Contact-Force-Sensing-Based Radiofrequency Catheter Ablation in Paroxysmal Supraventricular Tachycardias (COBRA-PATH): a randomized controlled trial

Tamas Geczy, Nawin L Ramdat Misier, Tamas Szili-Torok, Tamas Geczy, Nawin L Ramdat Misier, Tamas Szili-Torok

Abstract

Background: Multiple studies have demonstrated the importance of adequate catheter-tissue contact in the creation of effective lesions during radiofrequency catheter ablation. The development of contact force (CF)-sensing catheters has contributed significantly to improve clinical outcomes in atrial fibrillation. However, CF-sensing technology is not used in the ablation of paroxysmal supraventricular tachycardia (PSVT). The possible reason for this is that PSVT ablation with the conventional approach (i.e. nonirrigated, non-CF-sensing catheters) is considered a relatively low-risk procedure with fairly high success rates (short and long term). The aim of this study is to determine whether CF sensing can further improve the outcomes of PSVT ablation.

Methods/design: The COBRA-PATH study is a single-center, two-armed, randomized controlled trial. Patients without structural heart disease being referred for electrophysiology study, because of PSVT and potential treatment with radiofrequency (RF) catheter ablation, will be randomly assigned to either manual ablation with standard nonirrigated ablation catheters or manual ablation with an open-irrigated ablation catheter equipped with CF sensing (used in a virtual nonirrigated modus). The primary study endpoint is the difference in the number of RF applications during the ablation of atrioventricular nodal re-entry tachycardia, and that of Wolff-Parkinson-White syndrome and atrioventricular re-entrant tachycardia. Secondary outcome parameters include acute and long-term procedural success rates, overall duration of RF applications, procedure/fluoroscopy durations and safety parameters.

Discussion: We expect to see a reduced number/duration of RF applications required to achieve effective lesion creation, and consequently a decrease in total procedure/fluoroscopy times. Although a significant improvement in procedural success rates (acute/long term) might not be feasible to demonstrate (given the relatively high success rate of the standard ablation method), the possible decrease in procedure duration and the consequential reduction of radiation exposure has important clinical implications for both operators and patients undergoing the procedure.

Trial registration: ClinicalTrials, NCT04078685. Retrospectively registered on 2 September 2019.

Keywords: Ablation; Contact force sensing; Supraventricular tachycardia.

Conflict of interest statement

The authors declare that they have no competing interests.

Figures

Fig. 1
Fig. 1
Contact-Force-Sensing-Based Radiofrequency Catheter Ablation in Paroxysmal Supraventricular Tachycardias (COBRA-PATH) study flow chart. *Symptoms highly suggestive of PSVT: sudden onset of termination of rapid (and regular) palpitations. The palpitations can usually be terminated with the Valsalva maneuver or carotid massage. The patients have no evidence for an underlying structural heart disease. #For a detailed explanation, see “Study population”. $For a detailed explanation, see “Preprocedural preparation and electrophysiology (EP) study”. &This follow-up can be scheduled any time during the 12-month follow-up period, in case the patient experiences recurrent symptoms (symptom-based follow-up). AE adverse event, AVNRT atrioventricular nodal re-entry tachycardia, CF contact force, FU follow-up, I/E inclusion/exclusion, PVST paroxysmal supraventricular tachycardia, SAE serious adverse event, WPW-AVRT Wolff–Parkinson–White syndrome–atrioventricular re-entrant tachycardia

