Estimating Patient Size From a Single Radiograph (VocMepAdar)
Validation of a Computational Model to Estimate Patient Anterior-posterior Dimension From an Abdominal Radiograph
A computational model has been created to estimate the abdominal depth of a patient from a single x-ray image. The model has been tested using phantoms and found to be accurate; this study aims to test the accuracy of the model with patients and in a clinical setting.
This will be achieved by enrolling patient's who have already been referred for an anterior-posterior abdomen x-ray examination to the trial, taking a physical measurement of their anterior-posterior abdominal depth and then comparing this measured value with a value as estimated using the computational model based on the patient's x-ray image.
Обзор исследования
Статус
Вмешательство/лечение
Подробное описание
A non-commercial computational model has been developed in-house to estimate the patient's anterior-posterior or lateral depth using the radiographic image and the known exposure factors with which it was undertaken. This model has been tested using single composition phantoms and found to be accurate. If it was found to be accurate for real clinical examinations, this would automate the measurement of patient size and give institutions the estimate of patient size required for local paediatric patient dose audit. In turn, this would provide the national data required to propose national reference values for paediatric x-ray examinations, which would give all institutions an important comparator for their performance. This would lead to optimisation in those sites most requiring it; nationally, paediatric x-ray imaging would improve in time.
This pilot study is necessary to determine if the computational model is accurate enough to be relied upon. Accuracy will be determined by comparing the estimate made by the computational model for each patient with an actual measurement of the patient's anterior-posterior abdomen depth made at the time of the examination.
Тип исследования
Регистрация (Действительный)
Контакты и местонахождение
Места учебы
-
-
Angus
-
Dundee, Angus, Соединенное Королевство, DD1 9SY
- NHS Tayside
-
-
Критерии участия
Критерии приемлемости
Возраст, подходящий для обучения
Принимает здоровых добровольцев
Полы, имеющие право на обучение
Метод выборки
Исследуемая популяция
Описание
Inclusion Criteria:
- Adult
- Referred to Ninewells Hospital for an anterior-posterior abdomen x-ray examination
Exclusion Criteria:
- Patients unable to give consent
- Patients who have had a contrast injection in the previous 24 hours
- Patients suffering abdominal pain at the time of the examination
Учебный план
Как устроено исследование?
Детали дизайна
Когорты и вмешательства
Группа / когорта |
Вмешательство/лечение |
|---|---|
|
Patient cohort
20 patients referred to Ninewells Hospital radiology department for an anterior-posterior abdomen x-ray examination.
All of these patients will have a measurement of their anterior-posterior depth before undergoing their x-ray examination.
An estimate of their anterior-posterior depth will then be made from their x-ray image using the computational model.
|
A single measurement of the patient's anterior-posterior abdominal depth
|
Что измеряет исследование?
Первичные показатели результатов
Мера результата |
Мера Описание |
Временное ограничение |
|---|---|---|
|
Accuracy of the Computational Model
Временное ограничение: 2 months
|
The computational model was used to estimate the patient's anterior-posterior abdominal depth using the digital radiographic image, the exposure factors with which it was acquired and a priori knowledge relating to the x-ray unit and digital detector. The outcome measure was the accuracy with which the computational model estimates the patient's anterior-posterior abdominal depth. It was determined by comparing the estimate to measured anterior-posterior abdominal depth (measured at the time of the x-ray examination). Results are expressed as a percentage deviation; a low % deviation is more accurate, a high % deviation less accurate. |
2 months
|
Соавторы и исследователи
Спонсор
Соавторы
Следователи
- Учебный стул: Sarah Vinnicombe, MD, University of Dundee
Даты записи исследования
Изучение основных дат
Начало исследования (Действительный)
Первичное завершение (Действительный)
Завершение исследования (Действительный)
Даты регистрации исследования
Первый отправленный
Впервые представлено, что соответствует критериям контроля качества
Первый опубликованный (Действительный)
Обновления учебных записей
Последнее опубликованное обновление (Действительный)
Последнее отправленное обновление, отвечающее критериям контроля качества
Последняя проверка
Дополнительная информация
Термины, связанные с этим исследованием
Другие идентификационные номера исследования
- 2017RA01
Планирование данных отдельных участников (IPD)
Планируете делиться данными об отдельных участниках (IPD)?
Информация о лекарствах и устройствах, исследовательские документы
Изучает лекарственный продукт, регулируемый FDA США.
Изучает продукт устройства, регулируемый Управлением по санитарному надзору за качеством пищевых продуктов и медикаментов США.
Эта информация была получена непосредственно с веб-сайта clinicaltrials.gov без каких-либо изменений. Если у вас есть запросы на изменение, удаление или обновление сведений об исследовании, обращайтесь по адресу register@clinicaltrials.gov. Как только изменение будет реализовано на clinicaltrials.gov, оно будет автоматически обновлено и на нашем веб-сайте. .