CareTaker Self-Calibration Validation (Self_Cal)
The goal of the study is to test the capability of the CareTaker monitor to calibrate its continuous blood pressure readings independently. Currently the device requires another approved blood pressure monitor to provide a starting calibration.
The new control module enables the CareTaker to perform a pressure sweep of the internal pressure in the finger cuff. The resulting data is analyzed using a combination of pulse analysis/oscillometry approach.
Studienübersicht
Status
Bedingungen
Intervention / Behandlung
Detaillierte Beschreibung
The specific objective of this project was to demonstrate the CareTaker's ability to perform self-calibration to the same performance standard as its predicate device. That device demonstrated efficacy by showing substantial equivalence to the performance of classical sphygmomanometry using a Riva-Rocci/Korotkoff (RRK) measurement on the brachial artery, meeting a bias of at least 5 mmHg and a standard deviation of 8 mmHg across 3 paired readings from at least 85 patients, for a total of at least 255 paired readings.
The study was performed at the University of Virginia Medical Center in Charlottesville, Virginia. UVA's Institutional Review Board approved (#18686) and supervised the study for subjects who were University of Virginia hospital patients and staff > 18 years of age and who were able to give verbal consent.
Studientyp
Einschreibung (Tatsächlich)
Teilnahmekriterien
Zulassungskriterien
Studienberechtigtes Alter
Akzeptiert gesunde Freiwillige
Studienberechtigte Geschlechter
Probenahmeverfahren
Studienpopulation
Beschreibung
Inclusion Criteria:
- > 18 years of age, who are able and willing to participate and have given verbal assent
Exclusion Criteria:
- Unable to give verbal assent
- <18 years of age
- No or poor finger pulse, as determined through visual inspection for ischemic hands
Studienplan
Wie ist die Studie aufgebaut?
Designdetails
Was misst die Studie?
Primäre Ergebnismessungen
Ergebnis Maßnahme |
Maßnahmenbeschreibung |
Zeitfenster |
|---|---|---|
|
Performance comparison
Zeitfenster: 8 minutes for 3 paired readings
|
Comparison with performance of classical sphygmomanometry using a Riva-Rocci/Korotkov (RRK) measurement on the brachial artery, meeting a bias of at least 5 mmHg and a standard deviation of 8 mmHg
|
8 minutes for 3 paired readings
|
Mitarbeiter und Ermittler
Sponsor
Mitarbeiter
Studienaufzeichnungsdaten
Haupttermine studieren
Studienbeginn
Primärer Abschluss (Tatsächlich)
Studienabschluss (Tatsächlich)
Studienanmeldedaten
Zuerst eingereicht
Zuerst eingereicht, das die QC-Kriterien erfüllt hat
Zuerst gepostet (Schätzen)
Studienaufzeichnungsaktualisierungen
Letztes Update gepostet (Schätzen)
Letztes eingereichtes Update, das die QC-Kriterien erfüllt
Zuletzt verifiziert
Mehr Informationen
Begriffe im Zusammenhang mit dieser Studie
Schlüsselwörter
Andere Studien-ID-Nummern
- SELF_CAL_001
Plan für individuelle Teilnehmerdaten (IPD)
Planen Sie, individuelle Teilnehmerdaten (IPD) zu teilen?
Diese Informationen wurden ohne Änderungen direkt von der Website clinicaltrials.gov abgerufen. Wenn Sie Ihre Studiendaten ändern, entfernen oder aktualisieren möchten, wenden Sie sich bitte an register@clinicaltrials.gov. Sobald eine Änderung auf clinicaltrials.gov implementiert wird, wird diese automatisch auch auf unserer Website aktualisiert .