이 페이지는 자동 번역되었으며 번역의 정확성을 보장하지 않습니다. 참조하십시오 영문판 원본 텍스트의 경우.

Simultaneous Measurement and Responsive Treatment - Part 2 (SMART02)

2026년 6월 12일 업데이트: ClinSurge Research

This research study is testing an investigational dual-port insulin patch pump that integrates a continuous glucose monitor (CGM) in adults with type 1 diabetes. The goal of the study is to better understand how insulin delivery near a CGM sensor affects glucose readings and to collect data to support development of a combined insulin pump and CGM system.

People with type 1 diabetes require lifelong insulin therapy. Many use insulin pumps and CGMs, but these systems usually involve wearing multiple devices at different body sites. Managing several devices can increase treatment burden and may contribute to skin irritation, device failures, and challenges with glucose control.

This study is conducted in two in-patient parts. In Part A, participants will wear three investigational devices at the same time while glucose levels are closely monitored using laboratory blood tests and a commercial CGM. This part of the study is designed to measure how basal and bolus insulin delivery near the CGM sensor affects sensor accuracy and how quickly the sensor signal recovers after insulin delivery.

In Part B, participants will wear one investigational device while trained study staff use CGM information from the integrated sensor to guide insulin delivery recommendations generated by an automated glucose control algorithm. Insulin delivery decisions will be closely supervised, and glucose levels will be frequently monitored.

Participants will stay at the clinical research center for short, controlled study visits. Safety will be monitored throughout the study, with predefined procedures for treating low or high blood sugar. The information collected will be used to support further development of an integrated insulin pump and CGM system for people with type 1 diabetes.

연구 개요

상세 설명

Current insulin pump and continuous glucose monitoring systems have improved glucose management for people with type 1 diabetes but typically require multiple devices worn at different body sites. This increases treatment burden and may contribute to device-related complications, including skin irritation, site failures, and challenges with coordination between devices. Integrating insulin delivery and glucose sensing into a single system may reduce these burdens and improve usability, but placing insulin infusion and glucose sensing close together may affect CGM accuracy.

This prospective, single-arm, early feasibility study is designed to evaluate an investigational Dual Port Pump (DPP) System, which integrates an insulin infusion patch pump with a CGM sensor positioned near the insulin delivery site. The objectives of the study are to characterize the effects of insulin bolus and basal infusion on CGM sensor readings, assess glucose sensing accuracy, and collect data to support development and optimization of a sensor recovery monitoring algorithm. In Part B, the study also evaluates the feasibility and performance of an automated glycaemic control algorithm used with the integrated system.

The study is conducted in two sequential in-patient parts and enrolls adults with type 1 diabetes.

Part A (Sensor Characterization) consists of a 1.5-day in-patient admission during which participants will wear three DPP systems simultaneously. One system will deliver varying basal insulin rates, while the other two systems will deliver predefined insulin boluses. Glucose levels will be intensively monitored using reference plasma glucose measurements obtained with a Yellow Springs Instruments (YSI) glucose analyzer, along with data from a commercially available CGM. Data from the integrated CGM sensor will not be used for therapy decisions in Part A.

Part B (Automated Glycaemic Control Feasibility) consists of a 2.5-day in-patient admission during which participants will wear one DPP system. CGM values from the integrated sensor will be manually entered by trained study staff into an automated glycaemic control algorithm at regular intervals. Insulin dose recommendations generated by the algorithm will be manually applied using the pump application. Participants will consume standardized meals and complete standardized exercise sessions to evaluate glucose control under controlled conditions.

Throughout both parts of the study, glucose levels will be closely monitored, and predefined procedures are in place for the treatment of hypoglycemia and hyperglycemia. Safety data and device performance data collected in this study will be used to inform further development of an integrated insulin delivery and glucose sensing system and to support future clinical studies.

연구 유형

중재적

등록 (추정된)

40

단계

  • 초기 1단계

연락처 및 위치

이 섹션에서는 연구를 수행하는 사람들의 연락처 정보와 이 연구가 수행되는 장소에 대한 정보를 제공합니다.

연구 연락처

연구 연락처 백업

연구 장소

    • Ontario
      • Toronto, Ontario, 캐나다, M4G 3E8
        • 모병
        • ClinSurge Research
        • 연락하다:
        • 수석 연구원:
          • Ronnie Aronson, M.D.

