- ICH GCP
- Registro de ensayos clínicos de EE. UU.
- Ensayo clínico NCT03419390
A Pilot Study for the Evaluation of the Safety and Performance of a Combined OCT System (Hydra01)
A Prospective,Open Label,Non-randomized,Single-centre,Exploratory,Stratified Pilot Study for the Evaluation of the Safety and Performance of a Combined Coaxial Optical Coherence Tomography (OCT) System,to Image Diseases of the Posterior and Anterior Segment of the Eye
The novel Coaxial Optical Coherence Tomography (OCT) System is designed to visualize both structures of the retina and choroid.
The purpose of this research is the safety and evaluation of a combined coaxial optical coherence tomography (OCT) system, to image diseases of the posterior and anterior segment of the eye.
Descripción general del estudio
Estado
Condiciones
Intervención / Tratamiento
Descripción detallada
Background and Rationale:
Many posterior segment ocular diseases involve the retinal and choroidal vasculature. A new technology to noninvasively and simultaneously visualize vascular pathology is of utmost importance to reduce patient exposure to different diagnosis methods like e.g. fluorescence angiography. The novel Coaxial Optical Coherence Tomography (OCT) System is designed to visualize both structures of the retina and choroid.
Furthermore, it may be used to image other structures of the eye.
Objective(s):
The purpose of this research is the safety and evaluation of a combined coaxial optical coherence tomography (OCT) system, to image diseases of the posterior and anterior segment of the eye. This device may be useful for the early diagnosis and monitoring of a variety of diseases involving the eye, such as Glaucoma, Diabetic Retinopathy or Age-Related Macular Degeneration. Measurements are compared to normal controls.
Measurements and procedures:
After screening, all participants will undergo a scanning procedure on the same day.
Only one eye of each participant will be scanned with the investigational device. For patients, the eye showing the disease will be chosen, for healthy volunteers, the left or right eye will be scanned according to a randomisation list.
The image quality is rated by the investigator and an independent assessor on a VAS (visual analogue scale). A subgroup of each stratum will have a comparative measurement of the choroid thickness against a SS-OCT (swept-source-OCT) device (Topcon) and a retina thickness against the Spectralis data. The subgroup will consist of half planned sample size. The measurement is taken from one eye, ie. every second participant per stratum will be attributed to the thickness measurement subgroup. The endpoint is a true/false criterion on whether the measurement of the retina and choroid thickness was successful. The thickness is measured in μm. The measurement is considered a "success" if the value is no more than +/- 10% off the reference measurement.
Tipo de estudio
Inscripción (Actual)
Fase
- No aplica
Criterios de participación
Criterio de elegibilidad
Edades elegibles para estudiar
Acepta Voluntarios Saludables
Géneros elegibles para el estudio
Descripción
Inclusion Criteria:
- Subjects with healthy eye, or has been diagnosed already with eye disease
- Patients > 18 years of age
- Informed Consent as documented by date and signature
Exclusion Criteria:
- Clinically significant concomitant disease that impair measurement
- Inability to follow the procedures of the study, e.g. due to language problems, psychological disorders, late stage Alzheimer disease, etc. of the participant,
- Subjects using implanted electronic medical devices (e.g. cochlear implant, pacemaker, defibrillator, infusion pump).
- Participation in another study with investigational drug within the 30 days preceding and during the present study
Plan de estudios
¿Cómo está diseñado el estudio?
Detalles de diseño
- Propósito principal: Diagnóstico
- Asignación: No aleatorizado
- Modelo Intervencionista: Asignación de un solo grupo
- Enmascaramiento: Ninguno (etiqueta abierta)
Armas e Intervenciones
Grupo de participantes/brazo |
Intervención / Tratamiento |
|---|---|
|
Otro: Healthy subjects
One eye of each participant will be scanned with the investigational device (= combined Coaxial Optical Coherence Tomography (OCT) System)
|
The investigational device is an OCT scanner prototype with two different wavelengths developed based on a CE marked OCT device (Spectralis® Heidelberg) at the Berner Fachhochschule, Biel
|
|
Otro: Diseased groups
lf one eye is affected, this will be chosen.
lf both eyes are be affected, the eye with the severest symptoms will be chosen.
Scanning with the investigational device (= combined Coaxial Optical Coherence Tomography (OCT) System)
|
The investigational device is an OCT scanner prototype with two different wavelengths developed based on a CE marked OCT device (Spectralis® Heidelberg) at the Berner Fachhochschule, Biel
|
|
Otro: Diseased subgroups
Every second subject will be allocated to the subgroup.
|
The investigational device is an OCT scanner prototype with two different wavelengths developed based on a CE marked OCT device (Spectralis® Heidelberg) at the Berner Fachhochschule, Biel
|
¿Qué mide el estudio?
Medidas de resultado primarias
Medida de resultado |
Medida Descripción |
Periodo de tiempo |
|---|---|---|
|
Visual acuity (VA) after investigation
Periodo de tiempo: Within 1 hour after investigation
|
Safety of the imaging investigation expressed as visual acuity before (baseline) and after imaging, and tolerability assessment questionnaire.
|
Within 1 hour after investigation
|
Medidas de resultado secundarias
Medida de resultado |
Medida Descripción |
Periodo de tiempo |
|---|---|---|
|
Scan quality
Periodo de tiempo: After completion of all measurement, an average of 2 years.
|
Comparing scan quality of investigational device to scan quality of a reference device, expressed as thickness of the retina and the thickness of the choroid
|
After completion of all measurement, an average of 2 years.
|
|
Procedural success
Periodo de tiempo: After completion of all measurement, an average of 2 years.
|
Procedural success that is defined as: Image Quality of anterior and posterior eye structures
|
After completion of all measurement, an average of 2 years.
|
|
Procedural safety and comfort assessed by questionnaire
Periodo de tiempo: Within 1 hour after investigation
|
Procedural comfort in the opinion of the patients assessed with a questionnaire
|
Within 1 hour after investigation
|
Colaboradores e Investigadores
Patrocinador
Investigadores
- Director de estudio: Hendrik PN Scholl, MD, University Hospital Basel, Dept. of Ophthalmology
- Investigador principal: Pascal Hasler, MD, University Hospital Basel, Dept. of Ophthalmology
Fechas de registro del estudio
Fechas importantes del estudio
Inicio del estudio (Actual)
Finalización primaria (Actual)
Finalización del estudio (Actual)
Fechas de registro del estudio
Enviado por primera vez
Primero enviado que cumplió con los criterios de control de calidad
Publicado por primera vez (Actual)
Actualizaciones de registros de estudio
Última actualización publicada (Actual)
Última actualización enviada que cumplió con los criterios de control de calidad
Última verificación
Más información
Términos relacionados con este estudio
Palabras clave
Otros números de identificación del estudio
- 085-HAP-2017-001
Información sobre medicamentos y dispositivos, documentos del estudio
Estudia un producto farmacéutico regulado por la FDA de EE. UU.
Estudia un producto de dispositivo regulado por la FDA de EE. UU.
Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .