Validation of the pulse decomposition analysis algorithm using central arterial blood pressure

Martin C Baruch, Kambiz Kalantari, David W Gerdt, Charles M Adkins, Martin C Baruch, Kambiz Kalantari, David W Gerdt, Charles M Adkins

Abstract

Background: There is a significant need for continuous noninvasive blood pressure (cNIBP) monitoring, especially for anesthetized surgery and ICU recovery. cNIBP systems could lower costs and expand the use of continuous blood pressure monitoring, lowering risk and improving outcomes.The test system examined here is the CareTaker® and a pulse contour analysis algorithm, Pulse Decomposition Analysis (PDA). PDA's premise is that the peripheral arterial pressure pulse is a superposition of five individual component pressure pulses that are due to the left ventricular ejection and reflections and re-reflections from only two reflection sites within the central arteries.The hypothesis examined here is that the model's principal parameters P2P1 and T13 can be correlated with, respectively, systolic and pulse pressures.

Methods: Central arterial blood pressures of patients (38 m/25 f, mean age: 62.7 y, SD: 11.5 y, mean height: 172.3 cm, SD: 9.7 cm, mean weight: 86.8 kg, SD: 20.1 kg) undergoing cardiac catheterization were monitored using central line catheters while the PDA parameters were extracted from the arterial pulse signal obtained non-invasively using CareTaker system.

Results: Qualitative validation of the model was achieved with the direct observation of the five component pressure pulses in the central arteries using central line catheters. Statistically significant correlations between P2P1 and systole and T13 and pulse pressure were established (systole: R square: 0.92 (p < 0.0001), diastole: R square: 0.78 (p < 0.0001). Bland-Altman comparisons between blood pressures obtained through the conversion of PDA parameters to blood pressures of non-invasively obtained pulse signatures with catheter-obtained blood pressures fell within the trend guidelines of the Association for the Advancement of Medical Instrumentation SP-10 standard (standard deviation: 8 mmHg(systole: 5.87 mmHg, diastole: 5.69 mmHg)).

Conclusions: The results indicate that arterial blood pressure can be accurately measured and tracked noninvasively and continuously using the CareTaker system and the PDA algorithm. The results further support the physical model that all of the features of the pressure pulse envelope, whether in the central arteries or in the arterial periphery, can be explained by the interaction of the left ventricular ejection pressure pulse with two centrally located reflection sites.