참여기준

연구원은 적격성 기준이라는 특정 설명에 맞는 사람을 찾습니다. 이러한 기준의 몇 가지 예는 개인의 일반적인 건강 상태 또는 이전 치료입니다.

자격 기준

공부할 수 있는 나이

  • 성인
  • 고령자

건강한 자원 봉사자를 받아들입니다

아니

설명

Inclusion Criteria Part A:

  • Males and females ≥ 18 years of age.
  • Clinical diagnosis of type 1 diabetes for at least 12 months. The diagnosis of type 1 diabetes is based on the investigator's judgment; C peptide level and antibody determinations are not needed.
  • Undergoing multiple daily injection or continuous subcutaneous insulin infusion therapy for at least 3 months. Those using an automated insulin delivery system can also participate.
  • Total daily insulin dose (TDD) between 30 and 100 IU.

Inclusion Criteria Part B:

  • Males and females ≥ 18 years of age.
  • Clinical diagnosis of type 1 diabetes for at least 12 months. The diagnosis of type 1 diabetes is based on the investigator's judgment; C peptide level and antibody determinations are not needed.
  • Undergoing continuous subcutaneous insulin infusion therapy for at least 3 months. Those using an automated insulin delivery system can also participate.
  • Totally daily insulin dose (TDD) between 10 30 and 100 IU.

Exclusion Criteria (A and B):

  • Serious medical illness likely to interfere with study participation or with the ability to complete the trial by the judgment of the investigator.
  • Failure to comply with the study protocol or with the team's recommendations.
  • Current or recent use of any anti-hyperglycemic agent other than insulin (≤ one month for GLP1-RA, ≤ one week for all others).
  • Female participants of childbearing potential who are pregnant, breastfeeding, or unwilling to use effective contraception during the study. Pregnancy will be verified by urine dipstick testing at the time of admission visit.
  • Severe hypoglycemic episode within one month of admission.
  • Severe diabetic ketoacidosis episode within one month of admission.
  • Clinically significant nephropathy, neuropathy or retinopathy as judged by the investigator.
  • Recent (<6 months) acute macrovascular event e.g., acute coronary syndrome or cardiac surgery.
  • Other serious medical illness likely to interfere with study participation or with the ability to complete the trial by the judgment of the investigator.
  • Current or ≤ one month use of supraphysiological doses of systemic glucocorticoids
  • Pronounced lipohypertrophy in the abdominal subcutaneous adipose tissue, which may impair sensor function or insulin infusion.
  • Insufficient abdominal surface area to support the wearing of three DPP systems in Part A.

공부 계획

이 섹션에서는 연구 설계 방법과 연구가 측정하는 내용을 포함하여 연구 계획에 대한 세부 정보를 제공합니다.

연구는 어떻게 설계됩니까?

디자인 세부사항

  • 주 목적: 다른
  • 할당: 해당 없음
  • 중재 모델: 순차적 할당
  • 마스킹: 없음(오픈 라벨)

무기와 개입

참가자 그룹 / 팔
개입 / 치료
실험적: Sensor Characterization and Automated Glycaemic Control
Sensor characterization, Part "A": Participants will undergo a 1.5-day in-patient stay using the DPP System. The pump and the CGM functions of the DPP System are controlled by separate smartphone applications. Blood glucose will be monitored with a Yellow Springs Instruments (YSI) glucose analyzer and a separate commercial CGM (Dexcom G7 CGM System). Participants will also undergo euglycemic glucose clamp testing to assess sensor response to basal and bolus insulin delivery. Automated glycaemic control, "Part B": Participants will undergo a 2.5-day in-patient stay using the DPP System and an automated glycaemic control (AGC) algorithm. During this period, they will consume standardized meals and engage in standardized exercise to evaluate glucose control.
Participants will use the DPP System, an integrated insulin pump and continuous glucose monitoring-based device, during inpatient study visits. The system will be evaluated under multiple study conditions, including sensor characterization procedures and automated glycaemic control during standardized meals and exercise.

연구는 무엇을 측정합니까?

