Performance Evaluation of Mpox Molecular POC Diagnostics
Clinical Performance Evaluation of Novel Rapid Molecular Point-of-care (POC) Diagnostics for Mpox Virus
Panoramica dello studio
Stato
Condizioni
Intervento / Trattamento
Descrizione dettagliata
PP016 is a FIND-sponsored retrospective, non-interventional clinical performance evaluation study designed to assess the diagnostic accuracy and operational suitability of up to three rapid molecular point-of-care (mPOC) tests for the detection of Mpox virus (MPXV) using archived lesion swab specimens collected in Uganda. The study, titled "Clinical Performance Evaluation of Novel Rapid Molecular Point-of-Care Diagnostics for Mpox Virus," is being conducted at the Central Public Health Laboratory (CPHL) in Kampala, Uganda. The study was developed in response to the growing public health importance of mpox, particularly following the 2022 and 2024 WHO Public Health Emergencies of International Concern (PHEICs), and the ongoing outbreaks affecting several African countries including Uganda, where thousands of confirmed cases have been reported. Because conventional laboratory PCR testing remains centralized and difficult to access in many low-resource or decentralized settings, the study aims to determine whether rapid molecular POC assays can provide accurate and scalable diagnostic alternatives to improve outbreak response, surveillance, patient management, and contact tracing. The study evaluates three investigational molecular diagnostic platforms: the VZV-Q Real-Time PCR Kit for OnePCR (Genes2Me, India), the RADI-ONE Mpox Detection Kit (KH Medical, South Korea), and the STANDARD M10 MPX/OPX assay (SD Biosensor, South Korea). Their performance is compared against the laboratory-based BioPerfectus Monkeypox Virus Real Time PCR Kit, which serves as the reference standard and is currently listed under the African Medicines Regulatory Harmonisation Emergency Use Listing for mpox diagnosis. The study uses archived lesion swab samples stored in viral or universal transport medium from suspected mpox cases collected since 2020 under routine surveillance or prior research activities. Eligible samples must have known PCR status, adequate volume for both reference and index testing, and acceptable storage conditions.
The primary objective is to determine the clinical sensitivity/positive percent agreement (PPA) and specificity/negative percent agreement (NPA) of each investigational assay relative to the reference PCR, while secondary objectives include assessing performance according to PCR cycle threshold (Ct) values as a proxy for viral load and evaluating potential differences in performance by circulating mpox virus clades if such information is available. Statistical analyses will use Wilson score confidence intervals, with subgroup analyses by age, sex, Ct value, and viral clade. The study follows a blinded case-control design in which specimens are de-identified, shuffled, relabeled with unique FIND study identifiers, and tested independently to minimize bias. Reference PCR testing is performed first to confirm specimen quality and classification before testing on the investigational devices. Invalid or inconclusive results are repeated once and documented according to predefined procedures. No patient recruitment or intervention occurs, and test results are not used for clinical management, making the study low risk from an ethical and safety perspective.
Data are collected using OpenClinica Enterprise Edition, a validated electronic data capture system with audit trails and secure cloud hosting. FIND oversees monitoring, quality assurance, operator training, and regulatory compliance throughout the study.
The study is funded by FIND with support from the Government of the Netherlands and is expected to generate independent evidence to support regulatory submissions, procurement decisions, WHO evaluation processes, and future implementation of rapid mpox diagnostics in low- and middle-income countries.
Tipo di studio
Iscrizione (Stimato)
Criteri di partecipazione
Criteri di ammissibilità
Età idonea allo studio
- Bambino
- Adulto
- Adulto più anziano
Accetta volontari sani
Metodo di campionamento
Popolazione di studio
Descrizione
Inclusion Criteria:
- Lesion swabs stored in VTM/UTM (non-inactivating)
- Aliquot with sufficient volume to run the reference PCR and at least one index test (≥2600 μL per pair reference-index tests: ideally at least 1400 μL for 3 index tests and 1200 μL for reference PCR, including 1 possible repeat per test)
- Available PCR result for specimen selection (i.e. confirmed MPXV positive or negative)
- Samples collected from 2020 onwards.
Exclusion Criteria:
- Unknown date of collection
- >3 freeze-thaw cycles of sample.
- Samples collected with expired swab collection devices or UTM/VTM
Piano di studio
Come è strutturato lo studio?
