Determination of Specificity and Sensitivity in Detecting Myocardial Infarction Using the EDE CARDIO BIO-MAP Machine (CARDIO BIO-MAP)
Determination of Specificity and Sensitivity in Detecting Myocardial Infarction Using the EDE CARDIO BIO-MAP Machine, the EDE CARDIO BIO-MAP Study
The goal of this observational study is to learn whether a non-invasive electromagnetic wave scanner (EDE CARDIO BIO-MAP) can help diagnose myocardial infarction (heart attack) in adults (18 years and older) who come to the emergency department with chest pain and suspected myocardial infarction. The main questions it aims to answer are:
- How sensitive and specific is the EDE scanner in diagnosing myocardial infarction (MI) compared with standard blood tests (e.g., cardiac troponin)?
- How sensitive and specific are the immediate blood measurements from a portable lab device compared with conventional hospital laboratory measurements? Researchers will compare the results of the EDE scanner and the portable lab with standard hospital blood test results and the final clinical diagnosis to see how accurately each detects MI.
Participants will:
- Have a research visit (30 minutes to 1 hour) during their emergency care, including a quick scan with the Exponential Deep Examination (EDE) device (attached to a phone or tablet) and a small research blood sample tested on a portable lab device
- Repeat the scan and blood sample 4 to 6 hours later, only if a second blood draw is already needed for their clinical care
- Answer questions about their medical and social history, and allow the study team to collect relevant results from their medical record Scanner results will not be shared with participants or used for their clinical care.
Studie Overzicht
Toestand
Conditie
Gedetailleerde beschrijving
Myocardial infarction is a leading cause of death and disability worldwide. Current diagnosis relies on clinical assessment, electrocardiogram, and blood measurement of cardiac biomarkers such as troponin. Faster, non-invasive diagnostic tools could be especially valuable in the emergency setting, where timely intervention is critical.
The Exponential Deep Examination Cardio Bio-Map is a non-invasive device, classified as Class 1 under the European Conformity marking system, that analyzes passive electromagnetic signals (0.1 to 300 gigahertz) using artificial intelligence to measure cardiac biomarkers without a blood draw. Results are available within 50 milliseconds. The device does not collect or store patient data. It produces an immediate printed report, and all data processing occurs locally within Qatar. During the COVID-19 pandemic, the same technology detected the SARS-CoV-2 virus with a specificity of 95.62% and a sensitivity of 93.71% compared with polymerase chain reaction testing and antibody testing. This study will evaluate whether the device can screen for myocardial infarction through measurement of cardiac biomarkers, both in biological samples (in vitro, using a portable laboratory device) and directly in patients (in vivo, using the scanner).
Studietype
Inschrijving (Geschat)
Contacten en locaties
Studiecontact
- Naam: Charbel Abi Khalil, MD,PhD
- Telefoonnummer: +974 4492 8484
- E-mail: cha2022@qatar-med.cornell.edu
Deelname Criteria
Geschiktheidscriteria
Leeftijden die in aanmerking komen voor studie
Accepteert gezonde vrijwilligers
Geslachten die in aanmerking komen voor studie
Bemonsteringsmethode
Studie Bevolking
Participants will be recruited at Hamad Medical Corporation
Beschrijving
Inclusion Criteria:
- Age ≥18 years.
- Adults presenting to the emergency room for acute chest pain.
- English or Arabic speakers.
Exclusion Criteria:
- Pregnancy (self-reported).
- Adults with cognitive disorders.
- Unconscious patients.
- Citizens from countries where the General Data Protection Regulation (GDPR) applies, due to regulatory constraints.
Studie plan
Hoe is de studie opgezet?
Ontwerpdetails
- Observatiemodellen: COHORT
- Tijdsperspectieven: PROSPECTIVE
Wat meet het onderzoek?
