A Pilot Study for the Evaluation of the Safety and Performance of a Combined OCT System (Hydra01)

2020년 3월 31일 업데이트: University Hospital, Basel, Switzerland

A Prospective,Open Label,Non-randomized,Single-centre,Exploratory,Stratified Pilot Study for the Evaluation of the Safety and Performance of a Combined Coaxial Optical Coherence Tomography (OCT) System,to Image Diseases of the Posterior and Anterior Segment of the Eye

The novel Coaxial Optical Coherence Tomography (OCT) System is designed to visualize both structures of the retina and choroid.

The purpose of this research is the safety and evaluation of a combined coaxial optical coherence tomography (OCT) system, to image diseases of the posterior and anterior segment of the eye.

연구 개요

상세 설명

Background and Rationale:

Many posterior segment ocular diseases involve the retinal and choroidal vasculature. A new technology to noninvasively and simultaneously visualize vascular pathology is of utmost importance to reduce patient exposure to different diagnosis methods like e.g. fluorescence angiography. The novel Coaxial Optical Coherence Tomography (OCT) System is designed to visualize both structures of the retina and choroid.

Furthermore, it may be used to image other structures of the eye.

Objective(s):

The purpose of this research is the safety and evaluation of a combined coaxial optical coherence tomography (OCT) system, to image diseases of the posterior and anterior segment of the eye. This device may be useful for the early diagnosis and monitoring of a variety of diseases involving the eye, such as Glaucoma, Diabetic Retinopathy or Age-Related Macular Degeneration. Measurements are compared to normal controls.

Measurements and procedures:

After screening, all participants will undergo a scanning procedure on the same day.

Only one eye of each participant will be scanned with the investigational device. For patients, the eye showing the disease will be chosen, for healthy volunteers, the left or right eye will be scanned according to a randomisation list.

The image quality is rated by the investigator and an independent assessor on a VAS (visual analogue scale). A subgroup of each stratum will have a comparative measurement of the choroid thickness against a SS-OCT (swept-source-OCT) device (Topcon) and a retina thickness against the Spectralis data. The subgroup will consist of half planned sample size. The measurement is taken from one eye, ie. every second participant per stratum will be attributed to the thickness measurement subgroup. The endpoint is a true/false criterion on whether the measurement of the retina and choroid thickness was successful. The thickness is measured in μm. The measurement is considered a "success" if the value is no more than +/- 10% off the reference measurement.

연구 유형

중재적

등록 (실제)

30

단계

  • 해당 없음

참여기준

연구원은 적격성 기준이라는 특정 설명에 맞는 사람을 찾습니다. 이러한 기준의 몇 가지 예는 개인의 일반적인 건강 상태 또는 이전 치료입니다.

자격 기준

공부할 수 있는 나이

18년 이상 (성인, 고령자)

건강한 자원 봉사자를 받아들입니다

연구 대상 성별

모두

설명

Inclusion Criteria:

  • Subjects with healthy eye, or has been diagnosed already with eye disease
  • Patients > 18 years of age
  • Informed Consent as documented by date and signature

Exclusion Criteria:

  • Clinically significant concomitant disease that impair measurement
  • Inability to follow the procedures of the study, e.g. due to language problems, psychological disorders, late stage Alzheimer disease, etc. of the participant,
  • Subjects using implanted electronic medical devices (e.g. cochlear implant, pacemaker, defibrillator, infusion pump).
  • Participation in another study with investigational drug within the 30 days preceding and during the present study

공부 계획

이 섹션에서는 연구 설계 방법과 연구가 측정하는 내용을 포함하여 연구 계획에 대한 세부 정보를 제공합니다.

연구는 어떻게 설계됩니까?

