Recommendations for determining the validity of consumer wearable heart rate devices: expert statement and checklist of the INTERLIVE Network

Jan M Mühlen, Julie Stang, Esben Lykke Skovgaard, Pedro B Judice, Pablo Molina-Garcia, William Johnston, Luís B Sardinha, Francisco B Ortega, Brian Caulfield, Wilhelm Bloch, Sulin Cheng, Ulf Ekelund, Jan Christian Brønd, Anders Grøntved, Moritz Schumann, Jan M Mühlen, Julie Stang, Esben Lykke Skovgaard, Pedro B Judice, Pablo Molina-Garcia, William Johnston, Luís B Sardinha, Francisco B Ortega, Brian Caulfield, Wilhelm Bloch, Sulin Cheng, Ulf Ekelund, Jan Christian Brønd, Anders Grøntved, Moritz Schumann

Abstract

Assessing vital signs such as heart rate (HR) by wearable devices in a lifestyle-related environment provides widespread opportunities for public health related research and applications. Commonly, consumer wearable devices assessing HR are based on photoplethysmography (PPG), where HR is determined by absorption and reflection of emitted light by the blood. However, methodological differences and shortcomings in the validation process hamper the comparability of the validity of various wearable devices assessing HR. Towards Intelligent Health and Well-Being: Network of Physical Activity Assessment (INTERLIVE) is a joint European initiative of six universities and one industrial partner. The consortium was founded in 2019 and strives towards developing best-practice recommendations for evaluating the validity of consumer wearables and smartphones. This expert statement presents a best-practice validation protocol for consumer wearables assessing HR by PPG. The recommendations were developed through the following multi-stage process: (1) a systematic literature review based on the Preferred Reporting Items for Systematic Reviews and Meta-Analyses, (2) an unstructured review of the wider literature pertaining to factors that may introduce bias during the validation of these devices and (3) evidence-informed expert opinions of the INTERLIVE Network. A total of 44 articles were deemed eligible and retrieved through our systematic literature review. Based on these studies, a wider literature review and our evidence-informed expert opinions, we propose a validation framework with standardised recommendations using six domains: considerations for the target population, criterion measure, index measure, testing conditions, data processing and the statistical analysis. As such, this paper presents recommendations to standardise the validity testing and reporting of PPG-based HR wearables used by consumers. Moreover, checklists are provided to guide the validation protocol development and reporting. This will ensure that manufacturers, consumers, healthcare providers and researchers use wearables safely and to its full potential.

Keywords: cardiology; consensus statement; public health; sports and exercise medicine; sports medicine.

Conflict of interest statement

Competing interests: None declared.

© Author(s) (or their employer(s)) 2021. Re-use permitted under CC BY-NC. No commercial re-use. See rights and permissions. Published by BMJ.

Figures

Figure 1
Figure 1
The six domains identified as important factors to be considered during validity testing of wearable devices assessing heart rate by photoplethysmography.