Primaire uitkomstmaten
Uitkomstmaat |
Maatregel Beschrijving |
Tijdsspanne |
|---|---|---|
|
Sensitivity and specificity of the EDE CARDIO BIO-MAP scanner(in vivo) for detecting myocardial infarction
Tijdsspanne: Baseline
|
Sensitivity (true positives / [true positives + false negatives]) and specificity (true negatives / [true negatives + false positives]) of the scanner in emergency room patients with suspected myocardial infarction, compared with standard hospital laboratory results (cardiac troponin, creatine kinase-myocardial band, myoglobin, B-type natriuretic peptide, N-terminal pro-B-type natriuretic peptide) and the final clinical diagnosis in the medical record. |
Baseline
|
Secundaire uitkomstmaten
Uitkomstmaat |
Maatregel Beschrijving |
Tijdsspanne |
|---|---|---|
|
Sensitivity and specificity of the portable lab's immediate blood measurements (in vitro)
Tijdsspanne: Baseline
|
Sensitivity and specificity of the portable lab's immediate blood measurements for myocardial infarction, compared with conventional hospital laboratory measurements, using a confusion matrix. |
Baseline
|
|
Accuracy of the EDE scanner
Tijdsspanne: Baseline
|
Accuracy = (true positives + true negatives) / all results; |
Baseline
|
|
Misclassification rate of the EDE scanner
Tijdsspanne: Baseline
|
Misclassification rate = (false positives + false negatives) / all results; |
Baseline
|
|
Precision of the EDE scanner
Tijdsspanne: Baseline
|
precision = true positives / (true positives + false positives) |
Baseline
|
|
Accuracy of the portable lab
Tijdsspanne: Baseline
|
Accuracy = (true positives + true negatives) / all results; |
Baseline
|
|
Misclassification rate of the portable lab
Tijdsspanne: Baseline
|
Misclassification rate = (false positives + false negatives) / all results; |
Baseline
|
|
Precision of the portable lab
Tijdsspanne: Baseline
|
precision = true positives / (true positives + false positives) |
Baseline
|
|
Correlation Between EDE Scanner and Portable Laboratory Values for Cardiac Biomarkers
Tijdsspanne: Baseline
|
Pearson or Spearman correlation coefficient between scanner and portable lab values for cardiac biomarkers : Troponin T, Troponin I, Myoglobin, B-type Natriuretic Peptide, NT-proBNP, and Creatine Kinase-MB. |
Baseline
|
|
Predictive power of each cardiac biomarker for myocardial infarction
Tijdsspanne: Baseline
|
Logistic regression, area under the receiver operating characteristic curve, and optimal diagnostic cutoffs for each cardiac biomarker: Troponin T, Troponin I, Myoglobin, B-type Natriuretic Peptide, NT-proBNP, and Creatine Kinase-MB. |
Baseline
|
Andere uitkomstmaten
Uitkomstmaat |
Maatregel Beschrijving |
Tijdsspanne |
|---|---|---|
|
Change in cardiac biomarker levels between initial assessment and 4-6 hour follow-up
Tijdsspanne: Baseline and 4-6 hours after initial assessment
|
Change in cardiac biomarker levels( Troponin T, Troponin I, Myoglobin, B-type Natriuretic Peptide, NT-proBNP, and Creatine Kinase-MB) between the initial assessment and the follow-up, in participants who have a second clinical blood draw. |
Baseline and 4-6 hours after initial assessment
|
|
Diagnostic performance of the scanner by sex
Tijdsspanne: Baseline
|
Subgroup analysis of scanner sensitivity and specificity by sex |
Baseline
|
|
Diagnostic performance of the scanner by age group
Tijdsspanne: Baseline
|
Subgroup analysis of scanner sensitivity and specificity by age group |
Baseline
|
|
Diagnostic performance of the scanner by symptom profile
Tijdsspanne: Baseline
|
Subgroup analysis of scanner sensitivity and specificity by symptom profile. |
Baseline
|
Medewerkers en onderzoekers
Medewerkers
Onderzoekers
- Hoofdonderzoeker: Charbel Abi Khalil, MD,PhD, Weill Cornell Medicine-Qatar
Studie record data
Bestudeer belangrijke data
Studie start (Geschat)
Primaire voltooiing (Geschat)
Studie voltooiing (Geschat)
Studieregistratiedata
Eerst ingediend
Eerst ingediend dat voldeed aan de QC-criteria
Eerst geplaatst (Werkelijk)
Updates van studierecords
Laatste update geplaatst (Werkelijk)
Laatste update ingediend die voldeed aan QC-criteria
Laatst geverifieerd
Meer informatie
Termen gerelateerd aan deze studie
Aanvullende relevante MeSH-voorwaarden
Andere studie-ID-nummers
- 23-00017
Plan Individuele Deelnemersgegevens (IPD)
Bent u van plan om gegevens van individuele deelnemers (IPD) te delen?
Beschrijving IPD-plan
ongoing discussion with the sponsor
Informatie over medicijnen en apparaten, studiedocumenten
Bestudeert een door de Amerikaanse FDA gereguleerd geneesmiddel
Bestudeert een door de Amerikaanse FDA gereguleerd apparaatproduct
Deze informatie is zonder wijzigingen rechtstreeks van de website clinicaltrials.gov gehaald. Als u verzoeken heeft om uw onderzoeksgegevens te wijzigen, te verwijderen of bij te werken, neem dan contact op met register@clinicaltrials.gov. Zodra er een wijziging wordt doorgevoerd op clinicaltrials.gov, wordt deze ook automatisch bijgewerkt op onze website .