디자인 세부사항

  • 주 목적: 특수 증상
  • 할당: 무작위화되지 않음
  • 중재 모델: 단일 그룹 할당
  • 마스킹: 없음(오픈 라벨)

무기와 개입

참가자 그룹 / 팔
개입 / 치료
다른: Healthy subjects
One eye of each participant will be scanned with the investigational device (= combined Coaxial Optical Coherence Tomography (OCT) System)
The investigational device is an OCT scanner prototype with two different wavelengths developed based on a CE marked OCT device (Spectralis® Heidelberg) at the Berner Fachhochschule, Biel
다른: Diseased groups
lf one eye is affected, this will be chosen. lf both eyes are be affected, the eye with the severest symptoms will be chosen. Scanning with the investigational device (= combined Coaxial Optical Coherence Tomography (OCT) System)
The investigational device is an OCT scanner prototype with two different wavelengths developed based on a CE marked OCT device (Spectralis® Heidelberg) at the Berner Fachhochschule, Biel
다른: Diseased subgroups

Every second subject will be allocated to the subgroup.

  1. Scanning with the investigational device (= combined Coaxial Optical Coherence Tomography (OCT) System)
  2. Thickness measurement with reference medical device.
The investigational device is an OCT scanner prototype with two different wavelengths developed based on a CE marked OCT device (Spectralis® Heidelberg) at the Berner Fachhochschule, Biel

연구는 무엇을 측정합니까?

주요 결과 측정

결과 측정
측정값 설명
기간
Visual acuity (VA) after investigation
기간: Within 1 hour after investigation
Safety of the imaging investigation expressed as visual acuity before (baseline) and after imaging, and tolerability assessment questionnaire.
Within 1 hour after investigation

2차 결과 측정

결과 측정
측정값 설명
기간
Scan quality
기간: After completion of all measurement, an average of 2 years.
Comparing scan quality of investigational device to scan quality of a reference device, expressed as thickness of the retina and the thickness of the choroid
After completion of all measurement, an average of 2 years.
Procedural success
기간: After completion of all measurement, an average of 2 years.
Procedural success that is defined as: Image Quality of anterior and posterior eye structures
After completion of all measurement, an average of 2 years.
Procedural safety and comfort assessed by questionnaire
기간: Within 1 hour after investigation
Procedural comfort in the opinion of the patients assessed with a questionnaire
Within 1 hour after investigation

공동 작업자 및 조사자

여기에서 이 연구와 관련된 사람과 조직을 찾을 수 있습니다.

수사관

  • 연구 책임자: Hendrik PN Scholl, MD, University Hospital Basel, Dept. of Ophthalmology
  • 수석 연구원: Pascal Hasler, MD, University Hospital Basel, Dept. of Ophthalmology

연구 기록 날짜

이 날짜는 ClinicalTrials.gov에 대한 연구 기록 및 요약 결과 제출의 진행 상황을 추적합니다. 연구 기록 및 보고된 결과는 공개 웹사이트에 게시되기 전에 특정 품질 관리 기준을 충족하는지 확인하기 위해 국립 의학 도서관(NLM)에서 검토합니다.

연구 주요 날짜

연구 시작 (실제)

2016년 12월 1일

기본 완료 (실제)

2017년 3월 1일

연구 완료 (실제)

2019년 12월 1일

연구 등록 날짜

최초 제출

2017년 11월 17일

QC 기준을 충족하는 최초 제출

2018년 1월 25일

처음 게시됨 (실제)

2018년 2월 5일

연구 기록 업데이트

마지막 업데이트 게시됨 (실제)

2020년 4월 1일

QC 기준을 충족하는 마지막 업데이트 제출

2020년 3월 31일

마지막으로 확인됨

2020년 3월 1일

추가 정보

이 연구와 관련된 용어

키워드

기타 연구 ID 번호

  • 085-HAP-2017-001

약물 및 장치 정보, 연구 문서

미국 FDA 규제 의약품 연구

아니

미국 FDA 규제 기기 제품 연구

아니

이 정보는 변경 없이 clinicaltrials.gov 웹사이트에서 직접 가져온 것입니다. 귀하의 연구 세부 정보를 변경, 제거 또는 업데이트하도록 요청하는 경우 register@clinicaltrials.gov. 문의하십시오. 변경 사항이 clinicaltrials.gov에 구현되는 즉시 저희 웹사이트에도 자동으로 업데이트됩니다. .