References

    1. Keith A, Flack M. The Form and Nature of the Muscular Connections between the Primary Divisions of the Vertebrate Heart. J Anat Physiol 1907;41:172–89.
    1. Fox K, Borer JS, Camm AJ, et al. . Resting heart rate in cardiovascular disease. J Am Coll Cardiol 2007;50:823–30. 10.1016/j.jacc.2007.04.079
    1. Åstrand P-O, Ryhming I. A nomogram for calculation of aerobic capacity (physical fitness) from pulse rate during sub-maximal work. J Appl Physiol 1954;7:218–21. 10.1152/jappl.1954.7.2.218
    1. Kim H-G, Cheon E-J, Bai D-S, et al. . Stress and Heart Rate Variability: A Meta-Analysis and Review of the Literature. Psychiatry Investig 2018;15:235–45. 10.30773/pi.2017.08.17
    1. Ross R, Blair SN, Arena R, et al. . Importance of Assessing Cardiorespiratory Fitness in Clinical Practice: A Case for Fitness as a Clinical Vital Sign: A Scientific Statement From the American Heart Association. Circulation 2016;134:e653–99. 10.1161/CIR.0000000000000461
    1. Foster C, Rodriguez-Marroyo JA, de Koning JJ. Monitoring Training Loads: The Past, the Present, and the Future. Int J Sports Physiol Perform 2017;12:S22–8. 10.1123/IJSPP.2016-0388
    1. Buchheit M. Monitoring training status with HR measures: do all roads lead to Rome? Front Physiol 2014;5:73. 10.3389/fphys.2014.00073
    1. Cheung CC, Krahn AD, Andrade JG. The Emerging Role of Wearable Technologies in Detection of Arrhythmia. Can J Cardiol 2018;34:1083–7. 10.1016/j.cjca.2018.05.003
    1. Pereira T, Tran N, Gadhoumi K, et al. . Photoplethysmography based atrial fibrillation detection: a review. NPJ Digit Med 2020;3:3. 10.1038/s41746-019-0207-9
    1. Piwek L, Ellis DA, Andrews S, et al. . The Rise of Consumer Health Wearables: Promises and Barriers. PLoS Med 2016;13:e1001953. 10.1371/journal.pmed.1001953
    1. Raja JM, Elsakr C, Roman S, et al. . Apple Watch, Wearables, and heart rhythm: where do we stand? Ann Transl Med 2019;7:417. 10.21037/atm.2019.06.79
    1. Sartor F, Gelissen J, van Dinther R, et al. . Wrist-worn optical and chest strap heart rate comparison in a heterogeneous sample of healthy individuals and in coronary artery disease patients. BMC Sports Sci Med Rehabil 2018;10:10. 10.1186/s13102-018-0098-0
    1. Turakhia MP, Desai M, Hedlin H, et al. . Rationale and design of a large-scale, app-based study to identify cardiac arrhythmias using a smartwatch: The Apple Heart Study. Am Heart J 2019;207:66–75. 10.1016/j.ahj.2018.09.002
    1. Fortino GG, R.; Galzarano S. Wearable Computing: From Modeling to Implementation of Wearable Systems Based on Body Sensor Networks. Wiley, 2018.
    1. Bunn JA, Navalta JW, Fountaine CJ, et al. . Current State of Commercial Wearable Technology in Physical Activity Monitoring 2015-2017. Int J Exerc Sci 2018;11:503–15.
    1. Sartor F, Papini G, Cox LGE, et al. . Methodological Shortcomings of Wrist-Worn Heart Rate Monitors Validations. J Med Internet Res 2018;20:e10108. 10.2196/10108
    1. Hardey MM. On the body of the consumer: performance-seeking with wearables and health and fitness apps. Sociol Health Illn 2019;41:991–1004. 10.1111/1467-9566.12879
    1. Nauman J, Janszky I, Vatten LJ, et al. . Temporal changes in resting heart rate and deaths from ischemic heart disease. JAMA 2011;306:2579–87. 10.1001/jama.2011.1826
    1. John D, Sasaki J, Hickey A, et al. . ActiGraph™ activity monitors: "the firmware effect". Med Sci Sports Exerc 2014;46:834–9. 10.1249/MSS.0000000000000145
    1. Consumer Technology Association . ANSI/CTA Standard. Physical Activity Monitoring for Heart Rate. ANSI/CTA-2065, 2018.
    1. Nelson BW, Low CA, Jacobson N, et al. . Guidelines for wrist-worn consumer wearable assessment of heart rate in biobehavioral research. NPJ Digit Med 2020;3:90. 10.1038/s41746-020-0297-4
