Respiration Rate V2.0 in a Hospital Setting
Evaluation of Respiration Rate Parameters in a Hospital Population Using a Nellcor Bedside Respiratory Patient Monitoring System With Respiration Rate Version 2.0
연구 개요
상태
정황
상세 설명
Volunteer subjects from a hospital setting (see specific inclusion/exclusion criteria) over the age of 18, male and female, who provide written informed consent prior to the procedure, and are without any contact skin allergies to adhesives found in standard pulse oximetry sensors. A cross section of demographics - cardiovascular, respiratory, and metabolic type cases should be considered for a broad distribution of medical and surgical patients.
The study consists of measuring parameters related to respiration in hospitalized volunteers by means of the Bedside Respiratory Patient Monitoring System with RR V2.0 (Nightingale). The production-equivalent Nightingale device and multiparameter reference devices will be attached to the subject via sensors, nasal cannulas and ECG leads and the information from all devices will be captured continuously for up to 40 minutes to allow for at least 30 minutes of artifact free data 30 minutes.
연구 유형
등록 (실제)
연락처 및 위치
연구 장소
-
-
Ohio
-
Columbus, Ohio, 미국, 43210
- The Ohio State University Wexner Medical Center
-
-
참여기준
자격 기준
공부할 수 있는 나이
건강한 자원 봉사자를 받아들입니다
연구 대상 성별
샘플링 방법
연구 인구
설명
Inclusion Criteria:
- Male or female.
- 18 years or older.
- A patient at the study site on a General Care Floor or General Surgery Care Floor.
- Subject is willing and able to provide written consent.
Exclusion Criteria:
- Subject is younger than 18.
- Subject has severe contact allergies that may cause a reaction to standard adhesive materials found in pulse oximetry sensors, ECG electrodes, respiration monitor electrodes or other medical sensors.
- Subject has an abnormality that may prevent proper application of the device.
- Subject is in atrial fibrillation.
- Subject has a documented history of frequent premature ventricular contractions (PVCs), defined as greater than 3 in 30 seconds or greater than 6 in 60 seconds.
- Subject has an implanted pacemaker.
- Subject is unwilling or unable to sign informed consent.
공부 계획
연구는 어떻게 설계됩니까?
디자인 세부사항
코호트 및 개입
그룹/코호트 |
|---|
|
치료 없음
치료 없음, 전향적 관찰
|
연구는 무엇을 측정합니까?
주요 결과 측정
결과 측정 |
측정값 설명 |
기간 |
|---|---|---|
|
Mean Error (ME) +/- 1 Breath Per Minute, RR Sensor
기간: up to 40 minutes of continuous monitoring
|
The software calculates respiration rate values via Adult Respiratory Sensor with a mean error of +/- 1 breath per minute relative to a capnography based reference.
The Mean Error value estimates the mean difference in simultaneous estimates of respiration rate (the Respiration Rate calculated from the sensor and the Respiration Rate calculated from capnography).
|
up to 40 minutes of continuous monitoring
|
2차 결과 측정
결과 측정 |
측정값 설명 |
기간 |
|---|---|---|
|
Mean Error (ME) +/- 1 Breath Per Minute, Max-N Sensor
기간: up to 40 minutes of continuous monitoring
|
The software calculates respiration rate values via the Max-N Sensor with a mean error of +/- 1 breath per minute relative to a capnography based reference.
The Mean Error value estimates the mean difference in simultaneous estimates of respiration rate (the Respiration Rate calculated from the sensor and the Respiration Rate calculated from capnography).
The error value estimates the mean difference in simultaneous estimates of respiration rate (the Respiration Rate calculated from the Max-N sensor minus the Respiration Rate calculated from capnography.
|
up to 40 minutes of continuous monitoring
|
공동 작업자 및 조사자
스폰서
연구 기록 날짜
연구 주요 날짜
연구 시작
기본 완료 (실제)
연구 완료 (실제)
연구 등록 날짜
최초 제출
QC 기준을 충족하는 최초 제출
처음 게시됨 (추정)
연구 기록 업데이트
마지막 업데이트 게시됨 (실제)
QC 기준을 충족하는 마지막 업데이트 제출
마지막으로 확인됨
추가 정보
이 연구와 관련된 용어
기타 연구 ID 번호
- COVMOPR0399
이 정보는 변경 없이 clinicaltrials.gov 웹사이트에서 직접 가져온 것입니다. 귀하의 연구 세부 정보를 변경, 제거 또는 업데이트하도록 요청하는 경우 register@clinicaltrials.gov. 문의하십시오. 변경 사항이 clinicaltrials.gov에 구현되는 즉시 저희 웹사이트에도 자동으로 업데이트됩니다. .