Reliability of a New Pulse Contour Technique for Diagnosing an Increase in Stroke Volume During a Fluid Challenge for Hemodynamic Optimization in Patients Scheduled for High-risk Abdominal Surgery: Comparison With Transthoracic Echocardiography: COMPARE Study (COMPARE)
Aperçu de l'étude
Statut
Les conditions
Intervention / Traitement
Description détaillée
In European countries, day-7 mortality rate could reach 1-5. %. Emergency conditions, American Society Assessment (ASA) and major surgery were associated with a higher risk of immediate mortality. Moreover, hemodynamic instability is well known to lead to postoperative complications and a higher risk of mortality.
Since, nearly 30 years, optimization of hemodynamic conditions, especially via an optimization of the fluid administration has been shown to improve immediate and long-term patient outcome. This strategy is currently widely accepted and recommended in medium and high-risk surgery. Therefore, optimizing cardiac output and one of its surrogate is proposed over intraoperative surgery. For assessing the cardiac output or the indexed stroke volume, esophageal Doppler and pulse contour technique have been proposed and have shown that they could improve patient outcome.
However, the reliability of esophageal Doppler, pulse contour and non-invasive techniques assessing cardiac output and its ability to detect a change in the initial value of CO have been challenged. Initially, the reliability of a technique for measuring CO has been described by using a correlation coefficient and by building Bland & Altman diagram. However, searching a correlation between two different techniques measuring the same parameter will lead to a correlation. In the same, Bland & Altman technique shows the mean difference between measurements of the same parameters; However, demonstrating a good reliability should be shown by a narrow limit of agreement with no definition of the narrowness. Indeed, a value of cardiac index (CI) = 3 l/min/m2 with a mean difference of 0 +/- 1.1 /min/m2 by Bland & Altman technique mean that the CI value could be within 1 and 5 l/min/m2 that could lead to different treatment such as fluid challenge, vasopressor or inotrope infusion.
An interesting method for comparing two techniques measuring the same parameter could be to challenge one technique versus the other one in decision-making. Applying this method to techniques measuring CO, we could search for the interest of a new technique for diagnosing an increase in CO after a fluid challenge for optimizing cardiac preload and hemodynamic status. Esophageal Doppler is classically recommended for optimizing hemodynamic patient in high-risk surgery.
A new device using pulse contour technology has been available since a few years. Therefore, the present study aims at assessing the ability of this device for diagnosing an increase in CO > 15% during patient hemodynamic optimization by fluid challenge.
Type d'étude
Inscription (Réel)
Contacts et emplacements
Lieux d'étude
-
-
-
Nîmes, France, 30029
- CHU de Nîmes
-
-
Critères de participation
Critère d'éligibilité
Âges éligibles pour étudier
Accepte les volontaires sains
Méthode d'échantillonnage
Population étudiée
La description
Inclusion Criteria:
- Adult patient ASA status 1-3 who were orally informed and did refuse to participate.
- Patient in whom general anesthesia with tracheal intubation were planned.
- Patient in whom an arterial catheter and a monitoring of cardiac output by esophageal Doppler were indicated for hemodynamic optimization
- Patient with cardiac sinusal mode
Exclusion Criteria:
- Patient < 18-year-old
- Cardiac arrythmia
- Patient with anomaly in oro-pharyngo-esophageal tractus
- Patients with hemostasis anomaly (PT < 30%, platelets < 50 000 elements/mm3)
- Patient in whom the cardiac output monitoring or measurement was not possible by Esophageal Doppler or echography
- Patients in whom cardiac arrythmia occurred during the procedure of hemodynamic optimization.
Plan d'étude
Comment l'étude est-elle conçue ?
Détails de conception
- Modèles d'observation: Cohorte
- Perspectives temporelles: Éventuel
Cohortes et interventions
Groupe / Cohorte |
Intervention / Traitement |
|---|---|
|
Hemodynamic optimization
Patients scheduled for intermediate and high-risk abdominal surgery were eligible to participate
|
The stroke volume will be measured before and after 15 minutes fluid challenge with crystalloids
|
Que mesure l'étude ?
Principaux critères de jugement
Mesure des résultats |
Description de la mesure |
Délai |
|---|---|---|
|
Ability of pulse contour for diagnosing a >15% increase in stroke volume
Délai: 15 minutes
|
Ability of pulse contour for diagnosing a >15% increase in stroke volume
|
15 minutes
|
Mesures de résultats secondaires
Mesure des résultats |
Description de la mesure |
Délai |
|---|---|---|
|
Comparison with the ability of other techniques
Délai: 15 minutes
|
Comparison with the ability of other techniques (Esophagial Doppler, other pulse contour techniques for diagnosing a >15% SV increase)
|
15 minutes
|
|
Ability for predicting a >15% SV increase
Délai: 15 minutes
|
Ability of Esophagial Doppler and Pulse contour techniques for predicting a >15% SV increase after rapid infusion of 100ml cristalloid over one minute (mini-fluid challenge)
|
15 minutes
|
Collaborateurs et enquêteurs
Dates d'enregistrement des études
Dates principales de l'étude
Début de l'étude (Réel)
Achèvement primaire (Réel)
Achèvement de l'étude (Réel)
Dates d'inscription aux études
Première soumission
Première soumission répondant aux critères de contrôle qualité
Première publication (Réel)
Mises à jour des dossiers d'étude
Dernière mise à jour publiée (Réel)
Dernière mise à jour soumise répondant aux critères de contrôle qualité
Dernière vérification
Plus d'information
Termes liés à cette étude
Termes MeSH pertinents supplémentaires
Autres numéros d'identification d'étude
- IRB 15.11.02
Plan pour les données individuelles des participants (IPD)
Prévoyez-vous de partager les données individuelles des participants (DPI) ?
Informations sur les médicaments et les dispositifs, documents d'étude
Étudie un produit pharmaceutique réglementé par la FDA américaine
Étudie un produit d'appareil réglementé par la FDA américaine
Ces informations ont été extraites directement du site Web clinicaltrials.gov sans aucune modification. Si vous avez des demandes de modification, de suppression ou de mise à jour des détails de votre étude, veuillez contacter register@clinicaltrials.gov. Dès qu'un changement est mis en œuvre sur clinicaltrials.gov, il sera également mis à jour automatiquement sur notre site Web .