Randomized crossover comparison of the laryngeal mask airway classic with i-gel laryngeal mask airway in the management of difficult airway in post burn neck contracture patients

Jeevan Singh, Manohar Kumar Yadav, Sujan Babu Marahatta, Bikash Lal Shrestha, Jeevan Singh, Manohar Kumar Yadav, Sujan Babu Marahatta, Bikash Lal Shrestha

Abstract

Purpose: The objective of the study was to compare the performance of i-gel supraglottic airway with cLMA in difficult airway management in post burn neck contracture patients and assess the feasibility of i-gel use for emergency airway management in difficult airway situation with reduced neck movement and limited mouth opening.

Methods: Prospective, crossover, randomized controlled trial was performed amongst forty eight post burn neck contracture patients with limited mouth opening and neck movement. i-gel and cLMA were placed in random order in each patient. Primary outcome was overall success rate. Other measurements were time to successful ventilation, airway leak pressure, fiberoptic glottic view, visualization of square wave pattern.

Results: Success rate for the i-gel was 91.7% versus 79.2% for the cLMA. i-gel required shorter insertion time (19.3 seconds vs. 23.5 seconds, P=0.000). Airway leak pressure difference was statistically significant (i-gel 21.2 cm H20; cLMA 16.9 cm H(2)0; P=0.00). Fiberoptic view through the i-gel showed there were less epiglottic downfolding and better fiberoptic view of the glottis than cLMA. Overall agreement in insertion outcome for i-gel was 22/24 (91.7%) successes and 2/24(8.3%) failure and for cLMA, 19/24 (79.16%) successes and 5/24 (16.7%) failure in the first attempt.

Conclusion: The i-gel is cheap, effective airway device which is easier to insert and has better clinical performance in the difficult airway management of the airway in the post burn contracture of the neck. Our study shows that i-gel is feasible for emergency airway management in difficult airway situation with reduced neck movement and limited mouth opening in post burn neck.

Keywords: cLMA; difficult airway management; i-gel; post burn neck contracture.

Conflict of interest statement

Conflict of Interest: None declared

Figures

Figure 1
Figure 1
i-gel
Figure 2
Figure 2
cLMA

References

    1. i-gel user guide. 7th ed. Workingham, UK: Intersurgical Ltd; 2009.
    1. Stone B, Leach AB, Alexander CA. The use of the laryngeal mask airway by nurses during cardiopulmonary resuscitation. Results of a multicentre trial. Anaesthesia. 1994;49:3–7.
    1. Blunt MC, Burchett KR. Varient Creutzfeldt-Jakob disease and disposable anaesthetic equipment-balancing the risks. Br J Anaesth. 2003;90:1–3.
    1. Esler MD, Baines LC, Wilkinson DJ, Langford RM. Decontamination of laryngoscopes: A survey of national practice. Anaesthesia. 1999;54:587–92.
    1. Jackson KM, Cook TM. Evaluation of four airway manikins as patient simulators for the insertion of eight types of supraglottic airway devices. Anaesthesia. 2007;62:388–93.
    1. Richez B, Saltel L, Banchereau F, Torrielli R, Cros AM. A new single use supraglottic airway device with a non inflatable cuff and esophageal vent: An observational study of the i-gel. Anesth Analg. 2008;106:1137–9.
    1. Gatward JJ, Cook TM, Seller C, Handel J, Simpson T, Vanek V, et al. Evaluation of the size 4 i-gel airway in one hundred non paralysed patients. Anaesthesia. 2008;63:1124–30.
    1. Janakiraman C, Chethan DB, Wilkes AR, Stacey MR, Goodwin N. A randomized crossover trial comparing the i-gel supraglottic airway and classic laryngeal mask airway. Anaesthesia. 2009;64:674–8.
    1. Caplan RA, Benumof JL, Berry FA, Blitt CD, Bode RH, Cheney FW, et al. Practice guidelines for management of the difficult airway: An updated report by the American Society of Anesthesiologists Task Force on Management of the Difficult Airway. Anesthesiology. 2003;98:1269–77.
    1. Nolan JP, Deakin CD, Soar J, Bottiger BW, Smith G. European Resuscitation Council guidelines for resuscitation 2005. Section 4. Adult advanced life support. Resuscitation. 2005;67(Suppl 1):39–86.
    1. Brimacombe J, Holyoake L, Keller C, Brimacombe N, Scully M, Barry J, et al. Pharyngolaryngeal, neck, and jaw discomfort after anesthesia with the face mask and laryngeal mask airway at high and low cuff volumes in males and females. Anesthesiology. 2000;93:26–31.
    1. Bamgbade OA, Macnab WR, Khalaf WM. Evaluation of the i-gel airway in 300 patients. Eur J Anaesthesiol. 2008;25:865–6.
    1. Shin WJ, Cheong YS, Yang HS, Nishiyama T. The supraglottic airway i-gel in comparison with ProSeal Laryngeal mask airway and classic laryngeal mask airway in anaesthetized patients. Eur J Anaesthesiol. 2010;27:588–601.
    1. Livitan R M, Kinkle W C. Initial anatomic investigations of the airway: A novel supraglotic airway without inflatable cuff. Anaesthesia. 2005;60:1022–6.

Source: PubMed

3
Abonner