Evaluation of Leak Integrity of the LapBox® Manual Containment System During Laparoscopic Surgery

This study to evaluate the leak integrity of the LapBox Manual Containment System for laparoscopic hysterectomy and myomectomy. The goal is to determine the integrity of this containment system in order to increase surgical safety. 30 adult females who require laparoscopic hysterectomy or myomectomy for management of uterine fibroids will be enrolled and will be on study the day of surgery.

Aperçu de l'étude

Statut

Pas encore de recrutement

Les conditions

Description détaillée

The goal is to determine the integrity of the LapBox Manual Containment system in order to increase surgical safety of morcellation of uterine fibroids during laparoscopic hysterectomy and myomectomy.

Containment systems are imperative for laparoscopic morcellation due to the potential for tissue dissemination, particularly in cases of undiagnosed malignancy or benign tissue fragments during morcellation of uterine fibroids.

Integrity of the LapBox Manual Containment System will be evaluated by indicating the presence or absence of leakage from testing with methylene blue dye after morcellation. The LapBox and standard of care devices will be moved away from the surgical field prior to methylene blue being added to determine if there is breakage. The methylene blue dye will not come in contact with the participant or spill onto the surgical field.

The hypothesis is that laparoscopic hysterectomies and myomectomies performed using the LapBox Manual Containment System will result in less leakage when compared to the control containment system.

The primary endpoint will be presence or absence of leakage from the LapBox and the control. This will be directly visualized intraoperatively after morcellation takes place.

The secondary endpoint will be the incidence of perforation or rupture incidence in the LapBox or standard of care device and descriptive comments from surgeons for usability feedback.

Type d'étude

Interventionnel

Inscription (Estimé)

24

Phase

  • N'est pas applicable

Contacts et emplacements

Cette section fournit les coordonnées de ceux qui mènent l'étude et des informations sur le lieu où cette étude est menée.

Coordonnées de l'étude

Lieux d'étude

    • Wisconsin
      • Madison, Wisconsin, États-Unis, 53715
        • Meriter

Critères de participation

Les chercheurs recherchent des personnes qui correspondent à une certaine description, appelée critères d'éligibilité. Certains exemples de ces critères sont l'état de santé général d'une personne ou des traitements antérieurs.

Critère d'éligibilité

Âges éligibles pour étudier

  • Adulte
  • Adulte plus âgé

Accepte les volontaires sains

Non

La description

Inclusion Criteria:

  • Participants who require laparoscopic hysterectomy or myomectomy for management of uterine fibroids

Exclusion Criteria:

-

Plan d'étude

Cette section fournit des détails sur le plan d'étude, y compris la façon dont l'étude est conçue et ce que l'étude mesure.

Comment l'étude est-elle conçue ?

Détails de conception

  • Objectif principal: Traitement
  • Répartition: Randomisé
  • Modèle interventionnel: Affectation parallèle
  • Masquage: Double

Armes et Interventions

Groupe de participants / Bras
Intervention / Traitement
Expérimental: LapBox Manual Containment System
N=12
The LapBox Manual introduces a double wall, pressure resistant system to maintain safe separation between the cutting instrument and chamber wall. The LapBox Manual is already FDA-approved for use so this will be a post-market analysis.
Autres noms:
  • LapBox Manual
Comparateur actif: Standard Specimen Bag
N=12
Meriter has an endoscopic specimen retrieval bag which will be the standard of care specimen bag depending on the size of the uterine fibroids. The bag can be deployed through a 15 mm port. The standard bag is selected depending on the size of the uterine fibroids.
Autres noms:
  • endoscopic specimen retrieval bag

Que mesure l'étude ?

Principaux critères de jugement

Mesure des résultats
Description de la mesure
Délai
Presence or Absence of Leakage
Délai: data collected intraoperatively (on day of surgery, only study visit)
Integrity will be evaluated by indicating the presence or absence of leakage after morcellation and subsequent testing with methylene blue dye during laparoscopic hysterectomy or myomectomy.
data collected intraoperatively (on day of surgery, only study visit)

Mesures de résultats secondaires

Mesure des résultats
Délai
Incidence of Perforation or Rupture
Délai: data collected intraoperatively (on day of surgery, only study visit)
data collected intraoperatively (on day of surgery, only study visit)

Collaborateurs et enquêteurs

C'est ici que vous trouverez les personnes et les organisations impliquées dans cette étude.

Les enquêteurs

  • Chercheur principal: Bala Bhagavath, MD, UW School of Medicine and Public Health

Dates d'enregistrement des études

Ces dates suivent la progression des dossiers d'étude et des soumissions de résultats sommaires à ClinicalTrials.gov. Les dossiers d'étude et les résultats rapportés sont examinés par la Bibliothèque nationale de médecine (NLM) pour s'assurer qu'ils répondent à des normes de contrôle de qualité spécifiques avant d'être publiés sur le site Web public.

Dates principales de l'étude

Début de l'étude (Estimé)

1 septembre 2026

Achèvement primaire (Estimé)

1 juillet 2027

Achèvement de l'étude (Estimé)

1 juillet 2027

Dates d'inscription aux études

Première soumission

11 août 2026

Première soumission répondant aux critères de contrôle qualité

11 août 2026

Première publication (Réel)

17 août 2026

Mises à jour des dossiers d'étude

Dernière mise à jour publiée (Réel)

17 août 2026

Dernière mise à jour soumise répondant aux critères de contrôle qualité

11 août 2026

Dernière vérification

1 août 2026

Plus d'information

Termes liés à cette étude

Autres numéros d'identification d'étude

  • 2026-0071
  • Protocol Version 6/22/26 (Autre identifiant: UW Madison)
  • SMPH | OBGYN Administration (Autre identifiant: UW Madison)
  • 238739 (Autre identifiant: OnCore ID)

Plan pour les données individuelles des participants (IPD)

Prévoyez-vous de partager les données individuelles des participants (DPI) ?

NON

Informations sur les médicaments et les dispositifs, documents d'étude

Étudie un produit pharmaceutique réglementé par la FDA américaine

Non

Étudie un produit d'appareil réglementé par la FDA américaine

Oui

produit fabriqué et exporté des États-Unis.

Oui

Ces informations ont été extraites directement du site Web clinicaltrials.gov sans aucune modification. Si vous avez des demandes de modification, de suppression ou de mise à jour des détails de votre étude, veuillez contacter register@clinicaltrials.gov. Dès qu'un changement est mis en œuvre sur clinicaltrials.gov, il sera également mis à jour automatiquement sur notre site Web .

S'abonner