Biomechanical Reconstruction of Three Different Hip Stem Designs in Hip Dysplasia Using a 3D CT-based Planning Software (CTdevice)

6 september 2021 bijgewerkt door: Istituto Ortopedico Rizzoli

Evaluation of the Reconstruction of Biomechanical Parameters Provided by Three Different Hip Stem Designs in Developmental Hip Dysplasia Using a 3D CT-based Software for Surgical Planning

A random population of 200 CT scans of pelvis and thigh in an adult population affected by hip dysplasia will be selected. The aim of this study is to evaluate the biomechanical reconstruction of the hip anatomy provided by three different hip stem designs in dysplastic cases, using a 3D CT based software for pre-operative planning.

Studie Overzicht

Toestand

Voltooid

Conditie

Interventie / Behandeling

Gedetailleerde beschrijving

Pre-operative planning in hip arthroplasty with dedicated software is a useful guide to provide a better reconstruction of the biomechanical parameters of the replaced hip and to optimize the choice of component, improving the geometric understanding of the hip anatomy and the interaction between the native morphological structure and the prosthetic component.

The use of a 3D CT-based software for pre-operative planning may provide a better knowledge of the hip anatomy, simulating the biomechanical parameters more closely than a 2D X-ray-based software. In addition, it is the only way to anticipate the correct interaction between the prosthetic components, or the combined anteversion, a key factor in preventing implant instability.

Severe hip deformities, like developmental hip dysplasia, may not be adequately reconstructed by every implant and 3D pre-operative CT may increase the three-dimensional anatomical knowledge of the hip, improving the choice of the correct implant and possibly reducing the possible consequences of intra and post-operative complications.

Therefore, the purpose of this study is to pre-operatively plan, in a random series of 200 native pelvis and thigh CT scans performed in patients with hip dysplasia, 3 different types of hip stem designs using the 3D Hip-Op software, with the aim of analyzing, on the CT simulations, the percentage of adequate reconstruction of the optimal biomechanical parameters (combined anteversion between 25°-50°, global offset not inferior to 12% of the native offset, leg lengthening not superior to 3 cm, sagittal and coronal tilt not superior to +/-5°, canal filling not inferior to 80%). The percentage of every stem design providing the optimal reconstruction in dysplastic hips (all the 5 parameters matched) will be assessed. The single parameters, especially the combined anteversion, the offset restoration and the leg lengthening, will be assessed for every simulation.

Studietype

Observationeel

Inschrijving (Werkelijk)

150

Contacten en locaties

In dit gedeelte vindt u de contactgegevens van degenen die het onderzoek uitvoeren en informatie over waar dit onderzoek wordt uitgevoerd.

Studie Locaties

      • Bologna, Italië, 40136
        • Chirurgia Protesica e dei Reimpianti di Anca e Ginocchio, IRCCS Istituto Ortopedico Rizzoli, Bologna, Italy

Deelname Criteria

Onderzoekers zoeken naar mensen die aan een bepaalde beschrijving voldoen, de zogenaamde geschiktheidscriteria. Enkele voorbeelden van deze criteria zijn iemands algemene gezondheidstoestand of eerdere behandelingen.

Geschiktheidscriteria

Leeftijden die in aanmerking komen voor studie

18 jaar en ouder (Volwassen, Oudere volwassene)

Accepteert gezonde vrijwilligers

Nee

Geslachten die in aanmerking komen voor studie

Allemaal

Bemonsteringsmethode

Kanssteekproef

Studie Bevolking

A random population of 200 CT scans (from the fourth lumbar vertebra to the tibial plateaus) involving adult patients with developmental hip dysplasia.

Beschrijving

Inclusion Criteria:

  • definition of developmental hip dysplasia according to Wiberg (center edge angle <20°)
  • pre-operative CT scan extended from the fourth lumbar vertebra to the tibial plateaus

Exclusion Criteria:

  • other types of hip pathologies
  • inadequate Ct scans

Studie plan

Dit gedeelte bevat details van het studieplan, inclusief hoe de studie is opgezet en wat de studie meet.

Hoe is de studie opgezet?

Ontwerpdetails

Cohorten en interventies

Groep / Cohort
Interventie / Behandeling
CLS stem
CLS Zimmer stem implantation (single wedge, tapered stem) using the 3D CT based software for surgical pre-operative planning
In every hip (with appropriate CT scan), 3 different stem designs will be positioned using a 3D Ct based software for surgical pre-operative planning.
Wagner cone stem
Wagner cone Zimmer stem implantation (conical tapered stem) using the 3D CT based software for surgical pre-operative planning
In every hip (with appropriate CT scan), 3 different stem designs will be positioned using a 3D Ct based software for surgical pre-operative planning.
Aptafix stem
Aptafix Ortho stem implantation (anatomical stem) using the 3D CT based software for surgical pre-operative planning
In every hip (with appropriate CT scan), 3 different stem designs will be positioned using a 3D Ct based software for surgical pre-operative planning.

Wat meet het onderzoek?

Primaire uitkomstmaten

Uitkomstmaat
Maatregel Beschrijving
Tijdsspanne
Correct Reconstruction CLS
Tijdsspanne: Day 0
Percentage of hips correctly reconstructed using CLS stem: the correctly reconstructed hip should have combined anteversion between 25° and 50°, global offset not inferior to 12% of the native offset, leg lengthening not superior to 3 cm, sagittal and coronal tilt not superior to +/-5°, canal filling not inferior to 80% (all the 5 parameters should be present at the same time). Minimum-maximum values (as percentage %): 0-100. 100 is the best result, 0 the worst
Day 0
Correct Reconstruction Wagner Cone
Tijdsspanne: Day 0
Percentage of hips correctly reconstructed using CLS stem: the correctly reconstructed hip should have combined anteversion between 25° and 50°, global offset not inferior to 12% of the native offset, leg lengthening not superior to 3 cm, sagittal and coronal tilt not superior to +/-5°, canal filling not inferior to 80% (all the 5 parameters should be present at the same time). Minimum-maximum values (as percentage %): 0-100. 100 is the best result, 0 the worst
Day 0
Correct Reconstruction Aptafix
Tijdsspanne: Day 0
Percentage of hips correctly reconstructed using CLS stem: the correctly reconstructed hip should have combined anteversion between 25° and 50°, global offset not inferior to 12% of the native offset, leg lengthening not superior to 3 cm, sagittal and coronal tilt not superior to +/-5°, canal filling not inferior to 80% (all the 5 parameters should be present at the same time). Minimum-maximum values (as percentage %): 0-100. 100 is the best result, 0 the worst
Day 0

Secundaire uitkomstmaten

Uitkomstmaat
Maatregel Beschrijving
Tijdsspanne
Combined Anteversion CLS
Tijdsspanne: Day 0
Percentage of hips with adequate combined anteversion according to Dorr (25-50°) and Widmer (<37°) using CLS stems. Minimum-maximum values (as percentage %): 0-100. 100 is the best result, 0 is the worst
Day 0
Combined Anteversion Wagner Cone
Tijdsspanne: Day 0
Percentage of hips with adequate combined anteversion according to Dorr (25-50°) and Widmer (<37°) using CLS stems. Minimum-maximum values (as percentage %): 0-100. 100 is the best result, 0 is the worst
Day 0
Combined Anteversion Aptafix
Tijdsspanne: Day 0
Percentage of hips with adequate combined anteversion according to Dorr (25-50°) and Widmer (<37°) using CLS stems. Minimum-maximum values (as percentage %): 0-100. 100 is the best result, 0 is the worst
Day 0
Offset CLS
Tijdsspanne: Day 0
Percentage of hips with acetabular+femoral offset not inferior to 12% of the native value using CLS stems. Minimum-maximum values (as percentage %): 0-100. 100 is the best result, 0 is the worst
Day 0
Offset Wagner Cone
Tijdsspanne: Day 0
Percentage of hips with acetabular+femoral offset not inferior to 12% of the native value using CLS stems. Minimum-maximum values (as percentage %): 0-100. 100 is the best result, 0 is the worst
Day 0
Offset Aptafix
Tijdsspanne: Day 0
Percentage of hips with acetabular+femoral offset not inferior to 12% of the native value using CLS stems. Minimum-maximum values (as percentage %): 0-100. 100 is the best result, 0 is the worst
Day 0
Leg Lengthening CLS
Tijdsspanne: Day 0
Percentage of hips with leg lengthening not superior to 3 cm using CLS stems. Minimum-maximum values (as percentages %): 0-100. 100 is the best result, 0 is the worst
Day 0
Leg Lengthening Wagner Cone
Tijdsspanne: Day 0
Percentage of hips with leg lengthening not superior to 3 cm using CLS stems. Minimum-maximum values (as percentages %): 0-100. 100 is the best result, 0 is the worst
Day 0
Leg Lengthening Aptafix
Tijdsspanne: Day 0
Percentage of hips with leg lengthening not superior to 3 cm using CLS stems. Minimum-maximum values (as percentages %): 0-100. 100 is the best result, 0 is the worst
Day 0

Medewerkers en onderzoekers

Hier vindt u mensen en organisaties die betrokken zijn bij dit onderzoek.

Onderzoekers

  • Studie directeur: Francesco Traina, Professor, IRCCS Istituto Ortopedico Rizzoli

Publicaties en nuttige links

De persoon die verantwoordelijk is voor het invoeren van informatie over het onderzoek stelt deze publicaties vrijwillig ter beschikking. Dit kan gaan over alles wat met het onderzoek te maken heeft.

Studie record data

Deze datums volgen de voortgang van het onderzoeksdossier en de samenvatting van de ingediende resultaten bij ClinicalTrials.gov. Studieverslagen en gerapporteerde resultaten worden beoordeeld door de National Library of Medicine (NLM) om er zeker van te zijn dat ze voldoen aan specifieke kwaliteitscontrolenormen voordat ze op de openbare website worden geplaatst.

Bestudeer belangrijke data

Studie start (Werkelijk)

10 mei 2021

Primaire voltooiing (Werkelijk)

10 juli 2021

Studie voltooiing (Werkelijk)

10 juli 2021

Studieregistratiedata

Eerst ingediend

21 mei 2021

Eerst ingediend dat voldeed aan de QC-criteria

24 mei 2021

Eerst geplaatst (Werkelijk)

27 mei 2021

Updates van studierecords

Laatste update geplaatst (Werkelijk)

9 september 2021

Laatste update ingediend die voldeed aan QC-criteria

6 september 2021

Laatst geverifieerd

1 februari 2021

Meer informatie

Termen gerelateerd aan deze studie

Plan Individuele Deelnemersgegevens (IPD)

Bent u van plan om gegevens van individuele deelnemers (IPD) te delen?

NEE

Informatie over medicijnen en apparaten, studiedocumenten

Bestudeert een door de Amerikaanse FDA gereguleerd geneesmiddel

Nee

Bestudeert een door de Amerikaanse FDA gereguleerd apparaatproduct

Nee

Deze informatie is zonder wijzigingen rechtstreeks van de website clinicaltrials.gov gehaald. Als u verzoeken heeft om uw onderzoeksgegevens te wijzigen, te verwijderen of bij te werken, neem dan contact op met register@clinicaltrials.gov. Zodra er een wijziging wordt doorgevoerd op clinicaltrials.gov, wordt deze ook automatisch bijgewerkt op onze website .