Early Feasibility Study 2 of Outpatient Control-to-Range - Testing System Efficacy (Italy)

September 29, 2014 updated by: University of Virginia

Early Feasibility Study 2 of Outpatient Control-to-Range - Testing System Efficacy

An unblinded, randomized, cross-over design with each patient participating in two 40-hour outpatient admissions: (a) Experimental involving automated Control to Range (CTR) and (b) Control using Continuous Glucose Monitor (CGM)- augmented insulin pump treatment outside of a hospital based clinical research center.

Study Overview

Status

Completed

Intervention / Treatment

Detailed Description

The principal goal is to validate a smart phone-based control-to-range (CTR) system for ambulatory use and to estimate the effect of CTR vs. sensor-augmented pump therapy, thereby providing justification for further larger home-based trials of CTR.

Study Type

Interventional

Enrollment (Actual)

5

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

      • Padova, Italy, 35131
        • University of Padova

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

21 years to 64 years (Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Description

Inclusion Criteria:

  1. ≥21 and <65 years old.
  2. Clinical diagnosis of type 1 diabetes mellitus. For an individual to be enrolled at least one criterion from each list must be met.

    o Criteria for documented hyperglycemia (at least 1 must be met): i. Fasting glucose ≥126 mg/dL - confirmed ii. Two-hour Oral Glucose Tolerance Test (OGTT) glucose ≥200 mg/dL - confirmed iii. HbA1c ≥6.5% documented - confirmed iv. Random glucose ≥200 mg/dL with symptoms v. No data at diagnosis is available but the participant has a convincing history of hyperglycemia consistent with diabetes

    o Criteria for requiring insulin at diagnosis (1 must be met): i. Participant required insulin at diagnosis and continually thereafter ii. Participant did not start insulin at diagnosis but upon investigator review likely needed insulin (significant hyperglycemia that did not respond to oral agents) and did require insulin eventually and used continually iii. Participant did not start insulin at diagnosis but continued to be hyperglycemic, had positive islet cell antibodies - consistent with latent autoimmune diabetes in adults (LADA) and did require insulin eventually and used continually

  3. Use of an insulin pump to treat his/her diabetes for at least 1 year.
  4. Familiarity with a bolus calculator with the current insulin pump with pre-defined parameters for carbohydrate (CHO) ratio, insulin sensitivity factor (ISF), target glucose and active insulin.
  5. HbA1c <9% as measured with DCA2000 or equivalent device.
  6. Not currently known to be pregnant, breast feeding, or intending to become pregnant (females).
  7. Demonstration of proper mental status and cognition for the study.
  8. Willingness to avoid consumption of acetaminophen-containing products during the study interventions involving CGM use.
  9. If on antihypertensive, thyroid, anti-depressant or lipid lowering medication, have stability on the medication for at least 2 months prior to enrollment in the study.

Exclusion Criteria:

  1. Severe hypoglycemia resulting in seizure, loss of consciousness, or diabetic ketoacidosis within the 12 months prior to enrollment.
  2. Pregnancy; breast feeding, or intention of becoming pregnant.
  3. Uncontrolled arterial hypertension (Resting diastolic blood pressure >90 mmHg and/or systolic blood pressure >160 mmHg).
  4. Conditions which may increase the risks associated with possible hypoglycemia, such as any active cardiac disorder/arrhythmia, uncontrolled coronary artery disease during the previous year (e.g. history of myocardial infarction, acute coronary syndrome, therapeutic coronary intervention, coronary bypass or stenting procedure, stable or unstable angina, episode of chest pain of cardiac etiology with documented EKG changes, or positive stress test or catheterization with coronary blockages >50%), congestive heart failure, history of cerebrovascular event, seizure disorder, syncope, adrenal insufficiency, neurologic disease or atrial fibrillation.
  5. Self-reported hypoglycemia unawareness.
  6. History of a systemic or deep tissue infection with methicillin-resistant staph aureus or Candida albicans.
  7. Use of a device that may pose electromagnetic compatibility issues and/or radiofrequency interference with the CGM (implantable cardioverter-defibrillator, electronic pacemaker, neurostimulator, intrathecal pump, and cochlear implants).
  8. Anticoagulant therapy other than aspirin.
  9. Oral steroids.
  10. Medical condition requiring use of an acetaminophen-containing medication that cannot be withheld for the study admissions.
  11. Psychiatric disorders that would interfere with study tasks (e.g. inpatient psychiatric treatment within 6 months prior to enrollment).
  12. Mental incapacity, unwillingness or language barriers precluding adequate understanding or cooperation.
  13. Known current or recent alcohol or drug abuse.
  14. Medical conditions that would make operating a CGM, the DiAs cell phone or insulin pump difficult (e.g. blindness, severe arthritis, immobility).
  15. Any skin condition that prevents sensor or pump placement on the abdomen or arm (e.g. bad sunburn, pre-existing dermatitis, intertrigo, psoriasis, extensive scarring, cellulitis).
  16. In adherence with the One Touch Ultra 2 User Guide, subjects with hematocrit levels less than 30% and above 55% will be excluded.
  17. Impaired hepatic function measured as alanine aminotransferase or aspartate aminotransferase ≥three times the upper reference limit.
  18. Impaired renal function measured as creatinine >1.2 times above the upper limit of normal.
  19. Uncontrolled microvascular (diabetic) complications, such as current proliferative diabetic retinopathy or macular edema, known diabetic nephropathy (other than microalbuminuria with normal creatinine) or neuropathy requiring treatment.
  20. Active gastroparesis requiring current medical therapy.
  21. If on antihypertensive, thyroid, anti-depressant or lipid lowering medication, lack of stability on the medication for the past 2 months prior to enrollment in the study.
  22. Uncontrolled thyroid disease.
  23. Known bleeding diathesis or dyscrasia.
  24. Known allergy to medical adhesives, components of the insulin pump insertion set or continuous glucose monitor sensor.
  25. Active enrollment in another clinical trial.
  26. Use of anti-diabetic agents other than continuous subcutaneous insulin infusion (CSII) including long-acting insulin, intermediate-acting insulin, metformin, sulfonylureas, meglitinides, thiazolidinediones, DPP-IV inhibitors, glucagon-like peptide 1 agonists, and alpha-glucosidase inhibitors
  27. Subjects with basal rates less than 0.01U/hr.

