Outcomes of a multicenter U.S. clinical trial of a new monofocal single-piece hydrophobic acrylic IOL

Steven C Schallhorn, Michael Bonilla, Seth M Pantanelli, Steven C Schallhorn, Michael Bonilla, Seth M Pantanelli

Abstract

Purpose: To evaluate the safety and effectiveness of the CT LUCIA 611P intraocular lens (IOL) in patients with cataracts.

Setting: 23 surgeons at 15 different clinical sites.

Design: Prospective single-arm clinical trial.

Methods: The study was conducted under an Investigational Device Exemption for premarket approval of a new hydrophobic acrylic IOL in the United States. Patients were followed for 12 months, and the main measured variables included uncorrected (UDVA) and corrected (CDVA) distance visual acuities, manifest refraction, and adverse events.

Results: In total, 339 eyes of 339 patients were implanted with the study device, of which 310 (91.4%) reached the 12-month visit. The percentage of eyes within ±0.50 diopter (D) and ±1.00 D of emmetropia was 85.8% (266/310) and 96.8% (300/310), respectively. Manifest refraction spherical equivalent (MRSE) remained stable over the first year with the mean 12-month MRSE of -0.03 ± 0.45 D. The mean 12-month UDVA and CDVA were 0.09 ± 0.15 (≈20/25) and -0.02 ± 0.09 (≈20/19) logMAR, respectively. Of all patients, 99.4% (308/310) achieved postoperative CDVA ≥20/40. The incidence of Nd:YAG capsulotomy within the first year was 3.5% (11/310). Only 2 eyes had IOL tilt present at the 12-month postoperative visit with no associated visual symptoms. There were 2 cases of IOL decentration; one required removal of the IOL, whereas the other had no visual side effects related to decentration. There were no findings of glistening at any visit.

Conclusions: The CT LUCIA 611P IOL demonstrated excellent safety, efficacy, and stability of refractive outcomes. No significant issues related to the biocompatibility of the IOL material were observed.

Trial registration: ClinicalTrials.gov NCT03451786.

Copyright © 2022 The Author(s). Published by Wolters Kluwer Health, Inc. on behalf of ASCRS and ESCRS.

Figures

Figure 1.
Figure 1.
CT LUCIA 611P IOL.
Figure 2.
Figure 2.
Blueject injector.
Figure 3.
Figure 3.
A: Postoperative UDVA vs postoperative CDVA. *Note that the percentage of patients achieving 20/40 or better CDVA (99.7% or 309/310) is different from the value presented in the ISO comparison in the text of the article (99.4% or 308/310). The ISO comparison is based on the number of patients achieving 0.30 logMAR or better, whereas the 20/40 calculation of visual acuity in this histogram was based on the range between 0.25 logMAR and 0.35 logMAR. B: Difference between postoperative UDVA and postoperative CDVA. C: Distribution of postoperative MRSE. D: Scattergram of attempted vs achieved MRSE. E: Stability of MRSE over time calculated for a consistent cohort of patients who attended all postoperative visits. F: Distribution of postoperative refractive astigmatism (absolute value). MRSE = manifest refraction spherical equivalent

