Respiration Rate V2.0 in a Hospital Setting
Evaluation of Respiration Rate Parameters in a Hospital Population Using a Nellcor Bedside Respiratory Patient Monitoring System With Respiration Rate Version 2.0
調査の概要
状態
条件
詳細な説明
Volunteer subjects from a hospital setting (see specific inclusion/exclusion criteria) over the age of 18, male and female, who provide written informed consent prior to the procedure, and are without any contact skin allergies to adhesives found in standard pulse oximetry sensors. A cross section of demographics - cardiovascular, respiratory, and metabolic type cases should be considered for a broad distribution of medical and surgical patients.
The study consists of measuring parameters related to respiration in hospitalized volunteers by means of the Bedside Respiratory Patient Monitoring System with RR V2.0 (Nightingale). The production-equivalent Nightingale device and multiparameter reference devices will be attached to the subject via sensors, nasal cannulas and ECG leads and the information from all devices will be captured continuously for up to 40 minutes to allow for at least 30 minutes of artifact free data 30 minutes.
研究の種類
入学 (実際)
連絡先と場所
研究場所
-
-
Ohio
-
Columbus、Ohio、アメリカ、43210
- The Ohio State University Wexner Medical Center
-
-
参加基準
適格基準
就学可能な年齢
健康ボランティアの受け入れ
受講資格のある性別
サンプリング方法
調査対象母集団
説明
Inclusion Criteria:
- Male or female.
- 18 years or older.
- A patient at the study site on a General Care Floor or General Surgery Care Floor.
- Subject is willing and able to provide written consent.
Exclusion Criteria:
- Subject is younger than 18.
- Subject has severe contact allergies that may cause a reaction to standard adhesive materials found in pulse oximetry sensors, ECG electrodes, respiration monitor electrodes or other medical sensors.
- Subject has an abnormality that may prevent proper application of the device.
- Subject is in atrial fibrillation.
- Subject has a documented history of frequent premature ventricular contractions (PVCs), defined as greater than 3 in 30 seconds or greater than 6 in 60 seconds.
- Subject has an implanted pacemaker.
- Subject is unwilling or unable to sign informed consent.
研究計画
研究はどのように設計されていますか?
デザインの詳細
コホートと介入
グループ/コホート |
|---|
|
治療なし
無治療、前向き観察
|
この研究は何を測定していますか?
主要な結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Mean Error (ME) +/- 1 Breath Per Minute, RR Sensor
時間枠:up to 40 minutes of continuous monitoring
|
The software calculates respiration rate values via Adult Respiratory Sensor with a mean error of +/- 1 breath per minute relative to a capnography based reference.
The Mean Error value estimates the mean difference in simultaneous estimates of respiration rate (the Respiration Rate calculated from the sensor and the Respiration Rate calculated from capnography).
|
up to 40 minutes of continuous monitoring
|
二次結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Mean Error (ME) +/- 1 Breath Per Minute, Max-N Sensor
時間枠:up to 40 minutes of continuous monitoring
|
The software calculates respiration rate values via the Max-N Sensor with a mean error of +/- 1 breath per minute relative to a capnography based reference.
The Mean Error value estimates the mean difference in simultaneous estimates of respiration rate (the Respiration Rate calculated from the sensor and the Respiration Rate calculated from capnography).
The error value estimates the mean difference in simultaneous estimates of respiration rate (the Respiration Rate calculated from the Max-N sensor minus the Respiration Rate calculated from capnography.
|
up to 40 minutes of continuous monitoring
|
協力者と研究者
スポンサー
研究記録日
主要日程の研究
研究開始
一次修了 (実際)
研究の完了 (実際)
試験登録日
最初に提出
QC基準を満たした最初の提出物
最初の投稿 (見積もり)
学習記録の更新
投稿された最後の更新 (実際)
QC基準を満たした最後の更新が送信されました
最終確認日
詳しくは
本研究に関する用語
その他の研究ID番号
- COVMOPR0399
この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。