References

    1. Ullah W, Hunter RJ, Haldar S, McLean A, Dhinoja M, Sporton S, et al. Comparison of robotic and manual persistent AF ablation using catheter contact force sensing: an international multicenter registry study. Pacing Clin Electrophysiol. 2014;37(11):1427–1435. doi: 10.1111/pace.12501.
    1. Gerstenfeld EP. Contact force-sensing catheters: evolution or revolution in catheter ablation technology? Circ Arrhythm Electrophysiol. 2014;7(1):5–6. doi: 10.1161/CIRCEP.114.001424.
    1. Shurrab M, Di Biase L, Briceno DF, Kaoutskaia A, Haj-Yahia S, Newman D, et al. Impact of contact force technology on atrial fibrillation ablation: a meta-analysis. J Am Heart Assoc. 2015;4(9):e002476. doi: 10.1161/JAHA.115.002476.
    1. Martinek M, Lemes C, Sigmund E, Derndorfer M, Aichinger J, Winter S, et al. Clinical impact of an open-irrigated radiofrequency catheter with direct force measurement on atrial fibrillation ablation. Pacing Clin Electrophysiol. 2012;35(11):1312–1318. doi: 10.1111/j.1540-8159.2012.03503.x.
    1. Kuck KH, Reddy VY, Schmidt B, Natale A, Neuzil P, Saoudi N, et al. A novel radiofrequency ablation catheter using contact force sensing: Toccata study. Heart Rhythm. 2012;9(1):18–23. doi: 10.1016/j.hrthm.2011.08.021.
    1. Reddy VY, Shah D, Kautzner J, Schmidt B, Saoudi N, Herrera C, et al. The relationship between contact force and clinical outcome during radiofrequency catheter ablation of atrial fibrillation in the TOCCATA study. Heart Rhythm. 2012;9(11):1789–1795. doi: 10.1016/j.hrthm.2012.07.016.
    1. Neuzil P, Reddy VY, Kautzner J, Petru J, Wichterle D, Shah D, et al. Electrical reconnection after pulmonary vein isolation is contingent on contact force during initial treatment: results from the EFFICAS I study. Circ Arrhythm Electrophysiol. 2013;6(2):327–333. doi: 10.1161/CIRCEP.113.000374.
    1. Casella M, Dello Russo A, Russo E, Al-Mohani G, Santangeli P, Riva S, et al. Biomarkers of myocardial injury with different energy sources for atrial fibrillation catheter ablation. Cardiol J. 2014;21(5):516–523. doi: 10.5603/CJ.a2013.0153.
    1. Andrade JG, Monir G, Pollak SJ, Khairy P, Dubuc M, Roy D, et al. Pulmonary vein isolation using “contact force” ablation: the effect on dormant conduction and long-term freedom from recurrent atrial fibrillation—a prospective study. Heart Rhythm. 2014;11(11):1919–1924. doi: 10.1016/j.hrthm.2014.07.033.
    1. Kimura M, Sasaki S, Owada S, Horiuchi D, Sasaki K, Itoh T, et al. Comparison of lesion formation between contact force-guided and non-guided circumferential pulmonary vein isolation: a prospective, randomized study. Heart Rhythm. 2014;11(6):984–991. doi: 10.1016/j.hrthm.2014.03.019.
    1. Marijon E, Fazaa S, Narayanan K, Guy-Moyat B, Bouzeman A, Providencia R, et al. Real-time contact force sensing for pulmonary vein isolation in the setting of paroxysmal atrial fibrillation: procedural and 1-year results. J Cardiovasc Electrophysiol. 2014;25(2):130–137. doi: 10.1111/jce.12303.
    1. Natale A, Reddy VY, Monir G, Wilber DJ, Lindsay BD, McElderry HT, et al. Paroxysmal AF catheter ablation with a contact force sensing catheter: results of the prospective, multicenter SMART-AF trial. J Am Coll Cardiol. 2014;64(7):647–656. doi: 10.1016/j.jacc.2014.04.072.
    1. Sciarra L, Golia P, Natalizia A, De Ruvo E, Dottori S, Scara A, et al. Which is the best catheter to perform atrial fibrillation ablation? A comparison between standard ThermoCool, SmartTouch, and Surround Flow catheters. J Interv Card Electrophysiol. 2014;39(3):193–200.
    1. Wakili R, Clauss S, Schmidt V, Ulbrich M, Hahnefeld A, Schussler F, et al. Impact of real-time contact force and impedance measurement in pulmonary vein isolation procedures for treatment of atrial fibrillation. Clin Res Cardiol. 2014;103(2):97–106. doi: 10.1007/s00392-013-0625-7.
    1. Wutzler A, Huemer M, Parwani AS, Blaschke F, Haverkamp W, Boldt LH. Contact force mapping during catheter ablation for atrial fibrillation: procedural data and one-year follow-up. Arch Med Sci. 2014;10(2):266–272. doi: 10.5114/aoms.2014.42578.
    1. Jarman JW, Panikker S, Das M, Wynn GJ, Ullah W, Kontogeorgis A, et al. Relationship between contact force sensing technology and medium-term outcome of atrial fibrillation ablation: a multicenter study of 600 patients. J Cardiovasc Electrophysiol. 2015;26(4):378–384. doi: 10.1111/jce.12606.
