CO2 driven endotracheal tube cuff control in critically ill patients: A randomized controlled study

Gennaro De Pascale, Mariano Alberto Pennisi, Maria Sole Vallecoccia, Giuseppe Bello, Riccardo Maviglia, Luca Montini, Valentina Di Gravio, Salvatore Lucio Cutuli, Giorgio Conti, Massimo Antonelli, Gennaro De Pascale, Mariano Alberto Pennisi, Maria Sole Vallecoccia, Giuseppe Bello, Riccardo Maviglia, Luca Montini, Valentina Di Gravio, Salvatore Lucio Cutuli, Giorgio Conti, Massimo Antonelli

Abstract

Background: To determine the safety and clinical efficacy of an innovative integrated airway system (AnapnoGuard™ 100 system) that continuously monitors and controls the cuff pressure (Pcuff), while facilitating the aspiration of subglottic secretions (SS).

Methods: This was a prospective, single centre, open-label, randomized, controlled feasibility and safety trial. The primary endpoint of the study was the rate of device related adverse events (AE) and serious AE (SAE) as a result of using AnapnoGuard (AG) 100 during mechanical ventilation. Secondary endpoints were: (1) mechanical complications rate (2) ICU staff satisfaction; (3) VAP occurrence; (4) length of mechanical ventilation; (5) length of Intensive Care Unit stay and mortality; (6) volume of evacuated subglottic secretions. Sixty patients were randomized to be intubated with the AG endotracheal-tube (ETT) and connected to the AG 100 system allowing Pcuff adjustment and SS aspiration; or with an ETT combined with SS drainage and Pcuff controlled manually.

Results: No difference in adverse events rate was identified between the groups. The use of AG system was associated with a significantly higher incidence of Pcuff determinations in the safety range (97.3% vs. 71%; p<0.01) and a trend to a greater volume of aspirated SS secretions: (192.0[64-413] ml vs. 150[50-200], p = 0.19 (total)); (57.8[20-88.7] ml vs. 50[18.7-62] ml, p = 0.11 (daily)). No inter-group difference was detected using AG system vs. controls in terms of post-extubation throat pain level (0 [0-2] vs. 0 [0-3]; p = 0.7), hoarseness (42.9% vs. 75%; p = 0.55) and tracheal mucosa oedema (16.7% vs. 10%; p = 0.65). Patients enrolled in the AG group had a trend to reduced VAP risk of ventilator-associated pneumonia(VAP) (14.8% vs. 40%; p = 0.06), which were more frequently monomicrobial (25% vs. 70%; p = 0.03). No statistically significant difference was observed in duration of mechanical ventilation, ICU stay, and mortality.

Conclusions: The use AG 100 system and AG tube in critically ill intubated patients is safe and effective in Pcuff control and SS drainage. Its protective role against VAP needs to be confirmed in a larger randomized trial.

Trial registration: ClinicalTrials.gov NCT01550978. Date of registration: February 21, 2012.

Conflict of interest statement

Competing Interests: The authors have declared that no competing interests exist.

Figures

Fig 1. Flow chart of study inclusion…
Fig 1. Flow chart of study inclusion process.
ED, Emergency Department; BMI, Body Mass Index *Patients connected to either AG 110 system or subglottic suction system after 6 hours from tracheal intubation **The reasons for excluding patients at admission were: suspicion of pneumonia and lung contusion on chest X ray (n = 375); mechanical ventilation during previous three months (n = 58); facial, oropharingeal and neck trauma (n = 123); difficult intubation (n = 28); BMI >40 (n = 16).
Fig 2. Comparison between AG group and…
Fig 2. Comparison between AG group and control group according to total SS drained.
Fig 3. Comparison between AG group and…
Fig 3. Comparison between AG group and control group according to daily SS drained.
AG, Anapnoguard; Pcuff, Cuff Pressure; SS, Subglottic Secretions.
Fig 4. Cumulative rates of patients remaining…
Fig 4. Cumulative rates of patients remaining free of VAP in the AG group and control group, using the Kaplan-Meier method.
VAP, Ventilator-Associated Pneumonia; AG, Anapnoguard.
Fig 5. AnapnoGuard 100 system overview.
Fig 5. AnapnoGuard 100 system overview.
Fig 6. Automatic endotracheal tube cuff pressure…
Fig 6. Automatic endotracheal tube cuff pressure closed loop adjustment.
Fig 7. Subglottic secretions drainage based on…
Fig 7. Subglottic secretions drainage based on rinsing and suctioning.