주요 결과 측정

결과 측정
측정값 설명
기간
Maximum post-bolus difference between DPP CGM sensor glucose and YSI plasma glucose
기간: During Part A clamp period (Day 2; approximately 09:00-17:00), assessed after each bolus.
Highest (maximum) difference between DPP integrated CGM sensor glucose and reference YSI plasma glucose concentration following each predefined insulin bolus delivered by the two DPP systems assigned to bolus delivery.
During Part A clamp period (Day 2; approximately 09:00-17:00), assessed after each bolus.
Time to maximum post-bolus difference between DPP CGM sensor glucose and YSI plasma glucose
기간: During Part A clamp period (Day 2; approximately 09:00-17:00), assessed after each bolus.
Time from each predefined insulin bolus delivery to the time of the highest (maximum) difference between DPP integrated CGM sensor glucose and reference YSI plasma glucose concentration.
During Part A clamp period (Day 2; approximately 09:00-17:00), assessed after each bolus.
Time to full recovery of DPP CGM sensor signal following bolus insulin delivery
기간: During Part A clamp period (Day 2; approximately 09:00-17:00), assessed after each bolus.
Time from each predefined insulin bolus delivery until the DPP integrated CGM sensor signal is considered fully recovered following the transient post-bolus effect (as defined in the study analysis plan/protocol recovery criteria).
During Part A clamp period (Day 2; approximately 09:00-17:00), assessed after each bolus.
Percent of DPP CGM sensor values meeting 20/20 agreement criteria vs reference glucose
기간: Part A in-patient period (Day 1 to Day 2), with primary comparison during the clamp and intensive monitoring period.
Percent of DPP integrated CGM sensor values meeting 20/20 agreement criteria compared with (a) reference YSI plasma glucose and (b) Dexcom G7 interstitial glucose. 20/20 is defined as within ±20% of the reference value when reference glucose is ≥100 mg/dL (5.5 mmol/L) and within ±20 mg/dL (1.1 mmol/L) when reference glucose is <100 mg/dL (5.5 mmol/L). Reported including and excluding periods affected after bolus infusion, between bolus insulin deliveries, and during basal insulin deliveries.
Part A in-patient period (Day 1 to Day 2), with primary comparison during the clamp and intensive monitoring period.
Percent time in glucose ranges based on Dexcom G7 (Time in Range / Time Below / Time Above)
기간: Part B in-patient period (approximately 2.5 days; Day 1 and Day 2, and combined across Part B).

Percentage of time sensor glucose levels (Dexcom G7) are in each range:

3.9-7.8 mmol/L, 3.9-10.0 mmol/L, <3.9 mmol/L, <3.0 mmol/L, >7.8 mmol/L, >10.0 mmol/L, >13.9 mmol/L Calculated for Day 1 and Day 2 separately and for the entire Part B intervention combined.

Part B in-patient period (approximately 2.5 days; Day 1 and Day 2, and combined across Part B).
Mean absolute relative difference (MARD) of DPP CGM vs reference glucose
기간: Part B in-patient period (Day 1 and Day 2, and combined across Part B).
Mean absolute relative difference (MARD) of DPP integrated CGM values compared with (a) reference YSI plasma glucose and (b) Dexcom G7 interstitial glucose, expressed as a percentage. Calculated for Day 1, Day 2, and entire Part B combined.
Part B in-patient period (Day 1 and Day 2, and combined across Part B).
Mean absolute deviation (MAD) of DPP CGM vs reference glucose by basal rate
기간: Part B in-patient period (Day 1 and Day 2, and combined across Part B).
Mean absolute deviation of DPP integrated CGM values compared with (a) reference YSI plasma glucose and (b) Dexcom G7 interstitial glucose, expressed in mmol/L, summarized for each basal rate.
Part B in-patient period (Day 1 and Day 2, and combined across Part B).
Bias of DPP CGM vs reference glucose by basal rate
기간: Part B in-patient period (Day 1 and Day 2, and combined across Part B).
Bias (DPP CGM minus reference) in mmol/L between DPP integrated CGM values and (a) reference YSI plasma glucose and (b) Dexcom G7 interstitial glucose, summarized for each basal rate.
Part B in-patient period (Day 1 and Day 2, and combined across Part B).
Percent of DPP CGM sensor values meeting 20/20 agreement criteria vs YSI (excluding post-bolus affected periods)
기간: Part B in-patient period (Day 1 and Day 2, and combined across Part B).
Percent of DPP integrated CGM sensor values meeting 20/20 agreement criteria compared with reference YSI plasma glucose, excluding times when the sensor is affected after bolus infusion. 20/20 is defined as within ±20% of the reference value when reference glucose is ≥100 mg/dL (5.5 mmol/L) and within ±20 mg/dL (1.1 mmol/L) when reference glucose is <100 mg/dL (5.5 mmol/L).
Part B in-patient period (Day 1 and Day 2, and combined across Part B).

