SpO2 Infant and Pediatric Study (SpO2 Peds)
Assessment of TruSignal SpO2 Form, Fit and Clinical Performance With Infant and Child Patients
Studieoversigt
Status
Betingelser
Intervention / Behandling
Detaljeret beskrivelse
This prospective clinical study evaluates a pulse oximetry sensor in hospitalized infants (28 days to 23 months) and children (2-12 years) under routine clinical conditions. The primary objective is to assess the accuracy and agreement of SpO₂ measurements compared with reference SaO₂ values obtained from CO-oximetry analysis of simultaneously drawn arterial blood, using methods aligned with ISO 80601-2-61 (3rd edition CD draft) and FDA draft guidance.
Secondary objectives include measuring the proportion of time the device produces invalid or unusable data to assess reliability for clinical use. The safety objective is to collect safety information, including type and number of AEs, SAEs, and device issues. Results may support product development, regulatory submissions, and marketing activities.
Undersøgelsestype
Tilmelding (Anslået)
Fase
- Ikke anvendelig
Kontakter og lokationer
Studiesteder
-
-
Nebraska
-
Omaha, Nebraska, Forenede Stater, 68114
- University of Nebraska Medical Center Children's Hospital
-
Kontakt:
- Michael Facciolo, MD
- Telefonnummer: 402-671-6325
- E-mail: mfacciolo@childrensnebraska.org
-
Ledende efterforsker:
- Michael Facciolo, MD
-
-
Deltagelseskriterier
Berettigelseskriterier
Aldre berettiget til at studere
- Barn
Tager imod sunde frivillige
Beskrivelse
Inclusion Criteria:
- subjects are 1-23 months (infant cohort) or 2-12 years old (pediatric cohort)
- requiring arterial blood samples per the site's standard of care.
Exclusion Criteria:
- Fractional Methemoglobinemia (FMetHb >2%) or Fractional carboxyhemoglobinemia (FCOHb >3%), within 24 hours of arterial blood draw.
- Severe anemia (tHb <7g/dL) within 4 hours prior to arterial blood gas draw.
- Hematocrit (Hct) <25 within 4 hours prior to arterial blood gas draw.
- Patients with injuries, deformities, or abnormalities which may prevent proper application of the sensor.
- Patients on Extracorporeal Membrane Oxygenation (ECMO).
- Patients receiving point of care (POC) testing for blood gases.
- Patients with diseases/conditions that may cause differential (or reverse differential) cyanosis including, but not limited to:
- Patent ductus arteriosus with pulmonary arterial hypertension.
- Pulmonary hypertension and left-heart abnormalities aortic arch hypoplasia, interrupted aortic arch, critical coarctation, and critical aortic stenosis.
- Children with transposition of the great arteries (TGA) and coarctation of the aorta or interrupted aortic arch, and TGA with suprasystemic pulmonary vascular resistance.
- At the discretion of the principal investigator or designee due to the subject health condition.
Studieplan
Hvordan er undersøgelsen tilrettelagt?
Design detaljer
- Primært formål: Enhedens gennemførlighed
- Tildeling: N/A
- Interventionel model: Enkelt gruppeopgave
- Maskning: Ingen (Åben etiket)
Våben og indgreb
Deltagergruppe / Arm |
Intervention / Behandling |
|---|---|
|
Eksperimentel: All Subjects
Participants participating in this study will be connected to SpO2 sensors.
SpO2 data will be independently collected from each sensor and paired with simultaneous arterial blood sample SaO₂ values.
The protocol is consistent with the ISO 80601-2-61 pulse oximetry standard.
|
SpO2 sensors use red and infrared LED light to estimate the arterial oxygen saturation.
|
Hvad måler undersøgelsen?
Primære resultatmål
Resultatmål |
Foranstaltningsbeskrivelse |
Tidsramme |
|---|---|---|
|
Collection of SpO2 Percentage Sensor Data with simultaneous SaO2 values
Tidsramme: From enrollment to the end of standard of care arterial cannulation, approximately 1 week.
|
The collection of SpO2 percentage sensor data with simultaneous SaO2 values percentage from each subject enrolled in the study.
|
From enrollment to the end of standard of care arterial cannulation, approximately 1 week.
|
Andre resultatmål
Resultatmål |
Foranstaltningsbeskrivelse |
Tidsramme |
|---|---|---|
|
Forekomst af sikkerhedsbegivenheder
Tidsramme: 4 måneder
|
Analyse af bivirkninger vil blive opført pr. Emne, opsummeret med tællinger og procentdel af begivenheder og opsummeret med tællinger og procentdel af emner med begivenheder.
|
4 måneder
|
Samarbejdspartnere og efterforskere
Sponsor
Publikationer og nyttige links
Generelle publikationer
- Bland JM, Altman DG. Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat. 2007;17(4):571-82. doi: 10.1080/10543400701329422.
