Respiration Rate V2.0 in a Hospital Setting

June 19, 2017 updated by: Medtronic - MITG

Evaluation of Respiration Rate Parameters in a Hospital Population Using a Nellcor Bedside Respiratory Patient Monitoring System With Respiration Rate Version 2.0

The purpose of this study is to meet the respiration rate accuracy specifications for the study pulse oximeter monitoring system (study device) in a hospitalized volunteer population when compared to established technologies for measuring parameters related to respiration, such as Capnograph Respiration Rate (RR)(CO2).

Study Overview

Status

Completed

Conditions

Detailed Description

Volunteer subjects from a hospital setting (see specific inclusion/exclusion criteria) over the age of 18, male and female, who provide written informed consent prior to the procedure, and are without any contact skin allergies to adhesives found in standard pulse oximetry sensors. A cross section of demographics - cardiovascular, respiratory, and metabolic type cases should be considered for a broad distribution of medical and surgical patients.

The study consists of measuring parameters related to respiration in hospitalized volunteers by means of the Bedside Respiratory Patient Monitoring System with RR V2.0 (Nightingale). The production-equivalent Nightingale device and multiparameter reference devices will be attached to the subject via sensors, nasal cannulas and ECG leads and the information from all devices will be captured continuously for up to 40 minutes to allow for at least 30 minutes of artifact free data 30 minutes.

Study Type

Observational

Enrollment (Actual)

90

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • Ohio
      • Columbus, Ohio, United States, 43210
        • The Ohio State University Wexner Medical Center

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years and older (Adult, Older Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Sampling Method

Non-Probability Sample

Study Population

Volunteer subjects from a hospital setting on the general care floor or general surgical care floor

Description

Inclusion Criteria:

  • Male or female.
  • 18 years or older.
  • A patient at the study site on a General Care Floor or General Surgery Care Floor.
  • Subject is willing and able to provide written consent.

Exclusion Criteria:

  • Subject is younger than 18.
  • Subject has severe contact allergies that may cause a reaction to standard adhesive materials found in pulse oximetry sensors, ECG electrodes, respiration monitor electrodes or other medical sensors.
  • Subject has an abnormality that may prevent proper application of the device.
  • Subject is in atrial fibrillation.
  • Subject has a documented history of frequent premature ventricular contractions (PVCs), defined as greater than 3 in 30 seconds or greater than 6 in 60 seconds.
  • Subject has an implanted pacemaker.
  • Subject is unwilling or unable to sign informed consent.

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

Cohorts and Interventions

Group / Cohort
no treatment
no treatment, prospective observational

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Mean Error (ME) +/- 1 Breath Per Minute, RR Sensor
Time Frame: up to 40 minutes of continuous monitoring
The software calculates respiration rate values via Adult Respiratory Sensor with a mean error of +/- 1 breath per minute relative to a capnography based reference. The Mean Error value estimates the mean difference in simultaneous estimates of respiration rate (the Respiration Rate calculated from the sensor and the Respiration Rate calculated from capnography).
up to 40 minutes of continuous monitoring

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Mean Error (ME) +/- 1 Breath Per Minute, Max-N Sensor
Time Frame: up to 40 minutes of continuous monitoring
The software calculates respiration rate values via the Max-N Sensor with a mean error of +/- 1 breath per minute relative to a capnography based reference. The Mean Error value estimates the mean difference in simultaneous estimates of respiration rate (the Respiration Rate calculated from the sensor and the Respiration Rate calculated from capnography). The error value estimates the mean difference in simultaneous estimates of respiration rate (the Respiration Rate calculated from the Max-N sensor minus the Respiration Rate calculated from capnography.
up to 40 minutes of continuous monitoring

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start

March 1, 2013

Primary Completion (Actual)

August 1, 2013

Study Completion (Actual)

October 1, 2013

Study Registration Dates

First Submitted

April 17, 2013

First Submitted That Met QC Criteria

April 18, 2013

First Posted (Estimate)

April 23, 2013

Study Record Updates

Last Update Posted (Actual)

July 17, 2017

Last Update Submitted That Met QC Criteria

June 19, 2017

Last Verified

June 1, 2017

More Information

Terms related to this study

Other Study ID Numbers

  • COVMOPR0399

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.