2차 결과 측정

결과 측정
측정값 설명
기간
Overnight percentage of time in glucose ranges (Dexcom G7)
기간: Part B in-patient period (Day 1 and Day 2 nights, and combined across Part B).
Percentage of overnight time (23:00-08:00) during which sensor glucose levels measured by the Dexcom G7 are within the following ranges: 3.9-7.8 mmol/L; 3.9-10.0 mmol/L; <3.9 mmol/L; <3.0 mmol/L; >7.8 mmol/L; >10.0 mmol/L; and >13.9 mmol/L. Calculated for Day 1 and Day 2 separately and for the entire Part B intervention combined.
Part B in-patient period (Day 1 and Day 2 nights, and combined across Part B).
Daytime percentage of time in glucose ranges (Dexcom G7)
기간: Part B in-patient period (Day 1 and Day 2 daytime periods, and combined across Part B).
Percentage of daytime time (08:00-23:00) during which sensor glucose levels measured by the Dexcom G7 are within the following ranges: 3.9-7.8 mmol/L; 3.9-10.0 mmol/L; <3.9 mmol/L; <3.0 mmol/L; >7.8 mmol/L; >10.0 mmol/L; and >13.9 mmol/L. Calculated for Day 1 and Day 2 separately and for the entire Part B intervention combined.
Part B in-patient period (Day 1 and Day 2 daytime periods, and combined across Part B).
Total daily insulin dose
기간: Part B in-patient period (Day 1, Day 2, and combined across Part B).
Total daily insulin dose (TDD), expressed in international units (IU), calculated for Day 1 and Day 2 separately and for the entire Part B intervention combined.
Part B in-patient period (Day 1, Day 2, and combined across Part B).
Mean sensor glucose level
기간: Part B in-patient period (Day 1, Day 2, and combined across Part B).
Mean sensor glucose level measured by the Dexcom continuous glucose monitoring system, calculated for Day 1 and Day 2 separately and for the entire Part B intervention combined.
Part B in-patient period (Day 1, Day 2, and combined across Part B).

공동 작업자 및 조사자

여기에서 이 연구와 관련된 사람과 조직을 찾을 수 있습니다.

간행물 및 유용한 링크

연구에 대한 정보 입력을 담당하는 사람이 자발적으로 이러한 간행물을 제공합니다. 이것은 연구와 관련된 모든 것에 관한 것일 수 있습니다.

일반 간행물

연구 기록 날짜

이 날짜는 ClinicalTrials.gov에 대한 연구 기록 및 요약 결과 제출의 진행 상황을 추적합니다. 연구 기록 및 보고된 결과는 공개 웹사이트에 게시되기 전에 특정 품질 관리 기준을 충족하는지 확인하기 위해 국립 의학 도서관(NLM)에서 검토합니다.

연구 주요 날짜

연구 시작 (실제)

2026년 5월 19일

기본 완료 (추정된)

2026년 9월 1일

연구 완료 (추정된)

2026년 12월 1일

연구 등록 날짜

최초 제출

2026년 5월 21일

QC 기준을 충족하는 최초 제출

2026년 6월 12일

처음 게시됨 (실제)

2026년 6월 17일

연구 기록 업데이트

마지막 업데이트 게시됨 (실제)

2026년 6월 17일

QC 기준을 충족하는 마지막 업데이트 제출

2026년 6월 12일

마지막으로 확인됨

2026년 6월 1일

추가 정보

이 연구와 관련된 용어

개별 참가자 데이터(IPD) 계획

개별 참가자 데이터(IPD)를 공유할 계획입니까?

아니요

약물 및 장치 정보, 연구 문서

미국 FDA 규제 의약품 연구

아니

미국 FDA 규제 기기 제품 연구

아니

미국에서 제조되어 미국에서 수출되는 제품

아니

이 정보는 변경 없이 clinicaltrials.gov 웹사이트에서 직접 가져온 것입니다. 귀하의 연구 세부 정보를 변경, 제거 또는 업데이트하도록 요청하는 경우 register@clinicaltrials.gov. 문의하십시오. 변경 사항이 clinicaltrials.gov에 구현되는 즉시 저희 웹사이트에도 자동으로 업데이트됩니다. .

제1형 당뇨병에 대한 임상 시험

구독하다