    1. Bassett DR, Rowlands A, Trost SG. Calibration and validation of wearable monitors. Med Sci Sports Exerc 2012;44:S32–8. 10.1249/MSS.0b013e3182399cf7
    1. Keadle SK, Lyden KA, Strath SJ, et al. . A Framework to Evaluate Devices That Assess Physical Behavior. Exerc Sport Sci Rev 2019;47:206–14. 10.1249/JES.0000000000000206
    1. Freedson P, Bowles HR, Troiano R, et al. . Assessment of physical activity using wearable monitors: recommendations for monitor calibration and use in the field. Med Sci Sports Exerc 2012;44:S1–4. 10.1249/MSS.0b013e3182399b7e
    1. Welk GJ, McClain J, Ainsworth BE. Protocols for evaluating equivalency of accelerometry-based activity monitors. Med Sci Sports Exerc 2012;44:S39–49. 10.1249/MSS.0b013e3182399d8f
    1. Huang X, Lin J, Demner-Fushman D. Evaluation of PICO as a knowledge representation for clinical questions. AMIA Annu Symp Proc 2006;2006:359–63.
    1. Chatterjee S, Changawala N. Fragmented QRS complex: a novel marker of cardiovascular disease. Clin Cardiol 2010;33:68–71. 10.1002/clc.20709
    1. Inui T, Kohno H, Kawasaki Y, et al. . Use of a Smart Watch for Early Detection of Paroxysmal Atrial Fibrillation: Validation Study. JMIR Cardio 2020;4:e14857. 10.2196/14857
    1. Drexler M, Elsner C, Gabelmann V, et al. . Apple Watch detecting coronary ischaemia during chest pain episodes or an apple a day may keep myocardial infarction away. Eur Heart J 2020;41:2224. 10.1093/eurheartj/ehaa290
    1. Hsiu H, Hsu CL, Wu TL. Effects of different contacting pressure on the transfer function between finger photoplethysmographic and radial blood pressure waveforms. Proc Inst Mech Eng H 2011;225:575–83. 10.1177/0954411910396288
    1. Thomson EA, Nuss K, Comstock A, et al. . Heart rate measures from the Apple Watch, Fitbit Charge HR 2, and electrocardiogram across different exercise intensities. J Sports Sci 2019;37:1411–9. 10.1080/02640414.2018.1560644
    1. Stahl SE, An H-S, Dinkel DM, et al. . How accurate are the wrist-based heart rate monitors during walking and running activities? Are they accurate enough? BMJ Open Sport Exerc Med 2016;2:e000106. 10.1136/bmjsem-2015-000106
    1. Etiwy M, Akhrass Z, Gillinov L, et al. . Accuracy of wearable heart rate monitors in cardiac rehabilitation. Cardiovasc Diagn Ther 2019;9:262–71. 10.21037/cdt.2019.04.08
    1. Georgiou KE, Dimov RK, Boyanov NB, et al. . Feasibility of a New Wearable Device to Estimate Acute Stress in Novices During High-fidelity Surgical Simulation. Folia Med (Plovdiv) 2019;61:49–60. 10.2478/folmed-2019-0001
    1. Hahnen C, Freeman CG, Haldar N, et al. . Accuracy of Vital Signs Measurements by a Smartwatch and a Portable Health Device: Validation Study. JMIR Mhealth Uhealth 2020;8:e16811. 10.2196/16811
    1. Menghini L, Gianfranchi E, Cellini N, et al. . Stressing the accuracy: Wrist-worn wearable sensor validation over different conditions. Psychophysiology 2019;56:e13441. 10.1111/psyp.13441
    1. Dur O, Rhoades C, Ng MS, et al. . Design Rationale and Performance Evaluation of the Wavelet Health Wristband: Benchtop Validation of a Wrist-Worn Physiological Signal Recorder. JMIR Mhealth Uhealth 2018;6:e11040. 10.2196/11040
    1. Coca A, Roberge RJ, Williams WJ, et al. . Physiological monitoring in firefighter ensembles: wearable plethysmographic sensor vest versus standard equipment. J Occup Environ Hyg 2009;7:109–14. 10.1080/15459620903455722
    1. Pasadyn SR, Soudan M, Gillinov M, et al. . Accuracy of commercially available heart rate monitors in athletes: a prospective study. Cardiovasc Diagn Ther 2019;9:379–85. 10.21037/cdt.2019.06.05
    1. Dooley EE, Golaszewski NM, Bartholomew JB. Estimating Accuracy at Exercise Intensities: A Comparative Study of Self-Monitoring Heart Rate and Physical Activity Wearable Devices. JMIR Mhealth Uhealth 2017;5:e34. 10.2196/mhealth.7043