    1. Kautzner J, Neuzil P, Lambert H, Peichl P, Petru J, Cihak R, et al. EFFICAS II: optimization of catheter contact force improves outcome of pulmonary vein isolation for paroxysmal atrial fibrillation. Europace. 2015;17(8):1229–1235. doi: 10.1093/europace/euv057.
    1. Sigmund E, Puererfellner H, Derndorfer M, Kollias G, Winter S, Aichinger J, et al. Optimizing radiofrequency ablation of paroxysmal and persistent atrial fibrillation by direct catheter force measurement—a case-matched comparison in 198 patients. Pacing Clin Electrophysiol. 2015;38(2):201–208. doi: 10.1111/pace.12549.
    1. Reddy VY, Dukkipati SR, Neuzil P, Natale A, Albenque JP, Kautzner J, et al. Randomized, controlled trial of the safety and effectiveness of a contact force-sensing irrigated catheter for ablation of paroxysmal atrial fibrillation: results of the TactiCath Contact Force Ablation Catheter Study for Atrial Fibrillation (TOCCASTAR) Study. Circulation. 2015;132(10):907–915. doi: 10.1161/CIRCULATIONAHA.114.014092.
    1. Katritsis DG, Zografos T, Katritsis GD, Giazitzoglou E, Vachliotis V, Paxinos G, et al. Catheter ablation vs. antiarrhythmic drug therapy in patients with symptomatic atrioventricular nodal re-entrant tachycardia: a randomized, controlled trial. Europace. 2017;19(4):602–6.
    1. Cheng CH, Sanders GD, Hlatky MA, Heidenreich P, McDonald KM, Lee BK, et al. Cost-effectiveness of radiofrequency ablation for supraventricular tachycardia. Ann Intern Med. 2000;133(11):864–876. doi: 10.7326/0003-4819-133-11-200012050-00010.
    1. Goldberg AS, Bathina MN, Mickelsen S, Nawman R, West G, Kusumoto FM. Long-term outcomes on quality-of-life and health care costs in patients with supraventricular tachycardia (radiofrequency catheter ablation versus medical therapy) Am J Cardiol. 2002;89(9):1120–1123. doi: 10.1016/S0002-9149(02)02285-3.
    1. Larson MS, McDonald K, Young C, Sung R, Hlatky MA. Quality of life before and after radiofrequency catheter ablation in patients with drug refractory atrioventricular nodal reentrant tachycardia. Am J Cardiol. 1999;84(4):471–473. doi: 10.1016/S0002-9149(99)00338-0.
    1. Bathina MN, Mickelsen S, Brooks C, Jaramillo J, Hepton T, Kusumoto FM. Radiofrequency catheter ablation versus medical therapy for initial treatment of supraventricular tachycardia and its impact on quality of life and healthcare costs. Am J Cardiol. 1998;82(5):589–593. doi: 10.1016/S0002-9149(98)00416-0.
    1. Kalbfleisch SJ, Calkins H, Langberg JJ, el-Atassi R, Leon A, Borganelli M, et al. Comparison of the cost of radiofrequency catheter modification of the atrioventricular node and medical therapy for drug-refractory atrioventricular node reentrant tachycardia. J Am Coll Cardiol. 1992;19(7):1583–1587. doi: 10.1016/0735-1097(92)90621-S.
    1. Brugada J, Katritsis DG, Arbelo E, Arribas F, Bax JJ, Blomstrom-Lundqvist C, et al. 2019 ESC Guidelines for the management of patients with supraventricular tachycardiaThe Task Force for the management of patients with supraventricular tachycardia of the European Society of Cardiology (ESC) Eur Heart J. 2020;41(5):655–720. doi: 10.1093/eurheartj/ehz467.
    1. Veenhuyzen GD, Quinn FR, Wilton SB, Clegg R, Mitchell LB. Diagnostic pacing maneuvers for supraventricular tachycardia: part 1. Pacing Clin Electrophysiol. 2011;34(6):767–782. doi: 10.1111/j.1540-8159.2011.03076.x.
    1. Veenhuyzen GD, Quinn FR, Wilton SB, Clegg R, Mitchell LB. Diagnostic pacing maneuvers for supraventricular tachycardias: part 2. Pacing Clin Electrophysiol. 2012;35(6):757–769. doi: 10.1111/j.1540-8159.2012.03352.x.
    1. Page RL, Joglar JA, Caldwell MA, Calkins H, Conti JB, Deal BJ, et al. 2015 ACC/AHA/HRS Guideline for the Management of Adult Patients With Supraventricular Tachycardia: A Report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society. J Am Coll Cardiol. 2016;67(13):e27–e115. doi: 10.1016/j.jacc.2015.08.856.
    1. Calkins H, Yong P, Miller JM, Olshansky B, Carlson M, Saul JP, et al. Catheter ablation of accessory pathways, atrioventricular nodal reentrant tachycardia, and the atrioventricular junction: final results of a prospective, multicenter clinical trial. The Atakr Multicenter Investigators Group. Circulation. 1999;99(2):262–270. doi: 10.1161/01.CIR.99.2.262.
    1. Spector P, Reynolds MR, Calkins H, Sondhi M, Xu Y, Martin A, et al. Meta-analysis of ablation of atrial flutter and supraventricular tachycardia. Am J Cardiol. 2009;104(5):671–677. doi: 10.1016/j.amjcard.2009.04.040.

Source: PubMed

3
S'abonner