Biomechanical Reconstruction of Three Different Hip Stem Designs in Hip Dysplasia Using a 3D CT-based Planning Software (CTdevice)

2021年9月6日 更新者:Istituto Ortopedico Rizzoli

Evaluation of the Reconstruction of Biomechanical Parameters Provided by Three Different Hip Stem Designs in Developmental Hip Dysplasia Using a 3D CT-based Software for Surgical Planning

A random population of 200 CT scans of pelvis and thigh in an adult population affected by hip dysplasia will be selected. The aim of this study is to evaluate the biomechanical reconstruction of the hip anatomy provided by three different hip stem designs in dysplastic cases, using a 3D CT based software for pre-operative planning.

調査の概要

状態

完了

詳細な説明

Pre-operative planning in hip arthroplasty with dedicated software is a useful guide to provide a better reconstruction of the biomechanical parameters of the replaced hip and to optimize the choice of component, improving the geometric understanding of the hip anatomy and the interaction between the native morphological structure and the prosthetic component.

The use of a 3D CT-based software for pre-operative planning may provide a better knowledge of the hip anatomy, simulating the biomechanical parameters more closely than a 2D X-ray-based software. In addition, it is the only way to anticipate the correct interaction between the prosthetic components, or the combined anteversion, a key factor in preventing implant instability.

Severe hip deformities, like developmental hip dysplasia, may not be adequately reconstructed by every implant and 3D pre-operative CT may increase the three-dimensional anatomical knowledge of the hip, improving the choice of the correct implant and possibly reducing the possible consequences of intra and post-operative complications.

Therefore, the purpose of this study is to pre-operatively plan, in a random series of 200 native pelvis and thigh CT scans performed in patients with hip dysplasia, 3 different types of hip stem designs using the 3D Hip-Op software, with the aim of analyzing, on the CT simulations, the percentage of adequate reconstruction of the optimal biomechanical parameters (combined anteversion between 25°-50°, global offset not inferior to 12% of the native offset, leg lengthening not superior to 3 cm, sagittal and coronal tilt not superior to +/-5°, canal filling not inferior to 80%). The percentage of every stem design providing the optimal reconstruction in dysplastic hips (all the 5 parameters matched) will be assessed. The single parameters, especially the combined anteversion, the offset restoration and the leg lengthening, will be assessed for every simulation.

研究の種類

観察的

入学 (実際)

150

連絡先と場所

このセクションには、調査を実施する担当者の連絡先の詳細と、この調査が実施されている場所に関する情報が記載されています。

研究場所

      • Bologna、イタリア、40136
        • Chirurgia Protesica e dei Reimpianti di Anca e Ginocchio, IRCCS Istituto Ortopedico Rizzoli, Bologna, Italy

参加基準

研究者は、適格基準と呼ばれる特定の説明に適合する人を探します。これらの基準のいくつかの例は、人の一般的な健康状態または以前の治療です。

適格基準

就学可能な年齢

18年歳以上 (大人、高齢者)

健康ボランティアの受け入れ

いいえ

受講資格のある性別

全て

サンプリング方法

確率サンプル

調査対象母集団

A random population of 200 CT scans (from the fourth lumbar vertebra to the tibial plateaus) involving adult patients with developmental hip dysplasia.

説明

Inclusion Criteria:

  • definition of developmental hip dysplasia according to Wiberg (center edge angle <20°)
  • pre-operative CT scan extended from the fourth lumbar vertebra to the tibial plateaus

Exclusion Criteria:

  • other types of hip pathologies
  • inadequate Ct scans

研究計画

このセクションでは、研究がどのように設計され、研究が何を測定しているかなど、研究計画の詳細を提供します。

研究はどのように設計されていますか?

デザインの詳細

コホートと介入

グループ/コホート
介入・治療
CLS stem
CLS Zimmer stem implantation (single wedge, tapered stem) using the 3D CT based software for surgical pre-operative planning
In every hip (with appropriate CT scan), 3 different stem designs will be positioned using a 3D Ct based software for surgical pre-operative planning.
Wagner cone stem
Wagner cone Zimmer stem implantation (conical tapered stem) using the 3D CT based software for surgical pre-operative planning
In every hip (with appropriate CT scan), 3 different stem designs will be positioned using a 3D Ct based software for surgical pre-operative planning.
Aptafix stem
Aptafix Ortho stem implantation (anatomical stem) using the 3D CT based software for surgical pre-operative planning
In every hip (with appropriate CT scan), 3 different stem designs will be positioned using a 3D Ct based software for surgical pre-operative planning.

この研究は何を測定していますか?