    1. Gillinov S, Etiwy M, Wang R, et al. . Variable Accuracy of Wearable Heart Rate Monitors during Aerobic Exercise. Med Sci Sports Exerc 2017;49:1697–703. 10.1249/MSS.0000000000001284
    1. MJ L, Zhong WH, Liu YX, et al. . Sample Size for Assessing Agreement between Two Methods of Measurement by Bland-Altman Method. Int J Biostat 2016;12.
    1. Carstensen B. Comparing Clinical Measurement Methods: A Practical Guide. Wiley, 2010.
    1. Allen J, Murray A. Age-related changes in peripheral pulse timing characteristics at the ears, fingers and toes. J Hum Hypertens 2002;16:711–7. 10.1038/sj.jhh.1001478
    1. O’Driscoll R, Turicchi J, Hopkins M, et al. . The validity of two widely used commercial and research-grade activity monitors, during resting, household and activity behaviours. Health Technol 2020;10:637–48. 10.1007/s12553-019-00392-7
    1. Shcherbina A, Mattsson C, Waggott D, et al. . Accuracy in Wrist-Worn, Sensor-Based measurements of heart rate and energy expenditure in a diverse cohort. J Pers Med 2017;7:3. 10.3390/jpm7020003
    1. Chow H-W, Yang C-C. Accuracy of Optical Heart Rate Sensing Technology in Wearable Fitness Trackers for Young and Older Adults: Validation and Comparison Study. JMIR Mhealth Uhealth 2020;8:e14707. 10.2196/14707
    1. Dehghanojamahalleh S, Kaya M. Sex-Related Differences in Photoplethysmography Signals Measured From Finger and Toe. IEEE J Transl Eng Health Med 2019;7:1–7. 10.1109/JTEHM.2019.2938506
    1. Firooz A, Rajabi-Estarabadi A, Zartab H, et al. . The influence of gender and age on the thickness and echo-density of skin. Skin Res Technol 2017;23:13–20. 10.1111/srt.12294
    1. Horton JF, Stergiou P, Fung TS, et al. . Comparison of Polar M600 Optical Heart Rate and ECG Heart Rate during Exercise. Med Sci Sports Exerc 2017;49:2600–7. 10.1249/MSS.0000000000001388
    1. Tamura T, Chen W. Seamless Healthcare Monitoring: Advancements in Wearable, Attachable and Invisible Devices. Springer International Publishing, 2018.
    1. Wallen MP, Gomersall SR, Keating SE, et al. . Accuracy of Heart Rate Watches: Implications for Weight Management. PLoS One 2016;11:e0154420. 10.1371/journal.pone.0154420
    1. Claes J, Buys R, Avila A, et al. . Validity of heart rate measurements by the Garmin Forerunner 225 at different walking intensities. J Med Eng Technol 2017;41:480–5. 10.1080/03091902.2017.1333166
    1. Hendrikx J, Ruijs LS, Cox LG, et al. . Clinical Evaluation of the Measurement Performance of the Philips Health Watch: A Within-Person Comparative Study. JMIR Mhealth Uhealth 2017;5:e10. 10.2196/mhealth.6893
    1. Nelson BW, Allen NB. Accuracy of Consumer Wearable Heart Rate Measurement During an Ecologically Valid 24-Hour Period: Intraindividual Validation Study. JMIR Mhealth Uhealth 2019;7:e10828. 10.2196/10828
    1. Hermand E, Cassirame J, Ennequin G, et al. . Validation of a Photoplethysmographic Heart Rate Monitor: Polar OH1. Int J Sports Med 2019;40:462–7. 10.1055/a-0875-4033
    1. Khushhal A, Nichols S, Evans W, et al. . Validity and Reliability of the Apple Watch for Measuring Heart Rate During Exercise. Sports Med Int Open 2017;1:E206–11. 10.1055/s-0043-120195
    1. Spierer DK, Rosen Z, Litman LL, et al. . Validation of photoplethysmography as a method to detect heart rate during rest and exercise. J Med Eng Technol 2015;39:264–71. 10.3109/03091902.2015.1047536
    1. Konstantinou P, Trigeorgi A, Georgiou C, et al. . Comparing apples and oranges or different types of citrus fruits? Using wearable versus stationary devices to analyze psychophysiological data. Psychophysiology 2020;57:e13551. 10.1111/psyp.13551
    1. Brazendale K, Decker L, Hunt ET, et al. . Validity and Wearability of Consumer-based fitness Trackers in free-living children. Int J Exerc Sci 2019;12:471–82.
    1. Pope ZC, Lee JE, Zeng N, et al. . Validation of Four Smartwatches in Energy Expenditure and Heart Rate Assessment During Exergaming. Games Health J 2019;8:205–12. 10.1089/g4h.2018.0087
