Endotracheal Tube Cuff Inflation Pressure Varieties and Response to Education Among Anesthetists
Cuff pressure is essential in endotracheal tube management. Guidelines recommend a cuff pressure of 20 to 30 cm H2O. One study, for instance, found that cuff pressure exceeded 40 cm H2O in 40-to-90% of tested patients.
This study will investigate the endotracheal ETT cuff inflation pressure applied by the participating anaesthetists, and their response after being informed about the pressure the participants will apply.
研究概览
详细说明
Although the increasing uses of different supraglottic devices as an alternative, the traditional endotracheal intubation is still considered the commonly used method for managing the airway of a generally anaesthetized patient. After passing the endotracheal tube through the larynx, the anaesthetist inflates the cuff or asks the anaesthesia technician to inflate it. How much air is injected into the cuff is not a major concern for almost all anaesthetists and they usually depend on palpating the external cuff tense to judge is it too much, accurate or not enough? Cuff pressure is essential in endotracheal tube management. Guidelines recommend a cuff pressure of 20 to 30 cm H2O. Inflation of the cuff of pressure more than 30 cm H2O can damage the tracheal mucosa by compromising capillary perfusion. When pressures are greater than 50 cm H2O, total obstruction of tracheal blood flow occurs. For long times, without any evidence-based data, it has been believed that well-trained anaesthetists are capable of determining proper ETT cuff pressures. It is presumed that anaesthetists can detect appropriate inflation pressure and overinflated ETT cuff by palpating the ETT pilot balloon. Adequacy of cuff inflation is conventionally determined by palpation of the external balloon. Previous studies suggest that this approach is unreliable. One study, for instance, found that cuff pressure exceeded 40 cm H2O in 40-to-90% of tested patients. However, increased awareness of over-inflation risks may have improved recent clinical practice.
The aim of the work: This study will investigate the endotracheal cuff inflation pressure applied by the participating anaesthetists, and their response after being informed about the pressure the participants will apply. Material and Methods: Type of the study: Prospective controlled single-blind study. Settings: After obtaining approval by the Ethics Committee of the Suez Canal University Hospital, and written informed consent with an explanation regarding the purpose, effects, technique, and complications, all anaesthetists, working currently in the facility, are included.
研究类型
注册 (实际的)
阶段
- 不适用
联系人和位置
学习地点
-
-
-
Ismailia、埃及、2685
- Suez Canal University
-
-
参与标准
资格标准
适合学习的年龄
- 孩子
- 成人
- 年长者
接受健康志愿者
有资格学习的性别
描述
Inclusion Criteria:
- anaesthetists
- works at Suez canal university hospital
- resident, registrar, or senior registrar
Exclusion Criteria:
- refusal to participate
- who will not complete the 2 rounds 3 readings each
学习计划
研究是如何设计的?
设计细节
- 主要用途:预防
- 分配:非随机化
- 介入模型:顺序分配
- 屏蔽:单身的
武器和干预
参与者组/臂 |
干预/治疗 |
|---|---|
|
有源比较器:Before education
Before alarming the participant anaesthetists about the pressure values of their practice in inflating the ETT cuff. This is before applying the intervention which is informing and educating the participant anaesthetists about the proper ETT cuff pressure. |
Before applying the education program of the endotracheal cuff pressure inflation to the participating anaesthetists.
|
|
有源比较器:After education
After alarming the participant anaesthetists about the pressure values of their practice in inflating the ETT cuff. This is after applying the intervention which is informing and educating the participant anaesthetists about the proper ETT cuff pressure. |
After applying the education program of the endotracheal cuff pressure inflation to the participating anaesthetists.
|
研究衡量的是什么?
主要结果指标
结果测量 |
措施说明 |
大体时间 |
|---|---|---|
|
accuracy in inflating the cuff of the endotracheal tubes
大体时间:immediately after fixation of the endotracheal tube during general anaethesia
|
assessment of the Anesthetists' accuracy in inflating the cuff of the endotracheal tubes used to maintain the airway of generally anaesthetized patients the change following feedback.
|
immediately after fixation of the endotracheal tube during general anaethesia
|
合作者和调查者
研究记录日期
研究主要日期
学习开始 (实际的)
初级完成 (实际的)
研究完成 (实际的)
研究注册日期
首次提交
首先提交符合 QC 标准的
首次发布 (实际的)
研究记录更新
最后更新发布 (实际的)
上次提交的符合 QC 标准的更新
最后验证
更多信息
与本研究相关的术语
其他研究编号
- ETT cuff pressure