Intravenous contrast-enhanced cone beam computed tomography (IVCBCT) of intrahepatic tumors and vessels

Cynthia L Eccles, Regina V Tse, Maria A Hawkins, Mark T Lee, Douglas J Moseley, Laura A Dawson, Cynthia L Eccles, Regina V Tse, Maria A Hawkins, Mark T Lee, Douglas J Moseley, Laura A Dawson

Abstract

Purpose: Liver tumors are challenging to visualize on cone beam computed tomography (CBCT) without intravenous (IV) contrast. Image guidance for liver cancer stereotactic body ablative radiation therapy (SABR) could be improved with the direct visualization of hepatic tumors and vasculature. This study investigated the feasibility of the use of IV contrast-enhanced CBCT (IV-CBCT) as a means to improve liver target visualization.

Methods and materials: Patients on a liver SABR protocol underwent IV-CBCT before 1 or more treatment fractions in addition to a noncontrast CBCT. Image acquisition was initiated 0 to 30 seconds following injection and acquired over 60 to 120 seconds. "Stop and go" exhale breath-hold CBCT scans were used whenever feasible. Changes in mean CT number in regions of interest within visible vasculature, tumor, and adjacent liver were quantified between CBCT and IV-CBCT.

Results: Twelve pairs of contrast and noncontrast CBCTs were obtained in 7 patients. Intravenous-CBCT improved hepatic tumor visibility in breath-hold scans only for 3 patients (2 metastases, 1 hepatocellular carcinoma). Visible tumors ranged in volume from 124 to 564 mL. Small tumors in free-breathing patients did not show enhancement on IVCBT.

Conclusions: Intravenous-CBCT may enhance the visibility of hepatic vessels and tumor in CBCT scans obtained during breath hold. Optimization of IV contrast timing and reduction of artifacts to improve tumor visualization warrant further investigation.

Figures

Figure 1
Figure 1
Intravenous cone bean computed tomography (IV-CBCT) images demonstrating tumor enhancement in patient 1. (A) Axial CT slices comparing noncontrast CBCT (left) with simulation CT (middle) and IV-CBCT (right) from a single fraction. Arrows indicate gross tumor volume (top) and gross tumor volume contour (black) from planning CT shown on bottom. Note the artifact from clips in the CBCTs. (B) Sagittal CBCT slices with (left) and without (right) IV contrast demonstrating tumor visibility with contrast.
Figure 2
Figure 2
Intravenous cone bean computed tomography (IV-CBCT) images demonstrating normal tissue. (A) Coronal views of IV-CBCT demonstrating hepatic vasculature (patient 3). (B) Kidney architecture with (left) and without (right) IV contrast (patient 1).

