Validation of thoracic impedance cardiography by echocardiography in healthy late pregnancy

Jordan P R McIntyre, Kevin M Ellyett, Edwin A Mitchell, Gina M Quill, John Md Thompson, Alistair W Stewart, Robert N Doughty, Peter R Stone, Maternal Sleep in Pregnancy Study Group, Jordan P R McIntyre, Kevin M Ellyett, Edwin A Mitchell, Gina M Quill, John Md Thompson, Alistair W Stewart, Robert N Doughty, Peter R Stone, Maternal Sleep in Pregnancy Study Group

Abstract

Background: Assessment of stroke volume (SV) is often necessary in clinical and research settings. The clinically established method for SV assessment in pregnancy is echocardiography, but given its limitations, it is not always an appropriate measurement tool. Thoracic impedance cardiography (ICG) allows continuous, non-invasive SV assessment. However, SV determination relies on assumptions regarding the thoracic shape that may mean the algorithm is not valid in pregnancy. The available data regarding the validity of ICG against an established reference standard using modern SV algorithms are both limited and conflicting. We aimed to test the validity of ICG in a clinically realistic setting in late pregnancy using echocardiography.

Methods: Twenty-nine women in late pregnancy underwent standard echocardiography assessments with simultaneous ICG measurement. Agreement between devices was tested using Bland-Altman analysis.

Results: Bland-Altman analysis of the relationship between ICG and echocardiography demonstrated that the 95% limits of agreement exceeded acceptable or expected ranges. Measures of maternal and fetal anthropometry do not account for the lack of agreement.

Conclusions: Absolute values of SV as determined by ICG are not valid in pregnancy. Further work is required to examine the ability of ICG to assess relative changes in maternal haemodynamics in late pregnancy.

Figures

Figure 1
Figure 1
Bland-Altman plots; Echocardiography vs. Impedance Cardiography. Broken lines indicate mean ± 1.96 SD. HR, SV and CO are in beats per minute, mL, and L.min−1, respectively.