    1. Müller AM, Wang NX, Yao J, et al. . Heart Rate Measures From Wrist-Worn Activity Trackers in a Laboratory and Free-Living Setting: Validation Study. JMIR Mhealth Uhealth 2019;7:e14120. 10.2196/14120
    1. Sañudo B, De Hoyo M, Muñoz-López A, et al. . Pilot Study Assessing the Influence of Skin Type on the Heart Rate Measurements Obtained by Photoplethysmography with the Apple Watch. J Med Syst 2019;43:195. 10.1007/s10916-019-1325-2
    1. Bayès de Luna A. Clinical Electrocardiography: A Textbook. 4th edn. Wiley-Blackwell, 2012.
    1. Satija U, Ramkumar B, Manikandan MS. A Review of Signal Processing Techniques for Electrocardiogram Signal Quality Assessment. IEEE Rev Biomed Eng 2018;11:36–52. 10.1109/RBME.2018.2810957
    1. Giles D, Draper N, Neil W. Validity of the Polar V800 heart rate monitor to measure RR intervals at rest. Eur J Appl Physiol 2016;116:563–71. 10.1007/s00421-015-3303-9
    1. Tobon DP, Jayaraman S, Falk TH. Spectro-Temporal Electrocardiogram Analysis for Noise-Robust Heart Rate and Heart Rate Variability Measurement. IEEE J Transl Eng Health Med 2017;5:1–11. 10.1109/JTEHM.2017.2767603
    1. Kingsley M, Lewis MJ, Marson RE. Comparison of Polar 810s and an ambulatory ECG system for RR interval measurement during progressive exercise. Int J Sports Med 2005;26:39–44. 10.1055/s-2004-817878
    1. Nunan D, Donovan G, Jakovljevic DG, et al. . Validity and reliability of short-term heart-rate variability from the Polar S810. Med Sci Sports Exerc 2009;41:243–50. 10.1249/MSS.0b013e318184a4b1
    1. Caminal P, Sola F, Gomis P, et al. . Validity of the Polar V800 monitor for measuring heart rate variability in mountain running route conditions. Eur J Appl Physiol 2018;118:669–77. 10.1007/s00421-018-3808-0
    1. Weippert M, Kumar M, Kreuzfeld S, et al. . Comparison of three mobile devices for measuring R-R intervals and heart rate variability: Polar S810i, Suunto t6 and an ambulatory ECG system. Eur J Appl Physiol 2010;109:779–86. 10.1007/s00421-010-1415-9
    1. Wallén MB, Hasson D, Theorell T, et al. . Possibilities and limitations of the Polar RS800 in measuring heart rate variability at rest. Eur J Appl Physiol 2012;112:1153–65. 10.1007/s00421-011-2079-9
    1. force T. Heart rate variability. Standards of measurement, physiological interpretation, and clinical use. Task force of the European Society of Cardiology and the North American Society of Pacing and Electrophysiology. Eur Heart J 1996;17:354–81.
    1. Sassi R, Cerutti S, Lombardi F, et al. . Advances in heart rate variability signal analysis: joint position statement by the e-Cardiology ESC Working Group and the European Heart Rhythm Association co-endorsed by the Asia Pacific Heart Rhythm Society. Europace 2015;17:1341–53. 10.1093/europace/euv015
    1. Bent B, Goldstein BA, Kibbe WA, et al. . Investigating sources of inaccuracy in wearable optical heart rate sensors. NPJ Digit Med 2020;3:18. 10.1038/s41746-020-0226-6
    1. Castaneda D, Esparza A, Ghamari M, et al. . A review on wearable photoplethysmography sensors and their potential future applications in health care. Int J Biosens Bioelectron 2018;4:195–202. 10.15406/ijbsbe.2018.04.00125
    1. Lemay M, Bertschi M, Sola J, et al. . Chapter 2.3 - Application of Optical Heart Rate Monitoring. In: Sazonov E, Neuman MR, eds. Wearable Sensors. Oxford: Academic Press, 2014: 105–29.
    1. Zhang Y, Song S, Vullings R, et al. . Motion Artifact Reduction for Wrist-Worn Photoplethysmograph Sensors Based on Different Wavelengths. Sensors (Basel) 2019;19:673. 10.3390/s19030673
    1. Tăuţan A-M, Young A, Wentink E, et al. . Characterization and reduction of motion artifacts in photoplethysmographic signals from a wrist-worn device. Annu Int Conf IEEE Eng Med Biol Soc 2015;2015:6146-9. 10.1109/EMBC.2015.7319795
