1-1-8 one-step sevoflurane wash-in scheme for low-flow anesthesia: simple, rapid, and predictable induction

Sirirat Tribuddharat, Thepakorn Sathitkarnmanee, Naruemon Vattanasiriporn, Maneerat Thananun, Duangthida Nonlhaopol, Wilawan Somdee, Sirirat Tribuddharat, Thepakorn Sathitkarnmanee, Naruemon Vattanasiriporn, Maneerat Thananun, Duangthida Nonlhaopol, Wilawan Somdee

Abstract

Background: Sevoflurane is suitable for low-flow anesthesia (LFA). LFA needs a wash-in phase. The reported sevoflurane wash-in schemes lack simplicity, target coverage, and applicability. We proposed a one-step 1-1-8 wash-in scheme for sevoflurane LFA to be used with both N2O and Air. The objective of our study was to identify time for achieving each level of alveolar concentration of sevoflurane (FAS) from 1 to 3.5% in both contexts.

Methods: We recruited 199 adults requiring general anesthesia with endotracheal intubation and controlled ventilation-102 in group N2O and 97 in group Air. After induction and intubation, a wash-in was started using a fresh gas flow of O2:N2O or O2:Air at 1:1 L·min- 1 plus sevoflurane 8%. The ventilation was controlled to maintain end-tidal CO2 of 30-35 mmHg.

Results: The rising patterns of FAS and inspired concentration of sevoflurane (FIS) are similar, running parallel between the groups. The FAS/FIS ratio increased from 0.46 to 0.72 within 260 s in group N2O and from 0.42 to 0.69 within 286 s in group Air. The respective time to achieve an FAS of 1, 1.5, 2, 2.5, 3, and 3.5% was 1, 1.5, 2, 3, 3.5, and 4.5 min in group N2O and 1, 1.5, 2, 3, 4, and 5 min in group Air. The heart rate and blood pressure of both groups significantly increased initially then gradually decreased as FAS increased.

Conclusions: The 1-1-8 wash-in scheme for sevoflurane LFA has many advantages, including simplicity, coverage, swiftness, safety, economy, and that it can be used with both N2O and Air. A respective FAS of 1, 1.5, 2, 2.5, 3, and 3.5% when used with N2O and Air can be expected at 1, 1.5, 2, 3, 3.5, and 4.5 min and 1, 1.5, 2, 3, 4, and 5 min.

Trial registration: This study was retrospectively registered with ClinicalTrials.gov (NCT03510013) on June 8, 2018.

Keywords: Air; Low flow anesthesia; Nitrous oxide; Sevoflurane; Wash-in.

Conflict of interest statement

The authors declare that they have no competing interests.

Figures

Fig. 1
Fig. 1
Trajectories of FAS vs. time to achieve each FAS during wash-in of group N2O. FAS, alveolar concentration of sevoflurane
Fig. 2
Fig. 2
Trajectories of FAS vs. time to achieve each FAS during wash-in of group Air. FAS, alveolar concentration of sevoflurane
Fig. 3
Fig. 3
Rising pattern of FAS and FIS of group N2O and group Air. FAS, alveolar concentration of sevoflurane; FIS, inspired concentration of sevoflurane
Fig. 4
Fig. 4
Rising pattern of FAS/FIS ratio of group N2O and group Air. FAS, alveolar concentration of sevoflurane; FIS, inspired concentration of sevoflurane
Fig. 5
Fig. 5
Pattern of changes in heart rate and blood pressure of group N2O and group Air at each FAS. p < 0.001 for all values. FAS, alveolar concentration of sevoflurane; BPs, systolic blood pressure; BPd, diastolic blood pressure; HR, heart rate