Figures

Figure 1
Figure 1
Sketch of the aorta/arm complex arterial system and its effect on the arterial pressure pulse line shape that is observed at the radial/digital artery. Two reflection sites, one at the height of the renal arteries, the other one in the vicinity of the iliac bifurcation, give rise to the reflected pulse (gray) that trail the primary left ventricular ejection (black).
Figure 2
Figure 2
Different pressure pulse spectra of a 21-old male athlete. (A) Arterial pressure pulse obtained from digital integration of signal displayed in (B): derivative signal of arterial pressure pulse obtained with CareTaker physiological monitor (C); second derivative of arterial pressure pulse, obtained through differentiation of signal displayed in (B). Component pulses are identified with, respective, P1, P2, P3. Significance of arrows is explained in the text.
Figure 3
Figure 3
Different pressure pulse spectra of a 65-old male at two different blood pressures. (A) Arterial pressure pulses obtained from digital integration of signal displayed in (B): derivative signals of arterial pressure pulse obtained with CareTaker physiological monitor (C); second derivatives of arterial pressure pulse, obtained through differentiation of signal displayed in (B). Note the merging of P2 and P1 in the pulse envelopes (A) as well as the increase in the shoulder between data points 100 and 150 in the derivative signals (B). Significance of red arrow is explained in the text.
Figure 4
Figure 4
Overlap of raw and derived signals for patient 31. (A) Printout of the central line blood pressure data; (B) derivative signal of arterial pressure pulse obtained with CareTaker physiological monitor; (C) arterial pressure pulse obtained from digital integration of signal displayed in (B); (D) overlay of A-line systole, A-line pulse pressure, P2P1 ratio and T13 time interval.
Figure 5
Figure 5
Overlap of the inter-beat interval series obtained from central line catheter (red) and data obtained with the CareTaker physiological monitor (black) for patient 38.
Figure 6
Figure 6
Overlap of central systolic pressure (red) obtained from catheter signal and P2P1 ratio obtained from PDA analysis of non-invasively obtained arterial signal (black) for patient 38.
Figure 7
Figure 7
Overlap of central pulse pressure (red) obtained from catheter signal and T13 delay time obtained from PDA analysis of non-invasively obtained arterial signal (black) for patient 38.
Figure 8
Figure 8
Correlation of P2P1 parameter with central line systolic pressure, as shown in Figure 5, for patient 38.
Figure 9
Figure 9
Correlation of T13 parameter with central line pulse pressure, as shown if Figure 7, for patient 38.
Figure 10
Figure 10
Overall correlation of systolic blood pressures obtained through conversion of PDA parameters from non-invasively obtained arterial pulse signal, and central systole.
Figure 11
Figure 11
Overall correlation of diastolic blood pressures obtained through conversion of PDA parameters from non-invasively obtained arterial pulse signal, and central diastole.
Figure 12
Figure 12
Bland-Altman comparison of systolic pressure difference versus systolic pressure average. Standard deviation: 5.87 mmHg.
Figure 13
Figure 13
Bland-Altman comparison of diastolic pressure difference versus diastolic pressure average. Standard deviation: 5.69 mmHg.
Figure 14
Figure 14
Histogram distribution of the slope parameters of the linear conversions from P2P1 to systole. Blue marker is described in the text.
Figure 15
Figure 15
Histogram distribution of the slope parameters of the linear conversions from T13 to pulse pressure. Blue marker is described in the text.
Figure 16
Figure 16
Overlap of the central pressure pulse profiles collected with a catheter in a patient with left ventricular failure (#2) and normal heart function (#6). Due to the shortened LVET the individual component pulses are clearly resolved in the case of patient #2.
Figure 17
Figure 17
Central catheter pressure pulse profile for patient 38, where all of the component pulses (P1 through P5) are clearly resolved. Red vertical bars indicate positions of, respectively, P3, P4, P5.

