Deze pagina is automatisch vertaald en de nauwkeurigheid van de vertaling kan niet worden gegarandeerd. Raadpleeg de Engelse versie voor een brontekst.

Respiration Rate V2.0 in a Hospital Setting

19 juni 2017 bijgewerkt door: Medtronic - MITG

Evaluation of Respiration Rate Parameters in a Hospital Population Using a Nellcor Bedside Respiratory Patient Monitoring System With Respiration Rate Version 2.0

The purpose of this study is to meet the respiration rate accuracy specifications for the study pulse oximeter monitoring system (study device) in a hospitalized volunteer population when compared to established technologies for measuring parameters related to respiration, such as Capnograph Respiration Rate (RR)(CO2).

Studie Overzicht

Toestand

Voltooid

Gedetailleerde beschrijving

Volunteer subjects from a hospital setting (see specific inclusion/exclusion criteria) over the age of 18, male and female, who provide written informed consent prior to the procedure, and are without any contact skin allergies to adhesives found in standard pulse oximetry sensors. A cross section of demographics - cardiovascular, respiratory, and metabolic type cases should be considered for a broad distribution of medical and surgical patients.

The study consists of measuring parameters related to respiration in hospitalized volunteers by means of the Bedside Respiratory Patient Monitoring System with RR V2.0 (Nightingale). The production-equivalent Nightingale device and multiparameter reference devices will be attached to the subject via sensors, nasal cannulas and ECG leads and the information from all devices will be captured continuously for up to 40 minutes to allow for at least 30 minutes of artifact free data 30 minutes.

Studietype

Observationeel

Inschrijving (Werkelijk)

90

Contacten en locaties

In dit gedeelte vindt u de contactgegevens van degenen die het onderzoek uitvoeren en informatie over waar dit onderzoek wordt uitgevoerd.

Studie Locaties

    • Ohio
      • Columbus, Ohio, Verenigde Staten, 43210
        • The Ohio State University Wexner Medical Center

Deelname Criteria

Onderzoekers zoeken naar mensen die aan een bepaalde beschrijving voldoen, de zogenaamde geschiktheidscriteria. Enkele voorbeelden van deze criteria zijn iemands algemene gezondheidstoestand of eerdere behandelingen.

Geschiktheidscriteria

Leeftijden die in aanmerking komen voor studie

18 jaar en ouder (Volwassen, Oudere volwassene)

Accepteert gezonde vrijwilligers

Nee

Geslachten die in aanmerking komen voor studie

Allemaal

Bemonsteringsmethode

Niet-waarschijnlijkheidssteekproef

Studie Bevolking

Volunteer subjects from a hospital setting on the general care floor or general surgical care floor

Beschrijving

Inclusion Criteria:

  • Male or female.
  • 18 years or older.
  • A patient at the study site on a General Care Floor or General Surgery Care Floor.
  • Subject is willing and able to provide written consent.

Exclusion Criteria:

  • Subject is younger than 18.
  • Subject has severe contact allergies that may cause a reaction to standard adhesive materials found in pulse oximetry sensors, ECG electrodes, respiration monitor electrodes or other medical sensors.
  • Subject has an abnormality that may prevent proper application of the device.
  • Subject is in atrial fibrillation.
  • Subject has a documented history of frequent premature ventricular contractions (PVCs), defined as greater than 3 in 30 seconds or greater than 6 in 60 seconds.
  • Subject has an implanted pacemaker.
  • Subject is unwilling or unable to sign informed consent.

Studie plan

Dit gedeelte bevat details van het studieplan, inclusief hoe de studie is opgezet en wat de studie meet.

Hoe is de studie opgezet?

Ontwerpdetails

Cohorten en interventies

Groep / Cohort
geen behandeling
geen behandeling, prospectief observationeel

Wat meet het onderzoek?

Primaire uitkomstmaten

Uitkomstmaat
Maatregel Beschrijving
Tijdsspanne
Mean Error (ME) +/- 1 Breath Per Minute, RR Sensor
Tijdsspanne: up to 40 minutes of continuous monitoring
The software calculates respiration rate values via Adult Respiratory Sensor with a mean error of +/- 1 breath per minute relative to a capnography based reference. The Mean Error value estimates the mean difference in simultaneous estimates of respiration rate (the Respiration Rate calculated from the sensor and the Respiration Rate calculated from capnography).
up to 40 minutes of continuous monitoring

Secundaire uitkomstmaten

Uitkomstmaat
Maatregel Beschrijving
Tijdsspanne
Mean Error (ME) +/- 1 Breath Per Minute, Max-N Sensor
Tijdsspanne: up to 40 minutes of continuous monitoring
The software calculates respiration rate values via the Max-N Sensor with a mean error of +/- 1 breath per minute relative to a capnography based reference. The Mean Error value estimates the mean difference in simultaneous estimates of respiration rate (the Respiration Rate calculated from the sensor and the Respiration Rate calculated from capnography). The error value estimates the mean difference in simultaneous estimates of respiration rate (the Respiration Rate calculated from the Max-N sensor minus the Respiration Rate calculated from capnography.
up to 40 minutes of continuous monitoring

Medewerkers en onderzoekers

Hier vindt u mensen en organisaties die betrokken zijn bij dit onderzoek.

Studie record data

Deze datums volgen de voortgang van het onderzoeksdossier en de samenvatting van de ingediende resultaten bij ClinicalTrials.gov. Studieverslagen en gerapporteerde resultaten worden beoordeeld door de National Library of Medicine (NLM) om er zeker van te zijn dat ze voldoen aan specifieke kwaliteitscontrolenormen voordat ze op de openbare website worden geplaatst.

Bestudeer belangrijke data

Studie start

1 maart 2013

Primaire voltooiing (Werkelijk)

1 augustus 2013

Studie voltooiing (Werkelijk)

1 oktober 2013

Studieregistratiedata

Eerst ingediend

17 april 2013

Eerst ingediend dat voldeed aan de QC-criteria

18 april 2013

Eerst geplaatst (Schatting)

23 april 2013

Updates van studierecords

Laatste update geplaatst (Werkelijk)

17 juli 2017

Laatste update ingediend die voldeed aan QC-criteria

19 juni 2017

Laatst geverifieerd

1 juni 2017

Meer informatie

Termen gerelateerd aan deze studie

Andere studie-ID-nummers

  • COVMOPR0399

Deze informatie is zonder wijzigingen rechtstreeks van de website clinicaltrials.gov gehaald. Als u verzoeken heeft om uw onderzoeksgegevens te wijzigen, te verwijderen of bij te werken, neem dan contact op met register@clinicaltrials.gov. Zodra er een wijziging wordt doorgevoerd op clinicaltrials.gov, wordt deze ook automatisch bijgewerkt op onze website .

Abonneren