Febrile Whole Blood Specimen Collection and Testing
Febrile Whole Blood Specimen Collection and Testing Protocol to Demonstrate Product Specificity for the Applied Biosystems™ Bacillus Anthracis Detection Kit
Studieoversigt
Status
Betingelser
Intervention / Behandling
Detaljeret beskrivelse
Whole blood samples will be collected from febrile patients presenting with fever of unknown origin and flu-like syndromes. Collection sites will be advised of collection requirements specific to patient population with the goal of collecting samples form patient between the ages of 2-89 years of age. Samples collected will be de-identified and assigned a specific study number. Study numbers and labels will be provided to collection site by the study monitor. Samples must be shipped to MRIGlobal or designated study site on the day of collection.
Collection sites will consent patients and collect one (1) 4 mL whole blood sample from adults and either (1) 2 mL pediatric whole blood sample or 0.5 mL whole blood sample collected in a micro collection container from pediatric patients (anticoagulant: K2EDTA or defined alternative) from each patient enrolled into the study. Patients may not be enrolled in the study more than once. Patients considered for enrollment will be logged on an enrollment log. Log must be reviewed by the site prior to considering a patient for study enrollment to confirm enrollment status. Samples collected will be de-identified and assigned a specific study number. Study numbers and labels will be provided to collection site by the study monitor.
Study site will accept or reject samples based on shipping verification procedures and then process samples acceptable for testing within 24 hours of receipt. Sample testing must be performed in accordance with the Applied Biosystems™ Bacillus anthracis Detection Kit instructions for use.
At study start, study site will test whole blood samples received from collection sites daily (e.g. Tuesday, Wednesday, Thursday, and Friday) using the Applied Biosystems™ Bacillus anthracis Detection Kit. No reference testing will be required since the study is designed only to demonstrate product specificity. Samples are assumed to be negative for Bacillus anthracis. A minimum of 50, but likely 100 febrile whole blood samples will be tested at MRIGlobal.
Data generated will demonstrate product specificity when testing febrile whole blood samples.
Undersøgelsestype
Tilmelding (Faktiske)
Kontakter og lokationer
Studiesteder
-
-
Florida
-
Tarpon Springs, Florida, Forenede Stater, 33319
- Boca Biolistics
-
-
Massachusetts
-
Fall River, Massachusetts, Forenede Stater, 02720
- New England Center for Clinical Research
-
Franklin, Massachusetts, Forenede Stater, 02038
- Medical Research Network
-
-
Deltagelseskriterier
Berettigelseskriterier
Aldre berettiget til at studere
Tager imod sunde frivillige
Køn, der er berettiget til at studere
Prøveudtagningsmetode
Studiebefolkning
Inclusion criteria to be documented on data form provided.
- Fever of unknown origin
- Flu like syndrome
Age ranges of study participants is as follows:
2-15 Years of age 15-21 Years of age 22-40 Years of age >40 Years of age
Beskrivelse
Inclusion Criteria:
- Fever
Exclusion Criteria:
- None
Studieplan
Hvordan er undersøgelsen tilrettelagt?
Design detaljer
- Observationsmodeller: Kun etui
- Tidsperspektiver: Fremadrettet
Kohorter og interventioner
Gruppe / kohorte |
Intervention / Behandling |
|---|---|
|
Data Collection
|
The TaqMan® Bacillus anthracis Detection Kit is composed of the Applied Biosystems® 7500 Fast Dx Real-Time PCR Instrument with SDS Software, secondary interpretative software, and lyophilized reagents for the detection of gene targets for B. anthracis.
The test assay is a multiplexed assay targeting gene targets for B. anthracis, in addition to an internal positive control.
Reagents are lyophilized as a fully formulated Mastermix and are stable at room temperature for up to two years.
The kit is specifically designed for performing real-time PCR using the Applied Biosystems® 7500 Fast Dx Real-Time PCR Instrument with SDS Software , with nucleic acids extracted from clinical specimens using Qiagen's QIAamp® Blood Midi Kit manual extraction or Roche MagNAPure automated extraction methods.
|
Hvad måler undersøgelsen?
Primære resultatmål
Resultatmål |
Foranstaltningsbeskrivelse |
Tidsramme |
|---|---|---|
|
Product Specificity
Tidsramme: 1 day
|
Data generated will demonstrate product specificity when testing whole blood samples.
|
1 day
|
Samarbejdspartnere og efterforskere
Sponsor
Samarbejdspartnere
Datoer for undersøgelser
Studer store datoer
Studiestart (Faktiske)
Primær færdiggørelse (Faktiske)
Studieafslutning (Faktiske)
Datoer for studieregistrering
Først indsendt
Først indsendt, der opfyldte QC-kriterier
Først opslået (Faktiske)
Opdateringer af undersøgelsesjournaler
Sidste opdatering sendt (Faktiske)
Sidste opdatering indsendt, der opfyldte kvalitetskontrolkriterier
Sidst verificeret
Mere information
Begreber relateret til denne undersøgelse
Yderligere relevante MeSH-vilkår
Andre undersøgelses-id-numre
- MRI-2017-004/5
Plan for individuelle deltagerdata (IPD)
Planlægger du at dele individuelle deltagerdata (IPD)?
Lægemiddel- og udstyrsoplysninger, undersøgelsesdokumenter
Studerer et amerikansk FDA-reguleret lægemiddelprodukt
Studerer et amerikansk FDA-reguleret enhedsprodukt
Disse oplysninger blev hentet direkte fra webstedet clinicaltrials.gov uden ændringer. Hvis du har nogen anmodninger om at ændre, fjerne eller opdatere dine undersøgelsesoplysninger, bedes du kontakte register@clinicaltrials.gov. Så snart en ændring er implementeret på clinicaltrials.gov, vil denne også blive opdateret automatisk på vores hjemmeside .