    1. Tamura T, Maeda Y, Sekine M, et al. . Wearable Photoplethysmographic Sensors—Past and Present. Electronics 2014;3:282–302. 10.3390/electronics3020282
    1. Passler S, Müller N, Senner V. In-Ear Pulse Rate Measurement: A Valid Alternative to Heart Rate Derived from Electrocardiography? Sensors (Basel) 2019;19:3641. 10.3390/s19173641
    1. Zheng Y, Leung B, Sy S, et al. . A clip-free eyeglasses-based wearable monitoring device for measuring photoplethysmograhic signals. Annu Int Conf IEEE Eng Med Biol Soc 2012;2012:5022–5. 10.1109/EMBC.2012.6347121
    1. Jo E, Lewis K, Directo D, et al. . Validation of Biofeedback Wearables for Photoplethysmographic Heart Rate Tracking. J Sports Sci Med 2016;15:540–7.
    1. Allen J. Photoplethysmography and its application in clinical physiological measurement. Physiol Meas 2007;28:R1–39. 10.1088/0967-3334/28/3/R01
    1. Falter M, Budts W, Goetschalckx K, et al. . Accuracy of Apple Watch Measurements for Heart Rate and Energy Expenditure in Patients With Cardiovascular Disease: Cross-Sectional Study. JMIR Mhealth Uhealth 2019;7:e11889. 10.2196/11889
    1. Parak J, Uuskoski M, Machek J, et al. . Estimating Heart Rate, Energy Expenditure, and Physical Performance With a Wrist Photoplethysmographic Device During Running. JMIR Mhealth Uhealth 2017;5:e97. 10.2196/mhealth.7437
    1. Maeda Y, Sekine M, Tamura T. The advantages of wearable green reflected photoplethysmography. J Med Syst 2011;35:829–34. 10.1007/s10916-010-9506-z
    1. Abt G, Bray J, Benson AC. The validity and inter-device variability of the Apple Watch™ for measuring maximal heart rate. J Sports Sci 2018;36:1447–52. 10.1080/02640414.2017.1397282
    1. Cadmus-Bertram L, Gangnon R, Wirkus EJ, et al. . The Accuracy of Heart Rate Monitoring by Some Wrist-Worn Activity Trackers. Ann Intern Med 2017;166:610–2. 10.7326/L16-0353
    1. Hettiarachchi IT, Hanoun S, Nahavandi D, et al. . Validation of Polar OH1 optical heart rate sensor for moderate and high intensity physical activities. PLoS One 2019;14:e0217288. 10.1371/journal.pone.0217288
    1. Parak J, Korhonen I. Evaluation of wearable consumer heart rate monitors based on photopletysmography. Annu Int Conf IEEE Eng Med Biol Soc 2014;2014:3670–3. 10.1109/EMBC.2014.6944419
    1. Støve MP, Haucke E, Nymann ML, et al. . Accuracy of the wearable activity tracker Garmin Forerunner 235 for the assessment of heart rate during rest and activity. J Sports Sci 2019;37:895–901. 10.1080/02640414.2018.1535563
    1. Bai Y, Hibbing P, Mantis C, et al. . Comparative evaluation of heart rate-based monitors: Apple Watch vs Fitbit Charge HR. J Sports Sci 2018;36:1734–41. 10.1080/02640414.2017.1412235
    1. Reddy RK, Pooni R, Zaharieva DP, et al. . Accuracy of Wrist-Worn Activity Monitors During Common Daily Physical Activities and Types of Structured Exercise: Evaluation Study. JMIR Mhealth Uhealth 2018;6:e10338. 10.2196/10338
    1. Boudreaux BD, Hebert EP, Hollander DB, et al. . Validity of Wearable Activity Monitors during Cycling and Resistance Exercise. Med Sci Sports Exerc 2018;50:624–33. 10.1249/MSS.0000000000001471
    1. McArdle WKF, Katch V. Exercise Physiology: Energy, Nutrition and Human Performance. 8 edn. Lippincott Williams & Wilkins, 2014.
    1. Johnson JM. Physical training and the control of skin blood flow. Med Sci Sports Exerc 1998;30:382–6. 10.1097/00005768-199803000-00007
    1. Gorny AW, Liew SJ, Tan CS, et al. . Fitbit charge HR wireless heart rate monitor: validation study conducted under free-living conditions. JMIR Mhealth Uhealth 2017;5:e157. 10.2196/mhealth.8233
    1. Kroll RR, Boyd JG, Maslove DM. Accuracy of a Wrist-Worn Wearable Device for Monitoring Heart Rates in Hospital Inpatients: A Prospective Observational Study. J Med Internet Res 2016;18:e253. 10.2196/jmir.6025
