A Pilot Study for the Evaluation of the Safety and Performance of a Combined OCT System (Hydra01)
A Prospective,Open Label,Non-randomized,Single-centre,Exploratory,Stratified Pilot Study for the Evaluation of the Safety and Performance of a Combined Coaxial Optical Coherence Tomography (OCT) System,to Image Diseases of the Posterior and Anterior Segment of the Eye
The novel Coaxial Optical Coherence Tomography (OCT) System is designed to visualize both structures of the retina and choroid.
The purpose of this research is the safety and evaluation of a combined coaxial optical coherence tomography (OCT) system, to image diseases of the posterior and anterior segment of the eye.
Studienübersicht
Status
Bedingungen
Intervention / Behandlung
Detaillierte Beschreibung
Background and Rationale:
Many posterior segment ocular diseases involve the retinal and choroidal vasculature. A new technology to noninvasively and simultaneously visualize vascular pathology is of utmost importance to reduce patient exposure to different diagnosis methods like e.g. fluorescence angiography. The novel Coaxial Optical Coherence Tomography (OCT) System is designed to visualize both structures of the retina and choroid.
Furthermore, it may be used to image other structures of the eye.
Objective(s):
The purpose of this research is the safety and evaluation of a combined coaxial optical coherence tomography (OCT) system, to image diseases of the posterior and anterior segment of the eye. This device may be useful for the early diagnosis and monitoring of a variety of diseases involving the eye, such as Glaucoma, Diabetic Retinopathy or Age-Related Macular Degeneration. Measurements are compared to normal controls.
Measurements and procedures:
After screening, all participants will undergo a scanning procedure on the same day.
Only one eye of each participant will be scanned with the investigational device. For patients, the eye showing the disease will be chosen, for healthy volunteers, the left or right eye will be scanned according to a randomisation list.
The image quality is rated by the investigator and an independent assessor on a VAS (visual analogue scale). A subgroup of each stratum will have a comparative measurement of the choroid thickness against a SS-OCT (swept-source-OCT) device (Topcon) and a retina thickness against the Spectralis data. The subgroup will consist of half planned sample size. The measurement is taken from one eye, ie. every second participant per stratum will be attributed to the thickness measurement subgroup. The endpoint is a true/false criterion on whether the measurement of the retina and choroid thickness was successful. The thickness is measured in μm. The measurement is considered a "success" if the value is no more than +/- 10% off the reference measurement.
Studientyp
Einschreibung (Tatsächlich)
Phase
- Unzutreffend
Teilnahmekriterien
Zulassungskriterien
Studienberechtigtes Alter
Akzeptiert gesunde Freiwillige
Studienberechtigte Geschlechter
Beschreibung
Inclusion Criteria:
- Subjects with healthy eye, or has been diagnosed already with eye disease
- Patients > 18 years of age
- Informed Consent as documented by date and signature
Exclusion Criteria:
- Clinically significant concomitant disease that impair measurement
- Inability to follow the procedures of the study, e.g. due to language problems, psychological disorders, late stage Alzheimer disease, etc. of the participant,
- Subjects using implanted electronic medical devices (e.g. cochlear implant, pacemaker, defibrillator, infusion pump).
- Participation in another study with investigational drug within the 30 days preceding and during the present study
Studienplan
Wie ist die Studie aufgebaut?
Designdetails
- Hauptzweck: Diagnose
- Zuteilung: Nicht randomisiert
- Interventionsmodell: Einzelgruppenzuweisung
- Maskierung: Keine (Offenes Etikett)
Waffen und Interventionen
Teilnehmergruppe / Arm |
Intervention / Behandlung |
|---|---|
|
Sonstiges: Healthy subjects
One eye of each participant will be scanned with the investigational device (= combined Coaxial Optical Coherence Tomography (OCT) System)
|
The investigational device is an OCT scanner prototype with two different wavelengths developed based on a CE marked OCT device (Spectralis® Heidelberg) at the Berner Fachhochschule, Biel
|
|
Sonstiges: Diseased groups
lf one eye is affected, this will be chosen.
lf both eyes are be affected, the eye with the severest symptoms will be chosen.
Scanning with the investigational device (= combined Coaxial Optical Coherence Tomography (OCT) System)
|
The investigational device is an OCT scanner prototype with two different wavelengths developed based on a CE marked OCT device (Spectralis® Heidelberg) at the Berner Fachhochschule, Biel
|
|
Sonstiges: Diseased subgroups
Every second subject will be allocated to the subgroup.
|
The investigational device is an OCT scanner prototype with two different wavelengths developed based on a CE marked OCT device (Spectralis® Heidelberg) at the Berner Fachhochschule, Biel
|
Was misst die Studie?
Primäre Ergebnismessungen
Ergebnis Maßnahme |
Maßnahmenbeschreibung |
Zeitfenster |
|---|---|---|
|
Visual acuity (VA) after investigation
Zeitfenster: Within 1 hour after investigation
|
Safety of the imaging investigation expressed as visual acuity before (baseline) and after imaging, and tolerability assessment questionnaire.
|
Within 1 hour after investigation
|
Sekundäre Ergebnismessungen
Ergebnis Maßnahme |
Maßnahmenbeschreibung |
Zeitfenster |
|---|---|---|
|
Scan quality
Zeitfenster: After completion of all measurement, an average of 2 years.
|
Comparing scan quality of investigational device to scan quality of a reference device, expressed as thickness of the retina and the thickness of the choroid
|
After completion of all measurement, an average of 2 years.
|
|
Procedural success
Zeitfenster: After completion of all measurement, an average of 2 years.
|
Procedural success that is defined as: Image Quality of anterior and posterior eye structures
|
After completion of all measurement, an average of 2 years.
|
|
Procedural safety and comfort assessed by questionnaire
Zeitfenster: Within 1 hour after investigation
|
Procedural comfort in the opinion of the patients assessed with a questionnaire
|
Within 1 hour after investigation
|
Mitarbeiter und Ermittler
Ermittler
- Studienleiter: Hendrik PN Scholl, MD, University Hospital Basel, Dept. of Ophthalmology
- Hauptermittler: Pascal Hasler, MD, University Hospital Basel, Dept. of Ophthalmology
Studienaufzeichnungsdaten
Haupttermine studieren
Studienbeginn (Tatsächlich)
Primärer Abschluss (Tatsächlich)
Studienabschluss (Tatsächlich)
Studienanmeldedaten
Zuerst eingereicht
Zuerst eingereicht, das die QC-Kriterien erfüllt hat
Zuerst gepostet (Tatsächlich)
Studienaufzeichnungsaktualisierungen
Letztes Update gepostet (Tatsächlich)
Letztes eingereichtes Update, das die QC-Kriterien erfüllt
Zuletzt verifiziert
Mehr Informationen
Begriffe im Zusammenhang mit dieser Studie
Schlüsselwörter
Andere Studien-ID-Nummern
- 085-HAP-2017-001
Arzneimittel- und Geräteinformationen, Studienunterlagen
Studiert ein von der US-amerikanischen FDA reguliertes Arzneimittelprodukt
Studiert ein von der US-amerikanischen FDA reguliertes Geräteprodukt
Diese Informationen wurden ohne Änderungen direkt von der Website clinicaltrials.gov abgerufen. Wenn Sie Ihre Studiendaten ändern, entfernen oder aktualisieren möchten, wenden Sie sich bitte an register@clinicaltrials.gov. Sobald eine Änderung auf clinicaltrials.gov implementiert wird, wird diese automatisch auch auf unserer Website aktualisiert .