References

    1. Stauffer JL. Complications of translaryngeal intubation In: McGraw-Hill IE, editor. Principles and Practice of Mechanical Ventilation. First Edition ed. New York: MJ Tobin; 1994. p. 711–48.
    1. Estes RJ, Meduri GU. The pathogenesis of ventilator-associated pneumonia: I. Mechanisms of bacterial transcolonization and airway inoculation. Intensive care medicine. 1995;21(4):365–83. Epub 1995/04/01.
    1. Craven DE. Nosocomial pneumonia: new concepts on an old disease. Infection control and hospital epidemiology. 1988;9(2):57–8. Epub 1988/02/01.
    1. Duguet A, D'Amico L, Biondi G, Prodanovic H, Gonzalez-Bermejo J, Similowski T. Control of tracheal cuff pressure: a pilot study using a pneumatic device. Intensive care medicine. 2007;33(1):128–32. Epub 2006/10/26. 10.1007/s00134-006-0417-x
    1. Nseir S, Brisson H, Marquette CH, Chaud P, Di Pompeo C, Diarra M, et al. Variations in endotracheal cuff pressure in intubated critically ill patients: prevalence and risk factors. European journal of anaesthesiology. 2009;26(3):229–34. Epub 2009/02/28.
    1. Seegobin RD, van Hasselt GL. Endotracheal cuff pressure and tracheal mucosal blood flow: endoscopic study of effects of four large volume cuffs. Br Med J 1984;288(6422):965–8. Epub 1984/03/31.
    1. Deslee G, Brichet A, Lebuffe G, Copin MC, Ramon P, Marquette CH. Obstructive fibrinous tracheal pseudomembrane. A potentially fatal complication of tracheal intubation. American journal of respiratory and critical care medicine. 2000;162(3 Pt 1):1169–71. Epub 2000/09/16. 10.1164/ajrccm.162.3.9910047
    1. Rello J, Sonora R, Jubert P, Artigas A, Rue M, Valles J. Pneumonia in intubated patients: role of respiratory airway care. American journal of respiratory and critical care medicine. 1996;154(1):111–5. Epub 1996/07/01. 10.1164/ajrccm.154.1.8680665
    1. Guidelines for the management of adults with hospital-acquired, ventilator-associated, and healthcare-associated pneumonia. American journal of respiratory and critical care medicine. 2005;171(4):388–416. Epub 2005/02/09. 10.1164/rccm.200405-644ST
    1. Dezfulian C, Shojania K, Collard HR, Kim HM, Matthay MA, Saint S. Subglottic secretion drainage for preventing ventilator-associated pneumonia: a meta-analysis. The American journal of medicine. 2005;118(1):11–8. Epub 2005/01/11. 10.1016/j.amjmed.2004.07.051
    1. Pirrone M, Pinciroli R, Berra L. Microbiome, biofilms, and pneumonia in the ICU. Current opinion in infectious diseases. 2016;29(2):160–6. Epub 2016/02/09. 10.1097/QCO.0000000000000255
    1. Efrati S, Deutsch I, Weksler N, Gurman GM. Detection of endobronchial intubation by monitoring the CO level above the endotracheal cuff. Journal of clinical monitoring and computing. 2014. Epub 2014/05/30.
    1. Efrati S, Deutsch I, Gurman GM, Noff M, Conti G. Tracheal pressure and endotracheal tube obstruction can be detected by continuous cuff pressure monitoring: in vitro pilot study. Intensive care medicine. 2010;36(6):984–90. Epub 2010/03/17. 10.1007/s00134-010-1835-3
    1. Efrati S, Leonov Y, Oron A, Siman-Tov Y, Averbukh M, Lavrushevich A, et al. Optimization of endotracheal tube cuff filling by continuous upper airway carbon dioxide monitoring. Anesthesia and analgesia. 2005;101(4):1081–8. Epub 2005/09/30. 10.1213/01.ane.0000167641.64815.1a
    1. Le Gall JR, Lemeshow S, Saulnier F. A new Simplified Acute Physiology Score (SAPS II) based on a European/North American multicenter study. Jama. 1993;270(24):2957–63. Epub 1993/12/22.
    1. Ferreira FL, Bota DP, Bross A, Melot C, Vincent JL. Serial evaluation of the SOFA score to predict outcome in critically ill patients. Jama. 2001;286(14):1754–8. Epub 2001/10/12.
    1. File TM Jr. Recommendations for treatment of hospital-acquired and ventilator-associated pneumonia: review of recent international guidelines. Clinical infectious diseases: an official publication of the Infectious Diseases Society of America. 2010;51 Suppl 1:S42–7. Epub 2010/07/06.
    1. Dellinger RP, Levy MM, Rhodes A, Annane D, Gerlach H, Opal SM, et al. Surviving Sepsis Campaign: international guidelines for management of severe sepsis and septic shock, 2012. Intensive care medicine. 2013;39(2):165–228. Epub 2013/01/31. 10.1007/s00134-012-2769-8