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Treatment
  • Allocation: Randomized
  • Interventional Model: Crossover Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: Experimental Involving Automated CTR
Closed-Loop Control: Insulin delivery will be controlled by the Diabetes Assistant (DiAs) system running in Control to Range (CTR) or in Safety Only mode. Subjects will interact with the system through its Graphic User Interface (GUI). Subjects will not be allowed to administer correction boluses between meals and snacks as the DiAs will automatically be adjusting insulin to correct for hyperglycemia. The total doses recommended by the DiAs prior to meals and snacks includes the correction dose and Insulin on Board (IOB) calculated by the system.
A medical platform that uses a smart-phone to connect to a continuous glucose sensor to insulin pump and run closed-loop control. The cell phone runs the Control to Range and is connected to work with the insulin pump and continuous glucose monitor to help keep the blood sugar in a desired range (80-180 mg/dL during the day) and help avoid hypoglycemia during the night.
No Intervention: CGM-Augmented Insulin Pump Treatment
Open Loop Control: Insulin delivery will be controlled by the Diabetes Assistant (DiAs) system running in open-loop mode. Subjects will interact with the system through its Graphic User Interface (GUI). Subjects will be permitted to administer correction boluses at any time during the Control Admission, whether or not they are eating a scheduled meal or snack. DiAs will be initialized with the subject's typical insulin pump settings. Subjects will be reminded that all treatment decisions should be based on fingerstick values and not on continuous glucose monitor (CGM) values.

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Effect size of Control-to-Range (CTR) vs. Continuous Glucose Monitor (CGM)-augmented insulin pump treatment in an outpatient setting.
Time Frame: 40 hours
The investigators expect that compared to CGM-augmented insulin pump treatment, CTR will result in moderate effect size of approximately 0.4, in terms of reduction of the overnight risk for hypoglycemia as measured by the Low Blood Glucose Index computed from retrofitted CGM data. This effect is not expected to be statistically significant with the anticipated sample size but will be used to inform power analysis for the subsequent multi-center trial of CTR at home.
40 hours

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Time spent in target range
Time Frame: 40 hours
CTR will improve (non-significantly at the projected sample size of N=5 subjects/site) the time spent within the target range of 80-140 mg/dl overnight (computed from retrofitted CGM data) and will reduce the extent of postprandial glucose excursions during the day. These data will provide justification and design support for a subsequent larger multi-center trial of CTR at home.
40 hours

Other Outcome Measures

Outcome Measure
Measure Description
Time Frame
Patient comfort with the Diabetes Assistant (DiAs) user interface
Time Frame: 40 hours
Investigators want to ensure that subjects are comfortable using the cell phone to assess the current status of their diabetes management. Subjects will use the user guide as guidance on how to evaluate the DiAs' hypoglycemia and hyperglycemia warning lights, how to inform DiAs of carbohydrates consumed, and other important diabetes management issues.
40 hours
Reliability of DiAs remote monitoring
Time Frame: 40 hours
Assess the DiAs remote monitoring by medical personnel/technicians to confirm appropriate functioning outside of the hospital setting.
40 hours
Reliability of inter-device connections between DiAs and the CGM and between DiAs and the insulin pump.
Time Frame: 40 hours
Assess the functioning of the connections between DiAs, the continuous glucose sensor, and the insulin pump.
40 hours

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Investigators

  • Principal Investigator: Claudio Cobelli, Ph.D., University of Padova
  • Study Director: Boris P. Kovatchev, Ph.D., University of Virginia

Publications and helpful links

The person responsible for entering information about the study voluntarily provides these publications. These may be about anything related to the study.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start

January 1, 2013

Primary Completion (Actual)

May 1, 2013

Study Completion (Actual)

May 1, 2013

Study Registration Dates

First Submitted

November 8, 2012

First Submitted That Met QC Criteria

November 12, 2012

First Posted (Estimate)

November 16, 2012

Study Record Updates

Last Update Posted (Estimate)

September 30, 2014

Last Update Submitted That Met QC Criteria

September 29, 2014

Last Verified

September 1, 2014

More Information

Terms related to this study

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.