SpO2 Infant and Pediatric Study (SpO2 Peds)
Assessment of TruSignal SpO2 Form, Fit and Clinical Performance With Infant and Child Patients
Přehled studie
Postavení
Intervence / Léčba
Detailní popis
This prospective clinical study evaluates a pulse oximetry sensor in hospitalized infants (28 days to 23 months) and children (2-12 years) under routine clinical conditions. The primary objective is to assess the accuracy and agreement of SpO₂ measurements compared with reference SaO₂ values obtained from CO-oximetry analysis of simultaneously drawn arterial blood, using methods aligned with ISO 80601-2-61 (3rd edition CD draft) and FDA draft guidance.
Secondary objectives include measuring the proportion of time the device produces invalid or unusable data to assess reliability for clinical use. The safety objective is to collect safety information, including type and number of AEs, SAEs, and device issues. Results may support product development, regulatory submissions, and marketing activities.
Typ studie
Zápis (Odhadovaný)
Fáze
- Nelze použít
Kontakty a umístění
Studijní místa
-
-
Nebraska
-
Omaha, Nebraska, Spojené státy, 68114
- University of Nebraska Medical Center Children's Hospital
-
Kontakt:
- Michael Facciolo, MD
- Telefonní číslo: 402-671-6325
- E-mail: mfacciolo@childrensnebraska.org
-
Vrchní vyšetřovatel:
- Michael Facciolo, MD
-
-
Kritéria účasti
Kritéria způsobilosti
Věk způsobilý ke studiu
- Dítě
Přijímá zdravé dobrovolníky
Popis
Inclusion Criteria:
- subjects are 1-23 months (infant cohort) or 2-12 years old (pediatric cohort)
- requiring arterial blood samples per the site's standard of care.
Exclusion Criteria:
- Fractional Methemoglobinemia (FMetHb >2%) or Fractional carboxyhemoglobinemia (FCOHb >3%), within 24 hours of arterial blood draw.
- Severe anemia (tHb <7g/dL) within 4 hours prior to arterial blood gas draw.
- Hematocrit (Hct) <25 within 4 hours prior to arterial blood gas draw.
- Patients with injuries, deformities, or abnormalities which may prevent proper application of the sensor.
- Patients on Extracorporeal Membrane Oxygenation (ECMO).
- Patients receiving point of care (POC) testing for blood gases.
- Patients with diseases/conditions that may cause differential (or reverse differential) cyanosis including, but not limited to:
- Patent ductus arteriosus with pulmonary arterial hypertension.
- Pulmonary hypertension and left-heart abnormalities aortic arch hypoplasia, interrupted aortic arch, critical coarctation, and critical aortic stenosis.
- Children with transposition of the great arteries (TGA) and coarctation of the aorta or interrupted aortic arch, and TGA with suprasystemic pulmonary vascular resistance.
- At the discretion of the principal investigator or designee due to the subject health condition.
Studijní plán
Jak je studie koncipována?
Detaily designu
- Primární účel: Proveditelnost zařízení
- Přidělení: N/A
- Intervenční model: Přiřazení jedné skupiny
- Maskování: Žádné (otevřený štítek)
Zbraně a zásahy
Skupina účastníků / Arm |
Intervence / Léčba |
|---|---|
|
Experimentální: All Subjects
Participants participating in this study will be connected to SpO2 sensors.
SpO2 data will be independently collected from each sensor and paired with simultaneous arterial blood sample SaO₂ values.
The protocol is consistent with the ISO 80601-2-61 pulse oximetry standard.
|
SpO2 sensors use red and infrared LED light to estimate the arterial oxygen saturation.
|
Co je měření studie?
Primární výstupní opatření
Měření výsledku |
Popis opatření |
Časové okno |
|---|---|---|
|
Collection of SpO2 Percentage Sensor Data with simultaneous SaO2 values
Časové okno: From enrollment to the end of standard of care arterial cannulation, approximately 1 week.
|
The collection of SpO2 percentage sensor data with simultaneous SaO2 values percentage from each subject enrolled in the study.
|
From enrollment to the end of standard of care arterial cannulation, approximately 1 week.
|
Další výstupní opatření
Měření výsledku |
Popis opatření |
Časové okno |
|---|---|---|
|
Výskyt bezpečnostních událostí
Časové okno: 4 měsíce
|
Analýza nežádoucích účinků bude uvedena na subjekt, shrnuta s počty a procenty událostí a shrnuta s počty a procenty subjektů s událostmi.
|
4 měsíce
|
Spolupracovníci a vyšetřovatelé
Sponzor
Publikace a užitečné odkazy
Obecné publikace
- Bland JM, Altman DG. Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat. 2007;17(4):571-82. doi: 10.1080/10543400701329422.