References

    1. Evangelista A, Flachskampf F, Lancellotti P, Badano L, Aguilar R, Monaghan M, Zamorano J, Nihoyannopoulos P, European Association of E European Association of Echocardiography recommendations for standardization of performance, digital storage and reporting of echocardiographic studies. Eur J Echocardiogr. 2008;9(4):438–48. doi: 10.1093/ejechocard/jen174.
    1. Bernstein DP. A new stroke volume equation for thoracic electrical bioimpedance: theory and rationale. Crit Care Med. 1986;14(10):904–9. doi: 10.1097/00003246-198610000-00017.
    1. Smith DN. Bioimpedance measurement of cardiac output. Crit Care Med. 1994;22(9):1513–4. doi: 10.1097/00003246-199409000-00027.
    1. Burlingame J, Ohana P, Aaronoff M, Seto T. Noninvasive cardiac monitoring in pregnancy: impedance cardiography versus echocardiography. J Perinatol. 2013;33:675–80. doi: 10.1038/jp.2013.35.
    1. Secher NJ, Arnsbo P, Andersen LH, Thomsen A. Measurements of cardiac stroke volume in various body positions in pregnancy and during caesarean section: a comparison between thermodilution and impedance cardiography. Scand J Clin Lab Invest. 1979;39(6):569–76. doi: 10.3109/00365517909108835.
    1. Milsom I, Forssman L, Sivertsson R, Dottori O. Measurement of cardiac stroke volume by impedance cardiography in the last trimester of pregnancy. Acta Obstet Gynecol Scand. 1983;62(5):473–9. doi: 10.3109/00016348309154222.
    1. Easterling TR, Benedetti TJ, Carlson KL, Watts DH. Measurement of cardiac output in pregnancy by thermodilution and impedance techniques. Br J Obstet Gynaecol. 1989;96(1):67–9. doi: 10.1111/j.1471-0528.1989.tb01578.x.
    1. Masaki DI, Greenspoon JS, Ouzounian JG. Measurement of cardiac output in pregnancy by thoracic electrical bioimpedance and thermodilution: a preliminary report. Am J Obstet Gynecol. 1989;161(3):680–4. doi: 10.1016/0002-9378(89)90379-7.
    1. Clark SL, Southwick J, Pivarnik JM, Cotton DB, Hankins GDV, Phelan JP. A comparison of cardiac index in normal term pregnancy using thoracic electrical bio-impedance and oxygen extraction (Fick) techniques. Obstet Gynecol. 1994;83(5):669–72.
    1. Rosenberg P, Yancy CW. Noninvasive assessment of hemodynamics: an emphasis on bioimpedance cardiography. Curr Opin Cardiol. 2000;15(3):151–5. doi: 10.1097/00001573-200005000-00005.
    1. Quiñones MA, Otto CM, Stoddard M, Waggoner A, Zoghbi WA. Recommendations for quantification of Doppler echocardiography: a report from the Doppler quantification task force of the nomenclature and standards committee of the American Society of Echocardiography. J Am Soc Echocardiogr. 2002;15(2):167–84. doi: 10.1067/mje.2002.120202.
    1. Lang RM, Bierig M, Devereux RB, Flachskampf FA, Foster E, Pellikka PA, Picard MH, Roman MJ, Seward J, Shanewise JS, et al. Recommendations for chamber quantification: a report from the American Society of Echocardiography's Guidelines and Standards Committee and the Chamber Quantification Writing Group, developed in conjunction with the European Association of Echocardiography, a branch of the European Society of Cardiology. J Am Soc Echocardiography. 2005;18(12):1440–63. doi: 10.1016/j.echo.2005.10.005.
    1. Niccomo Impedance Cardiography. [].
    1. Bland MJ, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;327(8476):307–10. doi: 10.1016/S0140-6736(86)90837-8.
    1. Myles PS, Cui J. Using the Bland–Altman method to measure agreement with repeated measures (Editorial I) Br J Anaesth. 2007;99(3):309–11. doi: 10.1093/bja/aem214.
    1. Pot M, Sadler L, McDougall J, Harilall M, Battin M. National Women’s Annual Clinical Report 2010. In: National Women’s, Auckland District Health Board, editor. 2010.
    1. Moertl MG, Schlembach D, Papousek I, Hinghofer-Szalkay H, Weiss EM, Lang U, Lackner HK. Hemodynamic evaluation in pregnancy: limitations of impedance cardiography. Physiol Meas. 2012;33(6):1015. doi: 10.1088/0967-3334/33/6/1015.
    1. Stout K. Diagnosis and management of heart disease in pregnancy. In: Otto CM, editor. The practice of clinical echocardiography. 3. Philadelphia, PA: Saunders / Elsevier; 2007.
    1. Dennis A, Arhanghelschi I, Simmons S, Royse C. Prospective observational study of serial cardiac output by transthoracic echocardiography in healthy pregnant women undergoing elective caesarean delivery. Int J Obstet Anesth. 2010;19(2):142–8. doi: 10.1016/j.ijoa.2009.06.007.
    1. Desai DK, Moodley J, Naidoo DP. Echocardiographic assessment of cardiovascular hemodynamics in normal pregnancy. Obstet Gynecol. 2004;104(1):20–9. doi: 10.1097/01.AOG.0000128170.15161.1d.
    1. Morris R, Sunesara I, Rush L, Anderson B, Blake PG, Darby M, Sawardecker S, Novotny S, Bofill JA, Martin JNJ. Maternal hemodynamics by thoracic impedance cardiography for normal pregnancy and the postpartum period. Obstet Gynecol. 2014;123(2, PART 1):318–24. doi: 10.1097/AOG.0000000000000104.
    1. Robson SC, Murray A, Peart I, Heads A, Hunter S. Reproducibility of cardiac output measurement by cross sectional and Doppler echocardiography. Br Heart J. 1988;59(6):680–4. doi: 10.1136/hrt.59.6.680.
    1. Critchley LAH, Critchley JAJH. A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput. 1999;15(2):85–91. doi: 10.1023/A:1009982611386.
    1. Nguyen HB, Banta D, Stewart G, Kim T, Bansal R, Anholm J, Wittlake W, Corbett S. Cardiac index measurements by transcutaneous Doppler ultrasound and transthoracic echocardiography in adult and pediatric emergency patients. J Clin Monit Comput. 2010;24(3):237–47. doi: 10.1007/s10877-010-9240-6.
    1. Tomsin K, Mesens T, Molenberghs G, Gyselaers W. Impedance cardiography in uncomplicated pregnancy and pre-eclampsia: a reliability study. J Obstetrics Gynaecol. 2012;32(7):630–4. doi: 10.3109/01443615.2012.673036.
    1. Kubicek W, Karnegis J, Patterson R, Witsoe D, Mattson R. Development and evaluation of an impedance cardiac output system. Aerosp Med. 1966;37(12):1208–12.
    1. Jensen L, Yakimets J, Teo KK. A review of impedance cardiography. Heart Lung. 1995;24(3):183–93. doi: 10.1016/S0147-9563(05)80036-6.
    1. Fellahi J-L, Caille V, Charron C, Deschamps-Berger P-H, Vieillard-Baron A. Noninvasive assessment of cardiac index in healthy volunteers: a comparison between thoracic impedance cardiography and doppler echocardiography. Anesth Analg. 2009;108(5):1553–9. doi: 10.1213/ane.0b013e31819cd97e.
    1. Raaijmakers E, Faes TJC, Scholten RJPM, Goovaerts HG, Heethaar RM. A meta-analysis of published studies concerning the validity of thoracic impedance cardiography. Ann N Y Acad Sci. 1999;873(1):121–7. doi: 10.1111/j.1749-6632.1999.tb09458.x.

Source: PubMed

3
Abonner