References

    1. Martina JR(1), Westerhof BE, van Goudoever J, de Beaumont EM, Truijen J, Kim YS, Immink RV, Jöbsis DA, Hollmann MW, Lahpor JR, de Mol BA, van Lieshout JJ. Noninvasive continuous arterial blood pressure monitoring with Nexfin®. Anesthesiology. 2012;116(5):1092–1103.
    1. Imholz BP(1), Wieling W, van Montfrans GA, Wesseling KH. Fifteen years’ experience with finger arterial pressure monitoring: assessment of the technology. Cardiovasc Res. 1998;38(3):605–616.
    1. Hager H(1), Mandadi G, Pulley D, Eagon JC, Mascha E, Nutter B, Kurz A. A comparison of noninvasive blood pressure measurement on the wrist with invasive arterial blood pressure monitoring in patients undergoing bariatric surgery. Obes Surg. 2009;19(6):717–724.
    1. Meidert AS(1), Huber W, Müller JN, Schöfthaler M, Hapfelmeier A, Langwieser N, Wagner JY, Eyer F, Schmid RM, Saugel B. Radial artery applanation tonometry for continuous non-invasive arterial pressure monitoring in intensive care unit patients: comparison with invasively assessed radial arterial pressure. Br J Anaesth. 2014;112(3):521–528.
    1. Zorn EA(1), Wilson MB, Angel JJ, Zanella J, Alpert BS. Validation of an automated arterial tonometry monitor using Association for the Advancement of Medical Instrumentation standards. Blood Press Monit. 1997;2(4):185–188.
    1. Ott C, Haetinger S, Schneider MP, Pauschinger M, Schmieder RE. Comparison of two noninvasive devices for measurement of central systolic blood pressure with invasive measurement during cardiac catheterization. J Clin Hypertens (Greenwich) 2012;14(9):575–579.
    1. Dewhirst E(1), Corridore M, Klamar J, Beebe A, Rice J, Barry N, Tobias JD. Accuracy of the CNAP monitor, a noninvasive continuous blood pressure device, in providing beat-to-beat blood pressure readings in the prone position. J Clin Anesth. 2013;25(4):309–313.
    1. Agarwal R, Light RP. The effect of measuring ambulatory blood pressure on nighttime sleep and daytime activity–implications for dipping. Clin J Am Soc Nephrol. 2010;5(2):281–285.
    1. Dolan E(1), Stanton A, Thijs L, Hinedi K, Atkins N, McClory S, Den Hond E, McCormack P, Staessen JA, O’Brien E. Superiority of ambulatory over clinic blood pressure measurement in predicting mortality: the Dublin outcome study. Hypertension. 2005;46(1):156–161.
    1. Kim BK, Kim YM, Lee Y, Lim YH, Shin J. A reverse dipping pattern predicts cardiovascular mortality in a clinical cohort. J Korean Med Sci. 2013;28(10):1468–1473. doi: 10.3346/jkms.2013.28.10.1468. Epub 2013 Sep 25.
    1. Egan BM, Zhao Y, Axon RN. US trends in prevalence, awareness, treatment, and control of hypertension, 1988–2008. JAMA. 2010;303(20):2043.
    1. Williams B, Lacy PS, Yan P, Hwee CN, Liang C, Ting CM. Development and validation of a novel method to derive central aortic systolic pressure from the radial pressure waveform using an N-point moving average method. J Am Coll Cardiol. 2011;57(8):951–961.
    1. Adji A, O'Rourke MF. Determination of central aortic systolic and pulse pressure from the radial artery pressure waveform. Blood Press Monit. 2004;9(3):115–21.
    1. Miyashita H. Clinical Assessment of Central Blood Pressure. Curr Hypertens Rev. 2012;8(2):80–90.
    1. Takazawa K, Kobayashi H, Kojima I, Aizawa A, Kinoh M, Sugo Y, Shimizu M, Miyawaki Y, Tanaka N, Yamashina A, Avolio A. Estimation of central aortic systolic pressure using late systolic inflection of radial artery pulse and its application to vasodilator therapy. J Hypertens. 2012;30(5):908–916.
    1. Norton GR, Majane OH, Maseko MJ, Libhaber C, Redelinghuys M, Kruger D, Veller M, Sareli P, Woodiwiss AJ. Brachial blood pressure-independent relations between radial late systolic shoulder-derived aortic pressures and target organ changes. Hypertension. 2012;59(4):885–892.
    1. Baruch MC, Warburton DE, Bredin SS, Cote A, Gerdt DW, Adkins CM. Pulse Decomposition Analysis of the digital arterial pulse during hemorrhage simulation. Nonlinear Biomed Phys. 2011;5(1):1.
    1. Kríz J, Seba P. Force plate monitoring of human hemodynamics. Nonlinear Biomed Phys. 2008;2(1):1.
    1. Latham RD, Westerhof N, Sipkema P, Rubal BJ, Reuderink P, Murgo JP. Regional wave travel and reflections along the human aorta: a study with six simultaneous micromanometric pressures. Circulation. 1985;72:1257–1269.
    1. Nichols WW, O’Rourke MF. McDonald’s blood flow in arteries. Theoretical, experimental and clinical principles. London: Edward Arnold; 1999.
    1. Kriz J, Martinik K, Seba P, Tosnerova V. Force Plate Measurement of Human Hemodynamics. 2006. .
    1. Nichols WW, O’Rourke MF. McDonald’s blood flow in arteries. Theoretical, experimental and clinical principles. London: Edward Arnold; 1999. p. 469, figure 22.27 A.
    1. Nelson MR, Stepanek J, Cevette M, Covalciuc M, Hurst RT, Tajik AJ. Noninvasive measurement of central vascular pressures with arterial tonometry: clinical revival of the pulse pressure waveform? Mayo Clin Proc. 2010;85(5):460–472.
    1. O’Rourke MF, Yaginuma T. Wave reflections and the arterial pulse. Arch Intern Med. 1984;144(2):366–371.
    1. Greenwald SE, Carter AC, Berry CL. Effect of age on the in vitro reflection coefficient of the aortoiliac bifurcation in humans. Circulation. 1990;82(1):114–123.
    1. Anliker M, Histand MB, Ogden E. Dispersion and Attenuation of Small Artificial Pressure Waves in the Canine Aorta. Circ Res. 1968;23:539–551.