- Bickler PE, Feiner JR, Severinghaus JW. Effects of skin pigmentation on pulse oximeter accuracy at low saturation. Anesthesiology. 2005 Apr;102(4):715-9. doi: 10.1097/00000542-200504000-00004.
- Okunlola OE, Lipnick MS, Batchelder PB, Bernstein M, Feiner JR, Bickler PE. Pulse Oximeter Performance, Racial Inequity, and the Work Ahead. Respir Care. 2022 Feb;67(2):252-257. doi: 10.4187/respcare.09795. Epub 2021 Nov 12.
- Ruppel H, Makeneni S, Faerber JA, Lane-Fall MB, Foglia EE, O'Byrne ML, Bonafide CP. Evaluating the Accuracy of Pulse Oximetry in Children According to Race. JAMA Pediatr. 2023 May 1;177(5):540-543. doi: 10.1001/jamapediatrics.2023.0071.
Termíny studijních záznamů
Hlavní termíny studia
Začátek studia (Odhadovaný)
Primární dokončení (Odhadovaný)
Dokončení studie (Odhadovaný)
Termíny zápisu do studia
První předloženo
První předloženo, které splnilo kritéria kontroly kvality
První zveřejněno (Aktuální)
Aktualizace studijních záznamů
Poslední zveřejněná aktualizace (Aktuální)
Odeslaná poslední aktualizace, která splnila kritéria kontroly kvality
Naposledy ověřeno
Více informací
Termíny související s touto studií
Další identifikační čísla studie
- SA-000192
Plán pro data jednotlivých účastníků (IPD)
Plánujete sdílet data jednotlivých účastníků (IPD)?
Informace o lécích a zařízeních, studijní dokumenty
Studuje lékový produkt regulovaný americkým FDA
Studuje produkt zařízení regulovaný americkým úřadem FDA
produkt vyrobený a vyvážený z USA
Tyto informace byly beze změn načteny přímo z webu clinicaltrials.gov. Máte-li jakékoli požadavky na změnu, odstranění nebo aktualizaci podrobností studie, kontaktujte prosím register@clinicaltrials.gov. Jakmile bude změna implementována na clinicaltrials.gov, bude automaticky aktualizována i na našem webu .
Klinické studie na SpO2
-
Czech Technical University in PragueDokončenoArtefakty pohybu během měření SpO2 | Snížená perfuze během měření SpO2Česko
-
Masimo CorporationNábor
-
Masimo CorporationDokončeno
-
Philips Clinical & Medical Affairs GlobalDokončenoValidace měření SpO2 (periferní saturace kyslíkem).Spojené státy
-
National Taiwan University HospitalNáborKrevní tlak | SpO2Tchaj-wan
-
Philips Clinical & Medical Affairs GlobalUkončeno
-
GE HealthcareStaženoDýchání | Teplota | Neinvazivní krevní tlak | SpO2 | EKGSpojené státy
-
New York University Abu DhabiIQAir AG; LUT UniversityDokončenoSpát | Tepová frekvence | Dechová frekvence | Vystavení znečištění ovzduší | SpO2 | HRV | Klidová tepová frekvenceSpojené arabské emiráty
Klinické studie na SpO2 Sensor
-
University Hospital Inselspital, BerneDokončenoZařízení na podporu levé komoryŠvýcarsko
-
Vanderbilt University Medical CenterDokončeno
-
University of AlbertaAlberta Innovates Health Solutions; Heart and Stroke Foundation of CanadaDokončenoSrdeční selháníKanada
-
Northwestern UniversityDokončeno
-
Eunah Cho, MDDokončeno
-
University Hospital TuebingenFritz Stephan GmbHNeznámýVentilátor plíce; Novorozený | Syndrom dětské respirační tísněNěmecko
-
Léman Micro Devices SANeznámýV-Sensors for Vitals: Posouzení přesnosti senzoru vitálních funkcí namontovaného na chytrém telefonuKlinická přesnost senzoru tlaku vitálních funkcí
-
University of MiamiThe Gerber FoundationDokončenoApnoe předčasně narozených dětí | Periodické dýchání | Desaturace krve | Centrální apnoe | Obstrukční apnoe u novorozencůSpojené státy
-
Mayo ClinicDokončenoSpinopelvické vztahy ve vztahu k totální endoprotéze kyčleSpojené státy
-
Orthosensor, Inc.UkončenoOsteoartróza kolenaSpojené státy