- ICH GCP
- Registro de ensayos clínicos de EE. UU.
- Ensayo clínico NCT05240014
Mejora del movimiento voluntario en amplias poblaciones de pacientes con ortesis motorizadas modulares
The overall goal of this project is to develop modular, lower-limb, powered orthoses that fit to user-specific weakened joints and control force/torque in a manner that enhances voluntary motion in broad patient populations. This project aims to establish feasibility of assisting different populations with these modular powered orthoses. The investigators hypothesize that assisting lower-limb musculature with modular powered orthoses will improve 1) lifting/lowering posture in able-bodied subjects and 2) functional outcomes in elderly subjects.
Descripción general del estudio
Estado
Condiciones
Intervención / Tratamiento
Descripción detallada
The overall goal of this project is to develop modular, lower-limb, powered orthoses that fit to user-specific weakened joints and control force/torque in a manner that enhances voluntary motion in broad patient populations. Conventional orthoses tend to immobilize joints, and emerging powered orthoses constrain voluntary motion by using highly geared electric motors and/or control methods that force the user to follow a specific gait pattern. Consequently, these devices have not seen widespread success across populations with weakened voluntary control due to advanced age, musculoskeletal disorders, etc. These heterogeneous populations require partial, not full, assistance of user-specific muscle groups during daily activities. However, there is a fundamental gap in knowledge about how to design and control powered orthoses to assist the user without constraining their motion. The central hypothesis of this project is that high-torque, low-inertia motor systems controlled with energetic objectives will enable modular powered orthoses to partially assist the joints. High-torque electric motors combined with minimal transmissions can be freely rotated (i.e., backdriven) by human joints, allowing the use of an emerging torque control method called energy shaping to reduce the perceived weight/inertia of the body during any motion. By mounting these modular actuators to commercial orthoses, this technology will be easily prescribed/configured by clinicians. This project aims to establish feasibility of assisting different populations with modular powered orthoses. The investigators hypothesize that assisting lower-limb musculature with modular powered orthoses will improve 1) lifting/lowering posture in able-bodied subjects and 2) functional outcomes in elderly subjects.
Tipo de estudio
Inscripción (Actual)
Fase
- No aplica
Contactos y Ubicaciones
Ubicaciones de estudio
-
-
Michigan
-
Ann Arbor, Michigan, Estados Unidos, 48109
- Rehab Lab, University of Michigan
-
-
Criterios de participación
Criterio de elegibilidad
Edades elegibles para estudiar
Acepta Voluntarios Saludables
Géneros elegibles para el estudio
Descripción
Inclusion criteria for able-bodied, young participants will be:
- Aged between 18 to 65 years
- Weigh less than 250 lbs due to limitations in the design of the orthoses
- Ability to lift and lower a 10 kg weight using the neutral-spine squat technique for 10 repetitions
Exclusion criteria for able-bodied, young adult participants will be:
- Pregnant (self-report)
- Any significant neuromuscular or musculoskeletal disorder that would interfere with the study
- Prior history of chronic lower-back pain
- Unable to walk for 20 minutes
- History of any cardiovascular, vestibular, or visual diseases and/or impairments that may interfere with the study
- Cognitive deficits that would impair their ability to give informed consent or impair their ability to follow simple instructions during the experiments. Cognitive deficits will be determined by a Mini-Mental State Examination (MMSE) score of <22.
- Adults with a known allergy to medical grade tape
Inclusion criteria for elderly participants will be:
- Aged between 65 to 85 years
- Weigh less than 250 lbs due to limitations in the design of the orthoses
- Ability to walk 6 minutes without assistance from a person (may use walking aid)
Exclusion criteria for elderly participants will be:
- Pregnant (self-report)
- Significant pain due to arthritis or other joint problems that would limit their ability to walk
- Any recent lower-extremity fracture (within 3 months)
- Significant neurological (e.g., stroke), orthopedic, or cardiovascular disorder that may affect the ability to walk
- Advised by a physician not to walk or exercise
- Uncontrolled hypertension or diabetes
- Cognitive deficits or visual impairment that would impair their ability to give informed consent or impair their ability to follow simple instructions during the experiments. Cognitive deficits will be determined by a Mini-Mental State Examination (MMSE) score of <22.
