- ICH GCP
- Registro degli studi clinici negli Stati Uniti
- Sperimentazione clinica NCT07802119
Transjugular Approach for Implantation of Active Fixation Leadless Pacemakers
Transjugular Approach for Implantation of Active Fixation Leadless Pacemakers: A Feasibility Study
Panoramica dello studio
Stato
Intervento / Trattamento
Descrizione dettagliata
Background: Leadless pacemakers have emerged as an important alternative to conventional transvenous pacing systems, offering advantages such as reduced pocket-related complications and improved patient tolerance. The Aveir VR (Abbott) is an active-fixation leadless ventricular pacemaker with a retractable helix and retrievable design. While the femoral approach is standard, some patients have contraindications to femoral access (e.g., venous occlusion, severe obesity). The internal jugular vein approach offers a potential alternative, but clinical data-particularly from China-remain limited.
Objectives: To evaluate the feasibility and acute (intraoperative to 48h) safety of transjugular implantation of Aveir VR, and to assess short-term (1-3 months) pacing parameter stability and safety in a Chinese population.
Study Design: Prospective, single-arm, single-center feasibility study. Fifteen patients aged 18-85 years with symptomatic bradycardia requiring ventricular pacing will be enrolled. The procedure will be performed via the right internal jugular vein (or external jugular if needed) under local anesthesia. Pacing parameters (threshold, R-wave amplitude, impedance) will be tested intraoperatively and at 1-month and 3-month follow-up. Implant location (right ventricular septal vs. free wall) will be assessed by echocardiography or cardiac CT before discharge. Safety will be evaluated through adverse event monitoring from consent through 3 months post-procedure. Descriptive statistics will be used; no formal hypothesis testing will be performed given the exploratory feasibility nature of the study.
Tipo di studio
Iscrizione (Stimato)
Fase
- Non applicabile
Contatti e Sedi
Contatto studio
- Nome: Min Shen, MD, PhD
- Numero di telefono: +86-18099773953
- Email: 261726782@qq.com
Luoghi di studio
-
-
Shaanxi
-
Xi'an, Shaanxi, Cina, 710032
- Xijing Hospital, The First Affiliated Hospital of Air Force Medical University
-
Contatto:
- Min Shen, MD, PhD
- Numero di telefono: +86-18099773953
- Email: 261726782@qq.com
-
-
Criteri di partecipazione
Criteri di ammissibilità
Età idonea allo studio
- Adulto
- Adulto più anziano
Accetta volontari sani
Descrizione
Inclusion Criteria:
- Aged 18 to 85 years, regardless of gender.
Symptomatic bradyarrhythmia meeting the ventricular pacing indications recommended by the 2023 ACC/AHA/HRS Guideline for the Evaluation and Management of Bradyarrhythmias, the 2020 Chinese Expert Consensus on the Evaluation and Management of Patients with Bradycardia and Conduction Disorders, and the 2022 Chinese Expert Consensus on the Clinical Application of Leadless Pacemakers, including any of the following conditions:
- Third-degree or advanced atrioventricular (AV) block;
- Second-degree type II AV block, alternating bundle branch block, or bifascicular/trifascicular block with intermittent advanced/third-degree AV block;
- Adult congenital complete AV block; persistent irreversible advanced/third-degree AV block following myocardial infarction or cardiac surgery;
- Symptomatic bradycardia (e.g., dizziness, presyncope, syncope, fatigue, decreased exercise tolerance) or sick sinus syndrome (including chronotropic incompetence and tachy-brady syndrome);
- Drug-induced persistent symptomatic bradycardia with no alternative treatment options;
- Atrial fibrillation with slow ventricular response requiring long-term ventricular pacing;
- New-onset persistent complete left bundle branch block after TAVR, or new-onset advanced AV block after hypertrophic cardiomyopathy ablation;
- Neuromuscular disease, infiltrative cardiomyopathy, or adult congenital heart disease with progressive AV/branch conduction abnormalities and bradycardia-related symptoms;
- High infection risk populations (e.g., recurrent pacing system infection, diabetes mellitus, long-term corticosteroid use, end-stage renal disease on dialysis) or patients with abnormal conventional pacing access or at extremely high risk of conventional pacing lead complications.
Anatomical feasibility and vascular access assessment:
- Preoperative imaging confirms suitable right internal jugular vein access to accommodate the delivery sheath, without thrombosis, severe stenosis, or anatomical malformations;
- Right ventricular anatomy is suitable with adequate trabeculation to support active fixation of the leadless pacemaker, without severe structural abnormalities;
- Priority for enrollment: patients with limited femoral or inferior vena cava access (e.g., iliac/inferior vena cava stenosis, thrombosis, prior venous surgery, obesity precluding femoral access).
