Performance Evaluation of Mpox Molecular POC Diagnostics
Clinical Performance Evaluation of Novel Rapid Molecular Point-of-care (POC) Diagnostics for Mpox Virus
Studieoversikt
Status
Forhold
Intervensjon / Behandling
Detaljert beskrivelse
PP016 is a FIND-sponsored retrospective, non-interventional clinical performance evaluation study designed to assess the diagnostic accuracy and operational suitability of up to three rapid molecular point-of-care (mPOC) tests for the detection of Mpox virus (MPXV) using archived lesion swab specimens collected in Uganda. The study, titled "Clinical Performance Evaluation of Novel Rapid Molecular Point-of-Care Diagnostics for Mpox Virus," is being conducted at the Central Public Health Laboratory (CPHL) in Kampala, Uganda. The study was developed in response to the growing public health importance of mpox, particularly following the 2022 and 2024 WHO Public Health Emergencies of International Concern (PHEICs), and the ongoing outbreaks affecting several African countries including Uganda, where thousands of confirmed cases have been reported. Because conventional laboratory PCR testing remains centralized and difficult to access in many low-resource or decentralized settings, the study aims to determine whether rapid molecular POC assays can provide accurate and scalable diagnostic alternatives to improve outbreak response, surveillance, patient management, and contact tracing. The study evaluates three investigational molecular diagnostic platforms: the VZV-Q Real-Time PCR Kit for OnePCR (Genes2Me, India), the RADI-ONE Mpox Detection Kit (KH Medical, South Korea), and the STANDARD M10 MPX/OPX assay (SD Biosensor, South Korea). Their performance is compared against the laboratory-based BioPerfectus Monkeypox Virus Real Time PCR Kit, which serves as the reference standard and is currently listed under the African Medicines Regulatory Harmonisation Emergency Use Listing for mpox diagnosis. The study uses archived lesion swab samples stored in viral or universal transport medium from suspected mpox cases collected since 2020 under routine surveillance or prior research activities. Eligible samples must have known PCR status, adequate volume for both reference and index testing, and acceptable storage conditions.
The primary objective is to determine the clinical sensitivity/positive percent agreement (PPA) and specificity/negative percent agreement (NPA) of each investigational assay relative to the reference PCR, while secondary objectives include assessing performance according to PCR cycle threshold (Ct) values as a proxy for viral load and evaluating potential differences in performance by circulating mpox virus clades if such information is available. Statistical analyses will use Wilson score confidence intervals, with subgroup analyses by age, sex, Ct value, and viral clade. The study follows a blinded case-control design in which specimens are de-identified, shuffled, relabeled with unique FIND study identifiers, and tested independently to minimize bias. Reference PCR testing is performed first to confirm specimen quality and classification before testing on the investigational devices. Invalid or inconclusive results are repeated once and documented according to predefined procedures. No patient recruitment or intervention occurs, and test results are not used for clinical management, making the study low risk from an ethical and safety perspective.
Data are collected using OpenClinica Enterprise Edition, a validated electronic data capture system with audit trails and secure cloud hosting. FIND oversees monitoring, quality assurance, operator training, and regulatory compliance throughout the study.
The study is funded by FIND with support from the Government of the Netherlands and is expected to generate independent evidence to support regulatory submissions, procurement decisions, WHO evaluation processes, and future implementation of rapid mpox diagnostics in low- and middle-income countries.
Studietype
Registrering (Antatt)
Deltakelseskriterier
Kvalifikasjonskriterier
Alder som er kvalifisert for studier
- Barn
- Voksen
- Eldre voksen
Tar imot friske frivillige
Prøvetakingsmetode
Studiepopulasjon
Beskrivelse
Inclusion Criteria:
- Lesion swabs stored in VTM/UTM (non-inactivating)
- Aliquot with sufficient volume to run the reference PCR and at least one index test (≥2600 μL per pair reference-index tests: ideally at least 1400 μL for 3 index tests and 1200 μL for reference PCR, including 1 possible repeat per test)
- Available PCR result for specimen selection (i.e. confirmed MPXV positive or negative)
- Samples collected from 2020 onwards.
Exclusion Criteria:
- Unknown date of collection
- >3 freeze-thaw cycles of sample.
- Samples collected with expired swab collection devices or UTM/VTM
Studieplan
Hvordan er studiet utformet?
Designdetaljer
Kohorter og intervensjoner
Gruppe / Kohort |
Intervensjon / Behandling |
|---|---|
|
Positive samples
50 confirmed MPXV-positive samples
|
Each (negative and positive) sample will be tested on the standard of reference and the investigational IVD
|
|
Negative samples
30 confirmed MPXV-negative samples
|
Each (negative and positive) sample will be tested on the standard of reference and the investigational IVD
|
Hva måler studien?