Dettagli di progettazione
Coorti e interventi
Gruppo / Coorte |
Intervento / Trattamento |
|---|---|
|
Positive samples
50 confirmed MPXV-positive samples
|
Each (negative and positive) sample will be tested on the standard of reference and the investigational IVD
|
|
Negative samples
30 confirmed MPXV-negative samples
|
Each (negative and positive) sample will be tested on the standard of reference and the investigational IVD
|
Cosa sta misurando lo studio?
Misure di risultato primarie
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
|---|---|---|
|
Point estimates of sensitivity, specificity, positive and negative predictive value (PPV and NPV respectively) with 95% confidence intervals.
Lasso di tempo: July-August 2026
|
Point estimates of sensitivity, specificity, positive and negative predictive value (PPV and NPV respectively) with 95% confidence intervals.
|
July-August 2026
|
Misure di risultato secondarie
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
|---|---|---|
|
Point estimates of sensitivity/PPA stratified by Ct values of the reference PCR test.
Lasso di tempo: July-August 2026
|
Point estimates of sensitivity/PPA stratified by Ct values of the reference PCR test.
|
July-August 2026
|
|
Point estimates of sensitivity/PPA and specificity/NPA with 95% confidence intervals stratified by virus clade.
Lasso di tempo: July-August 2026
|
Point estimates of sensitivity/PPA and specificity/NPA with 95% confidence intervals stratified by virus clade.
|
July-August 2026
|
Collaboratori e investigatori
Studiare le date dei record
Studia le date principali
Inizio studio (Stimato)
Completamento primario (Stimato)
Completamento dello studio (Stimato)
Date di iscrizione allo studio
Primo inviato
Primo inviato che soddisfa i criteri di controllo qualità
Primo Inserito (Effettivo)
Aggiornamenti dei record di studio
Ultimo aggiornamento pubblicato (Effettivo)
Ultimo aggiornamento inviato che soddisfa i criteri QC
Ultimo verificato
Maggiori informazioni
Termini relativi a questo studio
Termini MeSH pertinenti aggiuntivi
Altri numeri di identificazione dello studio
- PP016
Informazioni su farmaci e dispositivi, documenti di studio
Studia un dispositivo regolamentato dalla FDA degli Stati Uniti
Queste informazioni sono state recuperate direttamente dal sito web clinicaltrials.gov senza alcuna modifica. In caso di richieste di modifica, rimozione o aggiornamento dei dettagli dello studio, contattare register@clinicaltrials.gov. Non appena verrà implementata una modifica su clinicaltrials.gov, questa verrà aggiornata automaticamente anche sul nostro sito web .
Prove cliniche su Vaiolo delle scimmie
-
Institut PasteurReclutamento
-
Children's Hospital of Fudan UniversityPeople's Hospital of Xinjiang Uygur Autonomous Region; Hainan women and children... e altri collaboratoriCompletatoGamma POX di neonati sani a diverse altitudini | Soglia di vaiolo per lo screening della CHD maggiore nei neonatiCina
Prove cliniche su MPOX IVD test
-
Samsung ElectronicsNexus DXTerminatoInsufficienza cardiaca congestiziaStati Uniti
-
Firalis SAReclutamento
-
Genlantis DiagnosticsNorthwestern UniversityNon ancora reclutamentoDimostrare le prestazioni cliniche del dispositivo diagnostico TriQuik Invitro
-
Istituto Oncologico Veneto IRCCSAttivo, non reclutanteCarcinoma, carcinoma polmonare non a piccole cellule (NSCLC)Italia
-
Abramson Cancer Center at Penn MedicineTranslational Breast Cancer Research Consortium; Danaher Inc.ReclutamentoCarcinoma mammario metastatico HER2-positivoStati Uniti
-
Fundación FLS de Lucha Contra el Sida, las Enfermedades...London School of Hygiene and Tropical Medicine; Hospital Clinic of Barcelona; Germans... e altri collaboratoriCompletatoMpoxSpagna, Perù, Chile, Panama
-
BioFire Defense LLCEmory University; Wake Forest University Health Sciences; George Washington University e altri collaboratoriReclutamentoCOVID | Coronavirus | Infezione respiratoriaStati Uniti
-
Chang Gung Memorial HospitalDr. Chip Biotechnology IncorporationCompletatoNeoplasia intraepiteliale cervicale | Infezione da HPV | Infezione da papillomavirus umanoTaiwan
-
Cyclacel Pharmaceuticals, Inc.TerminatoCarcinoma polmonare non a piccole celluleStati Uniti
-
Shepherd Medical CompanySconosciutoSterilizzazione maschileStati Uniti