Long Term Vital Parameter Monitoring (LAVIMO) (LAVIMO)
26. listopadu 2013 aktualizováno: RWTH Aachen University
Long Term Vital Parameter Monitoring Pilot Study to Evaluate Measurement Data Via In-ear Sensor in Comparison to Standard Polysomnography in Sleep Laboratory
The purpose of this study is to evaluate the feasibility of non-intrusive 24 hours non-invasive measurements with minimized LAVIMO sensor system.
The LAVIMO-system is a photoplethysmographic device for reflective mode measurements of the optical damping upper skin layers inside of the ear canal at two different wave lengths.
It consists of the in-ear sensor, the electronics and a PC.
Vital signs parameter like heart rate, heart rate variability, breathing and arterial blood oxygenation are monitored in patients with sleep apnoea and compared to measurements of standard polysomnography of these patients during one night in the sleep laboratory.
Přehled studie
Postavení
Dokončeno
Podmínky
Intervence / Léčba
Detailní popis
In 20 patients who routinely have to undergo a polysomnographic night in sleep laboratory LAVIMO measurements like heart rate, heart rate variability, breathing and arterial blood oxygenation are performed parallel to standard polysomnography measurements and subsequently compared to the data of the standard polysomnography.
The LAVIMO sensor system consists of the in- ear sensor, the electronics and a PC; the individually formed sensor is embedded into an ear mould for proper fit inside the ear canal.
Inside two LEDs emit light into the adjacent skin at 760 and 905 nm wavelengths.
The intensity of the backscattered light is detected by a photo detector.
In the electronics, the control signals for the LEDs are provided, as well as the photo detector signal is further processed and A-D converted.
The electronics also provide a wireless connection to the PC via bluetooth.
Data recording, post processing and displaying are tasks of the PC.
The feasibility of non- intrusive 24 hours and non-invasive measurements like heart rate, heart rate variability, breathing and oxygen saturation is proved via this minimized sensor and electronics placed at a suitable place at the body guaranteeing high wearing comfort.
This includes a wireless data communication for unrestricted mobility.
Subsequently the data are compared to the standard polysomnography data to extract vital signs or for example to recognize possible phases of sleep apnoea.
The goal in future refering to preventive medical monitoring will be early detection of risk factors for patient's health and improvement of living conditions.
Typ studie
Intervenční
Zápis (Aktuální)
8
Fáze
- Nelze použít
Kontakty a umístění
Tato část poskytuje kontaktní údaje pro ty, kteří studii provádějí, a informace o tom, kde se tato studie provádí.
Studijní místa
-
-
North Rhine Westphalia
-
Aachen, North Rhine Westphalia, Německo, 52074
- Neurological Clinic , University Hospital
-
-
Kritéria účasti
Výzkumníci hledají lidi, kteří odpovídají určitému popisu, kterému se říká kritéria způsobilosti. Některé příklady těchto kritérií jsou celkový zdravotní stav osoby nebo předchozí léčba.
Kritéria způsobilosti
Věk způsobilý ke studiu
18 let a starší (Dospělý, Starší dospělý)
Přijímá zdravé dobrovolníky
Ne
Pohlaví způsobilá ke studiu
Všechno
Popis
Inclusion Criteria:
- patients > 18 years who routinely have to undergo a polysomnography night
- sleep apnoea patient
- able to give informed consent
Exclusion Criteria:
- dementia
- cochlea implant
- severe lung diseases
- allergy against material of the sensor
- inflammation of the middle ear
- inflammation of auditory canal
- occlusion because of cerumen
- electric implants
Studijní plán
Tato část poskytuje podrobnosti o studijním plánu, včetně toho, jak je studie navržena a co studie měří.
Jak je studie koncipována?
Detaily designu
- Primární účel: Základní věda
- Přidělení: N/A
- Intervenční model: Přiřazení jedné skupiny
- Maskování: Žádné (otevřený štítek)
Zbraně a zásahy
Skupina účastníků / Arm |
Intervence / Léčba |
|---|---|
|
Experimentální: Device: in-ear sensor
Patients who routinely undergo a polysomnography night (1 night) are monitored via in-ear sensor which will be embedded in the auditory canal.
During this night vital signs parameters are monitored, processed and subsequently compared to the polysomnography data.
|
Long term vital signs parameters like heart rate, heart rate variability, breathing and oxygen saturation are measured via in- ear sensor system during 1 polysomnographic night
Ostatní jména:
|
Co je měření studie?
Primární výstupní opatření
Měření výsledku |
Popis opatření |
Časové okno |
|---|---|---|
|
Extraction of vital signs out of measurement data via individually formed in-ear sensor system in comparison to standard polysomnography data in sleep laboratory
Časové okno: one night
|
during one polysomnographic night in sleep laboratory the patient will receive the LAVIMO in- ear sensor system and vital parameters like heart rate, heart rate variability, breathing and oxygen saturation will be monitored.
Subsequently the data will be processed and compared to the standard polysomnographic data
|
one night
|
Spolupracovníci a vyšetřovatelé
Zde najdete lidi a organizace zapojené do této studie.
Sponzor
Vyšetřovatelé
- Vrchní vyšetřovatel: Johannes Schiefer, MD, Neurological Clinic, University Hospital Aachen
Termíny studijních záznamů
Tato data sledují průběh záznamů studie a předkládání souhrnných výsledků na ClinicalTrials.gov. Záznamy ze studií a hlášené výsledky jsou před zveřejněním na veřejné webové stránce přezkoumány Národní lékařskou knihovnou (NLM), aby se ujistily, že splňují specifické standardy kontroly kvality.
Hlavní termíny studia
Začátek studia
1. ledna 2012
Primární dokončení (Aktuální)
1. února 2013
Dokončení studie (Aktuální)
1. února 2013
Termíny zápisu do studia
První předloženo
20. června 2012
První předloženo, které splnilo kritéria kontroly kvality
20. června 2012
První zveřejněno (Odhad)
22. června 2012
Aktualizace studijních záznamů
Poslední zveřejněná aktualizace (Odhad)
27. listopadu 2013
Odeslaná poslední aktualizace, která splnila kritéria kontroly kvality
26. listopadu 2013
Naposledy ověřeno
1. listopadu 2013
Více informací
Termíny související s touto studií
Klíčová slova
Další relevantní podmínky MeSH
Další identifikační čísla studie
- CTC-A_10-016
- EK 231/10 (Jiný identifikátor: Ethics Committee University Hospital Aachen)
Tyto informace byly beze změn načteny přímo z webu clinicaltrials.gov. Máte-li jakékoli požadavky na změnu, odstranění nebo aktualizaci podrobností studie, kontaktujte prosím register@clinicaltrials.gov. Jakmile bude změna implementována na clinicaltrials.gov, bude automaticky aktualizována i na našem webu .
Klinické studie na LAVIMO in- ear sensor system
-
Przemyslaw KunertNaoX Technologies; Warsaw Medical University Clinical Center; Departments and...Zatím nenabírámeEpilepsie | Drogově rezistentní epilepsie | Fokální epilepsiePolsko
-
KK Women's and Children's HospitalActxa Pte LtdDokončenoGlukóza v krvi, vysoká | Hladina glukózy v krvi, nízkáSingapur