- Adults with a known allergy to medical grade tape
Plan de estudios
¿Cómo está diseñado el estudio?
Detalles de diseño
- Propósito principal: Otro
- Asignación: N / A
- Modelo Intervencionista: Asignación de un solo grupo
- Enmascaramiento: Ninguno (etiqueta abierta)
Armas e Intervenciones
Grupo de participantes/brazo |
Intervención / Tratamiento |
|---|---|
|
Experimental: Exoskeleton
Participants in this arm of the study will perform various tasks while wearing the modular powered orthosis |
This study will investigate modular, lower-limb, powered orthoses that fit to user-specific weakened joints and control force/torque in a manner that enhances voluntary motion in broad patient populations. The central hypothesis is that high-torque, low-inertia motor systems controlled with energetic objectives will enable modular powered orthoses to partially assist the joints. High-torque electric motors combined with minimal transmissions can be freely rotated (i.e., backdriven) by human joints, allowing the use of an emerging torque control method called energy shaping to reduce the perceived weight/inertia of the body during any motion. By mounting these modular actuators to commercial orthoses, this technology will be easily prescribed/configured by clinicians. |
¿Qué mide el estudio?
Medidas de resultado primarias
Medida de resultado |
Medida Descripción |
Periodo de tiempo |
|---|---|---|
|
Muscle Effort (% MVC) With and Without Powered Orthosis - Able Bodied 1
Periodo de tiempo: 1 day
|
For each orthosis module tested, electromyography (EMG) readings will be normalized per-muscle by the peak EMG observed during the no-orthosis condition, and then averaged over the cycle and across repetitions to obtain "normalized exertion" values. Data are presented as Muscle Effort (% Maximum Voluntary Contraction (MVC). Participants completed active movement trials and multi-terrain circuit loops under randomized conditions: 1) Baseline unassisted (No Exoskeleton) and 2) Assisted with active orthosis joint configurations running trajectory-free, task-invariant torque control via energy-shaping algorithms. |
1 day
|
|
Quadriceps Muscle Effort (% MVC) With and Without Powered Orthosis - Able Bodied 2
Periodo de tiempo: 1 day
|
We estimated quadriceps effort during six different lifting, lowering, and carrying (LLC) tasks (lifting-lowering (LL), level walking (LW), ramp ascent (RA), ramp descent (RD), stair ascent (SA), and stair descent (SD)) for the no-exo and exo conditions in a nonfatigued state. Muscle effort was calculated as the mean root mean square EMG, normalized to % maximum voluntary contraction (%MVC), over the gait/task cycle. Quadriceps effort was estimated by taking the weighted mean of the efforts of the vastus medialis oblique (VMO), vastus lateralis (VL), and rectus femoris (RF) based on their respective physiological cross-sectional areas. |
1 day
|
|
Time to Complete 10 Reps of Lifting/Lowering
Periodo de tiempo: 1 day, assessed per experimental condition (i.e. with orthosis and without orthosis)
|
Time to complete 10 reps of L&L will be measured post-fatigue in healthy subjects. The reported outcome is measured in seconds to completion. |
1 day, assessed per experimental condition (i.e. with orthosis and without orthosis)
|
|
Gait Speed
Periodo de tiempo: 1 day, assessed per experimental condition (i.e. with orthosis and without orthosis)
|
Gait speed will be determined by the time to complete a 10-meter walk test. This will be the primary outcome measure for elderly subjects. |
1 day, assessed per experimental condition (i.e. with orthosis and without orthosis)
|
|
Powered Orthosis Effect (Biological Torque)
Periodo de tiempo: 1 day, assessed per experimental condition (i.e. with orthosis and without orthosis)
|
For each orthosis module tested, peak biological torque (estimated by inverse dynamics) will be averaged across repetitions. Performance will be assessed by the difference between the orthosis condition and no-orthosis condition, averaging across tasks. Participants walked on a Bertec instrumented treadmill with a level slope at 1.0 and 1.2 m/s, a 5◦-inclined and 5◦-declined slope at 1.0 m/s, and a 10◦-inclined and 10◦-declined slope at 0.8 m/s. Participants also performed stair ascent and descent on a 5-step staircase with 7 in (178 mm) step height, and STS transitions from a knee-height chair. All activities were performed both with the powered exoskeleton and without. Results are presented as mean torque (Newton-meters per kilogram (Nm/kg). |
