subjects during exercise and hypoxia. Am Rev Respir Dis. 1991;144:1240–1244. doi: 10.1164/ajrccm/144.6.1240.
    1. Blanchet MA, Mercier G, Bouchard PA, Rousseau E, Lellouche F. Accuracy of pulse oximetry (SpO2) with different oximeters. Oxygap study. Intensive Care Med Exp. 2021;9(Suppl 1):51.
    1. International Organization for Standardization. Medical electrical equipment — Part 2–61: Particular requirements for basic safety and essential performance of pulse oximeter equipment (ISO 80601–2–61:2017). . Accessed 21 Apr 2022.
    1. Food and Drug Administration . Pulse oximeters – premarket notification submissions: guidance for industry and food and drug administration staff. 2013.
    1. Wong AK, Charpignon M, Kim H, Josef C, de Hond AA, Fojas JJ, et al. Analysis of discrepancies between pulse oximetry and arterial oxygen saturation measurements by race and ethnicity and association with organ dysfunction and mortality. JAMA Netw Open. 2021;4(11):e2131674. doi: 10.1001/jamanetworkopen.2021.31674.
    1. Silver SE. Skin color is not the same thing as race. Arch Dermatol. 2004;140(3):361. doi: 10.1001/archderm.140.3.361-a.
    1. Ware OR, Dawson JE, Shinohara MM, Taylor SC. Racial limitations of Fitzpatrick skin type. Cutis. 2020;105(2):77–80.
    1. Morrison A, Polisena J, Husereau D, Moulton K, Clark M, Fiander M, et al. The effect of English-language restriction on systematic review-based meta-analyses: a systematic review of empirical studies. Int J Technol Assess Health Care. 2012;28(2):138–144. doi: 10.1017/S0266462312000086.
    1. Nussbaumer-Streit B, Klerings I, Dobrescu AI, Persad E, Stevens A, Garritty C, et al. Excluding non-English publications from evidence-syntheses did not change conclusions: a meta-epidemiological study. J Clin Epidemiol. 2020;118:42–54. doi: 10.1016/j.jclinepi.2019.10.011.

Source: PubMed