    1. Loeser EA, Hodges M, Gliedman J, Stanley TH, Johansen RK, Yonetani D. Tracheal pathology following short-term intubation with low- and high-pressure endotracheal tube cuffs. Anesthesia and analgesia. 1978;57(5):577–9. Epub 1978/09/01.
    1. Nseir S, Duguet A, Copin MC, De Jonckheere J, Zhang M, Similowski T, et al. Continuous control of endotracheal cuff pressure and tracheal wall damage: a randomized controlled animal study. Critical care (London, England). 2007;11(5):R109. Epub 2007/10/05.
    1. Smulders K, van der Hoeven H, Weers-Pothoff I, Vandenbroucke-Grauls C. A randomized clinical trial of intermittent subglottic secretion drainage in patients receiving mechanical ventilation. Chest. 2002;121(3):858–62. Epub 2002/03/13.
    1. Berra L, De Marchi L, Panigada M, Yu ZX, Baccarelli A, Kolobow T. Evaluation of continuous aspiration of subglottic secretion in an in vivo study. Critical care medicine. 2004;32(10):2071–8. Epub 2004/10/16.
    1. Nseir S, Lorente L, Ferrer M, Rouze A, Gonzalez O, Bassi GL, et al. Continuous control of tracheal cuff pressure for VAP prevention: a collaborative meta-analysis of individual participant data. Annals of intensive care. 2015;5(1):43 Epub 2015/11/26. 10.1186/s13613-015-0087-3
    1. Muscedere J, Rewa O, McKechnie K, Jiang X, Laporta D, Heyland DK. Subglottic secretion drainage for the prevention of ventilator-associated pneumonia: a systematic review and meta-analysis. Critical care medicine. 2011;39(8):1985–91. Epub 2011/04/12. 10.1097/CCM.0b013e318218a4d9
    1. Valencia M, Ferrer M, Farre R, Navajas D, Badia JR, Nicolas JM, et al. Automatic control of tracheal tube cuff pressure in ventilated patients in semirecumbent position: a randomized trial. Critical care medicine. 2007;35(6):1543–9. Epub 2007/04/25. 10.1097/01.CCM.0000266686.95843.7D
    1. Nseir S, Zerimech F, Fournier C, Lubret R, Ramon P, Durocher A, et al. Continuous control of tracheal cuff pressure and microaspiration of gastric contents in critically ill patients. American journal of respiratory and critical care medicine. 2011;184(9):1041–7. Epub 2011/08/13. 10.1164/rccm.201104-0630OC
    1. Lorente L, Lecuona M, Jimenez A, Lorenzo L, Roca I, Cabrera J, et al. Continuous endotracheal tube cuff pressure control system protects against ventilator-associated pneumonia. Critical care (London, England). 2014;18(2):R77. Epub 2014/04/23.
    1. Vallecoccia MS, De Pascale G, Cutuli SL, Di Gravio V, Pennisi MA, Antonelli M. Endotracheal tubes cuff pressure control: does the CO2 matter? Minerva anestesiologica. 2015;81(3):352–3. Epub 2014/11/07.
    1. Lorente L, Lecuona M, Jimenez A, Mora ML, Sierra A. Influence of an endotracheal tube with polyurethane cuff and subglottic secretion drainage on pneumonia. American journal of respiratory and critical care medicine. 2007;176(11):1079–83. Epub 2007/09/18. 10.1164/rccm.200705-761OC
    1. Caroff DA, Li L, Muscedere J, Klompas M. Subglottic Secretion Drainage and Objective Outcomes: A Systematic Review and Meta-Analysis. Critical care medicine. 2016;44(4):830–40. Epub 2015/12/10.
    1. Zanella A, Scaravilli V, Isgro S, Milan M, Cressoni M, Patroniti N, et al. Fluid leakage across tracheal tube cuff, effect of different cuff material, shape, and positive expiratory pressure: a bench-top study. Intensive care medicine. 2011;37(2):343–7. Epub 2010/12/15. 10.1007/s00134-010-2106-z
    1. Dullenkopf A, Gerber A, Weiss M. Fluid leakage past tracheal tube cuffs: evaluation of the new Microcuff endotracheal tube. Intensive care medicine. 2003;29(10):1849–53. Epub 2003/08/19. 10.1007/s00134-003-1933-6
    1. Philippart F, Gaudry S, Quinquis L, Lau N, Ouanes I, Touati S, et al. Randomized intubation with polyurethane or conical cuffs to prevent pneumonia in ventilated patients. American journal of respiratory and critical care medicine. 2015;191(6):637–45. Epub 2015/01/15. 10.1164/rccm.201408-1398OC
    1. Chenelle CT, Fisher DF, Kacmarek RM, Berra L. Procedural Considerations on the Use of Polyurethane and/or Conical Cuffs. American journal of respiratory and critical care medicine. 2015;192(5):639–40. Epub 2015/11/13.

Source: PubMed

3