- New York Heart Association (NYHA) functional class ≤ III, and expected survival > 1 year.
- Able to cooperate with preoperative evaluation, intraoperative procedure, and 3-month follow-up.
- Voluntarily signs written informed consent and agrees to complete all scheduled observations per the study protocol.
Exclusion Criteria:
- Anatomical malformations, severe stenosis, occlusion, or thrombosis of the internal jugular vein or superior vena cava that preclude transjugular access for the procedure.
- Abnormal right ventricular anatomy, including but not limited to right ventricular hypoplasia, severe right ventricular dilation, or severe right ventricular myocardial fibrosis, preventing safe implantation of an active-fixation leadless ventricular pacemaker.
- Severe tricuspid regurgitation, or status post-mechanical tricuspid valve replacement, significantly affecting pacemaker implantation, device fixation, or function.
- NYHA functional class IV, or refractory heart failure, cardiogenic shock, rendering the patient unable to tolerate the procedure and perioperative management.
- Severe hepatic or renal impairment (eGFR < 30 mL/min/1.73m², severe liver dysfunction), coagulopathy, or active bleeding disorders constituting clear surgical contraindications.
- Confirmed allergy or severe intolerance to leadless pacemaker materials, intraoperative contrast media, or anesthetic agents.
- Prior implantation of other cardiovascular implantable electronic devices (e.g., pacemaker, ICD, CRT) or retained intracardiac leads/devices that would interfere with the current implantation or device function.
- Advanced malignancy, severe active infection, active autoimmune disease, or other serious systemic conditions with an expected survival < 12 months, precluding completion of the 3-month follow-up.
- Pregnant or lactating women, or women planning pregnancy during the follow-up period.
- Psychiatric disorders, cognitive impairment, or inability to cooperate with preoperative evaluation, intraoperative procedure, or regular postoperative follow-up due to personal or geographic reasons.
- Concurrent participation in another cardiovascular device clinical trial that may interfere with the outcome assessment of this study.
- Any other condition deemed by the investigator to make the participant unsuitable for enrollment.
Piano di studio
Come è strutturato lo studio?
Dettagli di progettazione
- Scopo principale: Trattamento
- Assegnazione: N / A
- Modello interventistico: Assegnazione di gruppo singolo
- Mascheramento: Nessuno (etichetta aperta)
Armi e interventi
Gruppo di partecipanti / Arm |
Intervento / Trattamento |
|---|---|
|
Sperimentale: Experimental: Aveir VR Leadless Pacemaker
Participants will undergo transjugular implantation of the Aveir VR active-fixation leadless ventricular pacemaker.
This experimental arm evaluates the procedure and device performance for ventricular pacing indications.
|
The Aveir VR leadless pacemaker is implanted via a transjugular (jugular vein) approach.
This active-fixation device is placed in the right ventricle for ventricular pacing indications.
|
Cosa sta misurando lo studio?
Misure di risultato primarie
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
|---|---|---|
|
Rate of Participants with Sustained Successful Aveir VR Lead Implantation (evaluated by fluoroscopy and clinical-electrophysiological follow-up)
Lasso di tempo: Through 3 months post-procedure
|
The percentage of participants with sustained successful Aveir VR lead implantation via trans-jugular venous access at 3-month follow-up.
Sustained implantation success is defined as intact lead position, stable pacing parameters, and absence of lead-related complications at the end of 3-month follow-up.
|
Through 3 months post-procedure
|
|
Acute adverse event rate
Lasso di tempo: Periprocedural and up to 48 hours
|
Incidence of adverse events within 48 hours post-procedure, measured as a percentage
|
Periprocedural and up to 48 hours
|
Misure di risultato secondarie
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
|---|---|---|
|
Pacing Threshold (measured in volts)
Lasso di tempo: Baseline, Post-procedure Day 1-3, 1 month, and 3 months post-procedure
|
Pacing threshold of the Aveir VR lead, measured in volts, to evaluate threshold stability across follow-up time points.
|
Baseline, Post-procedure Day 1-3, 1 month, and 3 months post-procedure
|
|
R-wave amplitude
Lasso di tempo: Baseline, Post-procedure Day 1-3, 1 month, and 3 months post-procedure
|
R-wave amplitude measured in millivolts.