References

    1. Jaffray D.A., Siewerdsen J.H., Wong J.W., Martinez A.A. Flat-panel cone-beam computed tomography for image-guided radiation therapy. Int J Radiat Oncol Biol Phys. 2002;53:1337–1349.
    1. Balter J.M., Dawson L.A., Kazanjian S. Determination of ventilatory liver movement via radiographic evaluation of diaphragm position. Int J Radiat Oncol Biol Phys. 2001;51(1):267–270.
    1. Hawkins M.A., Brock K.K., Eccles C., Moseley D., Jaffray D., Dawson L.A. Assessment of residual error in liver position using kV cone-beam computed tomography for liver cancer high-precision radiation therapy. Int J Radiat Oncol Biol Phys. 2006;66:610–619.
    1. Beddar A.S., Kainz K., Briere T.M. Correlation between internal fiducial tumor motion and external marker motion for liver tumors imaged with 4D-CT. Int J Radiat Oncol Biol Phys. 2007;67:630–638.
    1. Kirilova A., Lockwood G., Choi P. Three-dimensional motion of liver tumors using cine-magnetic resonance imaging. Int J Radiat Oncol Biol Phys. 2008;71:1189–1195.
    1. Baron R.L. Understanding and optimizing use of contrast material for CT of the liver. AJR Am J Roentgenol. 1994;163:323–331.
    1. Eccles C.L., Patel R., Simeonov A.K., Lockwood G., Haider M., Dawson L.A. Comparison of liver tumor motion with and without abdominal compression using cine-magnetic resonance imaging. Int J Radiat Oncol Biol Phys. 2011;79:602–608.
    1. Dawson L.A., Eccles C., Craig T. Individualized image guided iso-NTCP based liver cancer SBRT. Acta Oncol. 2006;45:856–864.
    1. Ning R., Chen B., Yu R., Conover D., Tang X., Ning Y. Flat panel detector-based cone-beam volume CT angiography imaging: System evaluation. IEEE Trans Med Imaging. 2000;19:949–963.
    1. Schafer S., Stayman J.W., Zbijewski W., Schmidgunst C., Kleinszig G., Siewerdsen J.H. Antiscatter grids in mobile C-arm cone-beam CT: Effect on image quality and dose. Med Phys. 2012;39:153–159.
    1. Brock K.K., Hawkins M., Eccles C. Improving image-guided target localization through deformable registration. Acta Oncol. 2008;47:1279–1285.
    1. Case R.B., Moseley D.J., Sonke J.J. Interfraction and intrafraction changes in amplitude of breathing motion in stereotactic liver radiotherapy. Int J Radiat Oncol Biol Phys. 2010;77:918–925.
    1. Dawson L.A., Eccles C., Bissonnette J.P., Brock K.K. Accuracy of daily image guidance for hypofractionated liver radiotherapy with active breathing control. Int J Radiat Oncol Biol Phys. 2005;62:1247–1252.
    1. Eccles C., Brock K.K., Bissonnette J.P., Hawkins M., Dawson L.A. Reproducibility of liver position using active breathing coordinator for liver cancer radiotherapy. Int J Radiat Oncol Biol Phys. 2006;64:751–759.
    1. Eccles C.L., Dawson L.A., Moseley J.L., Brock K.K. Interfraction liver shape variability and impact on GTV position during liver stereotactic radiotherapy using abdominal compression. Int J Radiat Oncol Biol Phys. 2011;80:938–946.
    1. Velec M., Moseley J.L., Eccles C.L. Effect of breathing motion on radiotherapy dose accumulation in the abdomen using deformable registration. Int J Radiat Oncol Biol Phys. 2011;80:265–272.
    1. Beddar A.S., Briere T.M., Balter P. 4D-CT imaging with synchronized intravenous contrast injection to improve delineation of liver tumors for treatment planning. Radiother Oncol. 2008;87:445–448.
    1. Mancosu P., Bettinardi V., Passoni P. Contrast enhanced 4D-CT imaging for target volume definition in pancreatic ductal adenocarcinoma. Radiother Oncol. 2008;87:339–342.
    1. Rodal J., Sovik S., Skogmo H.K., Knudtsen I.S., Malinen E. Feasibility of contrast-enhanced cone-beam CT for target localization and treatment monitoring. Radiother Oncol. 2010;97:521–524.
    1. Seppenwoolde Y., Wunderink W., Wunderink-van Veen S.R., Storchi P., Mendez Romero A., Heijmen B.J. Treatment precision of image-guided liver SBRT using implanted fiducial markers depends on marker-tumour distance. Phys Med Biol. 2011;56:5445–5468.
    1. Wurm R.E., Gum F., Erbel S. Image guided respiratory gated hypofractionated stereotactic body radiation therapy (H-SBRT) for liver and lung tumors: Initial experience. Acta Oncol. 2006;45:881–889.
    1. Patel N.V., Gounis M.J., Wakhloo A.K. Contrast-enhanced angiographic cone-beam CT of cerebrovascular stents: Experimental optimization and clinical application. AJNR Am J Neuroradiol. 2011;32:137–144.
    1. Miyayama S., Matsui O., Yamashiro M. Detection of hepatocellular carcinoma by CT during arterial portography using a cone-beam CT technology: Comparison with conventional CTAP. Abdom Imaging. 2009;34:502–506.
    1. Hirota S., Nakao N., Yamamoto S. Cone-beam CT with flat-panel-detector digital angiography system: Early experience in abdominal interventional procedures. Cardiovasc Intervent Radiol. 2006;29:1034–1038.
    1. Ishikura R., Ando K., Nagami Y. Evaluation of vascular supply with cone-beam computed tomography during intraarterial chemotherapy for a skull base tumor. Radiat Med. 2006;24:384–387.
    1. Coolens C., Driscoll B., Dawson L. Feasibility study of 4D perfusion CT for hepatocellular carcinoma patients treated with radiation and sorafenib. Practi Radiat Oncol. 2013;3(2 Suppl 1):S29.
    1. Rankine A.W., Lanzon P.J., Spry N.A. Effect of contrast media on megavoltage photon beam dosimetry. Med Dosim. 2008;33:169–174.
    1. Nasrollah J., Mikaeil M., Omid E., Mojtaba S.S., Ahad Z. Influence of the intravenous contrast media on treatment planning dose calculations of lower esophageal and rectal cancers. J Cancer Res Therapeutics. 2014;10:147–152.
    1. Letourneau D., Finlay M., O'Sullivan B. Lack of influence of intravenous contrast on head and neck IMRT dose distributions. Acta Oncol. 2008;47:90–94.
    1. Nakagawa K., Yamashita H., Igaki H., Terahara A., Shiraishi K., Yoda K. Contrast medium-assisted stereotactic image-guided radiotherapy using kilovoltage cone-beam computed tomography. Radiat Med. 2008;26:570–572.
    1. Sonke J.J., Zijp L., Remeijer P., van Herk M. Respiratory correlated cone beam CT. Med Phys. 2005;32:1176–1186.
    1. Case R.B., Sonke J.J., Moseley D.J., Kim J., Brock K.K., Dawson L.A. Inter- and intrafraction variability in liver position in non-breath-hold stereotactic body radiotherapy. Int J Radiat Oncol Biol Phys. 2009;75:302–308.
    1. Lagendijk J.J., Raaymakers B.W., Raaijmakers A.J. MRI/linac integration. Radiother Oncol. 2008;86:25–29.

Source: PubMed

3
S'abonner