SpO2 Infant and Pediatric Study (SpO2 Peds)
Assessment of TruSignal SpO2 Form, Fit and Clinical Performance With Infant and Child Patients
Aperçu de l'étude
Statut
Intervention / Traitement
Description détaillée
This prospective clinical study evaluates a pulse oximetry sensor in hospitalized infants (28 days to 23 months) and children (2-12 years) under routine clinical conditions. The primary objective is to assess the accuracy and agreement of SpO₂ measurements compared with reference SaO₂ values obtained from CO-oximetry analysis of simultaneously drawn arterial blood, using methods aligned with ISO 80601-2-61: 2026 3rd edition published in April 2026.
Secondary objectives include measuring the proportion of time the device produces invalid or unusable data to assess reliability for clinical use. The safety objective is to collect safety information, including type and number of AEs, SAEs, and device issues. Results may support product development, regulatory submissions, and marketing activities.
Type d'étude
Inscription (Estimé)
Phase
- N'est pas applicable
Contacts et emplacements
Lieux d'étude
-
-
Nebraska
-
Omaha, Nebraska, États-Unis, 68114
- University of Nebraska Medical Center Children's Hospital
-
Contact:
- Michael Facciolo, MD
- Numéro de téléphone: 402-671-6325
- E-mail: mfacciolo@childrensnebraska.org
-
Chercheur principal:
- Michael Facciolo, MD
-
-
Critères de participation
Critère d'éligibilité
Âges éligibles pour étudier
- Enfant
Accepte les volontaires sains
La description
Inclusion Criteria:
- subjects are 1-23 months (infant cohort) or 2-12 years old (child cohort)
- requiring arterial blood samples per the site's standard of care.
Exclusion Criteria:
- Fractional Methemoglobinemia (FMetHb >2%) or Fractional carboxyhemoglobinemia (FCOHb >3%), within 24 hours of arterial blood draw.
- Severe anemia (tHb <7g/dL) within 4 hours prior to arterial blood gas draw.
- Hematocrit (Hct) <25 within 4 hours prior to arterial blood gas draw.
- Patients with injuries, deformities, or abnormalities which may prevent proper application of the sensor.
- Patients on Extracorporeal Membrane Oxygenation (ECMO).
- Patients receiving point of care (POC) testing for blood gases.
- Patients with diseases/conditions that may cause differential (or reverse differential) cyanosis including, but not limited to:
- Patent ductus arteriosus with pulmonary arterial hypertension.
- Pulmonary hypertension and left-heart abnormalities aortic arch hypoplasia, interrupted aortic arch, critical coarctation, and critical aortic stenosis.
- Children with transposition of the great arteries (TGA) and coarctation of the aorta or interrupted aortic arch, and TGA with suprasystemic pulmonary vascular resistance.
- At the discretion of the principal investigator or designee due to the subject health condition.
Plan d'étude
Comment l'étude est-elle conçue ?
Détails de conception
- Objectif principal: Faisabilité de l'appareil
- Répartition: N / A
- Modèle interventionnel: Affectation à un seul groupe
- Masquage: Aucun (étiquette ouverte)
Armes et Interventions
Groupe de participants / Bras |
Intervention / Traitement |
|---|---|
|
Expérimental: All Subjects
Participants participating in this study will be connected to SpO2 sensors.
SpO2 data will be independently collected from each sensor and paired with simultaneous arterial blood sample SaO₂ values.
The protocol is consistent with the ISO 80601-2-61 pulse oximetry standard.
|
SpO2 sensors use red and infrared LED light to estimate the arterial oxygen saturation.
|
Que mesure l'étude ?
Principaux critères de jugement
Mesure des résultats |
Description de la mesure |
Délai |
|---|---|---|
|
Collection of SpO2 Percentage Sensor Data with simultaneous SaO2 values
Délai: From enrollment to the end of standard of care arterial cannulation, approximately 1 week.
|
The collection of SpO2 percentage sensor data with simultaneous SaO2 values percentage from each subject enrolled in the study.
|
From enrollment to the end of standard of care arterial cannulation, approximately 1 week.
|
Autres mesures de résultats
Mesure des résultats |
Description de la mesure |
Délai |
|---|---|---|
|
Incidence des événements de sécurité
Délai: 4 mois
|
L'analyse des événements indésirables sera répertoriée par sujet, résumée avec des dénombrements et des pourcentages d'événements, et résumées avec des dénombrements et des pourcentages de sujets avec des événements.
|
4 mois
|
Collaborateurs et enquêteurs
Parrainer
Publications et liens utiles
Publications générales
- Bland JM, Altman DG. Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat. 2007;17(4):571-82. doi: 10.1080/10543400701329422.
- Bickler PE, Feiner JR, Severinghaus JW. Effects of skin pigmentation on pulse oximeter accuracy at low saturation. Anesthesiology. 2005 Apr;102(4):715-9. doi: 10.1097/00000542-200504000-00004.
- Okunlola OE, Lipnick MS, Batchelder PB, Bernstein M, Feiner JR, Bickler PE. Pulse Oximeter Performance, Racial Inequity, and the Work Ahead. Respir Care. 2022 Feb;67(2):252-257. doi: 10.4187/respcare.09795. Epub 2021 Nov 12.
- Ruppel H, Makeneni S, Faerber JA, Lane-Fall MB, Foglia EE, O'Byrne ML, Bonafide CP. Evaluating the Accuracy of Pulse Oximetry in Children According to Race. JAMA Pediatr. 2023 May 1;177(5):540-543. doi: 10.1001/jamapediatrics.2023.0071.
Dates d'enregistrement des études
Dates principales de l'étude
Début de l'étude (Estimé)
Achèvement primaire (Estimé)
Achèvement de l'étude (Estimé)
Dates d'inscription aux études
Première soumission
Première soumission répondant aux critères de contrôle qualité
Première publication (Réel)
Mises à jour des dossiers d'étude
Dernière mise à jour publiée (Réel)
Dernière mise à jour soumise répondant aux critères de contrôle qualité
Dernière vérification
Plus d'information
Termes liés à cette étude
Autres numéros d'identification d'étude
- SA-000192
Plan pour les données individuelles des participants (IPD)
Prévoyez-vous de partager les données individuelles des participants (DPI) ?
Informations sur les médicaments et les dispositifs, documents d'étude
Étudie un produit pharmaceutique réglementé par la FDA américaine
Étudie un produit d'appareil réglementé par la FDA américaine
produit fabriqué et exporté des États-Unis.
Ces informations ont été extraites directement du site Web clinicaltrials.gov sans aucune modification. Si vous avez des demandes de modification, de suppression ou de mise à jour des détails de votre étude, veuillez contacter register@clinicaltrials.gov. Dès qu'un changement est mis en œuvre sur clinicaltrials.gov, il sera également mis à jour automatiquement sur notre site Web .