- Bickler PE, Feiner JR, Severinghaus JW. Effects of skin pigmentation on pulse oximeter accuracy at low saturation. Anesthesiology. 2005 Apr;102(4):715-9. doi: 10.1097/00000542-200504000-00004.
- Okunlola OE, Lipnick MS, Batchelder PB, Bernstein M, Feiner JR, Bickler PE. Pulse Oximeter Performance, Racial Inequity, and the Work Ahead. Respir Care. 2022 Feb;67(2):252-257. doi: 10.4187/respcare.09795. Epub 2021 Nov 12.
- Ruppel H, Makeneni S, Faerber JA, Lane-Fall MB, Foglia EE, O'Byrne ML, Bonafide CP. Evaluating the Accuracy of Pulse Oximetry in Children According to Race. JAMA Pediatr. 2023 May 1;177(5):540-543. doi: 10.1001/jamapediatrics.2023.0071.
Datoer for undersøgelser
Studer store datoer
Studiestart (Anslået)
Primær færdiggørelse (Anslået)
Studieafslutning (Anslået)
Datoer for studieregistrering
Først indsendt
Først indsendt, der opfyldte QC-kriterier
Først opslået (Faktiske)
Opdateringer af undersøgelsesjournaler
Sidste opdatering sendt (Faktiske)
Sidste opdatering indsendt, der opfyldte kvalitetskontrolkriterier
Sidst verificeret
Mere information
Begreber relateret til denne undersøgelse
Andre undersøgelses-id-numre
- SA-000192
Plan for individuelle deltagerdata (IPD)
Planlægger du at dele individuelle deltagerdata (IPD)?
Lægemiddel- og udstyrsoplysninger, undersøgelsesdokumenter
Studerer et amerikansk FDA-reguleret lægemiddelprodukt
Studerer et amerikansk FDA-reguleret enhedsprodukt
produkt fremstillet i og eksporteret fra U.S.A.
Disse oplysninger blev hentet direkte fra webstedet clinicaltrials.gov uden ændringer. Hvis du har nogen anmodninger om at ændre, fjerne eller opdatere dine undersøgelsesoplysninger, bedes du kontakte register@clinicaltrials.gov. Så snart en ændring er implementeret på clinicaltrials.gov, vil denne også blive opdateret automatisk på vores hjemmeside .
Kliniske forsøg med SpO2
-
Czech Technical University in PragueAfsluttetBevægelsesartefakter under SpO2-målinger | Reduceret perfusion under SpO2-målingerTjekkiet
-
Masimo CorporationRekruttering
-
Masimo CorporationAfsluttet
-
Philips Clinical & Medical Affairs GlobalAfsluttetSpO2-måling (perifer iltmætning) ValideringForenede Stater
-
National Taiwan University HospitalRekruttering
-
Philips Clinical & Medical Affairs GlobalAfsluttet
-
GE HealthcareTrukket tilbageRespiration | Temperatur | Ikke-invasivt blodtryk | SpO2 | EKGForenede Stater
-
New York University Abu DhabiIQAir AG; LUT UniversityAfsluttetSøvn | Hjerterytme | Åndedrætsfrekvens | Eksponering for luftforurening | SpO2 | HRV | HvilepulsForenede Arabiske Emirater
Kliniske forsøg med SpO2 Sensor
-
GE HealthcareAfsluttetSpO2 | Perifer iltmætningsmålingForenede Stater
-
University Hospital Inselspital, BerneAfsluttetVenstre ventrikulær hjælpeanordningSchweiz
-
Dormotech MedicalAfsluttet
-
Masimo CorporationRekruttering
-
Medtronic - MITGAfsluttet
-
Vanderbilt University Medical CenterAfsluttet
-
University of AlbertaAlberta Innovates Health Solutions; Heart and Stroke Foundation of CanadaAfsluttet
-
University Hospital TuebingenFritz Stephan GmbHUkendtVentilator Lunge; Nyfødt | Infantil Respiratory Distress SyndromeTyskland
-
University Hospital, AngersAfsluttet