References

    1. Khandelwal SS, Jun JJ, Mak S, Booth MS, Shekelle PG. Effectiveness of multifocal and monofocal intraocular lenses for cataract surgery and lens replacement: a systematic review and meta-analysis. Graefes Arch Clin Exp Ophthalmol 2019;257:863–875
    1. Braga-Mele R, Chang D, Dewey S, Foster G, Henderson BA, Hill W, Hoffman R, Little B, Mamalis N, Oetting T, Serafano D, Talley-Rostov A, Vasavada A, Yoo S; ASCRS Cataract Clinical Committee. Multifocal intraocular lenses: relative indications and contraindications for implantation. J Cataract Refract Surg 2014;40:313–322
    1. Cao K, Friedman DS, Jin S, Yusufu M, Zhang J, Wang J, Hou S, Zhu G, Wang B, Xiong Y, Li J, Li X, He H, Chai L, Wan XH. Multifocal versus monofocal intraocular lenses for age-related cataract patients: a system review and meta-analysis based on randomized controlled trials. Surv Ophthalmol 2019;64:647–658
    1. Guyton DL, Uozato H, Wisnicki HJ. Rapid determination of intraocular lens tilt and decentration through the undilated pupil. Ophthalmology 1990;97:1259–1264
    1. Christiansen G, Durcan FJ, Olson RJ, Christiansen K. Glistenings in the AcrySof intraocular lens: pilot study. J Cataract Refract Surg 2001;27:728–733
    1. Liou HL, Brennan NA. Anatomically accurate, finite model eye for optical modeling. J Opt Soc Am A Opt Image Sci Vis 1997;14:1684–1695
    1. Reinstein DZ, Archer TJ, Srinivasan S, Mamalis N, Kohnen T, Dupps WJ, Randleman JB. Standard for reporting refractive outcomes of intraocular lens-based refractive surgery. J Cataract Refract Surg 2017;43:435–439
    1. ISO 11979-7:2006. Ophthalmic implants—intraocular lenses—part 7: clinical investigations
    1. Wirtitsch MG, Findl O, Menapace R, Kriechbaum K, Koeppl C, Buehl W, Drexler W. Effect of haptic design on change in axial lens position after cataract surgery. J Cataract Refract Surg 2004;30:45–51
    1. Hienert J, Hirnschall N, Ruiss M, Ullrich M, Zwickl H, Findl O. Prospective study to compare axial position stability after fellow-eye implantation of 2 distinct intraocular lens designs. J Cataract Refract Surg 2021;47:999–1005
    1. Wormstone IM, Damm NB, Kelp M, Eldred JA. Assessment of intraocular lens/capsular bag biomechanical interactions following cataract surgery in a human in vitro graded culture capsular bag model. Exp Eye Res 2021;205:108487.
    1. Borkenstein AF, Borkenstein EM. Surgical experience with a redesigned, fully preloaded, hydrophobic acrylic intraocular lens in challenging cases of pseudoexfoliation syndrome, phacodonesis, and small pupils. Clin Ophthalmol 2019;13:199–206
    1. Gale RP, Saldana M, Johnston RL, Zuberbuhler B, McKibbin M. Benchmark standards for refractive outcomes after NHS cataract surgery. Eye (Lond) 2009;23:149–152
    1. Hahn U, Krummenauer F, Kölbl B, Neuhann T, Schayan-Araghi K, Schmickler S, von Wolff K, Weindler J, Will T, Neuhann I. Determination of valid benchmarks for outcome indicators in cataract surgery: a multicenter, prospective cohort trial. Ophthalmology 2011;118:2105–2112
    1. Lundström M, Dickman M, Henry Y, Manning S, Rosen P, Tassignon MJ, Young D, Stenevi U. Risk factors for refractive error after cataract surgery: analysis of 282 811 cataract extractions reported to the European Registry of Quality Outcomes for cataract and refractive surgery. J Cataract Refract Surg 2018;44:447–452
    1. Zhao Y, Yang K, Li J, Huang Y, Zhu S. Comparison of hydrophobic and hydrophilic intraocular lens in preventing posterior capsule opacification after cataract surgery: an updated meta-analysis. Medicine (Baltimore) 2017;96:e8301.
    1. Lehmann R, Maxwell A, Lubeck DM, Fong R, Walters TR, Fakadej A. Effectiveness and safety of the Clareon monofocal intraocular lens: outcomes from a 12-month single-arm clinical study in a large sample. Clin Ophthalmol 2021;15:1647–1657
    1. Summary of safety and effectiveness—Acrysof IQ Restor +3.0 D Multifocal Toric Intraocular Lens. Premarket approval study, decision date December 22, 2016. Available at: . Accessed April 1, 2022
    1. Nibourg LM, Gelens E, Kuijer R, Hooymans JM, van Kooten TG, Koopmans SA. Prevention of posterior capsular opacification. Exp Eye Res 2015;136:100–115
    1. Nanavaty MA, Spalton DJ, Boyce J, Brain A, Marshall J. Edge profile of commercially available square-edged intraocular lenses. J Cataract Refract Surg 2008;34:677–686
    1. Findl O, Buehl W, Bauer P, Sycha T. Interventions for preventing posterior capsule opacification. Cochrane Database Syst Rev 2010:CD003738.
    1. Nanavaty MA, Zukaite I, Salvage J. Edge profile of commercially available square-edged intraocular lenses: Part 2. J Cataract Refract Surg. 2019;45:847–853
    1. Ling R, Borkenstein EM, Borkenstein AF. Evaluation of Nd:YAG laser capsulotomy rates in a real-life population. Clin Ophthalmol 2020;14:3249–3257
    1. Borkenstein AF, Borkenstein EM. Long-term clinical results and scanning electron microscopic analysis of the aspheric, hydrophobic, acrylic intraocular lens CT LUCIA 611P(Y). Clin Ophthalmol 2018;12:1219–1227
    1. Yildirim TM, Schickhardt SK, Wang Q, Friedmann E, Khoramnia R, Auffarth GU. Quantitative evaluation of microvacuole formation in five intraocular lens models made of different hydrophobic materials. PLoS One 2021;16:e0250860.
    1. Henriksen BS, Kinard K, Olson RJ. Effect of intraocular lens glistening size on visual quality. J Cataract Refract Surg 2015;41:1190–1198
    1. DeHoog E, Doraiswamy A. Evaluation of the impact of light scatter from glistenings in pseudophakic eyes. J Cataract Refract Surg 2014;40:95–103
    1. Liu Y, Zhao J, Hu Y, Li B, Wang J, Zhang J. Comparison of the visual performance after implantation of three aberration-correcting aspherical intraocular lens. Curr Eye Res 2021;46:333–340

Source: PubMed

3
Abonner