|
Baseline, Post-procedure Day 1-3, 1 month, and 3 months post-procedure
|
|
Lead impedance
Lasso di tempo: Baseline, Post-procedure Day 1-3, 1 month, and 3 months post-procedure
|
Lead impedance measured in ohms.
|
Baseline, Post-procedure Day 1-3, 1 month, and 3 months post-procedure
|
|
Technical success rate
Lasso di tempo: Periprocedural
|
Rate of technical success measured as a percentage.
|
Periprocedural
|
|
Duration of the Implantation Procedure (measured in minutes)
Lasso di tempo: Periprocedural
|
Total time duration of the leadless pacemaker implantation procedure, measured in minutes.
|
Periprocedural
|
|
Fluoroscopy time
Lasso di tempo: Periprocedural
|
Duration of fluoroscopy measured in minutes.
|
Periprocedural
|
|
Number of puncture attempts
Lasso di tempo: Periprocedural
|
Count of puncture attempts measured in number of times.
|
Periprocedural
|
|
Number of lead fixation attempts
Lasso di tempo: Periprocedural
|
Count of fixation attempts measured in number of times.
|
Periprocedural
|
|
Implantation site
Lasso di tempo: Periprocedural
|
Anatomical location of implantation recorded as a categorical variable.
|
Periprocedural
|
|
Correlation between site and parameters
Lasso di tempo: Up to 3 months
|
Pacing parameters by site
|
Up to 3 months
|
|
Adverse event incidence
Lasso di tempo: Through 3 months post-procedure
|
Overall incidence of adverse events measured as a percentage.
|
Through 3 months post-procedure
|
|
Serious adverse event incidence
Lasso di tempo: Through 3 months post-procedure
|
Incidence of serious adverse events measured as a percentage.
|
Through 3 months post-procedure
|
|
Device-related complication rate
Lasso di tempo: Through 3 months post-procedure
|
Rate of device-related complications measured as a percentage.
|
Through 3 months post-procedure
|
Collaboratori e investigatori
Sponsor
Investigatori
- Investigatore principale: Min Shen, MD, PhD, Department of Cardiovascular Medicine, First Affiliated Hospital of Air Force Medical University
Pubblicazioni e link utili
Pubblicazioni generali
- Buchanan EB, Suresh SP, Waseem N, Paul G, Yoo D. Initial experience utilizing the right internal jugular vein access for implantation of retrievable leadless pacemakers: safety, efficacy, and efficiency. J Interv Card Electrophysiol. 2026 Jan;69(1):143-148. doi: 10.1007/s10840-025-02135-3. Epub 2025 Oct 16.
- Saleem-Talib S, van Driel VJ, Nikolic T, van Wessel H, Louman H, Borleffs CJW, van der Heijden J, Cox M, Ramanna H. The jugular approach for leadless pacing: A novel and safe alternative. Pacing Clin Electrophysiol. 2022 Oct;45(10):1248-1254. doi: 10.1111/pace.14587. Epub 2022 Sep 12.
Studiare le date dei record
Studia le date principali
Inizio studio (Stimato)
Completamento primario (Stimato)
Completamento dello studio (Stimato)
Date di iscrizione allo studio
Primo inviato
Primo inviato che soddisfa i criteri di controllo qualità
Primo Inserito (Effettivo)
Aggiornamenti dei record di studio
Ultimo aggiornamento pubblicato (Effettivo)
Ultimo aggiornamento inviato che soddisfa i criteri QC
Ultimo verificato
Maggiori informazioni
Termini relativi a questo studio
Parole chiave
Termini MeSH pertinenti aggiuntivi
Altri numeri di identificazione dello studio
- KY20262201-F-1
Piano per i dati dei singoli partecipanti (IPD)
Hai intenzione di condividere i dati dei singoli partecipanti (IPD)?
Descrizione del piano IPD
Informazioni su farmaci e dispositivi, documenti di studio
Studia un prodotto farmaceutico regolamentato dalla FDA degli Stati Uniti
Studia un dispositivo regolamentato dalla FDA degli Stati Uniti
prodotto fabbricato ed esportato dagli Stati Uniti
Queste informazioni sono state recuperate direttamente dal sito web clinicaltrials.gov senza alcuna modifica. In caso di richieste di modifica, rimozione o aggiornamento dei dettagli dello studio, contattare register@clinicaltrials.gov. Non appena verrà implementata una modifica su clinicaltrials.gov, questa verrà aggiornata automaticamente anche sul nostro sito web .