References

    1. Baum JA. Low-flow anesthesia: theory, practice, technical preconditions, advantages, and foreign gas accumulation. J Anesth. 1999;13:166–174. doi: 10.1007/s005400050050.
    1. Nickalls RW, Mapleson WW. Age-related iso-MAC charts for isoflurane, sevoflurane and desflurane in man. Br J Anaesth. 2003;91:170–174. doi: 10.1093/bja/aeg132.
    1. Lindqvist M, Jakobsson J. Minimal flow anesthesia for short elective day case surgery; high vaporiser settings are needed but still cost-effective. Ambul Surg. 2011;17:27–29.
    1. Horwitz M, Jakobsson JG. Desflurane and sevoflurane use during low- and minimal-flow anesthesia at fixed vaporizer settings. Minerva Anestesiol. 2016;82:180–185.
    1. Jakobsson P, Lindgren M, Jakobsson JG. Wash-in and wash-out of sevoflurane in a test-lung model: A comparison between Aisys and FLOW-i. F1000Res. 2017;6:389. doi: 10.12688/f1000research.11255.2.
    1. Young CJ, Apfelbaum JL. Inhalational anesthetics: desflurane and sevoflurane. J Clin Anesth. 1995;7:564–577. doi: 10.1016/0952-8180(95)00129-8.
    1. Ryu HG, Lee JH, Lee KK, Gil NS, Kim CS, Sim SE, et al. The effect of low fresh gas flow rate on sevoflurane consumption. Korean J Anesthesiol. 2011;60:75–77. doi: 10.4097/kjae.2011.60.2.75.
    1. Gentz BA, Malan TP., Jr Renal toxicity with sevoflurane: a storm in a teacup? Drugs. 2001;61:2155–2162. doi: 10.2165/00003495-200161150-00001.
    1. Marini F, Bellugi I, Gambi D, Pacenti M, Dugheri S, Focardi L, et al. Compound A, formaldehyde and methanol concentrations during low-flow sevoflurane anesthesia: comparison of three carbon dioxide absorbers. Acta Anaesthesiol Scand. 2007;51:625–632. doi: 10.1111/j.1399-6576.2007.01278.x.
    1. Honemann C, Hagemann O, Doll D. Inhalational anesthesia with low fresh gas flow. Indian J Anesth. 2013;57:345–350. doi: 10.4103/0019-5049.118569.
    1. Korman B, Mapleson WW. Concentration and second gas effects: can the accepted explanation be improved? Br J Anesth. 1997;78:618–625. doi: 10.1093/bja/78.5.618.
    1. Leijonhufvud F, Joneby F, Jakobsson JG. The impact of fresh gas flow on wash-in, wash-out time and gas consumption for sevoflurane and desflurane, comparing two anesthesia machines, a test-lung study. F1000Res. 2017;6:1997. doi: 10.12688/f1000research.13064.1.
    1. Shin HW, Yu HN, Bae GE, Huh H, Park JY, Kim JY. The effect of fresh gas flow rate and type of anesthesia machine on time to reach target sevoflurane concentration. BMC Anesthesiol. 2017;17:10. doi: 10.1186/s12871-016-0294-y.
    1. Vereecke HEM, Proost JH, Heyse B, Eleveld DJ, Katoh T, Luginbühl M, et al. Interaction between nitrous oxide, sevoflurane, and opioids: a response surface approach. Anesthesiology. 2013;118:894–902. doi: 10.1097/ALN.0b013e3182860486.
    1. Sathitkarnmanee T, Tribuddharat S, Suttinarakorn C, Nonlhaopol D, Thananun M, Somdee W, et al. 1-1-12 one-step wash-in scheme for desflurane-nitrous oxide low-flow anesthesia: rapid and predictable induction. Biomed Res Int. 2014;2014:867504. doi: 10.1155/2014/867504.
    1. Sathitkarnmanee T, Tribuddharat S, Nonlhaopol D, Thananun M, Somdee W. 1-1-12 one-step wash-in scheme for desflurane low flow anesthesia: performance without nitrous oxide. Drug Des Devel Ther. 2015;9:977–981. doi: 10.2147/DDDT.S78233.
    1. Hargasser S, Hipp R, Breinbauer B, Mielke L, Entholzner E, Rust M. A lower solubility recommends the use of desflurane more than isoflurane, halothane, and enflurane under low-flow conditions. J Clin Anesth. 1995;7:49–53. doi: 10.1016/0952-8180(94)00003-M.

Source: PubMed

3
Tilaa