Primære resultatmål
Resultatmål |
Tiltaksbeskrivelse |
Tidsramme |
|---|---|---|
|
Point estimates of sensitivity, specificity, positive and negative predictive value (PPV and NPV respectively) with 95% confidence intervals.
Tidsramme: July-August 2026
|
Point estimates of sensitivity, specificity, positive and negative predictive value (PPV and NPV respectively) with 95% confidence intervals.
|
July-August 2026
|
Sekundære resultatmål
Resultatmål |
Tiltaksbeskrivelse |
Tidsramme |
|---|---|---|
|
Point estimates of sensitivity/PPA stratified by Ct values of the reference PCR test.
Tidsramme: July-August 2026
|
Point estimates of sensitivity/PPA stratified by Ct values of the reference PCR test.
|
July-August 2026
|
|
Point estimates of sensitivity/PPA and specificity/NPA with 95% confidence intervals stratified by virus clade.
Tidsramme: July-August 2026
|
Point estimates of sensitivity/PPA and specificity/NPA with 95% confidence intervals stratified by virus clade.
|
July-August 2026
|
Samarbeidspartnere og etterforskere
Studierekorddatoer
Studer hoveddatoer
Studiestart (Antatt)
Primær fullføring (Antatt)
Studiet fullført (Antatt)
Datoer for studieregistrering
Først innsendt
Først innsendt som oppfylte QC-kriteriene
Først lagt ut (Faktiske)
Oppdateringer av studieposter
Sist oppdatering lagt ut (Faktiske)
Siste oppdatering sendt inn som oppfylte QC-kriteriene
Sist bekreftet
Mer informasjon
Begreper knyttet til denne studien
Ytterligere relevante MeSH-vilkår
Andre studie-ID-numre
- PP016
Legemiddel- og utstyrsinformasjon, studiedokumenter
Studerer et amerikansk FDA-regulert enhetsprodukt
Denne informasjonen ble hentet direkte fra nettstedet clinicaltrials.gov uten noen endringer. Hvis du har noen forespørsler om å endre, fjerne eller oppdatere studiedetaljene dine, vennligst kontakt register@clinicaltrials.gov. Så snart en endring er implementert på clinicaltrials.gov, vil denne også bli oppdatert automatisk på nettstedet vårt. .
Kliniske studier på Monkey Pox
-
PENTA FoundationUniversity College, London; European Commission; Servicio Madrileño de Salud... og andre samarbeidspartnereRekruttering
-
Institute of Tropical Medicine, BelgiumFullført
-
Institut PasteurCentre Médical de l'Institut PasteurRekruttering
-
University of MinnesotaFullførtMonkey PoxForente stater
-
Public Health Service of AmsterdamAvsluttetMonkeypox | Monkey PoxNederland
-
Beni-Suef UniversityFullført
-
SIGA TechnologiesNational Institutes of Health (NIH)FullførtKopper | Monkey Pox | Ortopoksviral sykdomForente stater
-
Robert ColebundersPatrick KototoHar ikke rekruttert ennå
-
Assistance Publique - Hôpitaux de ParisAktiv, ikke rekrutterendeMonkey Pox | Ape sykdommerFrankrike
-
xiaoliang HuangRekrutteringTykktarmskreft | Monkey PoxKina
Kliniske studier på MPOX IVD test
-
Samsung ElectronicsNexus DXAvsluttetKongestiv hjertesviktForente stater
-
Istituto Oncologico Veneto IRCCSAktiv, ikke rekrutterendeKarsinom, ikke-småcellet lungekreft (NSCLC)Italia
-
Genlantis DiagnosticsNorthwestern UniversityHar ikke rekruttert ennåFor å demonstrere den kliniske ytelsen til TriQuik Invitro Diagnostic Device
-
Firalis SARekruttering
-
Fundación FLS de Lucha Contra el Sida, las Enfermedades...London School of Hygiene and Tropical Medicine; Hospital Clinic of Barcelona og andre samarbeidspartnereFullført
-
Abramson Cancer Center at Penn MedicineTranslational Breast Cancer Research Consortium; Danaher Inc.RekrutteringHER2-positiv metastatisk brystkreftForente stater
-
Chang Gung Memorial HospitalDr. Chip Biotechnology IncorporationFullførtCervikal intraepitelial neoplasi | HPV-infeksjon | Infeksjon av humant papillomavirusTaiwan
-
BioFire Defense LLCEmory University; Wake Forest University Health Sciences; George Washington... og andre samarbeidspartnereRekrutteringCOVID | Koronavirus | LuftveisinfeksjonForente stater
-
Cyclacel Pharmaceuticals, Inc.AvsluttetIkke-småcellet lungekreftForente stater
-
Shepherd Medical CompanyUkjent