主要な結果の測定

結果測定
メジャーの説明
時間枠
Correct Reconstruction CLS
時間枠:Day 0
Percentage of hips correctly reconstructed using CLS stem: the correctly reconstructed hip should have combined anteversion between 25° and 50°, global offset not inferior to 12% of the native offset, leg lengthening not superior to 3 cm, sagittal and coronal tilt not superior to +/-5°, canal filling not inferior to 80% (all the 5 parameters should be present at the same time). Minimum-maximum values (as percentage %): 0-100. 100 is the best result, 0 the worst
Day 0
Correct Reconstruction Wagner Cone
時間枠:Day 0
Percentage of hips correctly reconstructed using CLS stem: the correctly reconstructed hip should have combined anteversion between 25° and 50°, global offset not inferior to 12% of the native offset, leg lengthening not superior to 3 cm, sagittal and coronal tilt not superior to +/-5°, canal filling not inferior to 80% (all the 5 parameters should be present at the same time). Minimum-maximum values (as percentage %): 0-100. 100 is the best result, 0 the worst
Day 0
Correct Reconstruction Aptafix
時間枠:Day 0
Percentage of hips correctly reconstructed using CLS stem: the correctly reconstructed hip should have combined anteversion between 25° and 50°, global offset not inferior to 12% of the native offset, leg lengthening not superior to 3 cm, sagittal and coronal tilt not superior to +/-5°, canal filling not inferior to 80% (all the 5 parameters should be present at the same time). Minimum-maximum values (as percentage %): 0-100. 100 is the best result, 0 the worst
Day 0

二次結果の測定

結果測定
メジャーの説明
時間枠
Combined Anteversion CLS
時間枠:Day 0
Percentage of hips with adequate combined anteversion according to Dorr (25-50°) and Widmer (<37°) using CLS stems. Minimum-maximum values (as percentage %): 0-100. 100 is the best result, 0 is the worst
Day 0
Combined Anteversion Wagner Cone
時間枠:Day 0
Percentage of hips with adequate combined anteversion according to Dorr (25-50°) and Widmer (<37°) using CLS stems. Minimum-maximum values (as percentage %): 0-100. 100 is the best result, 0 is the worst
Day 0
Combined Anteversion Aptafix
時間枠:Day 0
Percentage of hips with adequate combined anteversion according to Dorr (25-50°) and Widmer (<37°) using CLS stems. Minimum-maximum values (as percentage %): 0-100. 100 is the best result, 0 is the worst
Day 0
Offset CLS
時間枠:Day 0
Percentage of hips with acetabular+femoral offset not inferior to 12% of the native value using CLS stems. Minimum-maximum values (as percentage %): 0-100. 100 is the best result, 0 is the worst
Day 0
Offset Wagner Cone
時間枠:Day 0
Percentage of hips with acetabular+femoral offset not inferior to 12% of the native value using CLS stems. Minimum-maximum values (as percentage %): 0-100. 100 is the best result, 0 is the worst
Day 0
Offset Aptafix
時間枠:Day 0
Percentage of hips with acetabular+femoral offset not inferior to 12% of the native value using CLS stems. Minimum-maximum values (as percentage %): 0-100. 100 is the best result, 0 is the worst
Day 0
Leg Lengthening CLS
時間枠:Day 0
Percentage of hips with leg lengthening not superior to 3 cm using CLS stems. Minimum-maximum values (as percentages %): 0-100. 100 is the best result, 0 is the worst
Day 0
Leg Lengthening Wagner Cone
時間枠:Day 0
Percentage of hips with leg lengthening not superior to 3 cm using CLS stems. Minimum-maximum values (as percentages %): 0-100. 100 is the best result, 0 is the worst
Day 0
Leg Lengthening Aptafix
時間枠:Day 0
Percentage of hips with leg lengthening not superior to 3 cm using CLS stems. Minimum-maximum values (as percentages %): 0-100. 100 is the best result, 0 is the worst
Day 0

協力者と研究者

ここでは、この調査に関係する人々や組織を見つけることができます。

スポンサー

捜査官

  • スタディディレクター:Francesco Traina, Professor、IRCCS Istituto Ortopedico Rizzoli

出版物と役立つリンク

研究に関する情報を入力する責任者は、自発的にこれらの出版物を提供します。これらは、研究に関連するあらゆるものに関するものである可能性があります。

研究記録日

これらの日付は、ClinicalTrials.gov への研究記録と要約結果の提出の進捗状況を追跡します。研究記録と報告された結果は、国立医学図書館 (NLM) によって審査され、公開 Web サイトに掲載される前に、特定の品質管理基準を満たしていることが確認されます。

主要日程の研究

研究開始 (実際)

2021年5月10日

一次修了 (実際)

2021年7月10日

研究の完了 (実際)

2021年7月10日

試験登録日

最初に提出

2021年5月21日

QC基準を満たした最初の提出物

2021年5月24日

最初の投稿 (実際)

2021年5月27日

学習記録の更新

投稿された最後の更新 (実際)

2021年9月9日

QC基準を満たした最後の更新が送信されました

2021年9月6日

最終確認日

2021年2月1日

詳しくは

本研究に関する用語

個々の参加者データ (IPD) の計画

個々の参加者データ (IPD) を共有する予定はありますか?

いいえ

医薬品およびデバイス情報、研究文書

米国FDA規制医薬品の研究

いいえ

米国FDA規制機器製品の研究

いいえ

この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。

購読する