SpO2 Infant and Pediatric Study (SpO2 Peds)
Assessment of TruSignal SpO2 Form, Fit and Clinical Performance With Infant and Child Patients
研究概览
详细说明
This prospective clinical study evaluates a pulse oximetry sensor in hospitalized infants (28 days to 23 months) and children (2-12 years) under routine clinical conditions. The primary objective is to assess the accuracy and agreement of SpO₂ measurements compared with reference SaO₂ values obtained from CO-oximetry analysis of simultaneously drawn arterial blood, using methods aligned with ISO 80601-2-61: 2026 3rd edition published in April 2026.
Secondary objectives include measuring the proportion of time the device produces invalid or unusable data to assess reliability for clinical use. The safety objective is to collect safety information, including type and number of AEs, SAEs, and device issues. Results may support product development, regulatory submissions, and marketing activities.
研究类型
注册 (估计的)
阶段
- 不适用
联系人和位置
学习地点
-
-
Nebraska
-
Omaha、Nebraska、美国、68114
- University of Nebraska Medical Center Children's Hospital
-
接触:
- Michael Facciolo, MD
- 电话号码:402-671-6325
- 邮箱:mfacciolo@childrensnebraska.org
-
首席研究员:
- Michael Facciolo, MD
-
-
参与标准
资格标准
适合学习的年龄
- 孩子
接受健康志愿者
描述
Inclusion Criteria:
- subjects are 1-23 months (infant cohort) or 2-12 years old (child cohort)
- requiring arterial blood samples per the site's standard of care.
Exclusion Criteria:
- Fractional Methemoglobinemia (FMetHb >2%) or Fractional carboxyhemoglobinemia (FCOHb >3%), within 24 hours of arterial blood draw.
- Severe anemia (tHb <7g/dL) within 4 hours prior to arterial blood gas draw.
- Hematocrit (Hct) <25 within 4 hours prior to arterial blood gas draw.
- Patients with injuries, deformities, or abnormalities which may prevent proper application of the sensor.
- Patients on Extracorporeal Membrane Oxygenation (ECMO).
- Patients receiving point of care (POC) testing for blood gases.
- Patients with diseases/conditions that may cause differential (or reverse differential) cyanosis including, but not limited to:
- Patent ductus arteriosus with pulmonary arterial hypertension.
- Pulmonary hypertension and left-heart abnormalities aortic arch hypoplasia, interrupted aortic arch, critical coarctation, and critical aortic stenosis.
- Children with transposition of the great arteries (TGA) and coarctation of the aorta or interrupted aortic arch, and TGA with suprasystemic pulmonary vascular resistance.
- At the discretion of the principal investigator or designee due to the subject health condition.
学习计划
研究是如何设计的?
设计细节
- 主要用途:设备可行性
- 分配:不适用
- 介入模型:单组作业
- 屏蔽:无(打开标签)
武器和干预
参与者组/臂 |
干预/治疗 |
|---|---|
|
实验性的:All Subjects
Participants participating in this study will be connected to SpO2 sensors.
SpO2 data will be independently collected from each sensor and paired with simultaneous arterial blood sample SaO₂ values.
The protocol is consistent with the ISO 80601-2-61 pulse oximetry standard.
|
SpO2 sensors use red and infrared LED light to estimate the arterial oxygen saturation.
|
研究衡量的是什么?
主要结果指标
结果测量 |
措施说明 |
大体时间 |
|---|---|---|
|
Collection of SpO2 Percentage Sensor Data with simultaneous SaO2 values
大体时间:From enrollment to the end of standard of care arterial cannulation, approximately 1 week.
|
The collection of SpO2 percentage sensor data with simultaneous SaO2 values percentage from each subject enrolled in the study.
|
From enrollment to the end of standard of care arterial cannulation, approximately 1 week.
|
其他结果措施
结果测量 |
措施说明 |
大体时间 |
|---|---|---|
|
安全事件的发生率
大体时间:4个月
|
对不良事件的分析将列出每个受试者,并以事件的计数和百分比进行总结,并以事件的受试者的计数和百分比进行了总结。
|
4个月
|
合作者和调查者
出版物和有用的链接
一般刊物
- Bland JM, Altman DG. Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat. 2007;17(4):571-82. doi: 10.1080/10543400701329422.
- Bickler PE, Feiner JR, Severinghaus JW. Effects of skin pigmentation on pulse oximeter accuracy at low saturation. Anesthesiology. 2005 Apr;102(4):715-9. doi: 10.1097/00000542-200504000-00004.
- Okunlola OE, Lipnick MS, Batchelder PB, Bernstein M, Feiner JR, Bickler PE. Pulse Oximeter Performance, Racial Inequity, and the Work Ahead. Respir Care. 2022 Feb;67(2):252-257. doi: 10.4187/respcare.09795. Epub 2021 Nov 12.
- Ruppel H, Makeneni S, Faerber JA, Lane-Fall MB, Foglia EE, O'Byrne ML, Bonafide CP. Evaluating the Accuracy of Pulse Oximetry in Children According to Race. JAMA Pediatr. 2023 May 1;177(5):540-543. doi: 10.1001/jamapediatrics.2023.0071.
研究记录日期
研究主要日期
学习开始 (估计的)
初级完成 (估计的)
研究完成 (估计的)
研究注册日期
首次提交
首先提交符合 QC 标准的
首次发布 (实际的)
研究记录更新
最后更新发布 (实际的)
上次提交的符合 QC 标准的更新
最后验证
更多信息
与本研究相关的术语
其他研究编号
- SA-000192
计划个人参与者数据 (IPD)
计划共享个人参与者数据 (IPD)?
药物和器械信息、研究文件
研究美国 FDA 监管的药品
研究美国 FDA 监管的设备产品
在美国制造并从美国出口的产品
此信息直接从 clinicaltrials.gov 网站检索,没有任何更改。如果您有任何更改、删除或更新研究详细信息的请求,请联系 register@clinicaltrials.gov. clinicaltrials.gov 上实施更改,我们的网站上也会自动更新.