1 day, assessed per experimental condition (i.e. with orthosis and without orthosis)
|
|
Minimum Chair Height for Successful Sit-to-stand
Periodo de tiempo: 1 day, assessed per experimental condition (i.e. with orthosis and without orthosis)
|
We will evaluate the minimum chair height from which elderly participants can successfully rise with and without the knee exoskeleton. |
1 day, assessed per experimental condition (i.e. with orthosis and without orthosis)
|
|
Joint Power
Periodo de tiempo: 1 day, assessed per experimental condition (i.e. with orthosis and without orthosis)
|
For elderly participants, we will evaluate the peak values of biological and total (exo+biological) joint power at the hip with and without the exoskeleton. Results presented as mean power in watts per kilogram (W/kg). |
1 day, assessed per experimental condition (i.e. with orthosis and without orthosis)
|
Medidas de resultado secundarias
Medida de resultado |
Medida Descripción |
Periodo de tiempo |
|---|---|---|
|
User Satisfaction
Periodo de tiempo: 1 day, assessed per experimental condition (i.e. with orthosis and without orthosis), with orthosis reported
|
User satisfaction / perception of device effectiveness over an ambulation circuit (sit-to-stand, walk up and down a ramp, walk up and down a staircase) will be measured post-fatigue in healthy subjects using a modified Quebec User Evaluation of Satisfaction with Assistive Technology (QUEST 2.0) survey. The purpose of the scale is to evaluate user device satisfaction and consists of 12 items asking the user to rate their satisfaction on a scale of 1 to 5, where 1 = "not satisfied at all" and 5 = "very satisfied". The categories subjects were asked to rate were: Lifting/Lowering (LL), Ramp Ascent (RA), Stair Ascent (SA), Level walking (LW), Ramp Descent (RD), Stair Descent (SD) |
1 day, assessed per experimental condition (i.e. with orthosis and without orthosis), with orthosis reported
|
|
Thorax Angle
Periodo de tiempo: 1 day, assessed per experimental condition (i.e. with orthosis and without orthosis)
|
This is the global thorax angle (with respect to vertical) during a lifting/lowering cycle. This will be a secondary outcome measure for able-bodied subjects. Extracted at peak trunk lean (bottom of the squat) across 10 post-fatigue cycles. Measured in degrees of decrease in global thorax angle compared with the no-exoskeleton condition. |
1 day, assessed per experimental condition (i.e. with orthosis and without orthosis)
|
|
Stair Ascent Gait Style
Periodo de tiempo: 1 day, assessed per experimental condition (i.e. with orthosis and without orthosis)
|
Step-over-step gait style will be evaluated for the exo condition vs no exo condition for elderly participants. |
1 day, assessed per experimental condition (i.e. with orthosis and without orthosis)
|
Colaboradores e Investigadores
Patrocinador
Publicaciones y enlaces útiles
Fechas de registro del estudio
Fechas importantes del estudio
Inicio del estudio (Actual)
Finalización primaria (Actual)
Finalización del estudio (Actual)
Fechas de registro del estudio
Enviado por primera vez
Primero enviado que cumplió con los criterios de control de calidad
Publicado por primera vez (Actual)
Actualizaciones de registros de estudio
Última actualización publicada (Actual)
Última actualización enviada que cumplió con los criterios de control de calidad
Última verificación
Más información
Términos relacionados con este estudio
Términos MeSH relevantes adicionales
Otros números de identificación del estudio
- HUM00201957
- R01EB031166 (Subvención/contrato del NIH de EE. UU.)
Plan de datos de participantes individuales (IPD)
¿Planea compartir datos de participantes individuales (IPD)?
Información sobre medicamentos y dispositivos, documentos del estudio
Estudia un producto farmacéutico regulado por la FDA de EE. UU.
Estudia un producto de dispositivo regulado por la FDA de EE. UU.
producto fabricado y exportado desde los EE. UU.
Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .