Biomechanical Reconstruction of Three Different Hip Stem Designs in Hip Dysplasia Using a 3D CT-based Planning Software (CTdevice)
Evaluation of the Reconstruction of Biomechanical Parameters Provided by Three Different Hip Stem Designs in Developmental Hip Dysplasia Using a 3D CT-based Software for Surgical Planning
調査の概要
詳細な説明
Pre-operative planning in hip arthroplasty with dedicated software is a useful guide to provide a better reconstruction of the biomechanical parameters of the replaced hip and to optimize the choice of component, improving the geometric understanding of the hip anatomy and the interaction between the native morphological structure and the prosthetic component.
The use of a 3D CT-based software for pre-operative planning may provide a better knowledge of the hip anatomy, simulating the biomechanical parameters more closely than a 2D X-ray-based software. In addition, it is the only way to anticipate the correct interaction between the prosthetic components, or the combined anteversion, a key factor in preventing implant instability.
Severe hip deformities, like developmental hip dysplasia, may not be adequately reconstructed by every implant and 3D pre-operative CT may increase the three-dimensional anatomical knowledge of the hip, improving the choice of the correct implant and possibly reducing the possible consequences of intra and post-operative complications.
Therefore, the purpose of this study is to pre-operatively plan, in a random series of 200 native pelvis and thigh CT scans performed in patients with hip dysplasia, 3 different types of hip stem designs using the 3D Hip-Op software, with the aim of analyzing, on the CT simulations, the percentage of adequate reconstruction of the optimal biomechanical parameters (combined anteversion between 25°-50°, global offset not inferior to 12% of the native offset, leg lengthening not superior to 3 cm, sagittal and coronal tilt not superior to +/-5°, canal filling not inferior to 80%). The percentage of every stem design providing the optimal reconstruction in dysplastic hips (all the 5 parameters matched) will be assessed. The single parameters, especially the combined anteversion, the offset restoration and the leg lengthening, will be assessed for every simulation.
研究の種類
入学 (実際)
連絡先と場所
研究場所
-
-
-
Bologna、イタリア、40136
- Chirurgia Protesica e dei Reimpianti di Anca e Ginocchio, IRCCS Istituto Ortopedico Rizzoli, Bologna, Italy
-
-
参加基準
適格基準
就学可能な年齢
健康ボランティアの受け入れ
受講資格のある性別
サンプリング方法
調査対象母集団
説明
Inclusion Criteria:
- definition of developmental hip dysplasia according to Wiberg (center edge angle <20°)
- pre-operative CT scan extended from the fourth lumbar vertebra to the tibial plateaus
Exclusion Criteria:
- other types of hip pathologies
- inadequate Ct scans
研究計画
研究はどのように設計されていますか?
デザインの詳細
コホートと介入
グループ/コホート |
介入・治療 |
|---|---|
|
CLS stem
CLS Zimmer stem implantation (single wedge, tapered stem) using the 3D CT based software for surgical pre-operative planning
|
In every hip (with appropriate CT scan), 3 different stem designs will be positioned using a 3D Ct based software for surgical pre-operative planning.
|
|
Wagner cone stem
Wagner cone Zimmer stem implantation (conical tapered stem) using the 3D CT based software for surgical pre-operative planning
|
In every hip (with appropriate CT scan), 3 different stem designs will be positioned using a 3D Ct based software for surgical pre-operative planning.
|
|
Aptafix stem
Aptafix Ortho stem implantation (anatomical stem) using the 3D CT based software for surgical pre-operative planning
|
In every hip (with appropriate CT scan), 3 different stem designs will be positioned using a 3D Ct based software for surgical pre-operative planning.
|
この研究は何を測定していますか?
主要な結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Correct Reconstruction CLS
時間枠:Day 0
|
Percentage of hips correctly reconstructed using CLS stem: the correctly reconstructed hip should have combined anteversion between 25° and 50°, global offset not inferior to 12% of the native offset, leg lengthening not superior to 3 cm, sagittal and coronal tilt not superior to +/-5°, canal filling not inferior to 80% (all the 5 parameters should be present at the same time).
Minimum-maximum values (as percentage %): 0-100.
100 is the best result, 0 the worst
|
Day 0
|
|
Correct Reconstruction Wagner Cone
時間枠:Day 0
|
Percentage of hips correctly reconstructed using CLS stem: the correctly reconstructed hip should have combined anteversion between 25° and 50°, global offset not inferior to 12% of the native offset, leg lengthening not superior to 3 cm, sagittal and coronal tilt not superior to +/-5°, canal filling not inferior to 80% (all the 5 parameters should be present at the same time).
Minimum-maximum values (as percentage %): 0-100.
100 is the best result, 0 the worst
|
Day 0
|
|
Correct Reconstruction Aptafix
時間枠:Day 0
|
Percentage of hips correctly reconstructed using CLS stem: the correctly reconstructed hip should have combined anteversion between 25° and 50°, global offset not inferior to 12% of the native offset, leg lengthening not superior to 3 cm, sagittal and coronal tilt not superior to +/-5°, canal filling not inferior to 80% (all the 5 parameters should be present at the same time).
Minimum-maximum values (as percentage %): 0-100.
100 is the best result, 0 the worst
|
Day 0
|
二次結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Combined Anteversion CLS
時間枠:Day 0
|
Percentage of hips with adequate combined anteversion according to Dorr (25-50°) and Widmer (<37°) using CLS stems.
Minimum-maximum values (as percentage %): 0-100.
100 is the best result, 0 is the worst
|
Day 0
|
|
Combined Anteversion Wagner Cone
時間枠:Day 0
|
Percentage of hips with adequate combined anteversion according to Dorr (25-50°) and Widmer (<37°) using CLS stems.
Minimum-maximum values (as percentage %): 0-100.
100 is the best result, 0 is the worst
|
Day 0
|
|
Combined Anteversion Aptafix
時間枠:Day 0
|
Percentage of hips with adequate combined anteversion according to Dorr (25-50°) and Widmer (<37°) using CLS stems.
Minimum-maximum values (as percentage %): 0-100.
100 is the best result, 0 is the worst
|
Day 0
|
|
Offset CLS
時間枠:Day 0
|
Percentage of hips with acetabular+femoral offset not inferior to 12% of the native value using CLS stems.
Minimum-maximum values (as percentage %): 0-100.
100 is the best result, 0 is the worst
|
Day 0
|
|
Offset Wagner Cone
時間枠:Day 0
|
Percentage of hips with acetabular+femoral offset not inferior to 12% of the native value using CLS stems.
Minimum-maximum values (as percentage %): 0-100.
100 is the best result, 0 is the worst
|
Day 0
|
|
Offset Aptafix
時間枠:Day 0
|
Percentage of hips with acetabular+femoral offset not inferior to 12% of the native value using CLS stems.
Minimum-maximum values (as percentage %): 0-100.
100 is the best result, 0 is the worst
|
Day 0
|
|
Leg Lengthening CLS
時間枠:Day 0
|
Percentage of hips with leg lengthening not superior to 3 cm using CLS stems.
Minimum-maximum values (as percentages %): 0-100.
100 is the best result, 0 is the worst
|
Day 0
|
|
Leg Lengthening Wagner Cone
時間枠:Day 0
|
Percentage of hips with leg lengthening not superior to 3 cm using CLS stems.
Minimum-maximum values (as percentages %): 0-100.
100 is the best result, 0 is the worst
|
Day 0
|
|
Leg Lengthening Aptafix
時間枠:Day 0
|
Percentage of hips with leg lengthening not superior to 3 cm using CLS stems.
Minimum-maximum values (as percentages %): 0-100.
100 is the best result, 0 is the worst
|
Day 0
|
協力者と研究者
捜査官
- スタディディレクター:Francesco Traina, Professor、IRCCS Istituto Ortopedico Rizzoli
出版物と役立つリンク
一般刊行物
- Taniguchi N, Jinno T, Koga D, Hagino T, Okawa A, Haro H. Cementless Hip Stem Anteversion in the Dysplastic Hip: A Comparison of Tapered Wedge vs Metaphyseal Filling. J Arthroplasty. 2017 May;32(5):1547-1552. doi: 10.1016/j.arth.2016.12.020. Epub 2016 Dec 22.
- Argenson JN, Ryembault E, Flecher X, Brassart N, Parratte S, Aubaniac JM. Three-dimensional anatomy of the hip in osteoarthritis after developmental dysplasia. J Bone Joint Surg Br. 2005 Sep;87(9):1192-6. doi: 10.1302/0301-620X.87B9.15928.
- Sugano N, Takao M, Sakai T, Nishii T, Miki H. Does CT-based navigation improve the long-term survival in ceramic-on-ceramic THA? Clin Orthop Relat Res. 2012 Nov;470(11):3054-9. doi: 10.1007/s11999-012-2378-4.
- Castagnini F, Valente G, Crimi G, Taddei F, Bordini B, Stea S, Toni A. Component positioning and ceramic damage in cementless ceramic-on-ceramic total hip arthroplasty. J Orthop Sci. 2019 Jul;24(4):643-651. doi: 10.1016/j.jos.2018.12.011. Epub 2019 Jan 4.
研究記録日
主要日程の研究
研究開始 (実際)
一次修了 (実際)
研究の完了 (実際)
試験登録日
最初に提出
QC基準を満たした最初の提出物
最初の投稿 (実際)
学習記録の更新
投稿された最後の更新 (実際)
QC基準を満たした最後の更新が送信されました
最終確認日
詳しくは
本研究に関する用語
その他の研究ID番号
- 29/2021/Oss/IOR
個々の参加者データ (IPD) の計画
個々の参加者データ (IPD) を共有する予定はありますか?
医薬品およびデバイス情報、研究文書
米国FDA規制医薬品の研究
米国FDA規制機器製品の研究
この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。