    1. Pelizzo G, Guddo A, Puglisi A, et al. . Accuracy of a Wrist-Worn Heart Rate Sensing Device during Elective Pediatric Surgical Procedures. Children (Basel) 2018;5:38. 10.3390/children5030038
    1. Rhudy M. Time Alignment Techniques for Experimental Sensor Data. IJCSES 2014;5:1–14. 10.5121/ijcses.2014.5201
    1. Coakley KJ, Hale P. Alignment of Noisy Signals. IEEE Trans Instrum Meas 2001;50:141–9. 10.1109/19.903892
    1. Ghaleb FA, Kamat MB, Salleh M, et al. . Two-stage motion artefact reduction algorithm for electrocardiogram using weighted adaptive noise cancelling and recursive Hampel filter. PLoS One 2018;13:e0207176. 10.1371/journal.pone.0207176
    1. Luo S, Johnston P. A review of electrocardiogram filtering. J Electrocardiol 2010;43:486–96. 10.1016/j.jelectrocard.2010.07.007
    1. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;1:307–10.
    1. Ludbrook J. Statistical techniques for comparing measurers and methods of measurement: a critical review. Clin Exp Pharmacol Physiol 2002;29:527–36. 10.1046/j.1440-1681.2002.03686.x
    1. Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res 1999;8:135–60. 10.1177/096228029900800204
    1. Kottner J, Audigé L, Brorson S, et al. . Guidelines for Reporting Reliability and Agreement Studies (GRRAS) were proposed. J Clin Epidemiol 2011;64:96–106. 10.1016/j.jclinepi.2010.03.002
    1. Ravizza A, De Maria C, Di Pietro L, et al. . Comprehensive Review on Current and Future Regulatory Requirements on Wearable Sensors in Preclinical and Clinical Testing. Front Bioeng Biotechnol 2019;7:313. 10.3389/fbioe.2019.00313
    1. Shah AD, Langenberg C, Rapsomaniki E, et al. . Type 2 diabetes and incidence of cardiovascular diseases: a cohort study in 1·9 million people. Lancet Diabetes Endocrinol 2015;3:105–13. 10.1016/S2213-8587(14)70219-0
    1. Andersson C, Vasan RS. Epidemiology of cardiovascular disease in young individuals. Nat Rev Cardiol 2018;15:230–40. 10.1038/nrcardio.2017.154
    1. Bansilal S, Castellano JM, Fuster V. Global burden of CVD: focus on secondary prevention of cardiovascular disease. Int J Cardiol 2015;201(Suppl 1):S1–7. 10.1016/S0167-5273(15)31026-3
    1. Massoomi MR, Handberg EM. Increasing and Evolving Role of Smart Devices in Modern Medicine. Eur Cardiol 2019;14:181–6. 10.15420/ecr.2019.02
    1. Elgendi M, Fletcher R, Liang Y, et al. . The use of photoplethysmography for assessing hypertension. NPJ Digit Med 2019;2:60. 10.1038/s41746-019-0136-7
    1. Rodin D, Kirby M, Sedogin N, et al. . Comparative accuracy of optical sensor-based wearable system for non-invasive measurement of blood glucose concentration. Clin Biochem 2019;65:15–20. 10.1016/j.clinbiochem.2018.12.014
    1. Vandenberk T, Stans J, Mortelmans C, et al. . Clinical validation of heart rate Apps: mixed-methods evaluation study. JMIR Mhealth Uhealth 2017;5:e129. 10.2196/mhealth.7254
    1. Wang Z, Fu S. Evaluation of a strapless heart rate monitor during simulated flight tasks. J Occup Environ Hyg 2016;13:185–92. 10.1080/15459624.2015.1101121
    1. Chellakumar PJ, Brumfield A, Kunderu K, et al. . Heart rate variability: comparison among devices with different temporal resolutions. Physiol Meas 2005;26:979–86. 10.1088/0967-3334/26/6/008
    1. Engström E, Ottosson E, Wohlfart B, et al. . Comparison of heart rate measured by polar RS400 and ECG, validity and repeatability. Adv Physiother 2012;14:115–22. 10.3109/14038196.2012.694118
    1. Gilgen-Ammann R, Schweizer T, Wyss T. RR interval signal quality of a heart rate monitor and an ECG Holter at rest and during exercise. Eur J Appl Physiol 2019;119:1525–32. 10.1007/s00421-019-04142-5
    1. Romagnoli M, Alis R, Guillen J, et al. . A novel device based on smart textile to control heart's activity during exercise. Australas Phys Eng Sci Med 2014;37:377–84. 10.1007/s13246-014-0271-z

Source: PubMed

3
S'abonner