Maestro Handheld Cardiac Monitor Validation
Descripción general del estudio
Descripción detallada
The purpose of this study is to determine if Maestro can be used to detect heart arrhythmias, specifically atrial fibrillation in humans. Maestro will be used to measure heart rate and rhythm at the same time that patients are being monitored with a standard 3, 5, or 12 lead electrocardiogram (ECG) in a clinical setting. The recorded Maestro results will be compared to the ECG results to assess Maestro's accuracy.
Data will be analyzed by a pass/fail criterial. The data will be downloaded and printed for the investigators' interpretation who will first review and code the data from Meastro, blinded to the ECG findings. Then the investigators will interpret the ECG findings and the 2 sets of data will be compared. The ECG result is the true value with which to compare the Maestro result.
Tipo de estudio
Inscripción (Actual)
Fase
- No aplica
Contactos y Ubicaciones
Ubicaciones de estudio
-
-
Michigan
-
Ann Arbor, Michigan, Estados Unidos, 48109
- University of Michigan Cardiovascular Center
-
-
Criterios de participación
Criterio de elegibilidad
Edades elegibles para estudiar
Acepta Voluntarios Saludables
Géneros elegibles para el estudio
Descripción
Inclusion Criteria:
- ability to provide informed consent
Exclusion Criteria:
- implanted pacemakers
- open wounds or abrasions on their hands
- prior damage to hands or thumbs precluding obtaining an ECG tracing
Plan de estudios
¿Cómo está diseñado el estudio?
Detalles de diseño
- Propósito principal: Diagnóstico
- Asignación: No aleatorizado
- Modelo Intervencionista: Asignación paralela
- Enmascaramiento: Ninguno (etiqueta abierta)
Armas e Intervenciones
Grupo de participantes/brazo |
Intervención / Tratamiento |
|---|---|
|
Otro: Patients in afib or flutter
Maestro ECG
|
Telemetry is obtained through the Maestro handheld ECG device.
|
|
Otro: Patients in sinus rhythm
Maestro ECG
|
Telemetry is obtained through the Maestro handheld ECG device.
|
¿Qué mide el estudio?
Medidas de resultado primarias
Medida de resultado |
Medida Descripción |
Periodo de tiempo |
|---|---|---|
|
Sensitivity and Specificity of atrial fibrillation detection
Periodo de tiempo: 6 seconds
|
Each sample acquired by the Maestro system will be automatically analyzed and rhythm categorized by the automated algorithm.
This will then be assessed for accuracy and sensitivity and specificity for detection of atrial fibrillation will be calculated.
|
6 seconds
|
Colaboradores e Investigadores
Patrocinador
Investigadores
- Investigador principal: Rakesh Latchamsetty, MD, University of Michigan
Fechas de registro del estudio
Fechas importantes del estudio
Inicio del estudio (Actual)
Finalización primaria (Actual)
Finalización del estudio (Actual)
Fechas de registro del estudio
Enviado por primera vez
Primero enviado que cumplió con los criterios de control de calidad
Publicado por primera vez (Estimar)
Actualizaciones de registros de estudio
Última actualización publicada (Actual)
Última actualización enviada que cumplió con los criterios de control de calidad
Última verificación
Más información
Términos relacionados con este estudio
Términos MeSH relevantes adicionales
Otros números de identificación del estudio
- HUM00074138
Plan de datos de participantes individuales (IPD)
¿Planea compartir datos de participantes individuales (IPD)?
Información sobre medicamentos y dispositivos, documentos del estudio
Estudia un producto farmacéutico regulado por la FDA de EE. UU.
Estudia un producto de dispositivo regulado por la FDA de EE. UU.
Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .